The DIASCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip and is intended for use as an in vitro diagnostic aid using urine specimens. The strip contains solid phase reagent areas affixed to a plastic strip and is provided in a dry reagent format. The strip provides qualitative and semi-quantitative tests for specific gravity, ketones, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen by visual comparison with a color chart for each concentration range. No additional reagents and laboratory equipment are required. The reagent strips are packaged in a plastic vial containing a desiccant. The test strips must be kept tightly capped in the vial to assure reagent reactivity. The directions must be followed exactly and it is necessary to use fresh, well-mixed, uncentrifuged urine for optimal results.
Device Story
DIASCREEN® Reagent Strips are plastic strips with affixed dry reagent test areas for urinalysis. Used by healthcare professionals in clinical settings; no additional laboratory equipment required. Input: fresh, well-mixed, uncentrifuged patient urine. Operation: dip-and-read method; reagent areas react with urine analytes; color change occurs. Output: visual comparison of strip color against a provided color chart. Healthcare providers interpret results to assess patient urine levels of specific gravity, ketones, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen. Results aid in clinical decision-making for patient health status. Benefit: provides rapid, semi-quantitative diagnostic information for urinalysis.
Clinical Evidence
Clinical laboratory evaluation conducted at Wichita State University compared two lots of DIASCREEN® strips against Bayer Multistix 10 SG using fresh urine samples. Results showed high correlation: 66-79% of results were in the same color block and 100% were within one color block for analytes including specific gravity, leukocyte, pH, blood, and nitrite. Bench testing confirmed performance consistency between lots.
Technological Characteristics
Plastic strips with dry reagent pads. Reagents include bromothymol blue, methyl red, sodium nitroprusside, glucose oxidase, peroxidase, potassium iodide, tetrabromphenol blue, 2,5-dimethylhexane-2,5-dihydroperoxide, 3,3',5,5'-tetramethylbenzidine, p-arsanilic acid, and diazonium salts. Visual read format; no energy source; no software. Storage: 15-30°C, protected from moisture/light.
Indications for Use
Indicated for health care professionals to perform semi-quantitative urinalysis on patient urine samples to detect specific gravity, ketones, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A leukocyte peroxidase test is a device used to distinguish certain myeloid cells derived from the bone marrow, i.e., neutrophils, eosinophils, and monocytes, from lymphoid cells of the lymphatic system and erythroid cells of the red blood cell series on the basis of their peroxidase activity as evidenced by staining. The results of this test are used in the differential diagnosis of the leukemias.
{0}------------------------------------------------
K971976
# JUN 17 1997
## Summary of Safety and Effectiveness
### May 27, 1997
The following information is provided as a summary of safety and effectiveness information for the DIASCREEN® Reagent Strip System.
- Common Name: Visual Reagent Test Strip for Urinalysis (a]
Trade/Proprietary Name: DIASCREEN® Reagent Strips, in many different configurations with as many different product codes. is the registered trademark of Dia-Screen DIASCREEN® Corporation.
- Establishment Registration Number: 9010673 [b]
Chung Do Pharmaceutical Co., Ltd. 312 Gaju-Dong Jin Hae-City Kyung Sang Nam-Do, 645-500 Korea.
- Vicki Frawley Contact Person: [c]
- [d] Intended Use
The DIASCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip for semi-quantitative urinalysis. DIASCREEN® Reagent Strips have been developed to allow health care professionals to test semi-quantitatively for patient urine levels of specific gravity. ketone, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen by visual comparison with a color chart.
- Product Description [e]
DIASCREEN® Reagent Strips are plastic strips to which are affixed reagent test areas. Several product configurations are included with this submission, but all configurations are limited to one, some or all of the following tests: specific gravity, ketone, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen.
The reagent test areas on the DIASCREEN® Reagent Strips are ready to use upon removal from the bottle and the entire reagent
{1}------------------------------------------------
strip is disposable after use. The strips are to be read visually, requiring no additional laboratory equipment for testing.
