K961375 · Genesis Labs, Inc. · JIL · Jun 18, 1996 · Clinical Chemistry
Device Facts
Record ID
K961375
Device Name
DIA SCREEN 10 WAY REAGENT STRIPS
Applicant
Genesis Labs, Inc.
Product Code
JIL · Clinical Chemistry
Decision Date
Jun 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1340
Device Class
Class 2
Indications for Use
The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip for semiquantitative urinalysis. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to test semiquantitatively for patient urine levels of specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen by visual comparison with a color chart.
Device Story
Disposable plastic reagent strips for semiquantitative urinalysis; used by healthcare professionals in clinical settings. Input: fresh, well-mixed, uncentrifuged urine. Operation: dip strip into urine; visual comparison of reagent color changes against provided color chart. Output: semiquantitative levels of specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen. No laboratory equipment required. Results assist clinicians in patient assessment and diagnostic decision-making.
Clinical Evidence
Independent laboratory evaluation conducted at University of Minnesota. Study compared two lots of DIA SCREEN strips against Ames Multistix 10 SG and Behring Rapignost using fresh, uncentrifuged urine samples (n=58). Discrepant results were resolved using standard confirmatory methods (e.g., refractometer, microscopic evaluation, pH meter). Results showed 96.5% agreement within one color block for all analytes across both lots.
Technological Characteristics
Plastic strips with affixed chemical reagent pads. Reagents include sodium nitroprusside, glucose oxidase, peroxidase, tetrabromphenol blue, 3,3',5,5'-tetramethylbenzidine, and various diazonium salts. Visual read-only; no electronic components. Storage: 15-30°C, protected from moisture/light. Disposable.
Indications for Use
Indicated for health care professionals to perform semiquantitative urinalysis on patient urine samples to detect specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen.
Regulatory Classification
Identification
A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.
{0}
JUN 18 1996
K.961375
# Summary of Safety and Effectiveness
April 4, 1996
The following information is provided as a summary of safety and effectiveness information for the DIA SCREEN® Reagent Strip System.
[a] Common Name: Visual Reagent Test Strip for Urinalysis
Trade/Proprietary Name: DIA SCREEN® Reagent Strips, in many different configurations with as many different product codes. DIA SCREEN® is the registered trademark of Dia-Screen Incorporated, a wholly owned subsidiary of Genesis Labs, Inc.
[b] Establishment Registration Number 2183670
Genesis Labs, Inc.
5182 West 76th Street
Minneapolis, MN 55439-2904
[c] Contact Person: Bert Walter
[d] Intended Use
The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip for semiquantitative urinalysis. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to test semiquantitatively for patient urine levels of specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen by visual comparison with a color chart.
[e] Product Description
DIA SCREEN® Reagent Strips are plastic strips to which are affixed reagent areas. Several product configurations are included with this submission, but all configurations are limited to one, some or all of the following tests: specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen.
The reagent test areas on the DIA SCREEN® Reagent Strips are ready to use upon removal from the bottle and the entire reagent strip is disposable after use. The strips are to be read visually, requiring no additional laboratory equipment for testing.
SS and E Page 1
{1}
The directions must be followed exactly. Accurate timing is essential to provide optimal results. The reagent strips must be kept in the original bottle containing the desiccant with the cap tightly closed to maintain reagent activity. To obtain optimal results, it is necessary to use fresh, well-mixed, uncentrifuged urine.
Dia-Screen Corporation, a subsidiary of Genesis Labs, Inc., has previously received determinations of substantial equivalence for complete lines of urine reagent strips which have been marketed under the Bioscan (K940043) and DIA SCREEN (K952971) label.
The DIA SCREEN® Reagent Strips will be manufactured by Genesis Labs for distribution by Dia-Screen. The intended use of the strips which are the subject of this new premarket notification is exactly the same as the strips which were reviewed previously (K940043 and K952971).
[f] Substantial Equivalence
The intended use of the strips which are the subject of this new premarket notification is exactly the same as the strips which were reviewed previously (K940043 and K952971).
The design and function of the new DIA SCREEN® Reagent Strips are the same as the design and function of both the Bioscan and DIA SCREEN Reagent Strips which have been previously reviewed under premarket notifications K940043 and K952971.
The DIA SCREEN® Reagent Strips for Urinalysis are substantially equivalent to Ames Multistix 10 SG, Behring Rapignost, Bioscan and Boehringer Mannheim Chemstrip urine test strip systems (90% of all observations are within one color block).
