K974360 · Becton Dickinson Immunocytometry Systems · GKZ · Mar 11, 1998 · Hematology
Device Facts
Record ID
K974360
Device Name
MULTITEST CD3/CD8/CD45/CD4
Applicant
Becton Dickinson Immunocytometry Systems
Product Code
GKZ · Hematology
Decision Date
Mar 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
MultiTEST CD3/CD8/CD45/CD4 reagent is a four-color reagent for identifying and enumerating CD3+ T Lymphocytes, CD3+CD4+ helper/inducer, and CD3+CD8+ suppresser/cytotoxic T-Lymphocyte subsets by direct immunofluorescence. Subsets of T Lymphocytes are useful in managing immunodeficiency diseases. To characterize and monitor congenital or acquired immunodeficiencies, such as SCID or AIDS.
Device Story
MultiTEST CD3/CD8/CD45/CD4 is a four-color direct immunofluorescence reagent kit for flow cytometry. Input: erythrocyte-lysed human whole blood samples. Reagent contains monoclonal antibodies conjugated to FITC, PE, PerCP, and APC fluorochromes. Operation: samples processed on FACSCalibur flow cytometers equipped with blue (488-nm) and red (635-nm) lasers; instrument setup requires CaliBRITE beads and FACSComp software. Data processing: MultiSET software performs single-tube fluorescence gating and automated quality control to identify and enumerate T-lymphocyte subsets (CD3+, CD3+CD4+, CD3+CD8+). Output: absolute counts or percentages of T-lymphocyte subsets. Used in clinical laboratories by trained technicians/pathologists. Results assist clinicians in characterizing and monitoring immunodeficiency and autoimmune disease states.
Clinical Evidence
Bench and clinical testing performed at multiple sites. Accuracy: 129 specimens (70 normal, 59 abnormal) compared 3-color vs 4-color reagents on FACSCalibur, demonstrating equivalence. Within-site reproducibility: 9 donors (3 normal, 6 abnormal) across low/medium/high cell counts, 10 aliquots each. Across-site reproducibility: 15 normal, 31 abnormal donors, 5 aliquots each. Stability: samples stable up to 48 hours post-draw and 24 hours post-staining. Linearity: linear response across 200-29,700 WBC/uL and 100-9,000 Lymphocytes/uL.
Technological Characteristics
Four-color direct immunofluorescence reagent kit. Monoclonal antibodies conjugated to FITC, PE, PerCP, and APC. Excitation via blue (488-nm) argon-ion laser and red (635-nm) diode laser. Requires FACSCalibur flow cytometer with FL4 option. Software-based gating and automated quality control. In vitro diagnostic use.
Indications for Use
Indicated for identification and enumeration of mature human T lymphocyte subsets in erythrocyte-lysed peripheral blood using flow cytometry. Used to characterize and monitor congenital or acquired immunodeficiencies (e.g., SCID, AIDS) and autoimmune diseases (e.g., lupus). For in vitro diagnostic use.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
# MAR 1 1 1998
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92.
The assigned 510(k) number is K974360.
### Submitter Information (21 CFR 807.92(a)(1))
Submirrer: Becton Dickinson Immunocyrometry Systems 2350 Qume Drive San Jose, CA 95131-1807
Contact: Cindy Morrow Sr. Regulatory Specialist (408) 954-2694
November 18, 1997 Summary date:
#### Name of Device and Classification (21 CFR 807.92(a)(2))
MultiTEST CD3/CD8/CD45/CD4 Name:
Classification: Class II Device
#### Predicate Device (21 CFR 807.92(a)(3))
MultiTEST CD3/CD8/CD45/CD4 reagent and TRUCOUNT absolute count tubes are substantially equivalent" to TriTEST CD3/CD8/CD45 (K970326), TriTEST CD3/CD4/CD45 (K965053), TRUCOUNT tubes (K970836) and MultiSET software (K963963).
#### Description of the Device (21 CFR 807.92(a)(4))
The Becton Dickinson CD3 fluorescein isothiocyanate (FITC)/CD8 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PerCP)/CD4 allophycocyanin (APC) reagent is a four-color, direct immunofluorescence reagent for identifying and enumerating percentages of CD3+ T Lymphocytes, CD3+CD4+ helper/inducer, and CD3+CD8+ suppresser/cyroroxic T-Lymphocyte subscts in erythrocyce-lysed whole blood. Each vial of reagent yields 50 tests. The reagent is intended for use on the Becton Dickinson FACSCalibur 10 flow cycomerers equipped with the FL4 Option, the Apple Macintosh Quadra or PowerPC computer, and CELLQuest or MultiSET software. Daily instrument set-up requires CaliBRITE beads (unlabeled, FITC, PE, Per CP, and APC) and FACSComp software.
" The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug, and Cosmetic Acr, as annended and as applied under 21 CFR 807, Subpart B under which a device can be marketed withour pre-marker approval or reclassification of substantial equivalency under this notification is not inrended to have any bearing whassever on the resolution of parent infringement suits or any other patent marrers. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Parent Laws or their application by the courts.
{1}------------------------------------------------
The four-color method permits the identification and enumeration of helper/inducer and suppresser/cytotoxic T-Lymphocyte subsets using MultiSET software's single-tube fluorescence gating and automated quality control algorithms. A lymphocyce gate is drawn for the CD45* leucocytes with low side scarter and the T-Lymphocyce subsers are provided as an absolute count or percentage of total lymphocytes.