The directions must be followed exactly. Accurate timing is essential to provide optimal results. The reagent strips must be kept in the original bottle containing the desiccant with the cap tightly closed to maintain reagent activity. To obtain optimal results it is necessary to use fresh, well-mixed, uncentrifuged urine.
The DIASCREEN® Reagent Strips will be manufactured in Korea by Chunq Do Pharmaceutical CO., Ltd. of Seoul. Korea. They will be marketed in the United States by Dia-Screen Corporation. Dia-Screen is the exclusive United States distributor for Chung Do.
Dia-Screen Corporation has previously submitted 510k's on behalf of Chung Do Pharmaceutical Co., Ltd and Bioscan Instruments, Inc.(now defunct) under Dia-Screen Corporation's former name, Genesis Labs Inc. These are K940043 and K943008. In addition, Dia-Screen Corporation, which is a FDA registered facility, has received two 510k's for its own manufactured urine reagent strips. They are K961375 and K952971.
The intended use of the strips that are the subject of this new premarket notification is exactly the same as the Bioscan (K940043, K943008) and DIASCREEN® (K952971, K961375) strips which were reviewed previously.
#### Substantial Equivalence િય
The intended use of the strips which are the subject of this new pre-market notification is exactly the same as the strips which were reviewed previously (K940043, K943008, K952971, K961375),
The design and function of the new DIASCREEN® Reagent Strips are the same as the design and function of both the Bioscan and DIASCREEN® Reagent Strips which have been previously reviewed under pre-market notifications K940043, K943008. K952971, and K961375.
{2}------------------------------------------------
The DIASCREEN® Reagent Strips for Urinalysis are substantially equivalent to Bayer Multistix 10 SG, Behring Rapignost, Bioscan, and Boehringer Mannheim Chemstrip urine test strip systems.
Characteristics of the Bayer Multistix 10 SG and the Bioscan systems are compared with the DIA SCREEN® system in the following table:
{3}------------------------------------------------
| | This 510 (k) | Prevoius 510 (k) | Bayer Multistix 10 SG |
|---------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Strip Name | DIASCREEN® | Bioscan | Bayer |
| Distributor | Dia-Screen | Dia-Screen | Bayer |
| Reagents for: | | | |
| Specific | Bromothymol blue | Bromthymol blue | Bromthymol blue |
| Gravity | Methyl red | Methyl vinyl ether/ | Poly (methyl vinyl-ether<br>maleic anhydride) |
| | Polyvinyl phosphate | maleic acid copolymer | |
| Leukocyte | 2-Phenylthiazole amino<br>acid ester<br>Diazonium salt | Not offered | Derivatized pyrrole amino<br>acid ester<br>Diazonium salt<br>Buffer |
| Ketones | Sodium nitroprusside | Sodium nitroprusside | Sodium nitroprusside<br>Buffer |
| Glucose | Glucose oxidase<br>Peroxidase<br>Potassium iodide | Glucose oxidase<br>Peroxidase<br>Potassium iodide | Glucose oxidase<br>Peroxidase<br>Potassium iodide<br>Buffer |
| Protein | Tetrabromphenol blue<br>Citric acid<br>Trisodium citrate | Tetrabromphenol blue<br>Citric acid<br>Trisodium citrate | Tetrabromphenol blue<br>Buffer |
| Blood | 2,5 Dimethylhexane-<br>2,5-dihydroperoxide<br>3,3',5,5'-<br>Tetramethylbenzidine | 2,5 Dimethylhexane-<br>2,5-dihydroperoxide<br>3,3',5,5'-<br>Tetramethylbenzidine | Diisopropylbenzene-<br>dihydroperoxide<br>3,3',5,5'-<br>Tetramethylbenzidine |
-
-
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
| Strip Name<br>Distributor<br>Reagents for: | This 510 (k)<br>DIASCREEN®<br>Dia-Screen | Prevoius 510 (k)<br>Bioscan<br>Dia-Screen | Bayer Multistix 10 SG<br>Bayer |
|--------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------------------------|