Characteristics of the Ames Multistix 10 SG and Behring Rapignost systems are compared with the DIA SCREEN® system in the following table:
SS and E Page 2
{2}
| Strip Name | This 510 (k)
DIA SCREEN®
DIA SCREEN® | Previous 510 (k)
DIA SCREEN®
DIA SCREEN® | Ames Multistix 10 SG
Ames | Rapignost
Behring |
| --- | --- | --- | --- | --- |
| Distributor
Reagents for: | | | | |
| Specific
Gravity | Bromthymol blue
Detergent | NA | Bromthymol blue
Poly (methyl vinyl-ether
maleic anhydride | NA |
| Ketone | Sodium nitroprusside | Sodium nitroprusside | Sodium nitroprusside
Buffer | Sodium nitroprusside
Glycine
Buffer |
| Glucose | Glucose oxidase
Peroxidase
Potassium iodide | Glucose oxidase
Peroxidase
Potassium iodide | Glucose oxidase
Peroxidase
Potassium iodide
Buffer | Glucose oxidase
Peroxidase
Tolidine hydrochloride
Buffer |
| Protein | Tetrabromphenol blue
Citric acid | Tetrabromphenol blue
Citric acid | Tetrabromphenol blue
Buffer | Tetrabromphenol blue
Buffer |
| Blood | 2,5 Dimethylhexane-
2,5-dihydroperoxide
3,3',5,5'-
Tetramethylbenzidine | NA | Diisopropylbenzene-
dihydroperoxide
3,3',5,5'-
Tetramethylbenzidine | Cumene hydroperoxide
Tetramethylbenzidine
dihydrochloride
Buffer |
| Nitrite | Hydroxy(3)-1,2,3-
tetrahydrobenzo-(h)quinoline
p-Arsanilic acid | NA | p-Arsanilic acid
1,2,3,4-Tetrahydro-
benzo(h)quinoline-3-ol
Buffer | 4-Arsanilic acid
N-(Naphthyl)-
ethylenediammonium
Dihydrochloride
Buffer |
| pH | Bromthymol blue
Methyl red | NA | Bromthymol blue
Methyl red | Bromthymol blue
Cresol red
Methyl red |
| Ascorbic Acid | Phosphomolybdic acid | | | 2,6-Dichloro-phenol-
indophenol
Buffer salts |
SS and E Page 3
{3}
| Strip Name | This 510 (k)
DIA SCREEN®
DIA SCREEN® | Previous 510 (k)
DIA SCREEN®
DIA SCREEN® | Ames Multistix 10 SG
Ames | Rapignost
Behring |
| --- | --- | --- | --- | --- |
| Distributor
Reagents for: | | | | |
| Bilirubin | 2,4 Dichlorobenzene-diazonium salt | NA | 2,4-Dichloroaniline-diazonium salt
Buffer | 2,4-Dichlorobenzene-diazonium salt
Buffer |
| Urobilinogen | Dimethylaminobenzaldehyde | NA | Diethylaminobenzaldehyde | Fluorodiazoniumtetrafluoroborate
Buffer |
SS and E Page 4
{4}
| Strip Name
Distributor | This 510 (k)
DIA SCREEN®
DIA SCREEN® | Previous 510 (k)
DIA SCREEN
DIA SCREEN | N - Multistix
Ames | Rapignost
Behring |
| --- | --- | --- | --- | --- |
| Packaged with
Desiccant | Yes | Yes | Yes | Yes |
| Control Available | Yes | Yes | Yes | Yes |
| Time required to
read strips | 30 to 60 seconds | 60 seconds | 30 to 60 seconds | 60 seconds |
| Storage | Between 15 - 30 °C (59°-86° F). Do Not Store in refrigerator or freezer. Do not expose to moisture, heat or light. | Between 15 - 30 °C (59°-86° F). Do Not Store in refrigerator or freezer. Do not expose to moisture, heat or light | Between 15°- 30° C (59°-86° F) Do Not Store in Direct Sunlight | Between 15 - 30° C (59°- 86° F). Do Not Store in refrigerator or freezer. Do not expose to moisture, heat or light. |
| 510 (k)
Expected Values: | This submission | K952971 | --- | --- |
| Specific Gravity | Random urines vary from 1.001-1.035. Twenty-four hour urines from normal adults with normal diets and fluid intake will have a specific gravity of 1.016-1.022 | NA | Random urines vary from 1.001-1.035. Twenty-four hour urines from normal adults with normal diets and fluid intake will have a specific gravity of 1.016-1.022 | NA |
| Ketone | Should not be detected in normal urine. | Should not be detected in normal urine. | Normal specimens yield negative results. | Should not be detected in normal urine |
| Glucose | None found in normal urine. Concentrations of 100 mg/dL may be considered abnormal. | None found in normal urine. Concentrations of 100 mg/dL may be considered abnormal. | None found in normal urine. Concentration of 100 mg/dL may be considered abnormal. | None found in normal urine. Concentrations of 100 mg/dL may be considered abnormal. |
SS and E Page 5
{5}
| Strip Name
Distributor
Expected values: | This 510 (k)
DIA SCREEN®
DIA SCREEN® | Previous 510 (k)
DIA SCREEN®
DIA SCREEN® | Ames Multistix 10 SG
Ames | Rapignost
Behring |
| --- | --- | --- | --- | --- |
| Protein | Normal specimens ordinarily contain some protein (0-4 mg/dL). Persistent results of trace or higher indicate significant proteinuria. | Normal specimens ordinarily contain some protein (0-4 mg/dL). Persistent results of trace or higher indicate significant proteinuria. | Normal specimens ordinarily contain some protein (0-4 mg/dL). A color matching any “+” block indicates significant proteinuria | Normal specimens ordinarily contain some protein (0-4 mg/dL). Persistent results of 30 mg/dL or higher indicate significant proteinuria |