Intended Use (21 CPR 807.92(a)(5))
MultiTEST CD3/CD8/CD45/CD4 reagent is a four-color reagent for identifying and enumerating CD3+ T Lymphocytes, CD3+CD4* helper/inducer, and CD3+CD8+ suppresser/cytoroxic T-Lymphocyte subsets by direct immunofluorescence. Subsets of T Lymphocytes are useful in managing immunodeficiency diseases. To characterize and monitor congenital or acquired immunodeficiencies, such as SCID or AIDS.
Technological Characteristics (21 CFR 807.92(a)(6))
MultiTEST CD3/CD8/CD45/CD4 is a four-color reagent and TriTEST CD3/CD8/CD45 and TriTEST CD3/CD4/CD45 are three-color reagents. The three and four-color reagents all employ the same monoclonal antibody clones, however the CD4 antibody is conjugated to PE in the three-color reagent and to APC in the four-color reagent. The three-color reagent includes FITC, PE, and PerCP fluorescent dyes that are excited by the flow cytometer's blue (488-nm) argon-ion laser. The four-color reagent contains an additional APC fluorescent dyc. The APC is excited by a red diode (635-am) laser provided in Bector Dickinson's FL4 Option for the FACSort and FACSCalibur flow cytometers.
Performance Data (21 CFR 807.92(b)(2))
Performance of the product was established by testing at Children's Memorial Hospiral, Chicago, Illinois; Covance Central Laboratory, Indiana; Cleveland Clinic Foundation Cleveland, Ohio; of Becton Dickinson Immunocycomerry Systems, San Jose, California.
Several studies were performed:
- · Accuracy was determined by comparing results from lysed whole blood (LWB) specimens stained with both the three- and four-color reagents on the FACSCalibur. A total of 129 specimens, including 70 normals and 59 abnormals were obtained and analyzed at two clinical sites. Accuracy data demonstrated that MultiTEST CD3/CD45/CD4 is equivalent to TriTEST CD3/CD8/CD45 and TriTEST CD3/CD4/CD45.
- · Within-site reproducibility was performed on LWB specimens from 3 normal and 6 abnormal donors across the range of low (100-350), medium (350-700), and high (5700) CD4+ T cell counts (cells/L). Each specimen was divided into 10 aliquots and then stained and analyzed using a FACSCalibur system. Results demonstrated acceptable within-site reproducibility.
{2}------------------------------------------------
- @ 005
- · Across-site reproducibility was performed at three clinical sites on LWB specimens from a total of 15 normal and 31 abnormal donors across the same range of low, medium, and high CD4+ T cell counts (cells/uL). Each specimen was divided into 5 aliquots and then stained and analyzed using a FACSCalibur system. Results demonstrated acceptable across site reproducibility.
- · Stability studies were conducted at one site using samples from 10 normal and 20 abnormal donors. The studies assessed changes associated with the storage of whole blood prior to staining, changes as a result of time between staining and dara acquisition, and the combined effect of time of storage and time after staining. Results demonstrate acceptable stability of samples prepared up to 48 hours after blood draw and stability of prepared samples up to 24 hours after preparation.
- Lincarity and recovery was determined using blood specimens from 3 normal donors diluted to 5 different concentrations, ranging from 200 to 29,700 WBC/uL and from 100 to 9000 Lymphocytes/uL. Results indicate linear response over this range and acceptable recovery.
Comparison to Predicate Device (21 CFR 807.92(b)(3))
The results of the clinical studies demonstrate that the device is as safe and effective as the predicate device. The four-color MultiTEST CD3/CD8/CD45/CD4 reagent is substantially equivalent to the predicate three-color TriTEST CD3/CD4/CD45 and TriTEST CD3/CD8/CD45 reagents, in that they share the same inrended uses and methodologies. Results demonstrate that the products yield essentially equivalent performance characteristics.
Cindy Monroe
Cindy Morrow Sr Regulatory Specialist
3/4/98
Date
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services-USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle with its wings spread, with three wavy lines below the eagle. The eagle and wavy lines are black, and the background is white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Cindy Morrow Sr. Regulatory Specialist Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, CA 95131-1807
MAR 1 1 1998
Re: K974360 Trade Name: Multitest CD3/CD8/CD45/CD4 Regulatory Class: II Product Code: GKZ Dated: December 17, 1997 Received: December 18, 1997
Dear Ms. Morrow:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours.
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Attachment A Indications for Use Statement
Device Name:
# MultiTEST CD3/CD8/CD45/CD4
## Indications For Use:
- For the FACS® family of flow cytometers equipped with a blue (488-nm) and a red t diode (635-nm) laser.
- A monoclonal antibody reagent for identification and enumeration of mature human T . lymphocyte subsers in human peripheral blood.
- For use with crythrocyte lysed whole blood. .
- To characterize and monitor forms of autoimmune diseases, such as lupus. .
- To characterize and monitof congenital or acquired immunodeficiencies, such as . SCID or AIDS.
- For in vitro diagnostic use.
Peter E. Mckin
(Division Sign-Off)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number ***_***
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.