| Nitrite | p-Arsanilic acid<br>N-(1-naphthyl)ethylene<br>diamine | p-Arsanilic acid<br>N-(1-naphthyl)ethylene<br>diamine | p-Arsanilic acid<br>1,2,3,4-Tetrahydro-<br>benzo(h)quinoline-3-ol<br>Buffer |
| pH | Bromthymol blue<br>Methyl red | Bromthymol blue<br>Methyl red | Bromthymol blue<br>Methyl red |
| Bilirubin | 2,4 Dichlorobenzene-<br>diazonium salt<br>Oxalic acid | 2,4 Dichlorobenzene-<br>diazonium salt<br>Oxalic acid | 2,4-Dichloroaniline-<br>diazonium salt<br>Buffer |
| Urobilinogen | 4-methoxybenzene-<br>diazonium salt<br>Citric acid | 4-methoxybenzene-<br>diazonium salt<br>Citric acid | Diethylaminobenzaldehyde |
1
ાં
ﺇ
-
{5}------------------------------------------------
| Strip Name<br>Distributor | This 510 (k)<br>DIA SCREEN® | Previous 510 (k)<br>Bioscan | Multistix 10 SG |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Distributor | Dia-Screen | Dia-Screen | Bayer |
| Packaged with<br>Desiccant | Yes | Yes | Yes |
| Control Available | Yes | Yes | Yes |
| Time required to<br>read strips | 10 to 120 seconds | 60 seconds | 30 to 120 seconds |
| Storage | Between 15 - 30 °C (59°-<br>86° F). Do Not Store in<br>refrigerator or freezer. Do<br>not expose to moisture,<br>heat or light. | Between 15 - 30 °C (59°-<br>86° F). Do Not Store in<br>refrigerator or freezer. Do<br>not expose to moisture,<br>heat or light | Between 15°- 30° C (59°-<br>86° F) Do Not Store in<br>Direct Sunlight |
t
{6}------------------------------------------------
| | This 510 (k) | Previous 510 (k) | Multistix 10 SG |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Strip Name | DIA SCREEN® | Bioscan | |
| Distributor | Dia-Screen | Dia-Screen | Bayer |
| Expected Values: | | | |
| Specific Gravity | Random urines vary from<br>1.001-1.035. Twenty-four<br>hour urines from normal<br>adults with normal diets<br>and fluid intake will have<br>a specific gravity of<br>1.016-1.022 | Random urines vary from<br>1.001-1.035. Twenty-four<br>hour urines from normal<br>adults with normal diets<br>and fluid intake will have<br>a specific gravity of<br>1.016-1.022 | Random urines vary from<br>1.001-1.035. Twenty-four<br>hour urines from normal<br>adults with normal diets<br>and fluid intake will have<br>a specific gravity of<br>1.016-1.022 |
| Ketones | Should not be detected in<br>normal urine. | Should not be detected in<br>normal urine. | Normal specimens yield<br>negative results. |
| Glucose | None found in normal<br>urine. Concentrations of<br>100 mg/dL may be<br>considered abnormal. | None found in normal<br>urine. Concentrations of<br>100 mg/dL may be<br>considered abnormal. | None found in normal<br>urine. Concentration of<br>100 mg/dL may be<br>considered abnormal. |
| Protein | Normal specimens<br>ordinarily contain some<br>protein (0-4 mg/dL).<br>Persistent results of trace<br>or higher indicate<br>significant proteinuria. | Normal specimens<br>ordinarily contain some<br>protein (0-4 mg/dL).<br>Persistent results of trace<br>or higher indicate<br>significant proteinuria. | Normal specimens<br>ordinarily contain some<br>protein (0-4 mg/dL). A<br>color matching any "+"<br>block indicates significant<br>proteinuria |
l
(
SS and E Page 7
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{7}------------------------------------------------
| Strip Name<br>Distributor | This 510 (k)<br>DIA SCREEN®<br>Dia-Screen | Previous 510 (k)<br>Bioscan<br>Dia-Screen | Multistix 10 SG<br>Bayer |
|---------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Expected Values: | | | |
| Blood | The practical detection<br>limit of this test is 5 to 10<br>erythrocytes per microliter<br>of urine | The practical detection<br>limit of this test is 5 to 10<br>erythrocytes per microliter<br>of urine | The significance of the<br>trace reaction may vary<br>among patients. Clinical<br>judgement is required |