| Blood | The practical detection limit of this test is 5 to 10 erythrocytes per microliter of urine | NA | The significance of the trace reaction may vary among patients. Clinical judgment is required | The practical detection limit of this test is 5 to 10 erythrocytes per microliter of urine |
| Nitrite | Any degree of pink color after 30 seconds indicates clinically significant bacteria | NA | Normally, no nitrite is detectable in urine | Concentrations between .05 and 0.1 mg/dL produce a pink color, indicating significant bacteria |
| pH | Normal urine has a pH of 6 and urine pH values generally range from 5 to 8 | NA | Both the normal and abnormal pH range is from 5 to 9 pH units | Normal urine has a pH of 6 and urine pH values generally range from 5 to 9 |
| Ascorbic Acid | Urinary excretion of ascorbic acid can result from the intake of vitamin C | NA | NA | Any color change after 60 seconds indicates a positive result |
| Bilirubin | No bilirubin is detectable in urine of healthy persons | NA | Normally, no bilirubin is detectable in healthy persons | No bilirubin is detectable in urine of healthy persons |
| Urobilinogen | Normal range is 0.1 to 1.0 mg/dL | NA | Normal range is 0.1 to 1.0 Ehrlich units per 100 mL | Normal range is 0.5 to 1.0 mg/dL |
SS and E Page 6
{6}
Clinical Laboratory Evaluation of DIA SCREEN® Strips
The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip intended for use as an in-vitro diagnostic aid using urine specimens. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to semiquantitatively determine levels of specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen in patient urine by visual comparison with a color chart of each concentration range. No additional reagents or laboratory equipment are required. These reagent strips are packaged in a plastic vial containing a desiccant. The test strips must be maintained tightly capped in the plastic vial to assure reagent activity. The directions-for-use must be followed exactly.
An independent laboratory evaluation of the new DIA SCREEN® Reagent Strips for Urinalysis was conducted under the direction of Nancy Brunzel, C.L.S., M.T., at the University of Minnesota Department of Medical Technology. The purpose of the laboratory evaluation was to establish the performance of the DIA SCREEN® Reagent Strip for Urinalysis when compared to "510 (k) approved" marketed urinalysis strip systems.
The study consisted of comparing the performance of two (2) manufactured lots of DIA SCREEN® Reagent Strips with one (1) lot of a marketed product (Ames Multistix 10 SG and Behring Rapignost).
Fresh urine samples were obtained at the medical facility. The comparison strips, Ames Multistix 10 SG and Behring Rapignost strips, and abnormal urine controls were furnished by Dia-Screen. When data between the two products did not agree within one color block, an alternate procedure (see table below) was used to test the sample.
| Analyte | Confirmation Test | Manufacturer |
| --- | --- | --- |
| Specific Gravity | Refractometer | NA |
| Ketone | Acetest | Ames |
| Glucose | Chemstrip Reagent Strip | Boehringer Mannheim |
| Protein | SSA | NA |
| Blood | Microscopic Evaluation | NA |
| Nitrite | Microscopic Evaluation | NA |
| pH | pH Meter | NA |
| Ascorbic Acid | Urispec Reagent Strip | Henry Schein |
| Bilirubin | Ictotest | Ames |
| Urobilinogen | Chemstrip Reagent Strip | Boehringer Mannheim |
SS and E Page 7
{7}
Fresh, well-mixed, and uncentrifuged urine from a mixed patient population was reacted with the DIA SCREEN® Reagent Strips and the results were read by visual comparison with the DIA SCREEN® color chart. Results of these readings were compared to results measured by Multistix 10 SG and Rapignost Strips.
A summary of the data obtained from the clinical evaluation described above is provided in the following tables (pages 9 and 10):
SS and E Page 8
{8}
# DIA SCREEN versus Ames and Rapignost Clinical Correlation
Tables I and II present the percentage of responses for the indicated lots of the DIA SCREEN and Ames and Rapignost Reagent Strips that are within the same color block and that are within plus or minus one color block. Table III presents a similar analysis for the two lots of the DIA SCREEN Reagent Strips.
