Sapphire Cooling IPL Hair Removal Device (Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B)

K262524 · ONE BEAUTY Technology Co., Ltd. · OHT · Sep 18, 2026 · General, Plastic Surgery

Device Facts

Record IDK262524
Device NameSapphire Cooling IPL Hair Removal Device (Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B)
ApplicantONE BEAUTY Technology Co., Ltd.
Product CodeOHT · General, Plastic Surgery
Decision DateSep 18, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sapphire Cooling IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.

Device Story

Sapphire Cooling IPL Hair Removal Device uses intense pulsed light (IPL) from a xenon arc flashlamp to reduce hair growth; operates below skin surface without cutting or pulling. Device includes main unit, external power adapter, and protective goggles. Features skin contact sensor to prevent pulse emission unless fully contacting skin; integrated sapphire cooling system cools treatment area during use. Operated by user via finger switch. Output intensity adjustable across 10 levels. Benefits include hair reduction with minimal pain and improved user comfort via cooling. Used in home environment.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), photobiological safety (IEC 62471), and software verification/validation.

Technological Characteristics

IPL hair removal device; xenon arc flashlamp source; 510-1200nm wavelength; 3cm² spot size; 4.67J/cm² max energy density; 10 intensity levels. Powered by external adapter (100-240V). Includes skin contact sensor and sapphire cooling element. Biocompatible materials per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471.

