Medical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75)

K262426 · Wuhan Pioon Technology Co., Ltd. · GEX · Sep 10, 2026 · General, Plastic Surgery

Device Facts

Record IDK262426
Device NameMedical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75)
ApplicantWuhan Pioon Technology Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medical Diode Laser (Model: S1Pro+) is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. The Medical Diode Laser (Model: S1Pro K15) is indicated for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. -Laser assisted lipolysis. The Medical Diode Laser (Model: S1Pro N6) is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The Medical Diode Laser (Model: S1Pro N75) is intended to incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue.

Device Story

Medical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) delivers infrared laser energy (980nm, 1470nm, or 1940nm) to soft tissue; utilizes solid-state diode laser source; includes 650nm red aiming beam. System comprises main unit, foot switch, power adapter, and protective goggles; features LCD touchscreen for parameter control. Laser energy transmitted via third-party optical fibers (SMA905 connectors, 400μm/600μm cores). Operated by clinicians in surgical settings; provides continuous, pulsed, or single-pulse output for tissue interaction. Output enables precise surgical cutting, coagulation, or ablation; aids in endovenous occlusion and lipolysis. Benefits include controlled tissue interaction and hemostasis during procedures.

Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1, 60601-1-2), electromagnetic immunity (AIM 7351731), and laser performance/safety (IEC 60825-1, IEC 60601-2-22).

Technological Characteristics

Solid-state diode laser; wavelengths 980nm, 1470nm, 1940nm; 650nm aiming beam. Air-cooled. Fiber delivery system (SMA905). Microprocessor-controlled via LCD touchscreen. Power: 100-240VAC. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.

