The AutoQUANT, Quantitative Perfusion SPECT (QPS), and Quantitative Gated SPECT (QGS) applications are intended to enable a fully automated review and quantification of Cardiac SPECT data.
Device Story
AutoQUANT is a software suite for automated review and quantification of Cardiac SPECT data. It functions as an independent application or integrates QPS and QGS modules. Inputs include myocardial rest and stress short-axis projections (standard and gated). The software identifies left ventricle orientation/extent, constructs 3D contour maps, and calculates heart volume, lung/heart ratio, and transient ischemic dilation (TID). QPS generates 2D/3D perfusion maps and calculates defect extent/volume. QGS analyzes gated data to produce wall motion, wall thickening, and regional ejection fraction maps. Physicians use these outputs to assess cardiac anatomical and physiological functionality and myocardial defects. The system supports clinical decision-making by providing standardized, automated measurements of cardiac performance.
Clinical Evidence
Bench testing only. Software verification testing demonstrated that each application functioned according to specifications, with actual results matching expected results.
Technological Characteristics
Software-based cardiac SPECT analysis suite. Performs automated quantification of perfusion, wall motion, wall thickening, and volumetric data. Operates on standard cardiac SPECT short-axis and gated short-axis data sets. Supports creation of user-definable normal databases. Connectivity and platform details not specified.
Indications for Use
Indicated for automated review and quantification of Cardiac SPECT data to assess myocardial perfusion, left ventricular volume, wall motion, and wall thickening in patients undergoing cardiac imaging.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
| AutoQUANT, Quantitative Perfusion SPECT (QPS), & Quantitative Gated SPECT (QGS) |
|---------------------------------------------------------------------------------|
| ADAC Laboratories |
| 510(k) Premarket Notification |
| Appendix VIII, 510(k) Summary of Safety and Effectiveness Data |
|----------------------------------------------------------------|
| Page 1 of 3 |
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA
- I. General Information
A. Submitted By:
K980715
ADAC Laboratories 540 Alder Drive Milpitas, CA 95035 Tel: (408) 321-9100 Fax: (408) 321-9686
Contact Person:
Dennis Henkelman at address above
AutoOUANT Device Trade Name: B. Quantitative Perfusion SPECT (QPS) Quantitative Gated SPECT (QGS) Gamma Camera System Common Name: System, Emission Computed Tomography Classification Name:
| C. Predicate Device: | ADAC CEQUAL®<br>Sopha Sophy NXT |
|----------------------|---------------------------------|
|----------------------|---------------------------------|
#### Device Description: D.
AutoQUANT is a software application designed to enable a fully automated, comprehensive review and quantification of Cardiac SPECT data within a suite of applications. AutoQUANT can operate as an independent single application or can be integrated into an application containing Quantitative Perfusion SPECT (QPS) and/or Quantitative Gated SPECT (QGS). AutoQUANT provides a tool to review and quantify all types of Cardiac SPECT data sets (perfusion and/or gated) to determine the location, orientation, and anatomical extent of the left ventricle of the heart, to construct 3D contour maps of the heart, and to calculate the heart volume (for the left ventricular wall), the lung/heart ratio, and transient ischemic dilation (TID). Physicians use this information to assess the anatomical and physiological functionality of the heart and analyze the presence of myocardial defects through comprehensive imaging modalities.
{1}------------------------------------------------
Quantitative Perfusion SPECT (QPS) is a software application designed for the review and quantification of myocardial perfusion SPECT short axis data. QPS can operate as an independent single application or as an embedded function packaged with AutoQUANT. Using the QPS application, myocardial rest and stress short axis projections (standard and gated Cardiac SPECT) can be reviewed and compared, the extent and volume of cardiac defects can be calculated, and 3D and 2D perfusion maps can be generated. The QPS application can also be used to create a user-definable Normal database (with definable normal limits) for review, quantification, and comparison.
Quantitative Gated SPECT (QGS) is a software application designed to view and analyze processed SPECT short axis or gated short axis data sets. QGS can operate as an independent single application or as an embedded function packaged with AutoQUANT. Using the QGS application, myocardial gated short axis projections can be reviewed and compared, the extent and severity of cardiac defects can be calculated, and 3D and 2D wall thickening, wall motion and regional ejection fraction maps can be generated. The application can be used to display the left ventricular endocardial and epicardial surfaces; polar maps indicating perfusion, wall thickening, wall motion, and regional ejection fraction: standard cardiac SPECT short axis and horizontal slices; and 3D cardiac surfaces and volumes from gated SPECT data. When using gated SPECT data, the left ventricular volume and left ventricular ejection fraction (LVEF) can be automatically calculated or regions may be manually specified for calculation. When using non-gated short axis SPECT data, the left ventricular volume can be calculated automatically or manually.
- E. Indications for Use:
The AutoQUANT , Quantitative Perfusion SPECT (QPS), and Quantitative Gated SPECT (QGS) applications are intended to enable a fully automated review and quantification of Cardiac SPECT data.
- F. Technological Comparison:
The AutoQUANT, OPS, QGS, ADAC CEQUAL, and Sopha Sophy NXT Cardiac Software applications have similar indications for use and utilize the same type of data sets for analysis and calculation of data. The ADAC QPS, OGS, CEQUAL, and Sophy NXT Cardiac Software perform the same calculations and analyses on the data sets.
{2}------------------------------------------------
#### II. Testing
Testing was conducted to demonstrate that each software application functioned as per its specifications. All tests passed with the actual results matching the expected results.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black and white, and the text is also in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
6 1998 MAY
Dennis W. Henkelman Director, Regulatory Affairs and Quality Assurance ADAC Laboratories 540 Alder Drive Milpitas, CA 95035
Re:
K980715 AutoQUANT (Gamma Camera System) Dated: February 23, 1998 Received: February 24, 1998 Regulatory class: II 21 CFR 892.1200/Procode: 90 KPS
Dear Mr. Henkelman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Thease note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: AutoQUANT Quantitative Perfusion SPECT (QPS) Quantitative Gated SPECT (QGS)
ADAC Laboratories Sponsor Name:
### Indications for Use
The AutoQUANT, Quantitative Perfusion SPECT (QPS), and Quantitative Gated SPECT (QGS) applications are intended to enable a fully automated review and quantification of Cardiac SPECT data.
Do Not Write Below This Line - Continue on Another Page if Needed
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use Over-The-Counter Use
David H. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological Dev 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.