Table I. DIA SCREEN Lot 00226A versus Ames Lot D397095 and Rapignost Lot 633962A
| Specific Gravity | Same Color Block | | Within 1 Color Block | |
| --- | --- | --- | --- | --- |
| | 24/58 | 41.4% | 53/58 | 91.4% |
| Ketone | 52/58 | 89.6% | 58/58 | 100% |
| Glucose | 46/58 | 79.3% | 58/58 | 100% |
| Protein | 41/57 | 71.9% | 57/57 | 100% |
| Blood | 45/58 | 77.6% | 53/58 | 91.4% |
| Nitrite | 56/57 | 98.2% | 57/57 | 100% |
| pH | 23/58 | 39.6% | 58/58 | 100% |
| Ascorbic Acid | 46/57 | 80.7% | 56/57 | 98.2% |
| Bilirubin (lot 0306A) | 44/48 | 91.7% | 46/48 | 95.8% |
| Urobilinogen | 49/54 | 90.7% | 50/54 | 92.6% |
Table II. DIA SCREEN Lot 00226B versus Ames Lot D397095 and Rapignost Lot 633962A
| Specific Gravity | Same Color Block | | Within 1 Color Block | |
| --- | --- | --- | --- | --- |
| | 25/58 | 43.1% | 52/58 | 89.6% |
| Ketone | 51/58 | 87.9% | 58/58 | 100% |
| Glucose | 49/58 | 84.5% | 58/58 | 100% |
| Protein | 43/57 | 75.4% | 57/57 | 100% |
| Blood | 46/58 | 79.3% | 53/58 | 91.4% |
| Nitrite | 56/57 | 98.2% | 57/57 | 100% |
| pH | 27/58 | 46.5% | 58/58 | 100% |
| Ascorbic Acid | 49/57 | 86.0% | 57/57 | 100% |
| Bilirubin (lot 0306B) | 44/48 | 91.7% | 46/48 | 95.8% |
| Urobilinogen | 49/54 | 90.7% | 50/54 | 92.6% |
SS and E Page 9
{9}
Table III. DIA SCREEN Lot 00226A versus DIA SCREEN Lot 00226B
| Specific Gravity | Same Color Block | | Within 1 Color Block | |
| --- | --- | --- | --- | --- |
| | 54/58 | 93.1% | 58/58 | 100% |
| Ketone | 57/58 | 98.3% | 58/58 | 100% |
| Glucose | 55/58 | 94.8% | 58/58 | 100% |
| Protein | 55/57 | 96.5% | 57/57 | 100% |
| Blood | 57/58 | 98.3% | 58/58 | 100% |
| Nitrite | 57/57 | 100% | 57/57 | 100% |
| pH | 52/58 | 89.6% | 58/58 | 100% |
| Ascorbic Acid | 54/57 | 94.7% | 56/57 | 98.2% |
| Bilirubin (lots 0306A/B) | 48/48 | 100% | 48/48 | 100% |
| Urobilinogen | 54/54 | 100% | 54/54 | 100% |
SS and E Page 10
{10}
# DIA SCREEN versus Ames and Rapignost Clinical Correlation
Adjusted to use confirmatory test results for discrepant results.
Tables IV and V present the percentage of responses of the DIA SCREEN strip within one color block of the Ames and Rapignost Test Strips or the confirmatory test method when discrepancies of greater than one color block arose. 96.5% agreement within one color block was obtained for all analytes.
Table IV. DIA SCREEN Lot 00226A versus Ames Lot D397095 and Rapignost Lot 633962A
| Specific Gravity | Within 1 Color Block | |
| --- | --- | --- |
| | 55/58 | 96.5% |
| Ketone | 58/58 | 100% |
| Glucose | 58/58 | 100% |
| Protein | 57/57 | 100% |
| Blood | 57/58 | 98.3% |
| Nitrite | 57/57 | 100% |
| pH | 58/58 | 100% |
| Ascorbic Acid | 56/57 | 98.2% |
| Bilirubin (lot 0306A) | 48/48 | 100% |
| Urobilinogen | 54/54 | 100% |
Table V. DIA SCREEN Lot 00226B versus Ames Lot D397095 and Rapignost Lot 633962A
| Specific Gravity | Within 1 Color Block | |
| --- | --- | --- |
| | 55/58 | 96.5% |
| Ketone | 58/58 | 100% |
| Glucose | 58/58 | 100% |
| Protein | 57/57 | 100% |
| Blood | 57/58 | 98.3% |
| Nitrite | 57/57 | 100% |
| pH | 58/58 | 100% |
| Ascorbic Acid | 57/57 | 100% |
| Bilirubin (lot 0306B) | 48/48 | 100% |
| Urobilinogen | 54/54 | 100% |
SS and E Page 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.