FIRST MIDCATH CATHETER

K972254 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Sep 12, 1997 · General Hospital

Device Facts

Record IDK972254
Device NameFIRST MIDCATH CATHETER
ApplicantBecton Dickinson Infusion Therapy Systems, Inc.
Product CodeFOZ · General Hospital
Decision DateSep 12, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The intended use is to provide access to the vascular system to sample blood or administer drug solutions, blood products, or other fluids intravenously.

Device Story

FIRST MIDCATH™ is a short-term, single-use intravascular midline catheter designed for venous access (e.g., basilic or cephalic veins). It is inserted into the vascular system via a splittable introducer catheter. The device is used by clinicians to sample blood or administer fluids/medications. The subject device represents a modification to the stylet and labeling of the predicate FIRST MIDCATH™ catheter.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Intravascular midline catheter; various French sizes and lengths; includes stylet; single-use; sterile. No specific material standards or software/electronic components described.

Indications for Use

Indicated for intravenous blood sampling or administration of drug solutions, blood products, and other fluids in any patient population, considering patient size, solution appropriateness, and therapy duration.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 772254 ## SUMMARY OF SAFETY AND EFFICACY - A. The submitter's name, address, telephone number, contact person, and date of preparation. Submitted by Becton Dickinson Infusion Therapy Systems Inc., 9450 South State Street, Sandy, UT 84070. Contact: C. J. Welle. Telephone: 801-565-2535. Prepared: June 11, 1997. - B. The name of the device including trade or proprietary name if applicable, the common or usual name, and the classification name. Name: Intravascular Catheter Brand: FIRST MIDCATH™ Common Name: Midline Catheter Classification Name: Intravascular Catheter (80 FOZ) - C. An identification of the predicate or legally marketed device to which substantial equivalence is claimed: Predicate Device: The Becton Dickinson Infusion Therapy Systems Inc. FIRST MIDCATH™ brand catheters. - D. A description of the device that is the subject of the Premarket Notification submission. The subject catheter is a short term, single use, intravascular catheter of various French sizes and lengths which is designed to provide access to large veins such as the baslic or cephalic veins. It is inserted into the vascular system through a spittable introducer catheter. - E. Statement of intended use of the device. The intended use is to provide access to the vascular system to sample blood or administer drug solutions, blood products, or other fluids intravenously. - F. A statement of how the technological characteristics compare to those of the predicate or legally marketed device identified in section C above. The proposed catheter differs from the FIRST MIDCATH™ brand catheter in that the stylet and labeling have been modified. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, facing right. The eagle is composed of three curved lines that suggest the shape of a bird in flight. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP / 2 1997 Mr. Charles Welle Manager, Requlatory Affairs Becton Dickinson Infusion Therapy Systems 9450 South State Street Sandy, Utah 84070 K972254 Re: Trade Name: First MIDCATH™ Catheter Requlatory Class: II Product Code: FOZ Dated: June 11, 1997 Received: June 16, 1997 Dear Mr. Welle: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions However, you are responsible to determine that the of Act. medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510 (k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the {2}------------------------------------------------ Page 2 - Mr. Welle requlation (21 CFR Part 820) and that, through periodic (QS) inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, the Food and Drug Administration (FDA) action. may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. In addition, we have determined that your device kit contains Iodophor swabsticks, alcohol swabsticks, and skin protectant swabsticks which are subject to regulation as drugs. Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact: Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0063 This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket {3}------------------------------------------------ Page 3 - Mr. Welle notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the fear responsible of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Patricex Cucurite ffor Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ June 11, 1996 ## INDICATIONS FOR USE 510(k) Number: K 9 7 225 4 Device Name: _FIRST MIDCATH™ Catheter Indications for Use: To sample blood or administer drug solutions, blood products, and other fluids intravenously. These catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. Patricia Crescentz (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number 697225 Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: X (Per 21 CFR 801.109) --- OR Over-The -Counter Use: _______________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...