| Nitrite | Any degree of pink color<br>after 30 seconds<br>indicates clinically<br>significant bacteria | Any degree of pink color<br>after 30 seconds indicates<br>clinically significant<br>bacteria NA | Normally, no nitrite is<br>detectable in urine |
| pH | Normal urine has a pH of<br>6 and urine pH values<br>generally range from 5 to<br>8 | Normal urine has a pH of<br>6 and urine pH values<br>generally range from 5 to<br>8NA | Both the normal and<br>abnormal pH range is<br>from 5 to 9 pH units |
| Bilirubin | No bilirubin is detectable<br>in urine of healthy<br>persons | No bilirubin is detectable<br>in urine of healthy<br>persons | Normally, no bilirubin is<br>detectable in healthy<br>persons |
| Urobilinogen | Normal range is 0.1 to 1.0<br>mg/dL | N Normal range is 0.1 to<br>1.0 mg/dL A | Normal range is 0.1 to<br>1.0 Ehrlich units per 100<br>ml |
-
{8}------------------------------------------------
## Clinical Laboratory Evaluation of DIASCREEN® Strips
The DIASCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip intended for use as an in-vitro diagnostic aid using urine specimens. DIASCREEN® Reagent Strips have been developed to allow health care professionals to semi-quantitatively determine levels of specific gravity, ketone, alucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and urobilinogen in patient urine by visual comparison with a color chart of each concentration range. No additional reagents or laboratory equipment are required. These reagent strips are packaged in a plastic vial containing a desiccant. The test strips must be maintained tightly capped in the plastic vial to assure reagent activity. The directions-for-use must be followed exactly.
An independent laboratory evaluation of the new DIASCREEN® Reagent Strips for Urinalysis was conducted under the direction of James Jackson MT(ASCP) CLS, Ph.D. at Wichita State University Department of Medical Technology. The purpose of the laboratory evaluation was to establish the performance of the leukocyte and specific gravity tests and to evaluate the pH, nitrite and blood for placement adjacent to these tests on the DIASCREEN® Reagent Strip for Urinalysis when compared to a "510 (k) approved" marketed urinalysis strip svstem.
The study consisted of comparing the performance of two (2) manufactured lots of DIASCREEN® Reagent Strips with the marketed product (Bayer Multistix 10 SG).
Fresh urine samples were obtained at the medical facility. The comparison strips. Baver Multistix 10 SG Reagent Strips, Schein Urispec Reagent Strips(for ascorbic acid detection), Boehringer Manneheim Corporation CHEMSTRIP Reagent Strips(for confirmatory testing), normal and abnormal urine controls were furnished by Dia-Screen. When data between the two products did not agree within one color block, an alternate procedure (see table below) was used to test the sample.
| Analyte | Confirmation Test | Manufacturer |
|------------------|-------------------------------------------------------------------|-------------------------------------------|
| Specific Gravity | Refractometer | NA |
| Leukocyte | Microscopic Evaluation<br>CHEMSTRIP | NA<br>Boehringer Mannheim |
| pH | pH Meter | NA |
| Blood | Microscopic Evaluation<br>CHEMSTRIP<br>Ascorbic Acid Test/Urispec | NA<br>Boehringer Mannheim<br>Henry Schein |
| Nitrite | Microscopic Evaluation<br>CHEMSTRIP | NA<br>Boehringer Mannheim |
{9}------------------------------------------------
Fresh, well-mixed, and uncentrifuged urine from a mixed patient population was reacted with the DIASCREEN® Reagent Strips and the results were read by visual comparison with the DIASCREEN® color chart. Results of these readings were compared to results measured by Multistix 10 SG Reagent Strips.
