Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
Device Story
Personal, light-based hair reduction device for over-the-counter use. Device utilizes Intense Pulsed Light (IPL) technology generated by a Xenon arc lamp to reduce hair growth without cutting or pulling. Powered by an external AC/DC adapter; operation controlled by a finger switch. Includes a built-in skin sensor to detect appropriate skin contact; light pulse emission is inhibited if contact is insufficient. Used by the end-user at home. Output is a light pulse delivered directly to the skin surface. Benefits include hair reduction with minimal pain. Healthcare providers do not interact with the device output; the device is intended for self-use.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, -10, -23), electrical safety and EMC testing (IEC 60601-1, -1-2, -1-11, -2-83), and photobiological safety testing (IEC 62471). Software verification and validation were performed for a moderate level of concern.
Technological Characteristics
IPL hair removal device using a Xenon arc lamp. Wavelength range: 510-1200nm. Energy density: 1.28-5.89 J/cm². Pulse duration: 10.0±3.0ms. Materials: PC+ABS plastic, POM, glass, aluminum alloy. Powered by external adapter (Input: AC100-240V, Output: DC12V 3A). Includes skin contact sensor. Software-controlled. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, and IEC 62471.
Indications for Use
Indicated for removal of unwanted body hair in individuals with Fitzpatrick skin types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
November 6, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen Greatro Electronic Technology Co., Ltd Lucy Yan Regulatory Manager East Side, 4/F Building A1, Xu'xing'da Industrial Park Shiyan Sub-district, Bao'an District Shenzhen, Guangdong 518108 China
Re: K242595
Trade/Device Name: Intense Pulse Light Therapeutic Apparatus (IPL-18 BF, IPL-18 FG) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 18, 2024 Received: August 30, 2024
Dear Lucy Yan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yan Fu -S
Digitally signed by Yan
Date: 2024.11.06
23:57:14-05'00'
Tanisha Hithe for Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K242595
Device Name
Intense Pulse Light Therapeutic Apparatus (IPL-18 BF, IPL-18 FG)
Indications for Use (Describe)
Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K242595
This summary of 510(k) information is submitted as required by requirements of SMDA and 21 CFR §807.92.
### 1 Administrative Information
| Submission Date | July. 15, 2024 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Submitter's Name: Shenzhen Greatro Electronic Technology<br>Co., Ltd<br>Address:<br>East Side, 4/F Building A1, Xu'xing'da Industrial Park, Shiyan<br>Sub-district, Bao'an District, Shenzhen, China. 518108<br>Contact person: Xinhua Yue<br>TEL: +86 139 2374 2027<br>E-Mail: janis@jrdmic.com |
| Submission<br>Correspondent | Contact person: Ms Lucy.Yan<br>E-Mail: lucy.yan@aivikon.com<br>Company: Shenzhen Greatro Electronic Technology Co., Ltd<br>Address:<br>East Side, 4/F Building A1, Xu'xing'da Industrial Park, Shiyan<br>Sub-district, Bao'an District, Shenzhen, China. 518108<br>Tel: +8613316972203 |
### 2 Device Information
| Common name of the device | IPL Hair Removal Device |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name of the device | Intense Pulse Light Therapeutic Apparatus |
| Type/Model of the device | IPL-18 BF, IPL-18 FG |
| Classification information | Classification panel: General & Plastic Surgery<br>Classification name: Laser surgical instrument for use in<br>general and plastic surgery and in dermatology<br>Regulation Number: 21 CFR 878.4810<br>Device Class: II<br>Product Code: OHT |
| 510(k) type of submission | Traditional |
### 3 Primary predicate Device Information and Reference predicate Device Information
{5}------------------------------------------------
| Predicate Device: | Intense Pulse Light Therapeutic Apparatus |
|-------------------|--------------------------------------------------------------------------------------------------------------|
| Sponsor: | Shenzhen Greatro Electronic Technology Co., Ltd. |
| Device: | Intense Pulse Light Therapeutic Apparatus, Model(s):<br>IPL-1 PRO, IPL-1, IPL-12, IPL-17 |
| 510(K) Number: | K232932 |
| Product code: | OHT |
| Approval date: | December 4, 2023 |
| Reference Device: | IPL Hair Removal Device |
| Sponsor: | Shenzhen Ulike Smart Electronics Co., Ltd |
| Device: | IPL Hair Removal Device, Model(s): UI06 PN, UI06 PL,<br>UI06 JL, UI06 BR, UI06 DB, UI06 PR, UI06 OG, UI06 RD |
| 510(K) Number: | K223618 |
| Product code: | OHT |
| Approval date: | February 28, 2023 |
# 4 Device Descriptions
The Intense Pulse Light Therapeutic Apparatus is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does notinvolve any cutting or pulling, reducing hair growth with minimal pain.The Intense Pulse Light Therapeutic Apparatus is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and its built-in skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to thetreated skin, the device will not emit the light pulse.There are IPL-18 BF and IPL-18 FG two models in this application. Their work principle, intended use, structure, appearance, size, and composition are the same, with differences being product function are slight differences, but these parameters are within the predicate device and do not affect or change the intended use of the device.
