K972147 · Dynamic Systems, Inc. · ITI · Aug 25, 1997 · Physical Medicine
Device Facts
Record ID
K972147
Device Name
PEACHTREE PROPORTIONAL HEAD CONTROL UNIT (PHC-2)
Applicant
Dynamic Systems, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Aug 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The PHC-2 is intended to provide mobility control to persons restricted to a seated position while operating a power-drive wheelchair.
Device Story
PHC-2 is a non-contact, electronic, two-dimensional head position sensing device for power wheelchair control. It utilizes a low-level, high-frequency electrical field (capacitance sensing) to detect head movements without physical attachment. The device is integrated into a headrest pad. Users operate the system by tilting their head to generate proportional control commands or by tapping the headrest to switch between 'control' (stop, recline, ECU) and 'drive' (reverse, forward low/high, turn, spin) modes. The output interfaces directly with the wheelchair's joystick input port. By enabling proportional, non-contact control, the device allows users with limited mobility to navigate and operate wheelchair functions, potentially increasing independence.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Headrest-mounted electronic control unit; capacitance sensing principle; utilizes low-level, high-frequency electrical field; two-dimensional proportional sensing; interfaces with standard wheelchair joystick input ports; non-contact operation.
Indications for Use
Indicated for persons restricted to a seated position requiring mobility control for power-drive wheelchairs.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
### 510(k) SUMMARY
## Dynamic Systems, Inc.'s PHC-2
K972147
C
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Dynamic Systems, Inc. 5002 North Royal Atlanta Drive, Suite P Tucker, GA 30084-3050 Phone: (770) 939-1122 Facsimile: (770) 850-9875
David Court Contact Person:
June 3, 1997 Date Prepared:
#### Name of Device and Name/Address of Sponsor
Dynamic Systems, Inc. 5002 North Royal Atlanta Drive, Suite P Tucker, GA 30084-3050 Phone: (770) 939-1122 Facsimile: (770) 850-9875
### Common or Usual Name
Peachtree Proportional Head Control Unit (PHC-2)
### Classification Name
Power Wheelchair Control Unit
#### Predicate Devices
The PHC-2 ("PHC-2 ") is substantially equivalent to Adaptive Switch Laboratories. Inc.'s ASL Head Array, Invacare's Sip & Puff Head Array, Invacare's Remote Joystick, and Invacares R.I.M. controles L
{1}------------------------------------------------
### Intended Use
The PHC-2 is intended to provide mobility control to persons restricted to a seated position while operating a power-drive wheelchair.
#### Technological Characteristics and Substantial Equivalence
#### A. Device Description
The PHC-2 consists of a headrest pad containing a control unit which provides invisible contact between the user and the wheelchair. The control unit is a remote, electronic, two dimensional head position sensing device that utilizes a low level, high frequency electrical field to provide contact between the user and the wheelchair.
The PHC-2 control unit does not require physical attachments to the user. The user simply tilts his or her head to provide control commands to the wheelchair. The PHC-2 has two operational modes: control and drive, the user enters the various modes and functions by gently tapping the head rest a designated number of times. On power up, the unit is always in control mode which includes the following functions: stop, recline, ECU, and drive. The stop function results in no activity, recline allows the user to recline the back of the wheelchair with head motions, and the ECU function controls an environmental control unit, if available. The second level is drive mode which includes: stop, reverse, forward low, and forward high. Within the drive modes the user can accelerate, brake, spin, and turn by utilizing non-contact head movements.
#### B. Substantial Equivalence
The PHC-2 has the same intended use, similar principles of operation, and similar technological characteristics as the previously cleared predicate devices. Although there are minor differences in the characteristics of the PHC-2 and its predicate devices, those differences do not raise new questions of safety or efficacy.
Although there are some differences between the PHC-2 and its predicates, these differences are minor and raise no new questions of safety and effectiveness. The only significant differences between the named predicate devices and the PHC-2 are in the method employed to acquire user control input. The PHC-2 utilizes a remote, electronic, fully proportional, two dimensional head position sensing device to generate wheelchair control commands whereas the predicate devices either use
{2}------------------------------------------------
a joystick requiring physical contact or a nonproportional switch control mechanism to generate the commands.
{3}------------------------------------------------
# Substantial Equivalence Table Comparing Dynamic Systems, Inc.'s PHC-2 with Predicate Devices
| | PHC-2 | ASL Head<br>Array | SIP &<br>PUFF | FLIGHT<br>LINK<br>JOYSTICK | R.I.M. |
|----------------------------|-------------------------------------------------------------------------------------------------------|-----------------------|-------------------------------------|----------------------------|---------------------------|
| Intended Use | Movement<br>Commande<br>d Driving<br>Control for<br>Operating a<br>Variety of<br>Power<br>Wheelchairs | Same | Same | Same | Same |
| Operating<br>Principle | Capacitance<br>Sensing | Fiberoptic<br>Sensing | Air Pressure<br>Sensing | Mechanical | Mechanical |
| Non-contact<br>control | Yes | Yes | No | No | No |
| Device<br>interface | Wheelchair<br>Joystick Input<br>Port | Same | Same | Same | Same |
| Proportional<br>control | Yes | No | No | Yes | Yes |
| Movement<br>control | Head | Head | Head | Hand | Head |
| Functions | Stop -<br>Forward-<br>Reverse -<br>Turn -<br>Spin - | Same | Same | Same | Same |
| Physical<br>Characteristic | Headrest | Headrest | Headrest<br>with Sip &<br>Puff Pipe | Joystick | Headrest<br>with Joystick |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that appears to be three faces in profile, one behind the other, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David R. Court ·Dynamic Systems, Inc. 5002 North Royal Atlanta Drive Suite P Tucker, Georgia 30084
AUG 2 5 1997
K972147 Re: Peachtree Proportional Head Control Unit (PHC-2) Regulatory Class: II Product Code: ITI Dated: June 3, 1997 Received: June 9, 1997
Dear Mr. Court:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it' may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಡ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2 - Mr. David R. Court
This letter will allow you to begin marketing your device as described in your 510 k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(K) Number: K972147
Device Name: Peachtree Proportional Head Control
(UNIT PHC-2)
Indications for Use: The Peachtree Proportional Head Control Unit is non-contact, fully proportional, head movement commanded driving control intended to provide mobility control to persons restricted to a seated position while operating a variety of available power-drive wheelchairs.
# PLEASE DO NOT WRITE BELOW THIS LINE (CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
| OR | Over-the-Counter Use |
|----|--------------------------|
| | (Optional Format 1-2-96) |
| | (Division Sign-Off) |
Division of General Restorative Devices 510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.