Indications for Use

Indicated for removal of unwanted body hair in individuals with Fitzpatrick skin types I-V. Intended for OTC use on limbs, armpits, back, chest, and bikini area.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 18, 2026 ONE BEAUTY Technology Co., Ltd. % Axiu Yang Registration specialist Feiying Drug & Medical Consulting Technical Service Group Rm.2401 Zhenye International Business Center, # 3101-90, Qianhai Rd. Shenzhen, Guangdong 518082 China Re: K262524 Trade/Device Name: Sapphire Cooling IPL Hair Removal Device (Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 22, 2026 Received: July 22, 2026 Dear Axiu Yang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262524 - Axiu Yang Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn {2} K262524 - Axiu Yang Page 3 (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.09.18 15:43:39 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K262524 Device Name Sapphire Cooling IPL Hair Removal Device Model(s): Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B Indications for Use (Describe) The Sapphire Cooling IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary # 510(k) Summary-K262524 "510(k) Summary" as required by 21 CFR Part 807.92. Date: 2026-09-14 # I. Submitter ONE BEAUTY Technology Co., Ltd. Room 2208, Shenye Center Building, 5045 Shennan East Road, Xinwei Community, Guiyuan Street, Luohu District, Shenzhen, Guangdong, China. Post code: 518000 Tel.: +86-755-28910840 Ninghua Pan Title: Quality Assurance Department Manager Tel.: +86-755-28910840 Email: lisa@onebeauty.com # II. Device Name of Device: Sapphire Cooling IPL Hair Removal Device Model(s): Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810 # III. Predicate Device & Reference Device Predicate and reference device: | Manufacturer | Predicate and Reference Device | 510(k) Number | Cleared Date | | --- | --- | --- | --- | | Shenzhen Greatro Electronic Technology Co., Ltd | Intense Pulse Light Therapeutic Apparatus (Model(s): IPL-18 BF, IPL-18 FG) | K242595 | November 11, 2024 | | Shenzhen Mlay Intelligent Technology Co., Ltd. | IPL Hair Removal Device (Model(s): T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C) | K251176 | August 14, 2025 | Page 1 of 9 {5} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary #### IV. Device Description Sapphire Cooling IPL Hair Removal Device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The Sapphire Cooling IPL Hair Removal Device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains of main unit, power adapter and protective goggles. The device contains a skin sensor to detect appropriate skin contact, if the light emitting window is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the Sapphire Cooling IPL Hair Removal Device has the cooling function, which will be activated throughout the whole hair removal process to cool down the treatment area's temperature by the Sapphire and provide the user with a better using experience. There are Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, and Ice 2.2B six models in this application. Their Intended use, working principle, circuit diagram, critical components, performance parameter etc., only appearance colour and Skin Tone Recognition Function are different. #### V. Indications for Use The Sapphire Cooling IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair. #### VI. Comparison of Technological Characteristics With the Predicate Device Sapphire Cooling IPL Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, specifications, and performance: Page 2 of 9 {6} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary | Comparison Elements | Subject Device | Predicate Device | Reference Device | Remark | | --- | --- | --- | --- | --- | | 510(k) Number | K262524 | K242595 | K251176 | / | | Trade name | Sapphire Cooling IPL Hair Removal Device | Intense Pulse Light Therapeutic Apparatus | IPL Hair Removal Device | / | | Manufacturer | ONE BEAUTY Technology Co., Ltd. | Shenzhen Greatro Electronic Technology Co., Ltd | Shenzhen Mlay Intelligent Technology Co., Ltd. | / | | Model | Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B | IPL-18 BF, IPL-18 FG | T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C | | | Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | | Product code | OHT | OHT | OHT | Same | | Device classification | Class II | Class II | Class II | Same | | Indication for use/ Intended use | The Sapphire Cooling IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair. | Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair. | The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. | Same | Page 3 of 9 {7} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary--- | Comparison Elements | Subject Device | Predicate Device | Reference Device | Remark | | --- | --- | --- | --- | --- | | | | | The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | | | Prescription or OTC | OTC | OTC | OTC | Same | | Applicable skin | Fitzpatrick skin types I-V | Fitzpatrick skin types I-V | Fitzpatrick Skin Types I-V | Similar | | Treatment area | Limbs, armpits, back, chest, bikini | / | Small areas such as underarm and facial hair below the chin line, and large areas such as arms and legs. | Similar | | Energy source | Supplied by external adapter | Supplied by external adapter | Supplied by external adapter | Same | | Power supply | 100-240V~, 50/60Hz, 1.8A | 100~240V, 50/60Hz | T14B, T16B,T19B: AC 100-240V, 50/60Hz, 2.5A MAX T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, | Same | Page 4 of 9 {8} ONE BEAUTY Technology Co., Ltd. 510(k)s -510(k) Summary | Comparison Elements | Subject Device | Predicate Device | Reference Device | Remark | | --- | --- | --- | --- | --- | | | | | T22B, T25B, T25C: AC 100-240V, 50/60Hz, 1.5A | | | Dimension | 204.7mm × 73.1mm × 54.2mm (L × W × H) | IPL-18 BF: 154*79*46mm; IPL-18 FG: 154*79*46mm | T14B:123.8*49.4*176.1mm T16B:122.0*46.8*175.0mm T19B:137.0*59.0*197.0mm T15B:163.0*66.0*43.0mm T17C:169.5*80.1*52.3mm T18B:171.0*65.2*45.5mm T21A:153.6*82.5*50.3mm T21B:153.6*82.5*50.3mm T21C:153.6*82.5*50.3mm T21D:153.6*82.5*50.3mm T22A:135.8*84.6*49.5mm T22B:135.8*84.6*49.5mm T25B:164.0*63.5*41.0mm T25C:164.0*63.5*41.0mm | Different | | Sterilization | Not required | Not required | Not required | Same | Page 5 of 9 {9} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary--- | Comparison Elements | Subject Device | Predicate Device | Reference Device | Remark | | --- | --- | --- | --- | --- | | Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same | | Energy medium | Xenon Arc flashlamp | Xenon Arc flashlamp | Xenon Arc flashlamp | Same | | Wavelength range | 510nm ~ 1200nm | 510-1200nm | 510~1200nm | Same | | Max. Energy density | 4.67J/cm^{2} | IPL-18 BF: 4.80J/cm^{2}; IPL-18 FG: 5.89J/cm^{2} | 5.90J/cm^{2} | Similar | | Max. Output energy | 14.00J | 13.0 | 3.1cm^{2}: 18.29 J 3.5cm^{2}: 20.65 J 3.9 cm^{2}: 23.01 J | Similar | | Spot size | 3cm^{2} | IPL-18 BF: (3.5±0.25) cm^{2}; IPL-18 FG: (2.9±0.25) cm^{2} | 3.1cm^{2}; 3.5cm^{2}; 3.9 cm^{2} | Similar | | Pulse duration | 0.4~2.40ms | 10.0±3.0ms | 0.4ms-5.5ms | Similar | | Pulsing control | Finger switch | Finger switch | Finger switch | Same | | Delivery device | Direct illumination to tissue | Direct illumination to tissue | Direct illumination to tissue | Same | | Output intensity level | 10 levels | 1-5 level | / | Different | Page 6 of 9 {10} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary | Comparison Elements | Subject Device | Predicate Device | Reference Device | Remark | | --- | --- | --- | --- | --- | | Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Same | | Electrical safety | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-11 IEC 60601-2-83 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 60601-1-6 IEC 60601-4-2 IEC 60601-2-57 | Same | | Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | Same | | Biocompatibility | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 | ISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-23 | Same | Page 7 of 9 {11} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary ## VII. Performance Data The following performance data were provided in support of the substantial equivalence determination. ### 1) Biocompatibility Safety The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 2023’”, as recommended by FDA. The following testing was performed to, and passed, including: - ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity - ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization - ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation ### 2) Electrical Safety and EMC Safety Electrical safety and Eye safety testing was performed to, and passed, the following standards: - IEC 60601-1-2:2014+AMD1:2020/ EN 60601-1-2:2015+A1:2021 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility - IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems - IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020; ANSI/AAMI ES60601-1:2005/A2:2021 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance - IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020 for use in conjunction with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - IEC 60601-2-83:2019, IEC 60601-2-83:2019/AMD1:2022 for use in conjunction with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical Electrical Equipment - Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment - IEC 62471:2006 Photobiological safety of lamps and lamp systems - IEC 60601-1-6:2010, IEC 60601-1-6:2010/AMD1:2013, IEC 60601-1- 6:2010/AMD2:2020 for use in conjunction with IEC 62366-1:2015, IEC 62366-1:2015/AMD1:2020, and IEC 60601-1:2005, IEC 60601- 1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability ### 3) Software Verification and Validation Page 8 of 9 {12} ONE BEAUTY Technology Co., Ltd. 510(k)s –510(k) Summary Software documentation consistent with Basic Documentation of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. #### 4) Summary Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device. #### VIII. Conclusions In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design and performance, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device. Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...