### 5 Intended Use/ Indications for Use
Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
### 6 SE Comparisons
The subject device Intense Pulse Light Therapeutic Apparatus has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the subject device may be found substantially equivalent to its predicate device.The subject device is compared with the following Predicate Devices in terms of intended use, design, specifications, and performance:
{6}------------------------------------------------
| Characteristics | Subject device | Predicate device<br>(K232932) | Reference Device<br>(K223618) | Remark |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Device Name | Intense Pulse Light<br>Therapeutic Apparatus | Intense Pulse Light<br>Therapeutic Apparatus | IPL Hair Removal<br>Device | NA |
| Device Model | IPL-18 BF, IPL-18 FG | IPL-1 PRO, IPL-1, IPL-12,<br>IPL-17 | UI06 PL, UI06 PN, UI06<br>JL, UI06 BR, UI06 DB,<br>UI06 PR, UI06 OG,<br>UI06 RD | NA |
| Manufacturer | Shenzhen Greatro<br>Electronic Technology<br>Co., Ltd | Shenzhen Greatro<br>Electronic Technology<br>Co., Ltd | Shenzhen Ulike Smart<br>Electronics Co., Ltd | NA |
| Product code | OHT | OHT | OHT | NA |
| Classification | II | II | II | NA |
| Intended Use/<br>Indication for<br>Use | Intense Pulse Light<br>Therapeutic Apparatus is<br>an over-the-counter<br>device intended for<br>removal of unwanted body<br>hair. | Intense Pulse Light<br>Therapeutic Apparatus is<br>an over-the-counter device<br>intended for removal<br>of unwanted body hair. | IPL Hair Removal<br>Device is indicated for<br>the removal of<br>unwanted hair. The<br>device is also indicated<br>for the permanent<br>reduction in hair<br>regrowth, defined as<br>the long-term, stable<br>reduction in the number<br>of hairs regrowing when<br>measured at6, 9 and 12<br>months after the<br>completion of a<br>treatment regime. | NA |
| Prescription or<br>OTC | OTC | OTC | OTC | Same |
| Applicable skin | Fitzpatrick skin types I-V | Fitzpatrick skin types I-V | Fitzpatrick skin types I-V | Same |
| Dimension | IPL-18 BF: 154*79*46mm;<br>IPL-18 FG: 154*79*46mm | IPL-<br>1PRO:133.5*83*48mm<br>IPL-1: 133.5*83*48mm<br>IPL-12:134*83.5*48.5mm<br>IPL-17: 140.5*78*45mm | 58x34x179 mm (W x H<br>x D) | Different<br>Note 01 |
| Weight | IPL-18 BF: 213g<br>IPL-18 FG: 201g | IPL-1 PRO:200g<br>IPL-1: 194g<br>IPL-12: 183g<br>IPL-17: 184g | / | |
| Power source | Supplied by external<br>adapter | Supplied by external<br>adapter | Supplied by external<br>adapter | Same |
| Power supply | Input:<br>AC100~240V50/60Hz<br>Output: DC12V 3A | Input:<br>AC100~240V50/60Hz<br>Output: DC12V 3A | 100~240V, 50/60Hz | Same |
| Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
| Energy medium | Xenon Arc lamp | Xenon Arc lamp | Xenon Arc lamp | Same |
### Table 1. Substantial Equivalence Comparison
{7}------------------------------------------------
| Shenzhen Greatro Electronic Technology Co., Ltd |
|-------------------------------------------------|
| 510(k)s_510(k) Summary K242595 |
| Wavelength<br>range | 510~1200nm | 510~1200nm | 560-1200mm | Same |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------|
| Energy density | IPL-18 BF:<br>1.28~4.80J/cm²;<br>IPL-18 FG:<br>1.52~5.89J/cm² | Max. 4.92 J/cm² | 3~6 J/cm² | |
| Output energy | Level 1: 6.0±1.2J<br>Level 2: 7.5±1.5J<br>Level 3: 9.5±1.9J<br>Level 4: 11.0±2.2J<br>Level 5: 13.0±2.6J<br>Error is ±20% | Level 1: 6.0J<br>Level 2: 7.5J<br>Level 3: 9.5J<br>Level 4: 11.0J<br>Level 5: 12.5J<br>Error is ±20% | 9.9~19.8J | Different<br>Note 02 |
| Output energy<br>level | 1 to 5 level | 1 to 5 level | 1 to 3 level | Same |
| Spot size | IPL-18 BF: (3.5±0.25)<br>cm²;<br>IPL-18 FG: (2.9±0.25) cm² | 3.3(±0.25) cm² | 3.3cm² | Different<br>Note 02 |
| Pulse duration | 10.0±3.0ms | 5.84~9.24ms | 1ms~7ms | Different<br>Note 03 |
| Pulsing control | Finger switch | Finger switch | Finger switch | Same |
| Delivery device | Direct illumination to<br>tissue | Direct illumination to tissue | Direct illumination to<br>tissue | Same |
| Availability of<br>skin sensors<br>(that used to<br>identify the<br>presence or<br>absence of skin<br>areas) | Presence | Presence | Presence | Same |
| Software/Firmw<br>are/Microproce<br>ss or Control? | Yes | Yes | Yes | Same |
| Skin-contacting<br>components | Device enclosure and<br>treatment panel | Device enclosure and light<br>outlet panel | Enclosure and<br>treatment window | Different<br>Note 04 |
| Materials of<br>skin-contacting<br>components | IPL-18 BF: PC+ABS<br>plastic, POM, Glass, and<br>Aluminum alloy;<br>IPL-18 FG: PC+ABS<br>plastic, POM, Glass | PC+ABS plastic, Glass | ABS, PC, Crystal<br>application | Different<br>Note 04 |
