CrossFAST Integrated Microcatheter Advanced Delivery System
Applicant
Vantis Vascular, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Sep 18, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The CrossFAST Integrated Microcatheter Advanced Delivery System is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
Device Story
CrossFAST (CFM60) is a percutaneous catheter system comprising an inner and outer catheter; used in cardiac catheterization labs by trained physicians. The outer catheter acts as a guide catheter extension; the inner catheter provides support during delivery. The user pre-loads the inner catheter into the outer catheter; once delivered to the target site, the inner catheter is removed to allow passage of interventional devices. The device is compatible with standard 0.014" guidewires and 6F guide catheters. It facilitates interventional device placement, potentially improving procedural success in complex vascular anatomy.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by bench testing, including visual/dimensional, tensile, track force, stent passage, kink/torque resistance, contrast delivery, corrosion, radiopacity, particulate, and coating characterization testing, alongside design validation and HFE summative testing. Biocompatibility, sterilization, and shelf-life data from the predicate were leveraged.
Technological Characteristics
Percutaneous catheter system; single-use; EO sterilized. Compatible with 0.014" guidewires and 6F guide catheters. Designed for vessels >2.5mm. No new patient-contacting materials compared to predicate.
Indications for Use
Indicated for patients requiring access to coronary or peripheral vasculature to facilitate placement of interventional devices. Not for use in neurovasculature or venous system. Designed for vessels >2.5mm.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM55) (K242276)
Submission Summary (Full Text)
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 18, 2026
Vantis Vascular, Inc.
Mai-Ly Wilcox
VP of Regulatory and Quality Affairs
2570 N First St.
Suite 200
San Jose, California 95131
Re: K262817
Trade/Device Name: CrossFAST Integrated Microcatheter Advanced Delivery System
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY
Dated: August 7, 2026
Received: August 7, 2026
Dear Mai-Ly Wilcox:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262817 - Mai-Ly Wilcox
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262817 - Mai-Ly Wilcox
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LYDIA S.
GLAW -S
Digitally signed by
LYDIA S. GLAW -S
Date: 2026.09.18
15:29:29 -04'00'
Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K262817
Device Name
CrossFAST Integrated Microcatheter Advanced Delivery System
Indications for Use (Describe)
The CrossFAST Integrated Microcatheter Advanced Delivery System is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k) #:
# 510(k) Summary
Prepared on: 2026-08-07
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Vantis Vascular, Inc. |
| --- | --- |
| Applicant Address | 2570 N First St. Suite 200 San Jose CA 95131 United States |
| Applicant Contact Telephone | 415-254-1930 |
| Applicant Contact | Ms. Mai-Ly Wilcox |
| Applicant Contact Email | mwilcox@vantisvascular.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM60) |
| --- | --- |
| Common Name | Percutaneous catheter |
| Classification Name | Catheter, Percutaneous |
| Regulation Number | 870.1250 |
| Product Code(s) | DQY |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K242276 | CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM55) | DQY |
Device Description Summary
21 CFR 807.92(a)(4)
CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM60) ("Device", "CrossFAST", "CrossFAST (CFM60)") is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices. It is intended for use in a cardiac catheterization laboratory by physicians trained in percutaneous, intravascular techniques and procedures.
CrossFAST (CFM60) consists of two catheters: an Inner Catheter and an Outer Catheter. The Outer Catheter is intended to act as an extension to a traditional guide catheter (i.e., "guide extension") and to facilitate delivery of interventional devices within the target vasculature. The Inner Catheter is intended to be pre-loaded into the Outer Catheter by the user and to provide additional support to the Outer Catheter during delivery. Once assembled, the Inner Catheter tip extends 3cm beyond the Outer Catheter tip. After the Outer Catheter has been delivered to the target vasculature, the Inner Catheter is removed from the Outer Catheter to allow for the passage of interventional devices.
The Device is designed for use in vessels larger than 2.5mm and is not for use in the neurovasculature or the venous system. The Device is single use and provided sterile via Ethylene Oxide (EO) sterilization. The Device is compatible with standard, off-the-shelf 0.014" lumen inner diameter (ID) and minimum 180cm length guidewires and a minimum 100cm length 6F guide catheters. The Device is intended for use with a hemostasis valve.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The CrossFAST™ Integrated Microcatheter Advanced Delivery System is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
{5}
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| The CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM60) has the same Indications for Use as the predicate device, CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM55). The modifications do not change the intended use or introduce a new intended use. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM60) has the same fundamental scientific technology, operating principle, materials of construction, manufacturing processes, sterilization method, and packaging configuration as the predicate device, CrossFAST™ Integrated Microcatheter Advanced Delivery System (CFM55). The subject device incorporates limited dimensional refinements while maintaining the same intended use and compatibility with a 6 Fr guide catheter. Design verification and validation testing demonstrated that these modifications do not raise different questions of safety or effectiveness. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The technological differences between the subject and predicate devices have been evaluated through design verification and validation testing and a biocompatibility assessment to provide evidence of substantial equivalence for the CrossFAST (CFM60). Testing previously conducted on the predicate device that is unaffected by the modifications was leveraged in support of this submission. The testing assessed for CrossFAST (CFM60) is listed below. | |
| Design Verification and Validation/In-Vitro (Bench) Testing: 1) Visual and Dimensional Testing 2) Tensile Testing (including Simulated Use) 3) Track Force Testing (including Simulated Use) 4) Performance Testing: • System Preparation & Inspection Testing • Stent Passage Testing (including Simulated Use) • Catheter Kink Resistance Testing • Catheter Torque Resistance Testing (including Simulated Use) • Contrast Delivery Testing • Corrosion Resistance Testing • Radiopacity Testing 5) Particulate Testing (including Simulated Use) 6) Coating Characterization • Baseline Coating Integrity • Post Simulated Use Coating Integrity 7) Design Validation & HFE Summative Testing | |
| Biocompatibility Assessment: A biocompatibility assessment was performed for the CrossFAST (CFM60). The subject device does not introduce new or changed patient-contacting materials, primary packaging, or manufacturing processing aids as compared to the predicate device. The assessment confirmed that the biocompatibility data previously established for the predicate device remain applicable to the subject device, and that no additional biocompatibility testing was required. | |
| Sterilization, Packaging, and Shelf Life: The sterilization method, packaging configuration, and shelf life of the subject device are unchanged from the predicate device. The sterilization, packaging, and shelf-life validations previously conducted on the predicate device were therefore leveraged, and no additional validation was required. | |
| No clinical testing was conducted to support this 510(k) Premarket Notification. | |
| The non-clinical testing demonstrated that the subject device functions as intended for its intended use and Indications for Use, which are unchanged from the predicate device, and confirmed that any technological differences between the subject device and the selected predicate device do not raise different questions of safety or effectiveness. The test results support that the subject device is substantially equivalent to the predicate device for the same intended use. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.