To estimate bone mineral content (BMC, grams) and bone mineral density (BMD, grams/cm²).
Device Story
The DTX-200 Bone Densitometer is a dual-energy X-ray absorptiometry (DXA) system used to estimate bone mineral content (BMC) and bone mineral density (BMD) in the forearm. The device consists of a computer, monitor, keyboard, printer, and scanner. During operation, the patient's forearm is submerged in a water bath (soft-tissue equivalent material). An X-ray generator (55 kV, 0.3 mAmps) with a 1 mm diameter source collimation directs X-rays through the forearm to a scintillating crystal and solid-state dual-energy sandwich detector. The system performs line-by-line calibration to ensure precision. The device produces high-quality images and quantitative BMC/BMD estimates in approximately 3.0 to 4.5 minutes. Healthcare professionals use these outputs to compare a patient's bone density against age-matched peers and young adult mean values, aiding in the assessment of bone health. The effective patient dose is approximately 0.1 micro Sievert per scan.
Clinical Evidence
Clinical data was collected to establish a reference database for normal American female Caucasians (ages 20-80). Two studies were conducted: Helen Hayes Hospital (protocol #94-12) and Tufts University (protocol #3895). Data included at least 50 persons per ten-year age group. Mean and standard deviations were calculated to create a reference database for comparing patient BMC and BMD against age-matched peers and young adult means.
Technological Characteristics
Dual X-ray photon absorptiometry (DXA) system. Components: X-ray generator (55 kV, 0.3 mAmps), scintillating crystal, solid-state dual-energy sandwich detector. Soft-tissue equivalent: water bath. Calibration: line-by-line. Source collimation: 1 mm diameter. Reference location: 8 mm Ulna-Radius gap. System includes computer, keyboard, monitor, and printer.
Indications for Use
Indicated for the estimation of bone mineral content (BMC) and bone mineral density (BMD) in the distal section of the forearm in patients. Prescription use only.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
K964562
# Appendix E
510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
21 CFR 807.92(a): AUG 22 1997
21 CFR 807.92(a)(1):
* Submitter's name and address:
Osteometer MediTech A/S
Glerupvej 2, DK-2160 Rodovre
Denmark
* Submitter's telephone number: 011 45 44 92 42 00
* Contact person:
Ms. Anja Koue
Official Correspondent
Osteometer MediTech A/S
Glerupvej 2, DK-2160 Rodovre
Denmark
* Date this 510(k) summary was prepared: November 5, 1996
21 CFR 807.92(a)(2):
* Trade/proprietary name of the device: DTX-200 Bone Densitometer
* Classification name: Bone densitometer
21 CFR 807.92(a)(3); Legally marketed device (predicate device) to which equivalence is claimed:
* Osteometer MediTech A/S model DTX-200 bone densitometer
21 CFR 807.92(a)(4); Description of the device that is the subject of this premarket notification:
The DTX-200 candidate device is a dual energy system, where the measured object, in this case the forearm, is submerged into a soft tissue equivalent water bath.
The X-Ray generator is driven at 55 Kv, with a current of 0.3
OM70
218
{1}
OM70
2
219
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
mAmps. The detector uses a scintillating crystal and a solid state dual energy sandwich detector. A highly stable X-Ray output, together with a source collimation of 1 mm diameter, and excellent pixel resolution, yields images of very high quality.
The typical time to estimate bone mineral content and bone mineral density is approximately 3.0 to 4.5 minutes. The effective patient dose is estimated to be 0.1 micro Sievert per scan.
## 21 CFR 807.92(a)(5); Intended use:
The intended use of the DTX-200 candidate device is the same as the labeled intended use of the predicate device to which equivalence is claimed; i.e., "To estimate bone mineral content (BMC, grams) and bone mineral density (BMD, grams/cm²).
## 21 CFR 807.92(a)(6); Technological characteristics:
The design, material, chemical composition, energy source and other technological characteristics of the subject device are considered to be the same as the technological characteristics of the predicate devices. A summary of the technological characteristics of the subject device in comparison to those of the predicate devices follows:
* X-ray transmission source:
Both the DTX-200 candidate and the DTX-200 predicate bone densitometer devices consist of a computer, keyboard, monitor, printer, and scanner system using an X-ray transmission source. The DTX-200 candidate device and the DTX-200 predicate device are identical and use dual X-ray photon absorptiometry (DXA) to estimate bone mineral content (BMC) and bone mineral density (BMD).