| Materials<br>statement | All user directly contacting<br>materials are compliance<br>with ISO10993-5,<br>ISO10993-10 and ISO<br>10993-23 requirements. | All user directly contacting<br>materials are compliance<br>with ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials are<br>compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | SE |
| EMC safety<br>statement | Comply with IEC60601-1-<br>2 requirements | Comply with IEC60601-1-<br>2 requirements | Comply with IEC60601-<br>1-2 requirements | Same |
| Electrical<br>safety<br>statement | Comply with IEC60601-1,<br>IEC60601-1-11,<br>IEC60601-2-83<br>requirements | Comply with IEC60601-1,<br>IEC60601-1-11,<br>IEC60601-2-83<br>requirements | Comply with IEC60601-<br>1, IEC60601-1-11, IEC<br>60601-2-83<br>requirements | Same |
| Photobiological | Comply with IEC62471 | Comply with IEC62471 | Comply with IEC62471 | Same |
| statement | requirements. | requirements. | requirements. | |
| Environment<br>for operation | Temperature:<br>5°C~30°CHumidity:<br>≤80%RH | Temperature:<br>5°C~30°CHumidity:<br>≤80%RH | / | SE |
| Environment<br>for storage | Temperature: -<br>10°C~60°CHumidity:<br>5%~90% | Temperature: -<br>10°C~60°CHumidity:<br>5%~90% | / | SE |
{8}------------------------------------------------
Shenzhen Greatro Electronic Technology Co., Ltd 510(k)s 510(k) Summary K242595
Note 01: The "Weight", "Dimension", is belonging to basic physical characteristics, although it is a little different from the predicate devices, it will not affect the main function and the intended use of the device. They all also comply with IEC60601-1 requirements. Besides, the subtle change of the physical characteristics will not affect the critical functions or normal use, and not raise any safety or effectiveness issues
Note 02: The max "Energy density" of the subject device is lower than the Reference device, so the subject device is substantially equivalent to the device. Although there is a minor difference in the spot size between the subject device and the predicate device, the spot size and output energy level which is related to energy density. The subject device complies with IEC60601-1 and IEC60601-2-83 and since the difference in energy density is not significant and be substantially equivalent, so this difference will not raise any safety or effectiveness issue.
Note03: The "Pulse duration" of the subject device is little different from the predicate device. But the subject device complies with IEC60601-1 and IEC60601-2-83; so this difference will not raise any safety or effectiveness issues.
Note04: Although the "skin-contact components" and the "Materials of skin-contacting components" of the subject device and predicate device are not exactly the same, the skincontacting components of the subject device has been tested and satisfied the standard requirements of ISO 10993-1. So this difference will not raise any safety/effectiveness problems.
These different technological characteristics of the subject devices do not raise different questions of safety and effectiveness. Thus, the subject device is Substantially Equivalent (SE) to the predicate devices which is legally marketed in US.
### 7 Performance Data
The following performance data were provided in support of the substantial equivalence determination.
- 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed and passed, including:
- ISO 10993-5:2009, Biological evaluation of medical devices -Part 5: Tests for in vitro や cytotoxicity
- や ISO 10993-10:2021, Biological evaluation of medical devices -Part 10: Tests for skin
{9}------------------------------------------------
sensitization
- や ISO 10993-23:2021, Biological evaluation of medical devices -Part 23: Tests for skin irritation
2) Electrical Safety and EMC
Electrical safety and EMC testing was performed and passed, the following standards:
- ゃ IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- ゃ IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- や IEC 60601-111 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- ゃ IEC 60601-2-83 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment 3) Eve Safetv
- ゃ IEC 62471 Photobiological safety of lamps and lamp systems
4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
### Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device.
### 8 Conclusions
The subject device:
Intense Pulse Light Therapeutic Apparatus is respectively substantially equivalent to the predicate device (Intense Pulse Light Therapeutic Apparatus) manufactured by Shenzhen Greatro Electronic Technology Co. (K232932) and reference device (IPL Hair Removal Device) manufactured by Shenzhen Ulike Smart Electronics Co., Ltd (K223618).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.