TRUCOUNT ABSOLUTE COUNT TUBES/COUNTING CONTROL BEADS
K970836 · Becton Dickinson Immunocytometry Systems · GKZ · Aug 26, 1997 · Hematology
Device Facts
Record ID
K970836
Device Name
TRUCOUNT ABSOLUTE COUNT TUBES/COUNTING CONTROL BEADS
Applicant
Becton Dickinson Immunocytometry Systems
Product Code
GKZ · Hematology
Decision Date
Aug 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
TRUCOUNT Absolute Count Tubes are intended for use as an accessory to TriTEST in vitro diagnostics, such as that described in K965053 (CD3/CD4/CD45), to allow computation of positive cells per known volume of blood, using flow cytometry. The TRUCOUNT Counting Control Beads are provided in three levels, and are intended to provide a means to check the reproducibility and accuracy of volumetric pipetting of blood samples. They are an accessory to TriTEST reagents, when used with TRUCOUNT Absolute Count Tubes to obtain absolute counts of positive cells using flow cytometry.
Device Story
TRUCOUNT Absolute Count Tubes contain fluorescent beads in known quantities; blood sample of known volume added to tube; reagent added; flow cytometry measures positive cell events vs. bead events; ratio used to compute positive cells per unit volume. TRUCOUNT Counting Control Beads (High, Medium, Low) used to verify pipetting accuracy by comparing experimental counts to expected label values. Used in clinical laboratories; operated by technicians/clinicians. Output allows healthcare providers to monitor immunodeficiency and autoimmune disease via absolute lymphocyte subtype counts.
Clinical Evidence
Bench testing only. Stability testing confirmed one-year shelf life for Absolute Count Tubes and six-month shelf life for Counting Control Beads. Accuracy and reproducibility of control beads verified against established performance ranges. Functionality confirmed for both manual use and use with FACS Loader.
Technological Characteristics
Fluorescent particles (beads) supplied as pellets (tubes) or suspensions (control beads). Designed for use with flow cytometry systems. Compatible with FACS Loader. No electronic components; purely chemical/mechanical accessory for volumetric calibration in flow cytometry.
Indications for Use
Indicated for use with erythrocyte lysed whole blood, FACS Loader, IVD immunophenotyping reagents, and flow cytometers with designated detection ranges to obtain absolute cell counts.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
K970836
# Attachment D
## Summary of Safety & Effectiveness
AUG 26 1997
**Submitter Information (21 CFR 807.92(a)(1))**
Submitter: Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, CA 95131-1807
Contact: Anna Longwell, Esq.
Director, Regulatory Affairs – Corporate
(408) 954-2254
Summary Date: February 24, 1997
**Name of Device and Classification (21 CFR 807.92(a)(2))**
Name:
TRUCOUNT™ Absolute Count Tubes
TRUCOUNT™ Counting Control Beads
Classification: Accessories to a Class II Product, Lymphocyte immunophenotyping reagent
**Predicate Device (21 CFR 807.92(a)(3))**
Lymphocyte immunophenotyping reagents, when used for absolute counts are substantially equivalent to Simultest™ reagents, cleared in the Simultest IMK-Lymphocyte Kit, (K913192), plus ADCC, or, in the case of TriTEST™ reagents CD3/CD4/CD45 and CD3/CD8/CD45, to FACSCount, K933486. The absolute count application is the subject of current lymphocyte immunophenotyping reagent submissions.
**Description of the Device (21 CFR 807.92(a)(4))**
TRUCOUNT Absolute Count Tubes are fluorescent particles supplied as a pellet in a tube. The pellet contains a known number of beads. A blood sample of known volume is added to the tube. Then the appropriate reagent is added to the mixture. By comparing the number of positive cell events to the number of bead events in the flow cytometry data output, positive cells per unit volume of blood sample can be computed.
TRUCOUNT Counting Control Beads are fluorescent particles supplied as suspensions in three bead concentrations, High, Medium and Low. The fluorescence of these beads is similar to that of the Absolute Count Tube in FL1 and FL3, but is different in FL2, so that both beads may be run in the same sample, and distinguished from one another. Control beads are added to samples of normal blood prepared with TriTEST reagents and Absolute Count Tubes. The Control Bead count is obtained by comparison to the count of the Absolute Count Tubes. These experimental counts are compared to the expected count on the Control Bead label, to check the accuracy of pipetting.
