CrossFAST™ Integrated Microcatheter Guide Extension System (CFM55)
Applicant
Vantis Vascular, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Sep 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The CrossFAST™ Integrated Microcatheter Guide Extension System is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
Device Story
CrossFAST™ Integrated Microcatheter Guide Extension System (GES) is a two-component catheter system (Inner and Outer Catheters) used in cardiac catheterization laboratories. The Outer Catheter acts as a guide extension to a traditional guide catheter; the Inner Catheter provides additional support during delivery. The system is pre-loaded; once the Outer Catheter reaches the target vasculature, the Inner Catheter is removed to allow passage of interventional devices. Operated by physicians trained in percutaneous, intravascular procedures. The device facilitates delivery of interventional devices to discrete regions of coronary/peripheral vasculature. Benefits include improved support and access to target vessels. Compatible with standard 0.014" guidewires and 6F guide catheters.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including packaging/transportation, visual/dimensional, tensile, simulated use, kink/torque resistance, contrast delivery, radiopacity, track force, particulate, and coating integrity testing. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility (thromboresistance, hemolysis, complement activation, and partial thromboplastin time).
Technological Characteristics
Percutaneous catheter system; single-use; EO sterilized. Compatible with 0.014" guidewires and 6F guide catheters. Features radiopaque markers. Designed for vessels >2.5mm. No software or electronic components.
Indications for Use
Indicated for use in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature and facilitate placement of interventional devices. Intended for use in vessels >2.5mm. Not for use in the venous system. For prescription use only by physicians trained in percutaneous, intravascular techniques.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
September 30, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vantis Vascular, Inc. Angela Thompson Vice President, Quality 2570 N First St., Suite 200 San Jose, California 95131
Re: K242276
Trade/Device Name: CrossFAST™ Integrated Microcatheter Guide Extension System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 1, 2024 Received: August 1, 2024
Dear Angela Thompson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Samuel G. Raben -S
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242276
Device Name
CrossFAST™ Integrated Microcatheter Guide Extension System (CFM55)
Indications for Use (Describe)
The CrossFAST™ Integrated Microcatheter Guide Extension System is intended to be used in conjunction with quide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K242276 |
|-----------|---------|
|-----------|---------|
| | 510(k) Summary | Prepared on: 2024-09-10 |
|--|----------------|-------------------------|
|--|----------------|-------------------------|
Contact Details
| | 21 CFR 807.92(a)(1) |
|-----------------------------|------------------------------------------------------------|
| Applicant Name | Vantis Vascular, Inc. |
| Applicant Address | 2570 N First St. Suite 200 San Jose CA 95131 United States |
| Applicant Contact Telephone | 510-673-0983 |
| Applicant Contact | Angela Thompson |
| Applicant Contact Email | athompson@vantisvascular.com |
Device Name
| | 21 CFR 807.92(a)(2) |
|---------------------|---------------------------------------------------------------------------------|
| Device Trade Name | CrossFAST <sup>TM</sup> Integrated Microcatheter Guide Extension System (CFM55) |
| Common Name | Percutaneous catheter |
| Classification Name | Catheter, Percutaneous |
| Regulation Number | 870.1250 |
| Product Code(s) | DQY |
Legally Marketed Predicate Devices
| | 21 CFR 807.92(a)(3) |
|--|---------------------|
|--|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K183353 | Telescope <sup>TM</sup> Guide Extension Catheter | DQY |
Device Description Summary
| | 21 CFR 807.92(a)(4) |
|--|---------------------|
|--|---------------------|
CrossFAST™ Integrated Microcatheter Guide Extension System (GES) ("Device" or "CrossFAST" (GES") is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices. It is intended for use in a cardiac catheterization laboratory by physicians trained in percutaneous, intravascular techniques and procedures.
CrossFAST™ GES consists of two catheters: an Inner Catheter. The Outer Catheter is intended to act as an extension to a traditional guide catheter (i.e., "guide extension") and to facilitate delivery of interventional devices within the target vasculature. The Inner Catheter is intended to be pre-loaded into the user and to provide additional support to the Outer Catheter during delivery. Once assembled, the Inneyond the Outer Catheter tip. After the Outer Catheter has been delivered to the target vasculature, the Inner Catheter is removed from the passage of interventional devices.
The Device is designed for use in vessels larger than 2.5mm and is not for use in the venous system. The Device is single use and provided sterile via Ethylene Oxide (EO) sterilization. The Device is compatible with standard, off-the-shelf 0.014" lumen inner diameter (ID) and minimum 180cm length guidewires and a minimum 100cm length 6F guide is intended for use with a hemostasis valve.
### Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The CrossFAST™ Integrated Microcatheter Guide Extension System is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.
{5}------------------------------------------------
## Indications for Use Comparison
### 21 CFR 807.92(a)(5)
CrossFAST™ Integrated Microcatheter Guide Extension System GES ("CrossFAST" GES") has the same intended use and Indications for Use as Telescope™ Guide Extension Catheter (K18353). Both devices are intended to be used in a cardiac catheterization laboratory by physicians trained in percutaneous, intravascular procedure. Both devices are for prescription use only. Both devices are similar in size and utilize radiopaque markers. Any differences between Cross AST™ GES and Telescope Guide Extension Catheter (K183353) do not raise any different questions of safety and effectiveness as confirmed through the Performance Testing.
### Technological Comparison
### 21 CFR 807.92(a)(6)
CrossFAST™ Integrated Microcatheter Guide Extension System GES ("CrossFAST™ GES") and Telescope Guide Extension Catheter (K183353) share the same or similar fundamental technological characteristics and principles of operation. Vantis Vascular, hc. has performed comprehensive performance testing to demonstrate that the device meets all required specifications. The performance testing demonstrated that any technological differences between CrossFAST™ GES and Telescope Guide Extension Catheter (K183353) do not raise any different questions of safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The technological differences between the subject and predicate device has been evaluated through design tests and biocompatibility tests to provide evidence of substantial equivalence for the CrossFAST™ GES. The following testing conducted is listed below.
Design Verification/In-Vitro (Bench) Testing:
- 1) Packaging/Transportation Testing
- 2) Visual and Dimensional Testing
- 3) Tensile Testing (including Simulated Use)
- 4) Performance Testing:
- · System Preparation & Inspection Testing
- Stent Passage Testing (including Simulated Use)
- · Catheter Kink Resistance Testing
- · Catheter Torque Resistance Testing (including Simulated Use)
- · Contrast Delivery Testing
- · Radiopacity Testing
- 5) Track Force Testing (including. Simulated Use)
- 6) Particulate Testing (including Simulated Use)
- 7) Coating Characterization
- · Baseline Coating Integrity
- · Post Simulated Use Coating Integrity
Biocompatibility Testing:
- 1) Cytotoxicity
- 2) Sensitization
- 3) Irritation/Intracutaneous Toxicity
- 4) Acute Systemic Toxicity
- 5) Material Mediated Pyrogenicity
- 6) Hemocompatibility
- In-vivo Thromboresistance
- ASTM Hemolysis
- Complement Activation
- 7) Partial Thromboplastin Time
These tests confirmed that any technological differences between the selected predicate device device does not raise different questions of safety or effectiveness. The test results support that the proposed device is at least as safe and as effective as the predicate device for the same intended use.
No clinical testing was conducted to support this 510(k) Premarket Notification. As such, this section is not applicable.
The tests demonstrated that the proposed device functions as intended use and Indications for Use. These results further support that the proposed device performs at least as well as the for the same intended use and confirm that any technological differences do not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.