A summary of the data obtained from the clinical evaluation described above is provided in the following tables.
{10}------------------------------------------------
## DIASCREEN® versus Bayer Clinical Correlation
Tables I and II present the percentage of responses for the indicated lots of the DIASCREEN® and Bayer Reagent Strips that are within the same color block and that are within plus or minus one color block. Table III presents a similar analysis for the two lots of the DIASCREEN® Reagent Strips.
DIASCREEN® Lot 1 versus Bayer Lot D639066 or 6K30C Table I.
| | Same Color Block | |
|------------------|------------------|------|
| Specific Gravity | 39/53 | 74% |
| Leukocyte | 42/53 | 79% |
| pH | 37/53 | 70% |
| Blood | 49/53 | 92% |
| Nitrite | 53/53 | 100% |
| Within 1 Color Block | |
|----------------------|------|
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
DIASCREEN® Lot 2 versus Bayer Lot D639066 or 6K30C Table II.
| | Same Color Block | |
|------------------|------------------|------|
| Specific Gravity | 35/53 | 66% |
| Leukocyte | 43/53 | 81% |
| pH | 35/53 | 66% |
| Blood | 44/53 | 83% |
| Nitrite | 53/53 | 100% |
| Within 1 Color Block | |
|----------------------|------|
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
### DIASCREEN® Lot 1 versus DIASCREEN® Lot 2 Table III.
| | Same Color Block | |
|------------------|------------------|------|
| Specific Gravity | 48/53 | 91% |
| Leukocyte | 47/53 | 89% |
| pH | 49/53 | 92% |
| Blood | 52/53 | 98% |
| Nitrite | 53/53 | 100% |
| Within 1 Color Block | |
|----------------------|------|
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
| 53/53 | 100% |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black against a white background. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 17 1997
John Murray President and Chief Executive Officer Dia-Screen Corporation -----5182 West 76th Street Minneapolis, Minnesota ਦ ਦੇ ਕੇ ਤੇ ਰ
K971976 Re : DIASCREEN® 10 Way Reagent Strips for Urinalysis (In Duplicate) Regulatory Class: I & II Product Code: LJX, JNA, KQO May 27, 1997 Dated: Received: May 29, 1997
Dear Mr. Murray:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{12}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
| 0(k) Number (if known): | K971976 |
|-------------------------|--------------------------|
| Device Name: | DIASCREEN Reagent Strips |
| Indications For Use: | |
The DIASCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip
and is intended for use as an *in vitro* diagnostic aid using urine specimens.
The strip contains solid phase reagent areas affixed to a plastic strip and is
provided in a dry reagent format.
The strip provides qualitative and semi-quantitative tests for specific gravity,
ketones, glucose, protein, blood, leukocytes, nitrite, pH, bilirubin, and
urobilinogen by visual comparison with a color chart for each concentration
range. No additional reagents and laboratory equipment are required. The
reagent strips are packaged in a plastic vial containing a desiccant. The
test strips must be kept tightly capped in the vial to assure reagent
reactivity. The directions must be followed exactly and it is necessary to use
fresh, well-mixed, uncentrifuged urine for optimal results.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Page 1 of 1
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <i>Veronica Calvin for Di Montgomery</i> |
|---------------|------------------------------------------|
| | (Division Sign-Off) |
| | Division of Clinical Laboratory Devices |
| 510(k) Number | K971976 |
| Prescription Use<br>(21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|--------------------------------------|----|--------------------------------------------------|
|--------------------------------------|----|--------------------------------------------------|
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
:
・・
...
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.