PERFECTA PS REVISION STEM

K962267 · Wrightmedicaltechnologyinc · JDI · Sep 10, 1996 · Orthopedic

Device Facts

Record IDK962267
Device NamePERFECTA PS REVISION STEM
ApplicantWrightmedicaltechnologyinc
Product CodeJDI · Orthopedic
Decision DateSep 10, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PERFECTA® Revision Hip System is intended to be used in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

Device Story

Monolithic femoral component for total hip arthroplasty; titanium alloy (ASTM F-136) construction with plasma spray coating; available in straight or bowed configurations; features calcar platform and coronal slot (select sizes); press-fit or cemented implantation; used by orthopedic surgeons in clinical/OR settings; provides structural support for hip joint reconstruction; benefits patients via pain relief and improved joint function.

Clinical Evidence

Bench testing only. Abrasion testing demonstrates plasma spray coating comparability to sintered bead coating. Fatigue testing confirms device withstands simulated in vivo loading.

Technological Characteristics

Material: Titanium alloy (ASTM F-136). Coating: Titanium plasma spray. Design: Monolithic femoral stem, straight or bowed, with calcar platform and coronal slot. Fixation: Press-fit or cemented. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients undergoing total hip arthroplasty requiring pain reduction, relief, or improved hip function.

Regulatory Classification

Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0} SENT BY: 9-8-96 :11:05AM :WRIGHT MEDICAL TECH. 301 594 2358;# 3/ 3 K962267 WRIGHT MEDICAL TECHNOLOGY, INC. 5677 AIRLINE ROAD ARLINGTON, TN 38002 901-867-9971 SEP 10 1996 Contact Person: Cristie Manuel Date: September 8, 1996 # 510(k) Summary Trade/Proprietary Name: PERFECTA® PS Revision Hip Stem 510(k) Number: K962267 Common Name: Plasma Spray Coated Femoral Hip Stem Classification: Class II Predicate Devices: PERFECTA® Hip System and PERFECTA® Revision Hip System manufactured by Wright Medical Technology, Inc. and Integral Total Hip System manufactured by Biomet This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. ## Description/Intended Use The PERFECTA® Revision Hip System is a monolithic design femoral component. The device is manufactured from titanium alloy (ASTM F-136) with a titanium plasma spray coating. The stem is available in two configurations: straight and bowed. All configurations have a calcar platform, from 2.0cm to 3.75cm. All sizes except the 12mm x 190mm have a coronal slot. The stem is designed to be press-fit or can be used with bone cement. The PERFECTA® Revision Hip System is intended to be used in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. ## Substantial Equivalence The PERFECTA® PS Revision Hip Stem is substantially equivalent to the PERFECTA® Hip System and the PERFECTA® Revision Hip System manufactured by Wright Medical Technology, Inc. and the Integral Total Hip System manufactured by Biomet. ## Testing - Submitted abrasion testing demonstrates that the plasma spray porous coating is comparable to sintered bead coating. - Submitted fatigue testing demonstrates that the PERFECTA® PS Revision Hip Stem was able to withstand test loading designed to simulate in vivo loading.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...