TRUCOUNT 510(k) Notification
Attachment D: Summary of Safety and Effectiveness
{1}
Summary of Safety and Effectiveness
## Intended Use (21 CFR 807.92(a)(5))
TRUCOUNT Absolute Count Tubes are intended for use as an accessory to TriTEST in vitro diagnostics, such as that described in K965053 (CD3/CD4/CD45), to allow computation of positive cells per known volume of blood, using flow cytometry.
The TRUCOUNT Counting Control Beads are provided in three levels, and are intended to provide a means to check the reproducibility and accuracy of volumetric pipetting of blood samples. They are an accessory to TriTEST reagents, when used with TRUCOUNT Absolute Count Tubes to obtain absolute counts of positive cells using flow cytometry.
## Indications for Use
- For erythrocyte lysed whole blood
- For use with FACS Loader
- For use with IVD immunophenotyping reagents
- For use on flow cytometers with designated detection ranges
## Clinical Utility
The determination of absolute counts of subtypes of lymphocytes has been found useful in monitoring some forms of immunodeficiency and autoimmune disease.
## Comparison to Predicate Device (21 CFR 807.92(a)(6))
Use of the TRUCOUNT Absolute Count Tubes with each of the TriTEST reagents was compared to either FACSCount, K933486 (for CD3/CD4/CD45 and CD3/CD8/CD45), or to the appropriate vial of IMK-Lymphocyte, K913192, plus ADCC. Substantial equivalence of the beads when used with other flow cytometry reagents is included in the respective reagent submission (e.g., TriTEST CD3/CD4/CD45, K965053). Data for these comparisons is not included in this submission.
Use of the TRUCOUNT Counting Control Beads does not change the safety and effectiveness of the above assays.
## Performance Data (21 CFR 807.92(b)(2))
Performance of the product was established by testing at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California.
- Stability of the Absolute Count Tube and Counting Control Bead was assessed and found to be acceptable to support a one year and six month shelf life, respectively.
- Accuracy and reproducibility of the Counting Control Beads was determined and found to be acceptable to support the ranges for acceptable performance cited in the Package Insert.
- Absolute Count Tubes were found to remain stable after the foil pouch is opened, as long as the pouch is stored closed at room temperature, and the desiccant remains blue.
- The Absolute Count Tubes and Counting Control Beads were seen to function as expected when used with the FACS Loader, as well as manually.
TRUCOUNT 510(k) Notification
Attachment D: Summary of Safety and Effectiveness
{2}
# Summary of Safety and Effectiveness
Performance Data—Conclusions (21 CFR 807.92(b)(3))
The results of the studies demonstrate that:
- Expiration dating claim for the TRUCOUNT Absolute Count Tube Product is one year.
- Expiration dating claim for the TRUCOUNT Counting Control Beads is six months.
- TRUCOUNT Absolute Count Tube and TRUCOUNT Counting Control Bead Product may be used with the FACS Loader.
The product does not change the substantial equivalence of the reagents it is used with, as demonstrated in the respective reagent submissions.
TRUCOUNT 510(k) Notification
Attachment D: Summary of Safety and Effectiveness
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Ms. Anna Longwell, Esq.
Director, Regulatory Affairs-Corporate
Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, California 95131-1807
AUG 26 1997
Re: K970836/S2
Trade Name: Becton Dickinson TruCount™ Absolute Count Tubes and TruCount™ Counting Control Beads
Regulatory Class: II
Product Code: GKZ
Dated: June 5, 1997
Received: June 10, 1997
Dear Ms. Longwell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{4}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number (if known): K970836
Device Name: Traquant Absolute Count Tubes
Traquant Counting Control Seeds
## Indications For Use:
- For use with FACS Loader
- For use with *in vitro* diagnostic immunophenotyping reagents
- For use in erythrocyte lysed whole blood
- For use to obtain absolute counts by flow cytometry
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.