* Soft-tissue equivalent material:
Both the candidate and predicate DTX-200 devices are identical and use water as soft tissue equivalent material.
* Calibration:
Both the DTX-200 candidate device and the DTX-200 predicate device use a line-by-line calibration system, ensuring good precision.
* Imaging technique, data acquisition, and quality control
{2}
OM70
3
220
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
### phantom:
These characteristics are the same on the DTX-200 candidate device as the DTX-200 predicate device.
### Source collimation:
The source collimation for both the DTX-200 candidate device and the DTX-200 predicate device is the 1 mm diameter.
### Reference location:
The reference location for both the DTX-200 candidate device and the DTX-200 predicate device is the 8 mm Ulna-Radius gap.
### Accuracy and precision error:
Accuracy and precision error of the DTX-200 candidate device is comparable to that of the DTX-200 predicate device.
## 21 CFR 807.92(b);
510(k) summaries for those premarket submissions in which a determination of substantial equivalence is also based on performance data shall also contain the following:
## 21 CFR 807.92(b)(1);
Brief discussion of the nonclinical tests submitted, referenced, or relied on in this premarket notification submission:
There were no nonclinical tests submitted, referenced, or relied on in this submission.
## 21 CFR 807.92(b)(2);
Brief discussion of the clinical tests submitted, referenced, or relied on in this premarket notification submission:
The results of two (2) clinical tests were used to obtain data to establish a reference data base for a reference population of normal American female caucasians ranging in age from 20 to 80 years. These clinical studies were conducted at:
1. Helen Hayes Hospital (West Haverstraw, NY); protocol # 94-12:
* Title: "Determination of Forearm Bone Density in a Normal
{3}
K964562
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
US Population and Comparison of Two New Procedures to Measure Bone Mineral Density".
* Conducted from February 1996 through September 1996.
2. United States Department of Agriculture Human Nutrition Research Center on Aging at Tufts University (Boston, MA); protocol # 3895
* Title: "Mineral Density in a Normal American Population".
* Conducted from March 1996 through October 1996.
Data in a reference population (normal, American, female, caucasians) of a statistically valid quantity of women participating in the two clinical studies were collected using a model DTX-200 bone densitometer to estimate bone mineral content and bone mineral density in the non-dominant forearm of a representative sample of the reference population and included at least 50 persons in ten year peer age groups. The mean and standard deviations of the data was calculated using appropriate statistical methods and was used to establish the reference data base for the reference population.
21 CFR 807.92(b)(3); The conclusions drawn from the nonclinical and clinical tests that demonstrate that the subject device is as safe, as effective, and performs as well as or better than the predicate device:
The standard deviation of estimated bone mineral content (BMC) and bone mineral density (BMD) of the reference population of participating female caucasians is statistically representative (as required by FDA's "Draft Guidance for Review of Bone Densitometer 510(k) Submissions"); i.e., at least 50 people in each ten year age group) of the reference population and the American reference data base (derived from this data) is suitable for use by a health care professional to compare BMC and BMD with age matched peers and for comparison with young adult mean value.
OM70
... END OF 510(k) SUMMARY ...
4
221
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
AUG 22 1997
Bert Hudson
Regulatory Consultant
Osteometer Meditech A/S
C/O Shotwell & Carr, Inc.
3003 LBJ Freeway Suite 100
Dallas, TX 75234
Re: K964562
DTX-200 Bone Densitometer
Dated: May 28, 1997
Received: May 29, 1997
Regulatory class: II
21 CFR 892.1170/Procode: 90 KGI
Dear Mr. Hudson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaniam.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
K 94 4562
# Appendix G1
FDA's New Form Describing "Indications for Use".
Page 1 of 1
510(k) Number (if known): Unknown; Not yet assigned by FDA.
Device Name: Model DTX-200 Bone Densitometer
Indications For Use for the device: Estimate bone mineral content (BMC in grams) and bone mineral density (BMD in grams/cm²) in the distal section of the forearm.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
OM67.3
David G. Sypn
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964562
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.