Indications for Use

Indicated for soft tissue surgical procedures including incision, excision, vaporization, ablation, hemostasis, and coagulation across various specialties (ENT, dental, general, plastic, dermatology, podiatry, urology, gynecology, etc.). Specific models indicated for endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux and laser-assisted lipolysis.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 10, 2026 Wuhan Pioon Technology Co., Ltd. Tracy Liu Regulatory Affairs 12th Floor, Bldg. 1, Innovative Unit, R&D Center Project, Marine World Shipyard Park, # 16 Fozuling 3rd Rd., E. Lake High-Tech Development Zone, Wuhan, Hubei 430205 China Re: K262426 Trade/Device Name: Medical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 15, 2026 Received: July 16, 2026 Dear Tracy Liu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262426 - Tracy Liu Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K262426 - Tracy Liu Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.09.10 13:50:23 -04'00' Tanisha L. Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K262426 Device Name Medical Diode Laser Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75 Indications for Use (Describe) The Medical Diode Laser (Model: S1Pro+) is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. The Medical Diode Laser (Model: S1Pro K15) is indicated for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. -Laser assisted lipolysis. The Medical Diode Laser (Model: S1Pro N6) is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The Medical Diode Laser (Model: S1Pro N75) is intended to incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} # **K262426** **510(k) Summary** In accordance with the content and format regulatory requirements of 21 CFR Part 807.92, the 510(k) Summary for the Medical Diode Laser is provided below. # **1. Submitter** Device Submitter: Wuhan Pioon Technology Co., Ltd. Address: 12th Floor, Building 1, Innovative Unit, R&D Center Project, Marine World Shipyard Park, No. 16 Fozuling 3rd Road, East Lake High-tech Development Zone, 430205, Wuhan, Hubei, China Tel: +86 27 81783687 Contact Person: Zhang Feng, Management Representative Phone: +86 18062448535 E-mail: zhangfeng@pioon.com Date Prepared: August 24, 2026 Official Correspondent: Tracy Liu, Wuhan Pioon Technology Co., Ltd. Phone: +86 15012997429 Email: tracy@pioon.com # **2. Device** Type of 510(k) submission: Traditional Device name: Medical Diode Laser (Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) Device Classification Name: Powered Laser Surgical Instrument Regulation: 878.4810 - Laser surgical instrument for use in general and plastic surgery and in dermatology Medical Specialty: General & Plastic Surgery Regulatory Class: II Product Code: GEX # **3. Predicate Device** | Item | For S1Pro+ | For S1Pro K15 | For S1Pro N6 | For S1Pro N75 | | --- | --- | --- | --- | --- | Page 1 of 9 {5} | | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | | --- | --- | --- | --- | --- | --- | | 510(k) Number | K252658 | K240179 | K241791 | K241791 | K230274 | | Device Name | Medical Diode Laser Systems (TR-B) | Medical Diode Laser(L2) | Medical Diode Laser Systems (GBOX-6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS Pro-30B15D, VELAS II-6H, VELAS II-15D) | Medical Diode Laser Systems (GBOX-6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS Pro-30B15D, VELAS II-6H, VELAS II-15D) | Medical Diode Laser(M2) | | Submitter | Baoding Te'anzhou Electronic Technology Co., Ltd. | Wuhan Pioon Technology Co.,Ltd. | Gigaalaser Company Ltd. | Gigaalaser Company Ltd. | Wuhan Pioon Technology Co.,Ltd. | | Product Code | GEX | GEX | GEX | GEX | GEX | ## 4. Device Description The Medical Diode Laser (Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N7) incorporates a solid state diode as laser energy source for producing its therapeutic effects, it is designed to deliver continuous or pulsed, infrared laser energy at wavelengths of 980nm and 1470nm respectively for S1Pro+, 1470nm for S1Pro K15 and 1940nm for S1Pro N6, S1Pro N75. The device also incorporates a red (650nm) aiming beam diode to indicate the area to be irradiated by the laser beam. The device is composed of the main unit, foot switch, power adapter, and protective goggles. The device incorporates an LCD touchscreen for the device user to set output parameters and for control of the device. The device transmits the laser output via third-party optical fibers with SMA905 connectors and with single cores of 400μm and 600μm diameter respectively. ## 5. Indications for Use The Medical Diode Laser (Model: S1Pro+) is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. Page 2 of 9 {6} -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. The Medical Diode Laser (Model: S1Pro K15) is indicated for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. -Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. -Laser assisted lipolysis. The Medical Diode Laser (Model: S1Pro N6) is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The Medical Diode Laser (Model: S1Pro N75) is intended to incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. ## 6. Comparison to the Predicate Device The Medical Diode Laser (Model: S1Pro+), has been compared to the predicate device Medical Diode Laser Systems (Model: TR-B)(K252658) as reference for substantial equivalence. The Medical Diode Laser (Model: S1Pro K15), has been compared to the predicate device Medical Diode Laser (Model: L2)(K240179) and predicate device Medical Diode Laser Systems (Model: GBOX-15D)(K241791) as reference for substantial equivalence. The Medical Diode Laser (Model: S1Pro N6), has been compared to the predicate device Medical Diode Laser Systems (Models: GBOX-6H, VELAS II-6H)(K241791) as reference for substantial equivalence. The Medical Diode Laser (Model: S1Pro N75) has been compared to the predicate device Medical Diode Laser (Model: M2)(K230274) as reference for substantial equivalence. A table comparing the predicate devices to the subject device is shown as the following: Page 3 of 9 {7} | Item | Proposed Device(this submission) | For S1Pro+ | For S1Pro K15 | | For S1Pro N6 | For S1Pro N75 | | --- | --- | --- | --- | --- | --- | --- | | | | Predicate Device(K252658) | Predicate Device(K240179) | Predicate Device(K241791) | Predicate Device(K241791) | Predicate Device(K230274) | | Trade Name | Medical Diode Laser | Medical Diode Laser Systems | Medical Diode Laser | Medical Diode Laser Systems | Medical Diode Laser Systems | Medical Diode Laser | | Model | S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75 | TR-B | L2 | GBOX-15D | GBOX-6H, VELAS II-6H | M2 | | Product Code | GEX | GEX | GEX | GEX | GEX | GEX | | Regulation NO. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | | Class | II | II | II | II | II | II | | Indications for Use | The Medical Diode Laser (Model: S1Pro+) is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, | The Medical Diode Laser Systems are intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, | The Medical Diode Laser (Model: L2) is indicated for:-Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux | The Medical Diode Laser Systems, model: GBOX-15D, VELASII-15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELASII-15D is further indicated for laser assisted lipolysis. | The Medical Diode Laser Systems, model: GBOX-6H, VELAS II-6H is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, | Medical Diode Laser is intended to incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. | Page 4 of 9 {8} | | gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: -Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue.-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.The Medical Diode Laser (Model: S1Pro K15) is indicated for:-Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue.-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.-Laser assisted lipolysis.The Medical Diode Laser | gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for: Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux. | | | Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. | | | --- | --- | --- | --- | --- | --- | --- | Page 5 of 9 {9} | | (Model: S1Pro N6) is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The Medical Diode Laser (Model: S1Pro N75) is intended to incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. | | | | | | | --- | --- | --- | --- | --- | --- | --- | Page 6 of 9 {10} | **Use of Device** | Rx only | Rx only | Rx only | Rx only | Rx only | Rx only | | --- | --- | --- | --- | --- | --- | --- | | **Laser Type** | Diode laser | Diode laser | Diode laser | Diode laser | Diode laser | Diode laser | | **Laser Classification** | Class IV | Class IV | Class IV | Class IV | Class IV | Class IV | | **Configuration** | Main Unit | Main Unit | Main Unit | Main Unit | Main Unit | Main Unit | | | Foot Control | Foot Control | Foot Control | Foot Control | Foot Control | Foot Control | | **Laser Wavelength** | S1Pro+: 980nm, 1470nm S1Pro K15: 1470nm S1Pro N6, S1Pro N75: 1940nm | 980nm 1470nm | 1470nm | 1470nm | 1940nm | 1940nm | | **Max Power** | S1Pro+: 20W-980nm, 15W-1470nm S1Pro K15: 15W S1Pro N6: Up to 6W S1Pro N75:Up to 7.5W | 20W-980nm 15W-1470nm | 20W | 15W | Up to 6W | Up to 7.5W | | **Operation Mode** | CW, Pulsed, Single Pulse | CW, Pulsed, Single Pulse | CW, Pulsed, Single Pulse | CW, Pulsed, Single Pulse | CW, pulsed, single pulse | CW and Pulse | | **Pulse Duration** | 10ms – 10s | 0.1ms - 10s | 10ms – 10s | 10ms – 10s | 10ms – 10s | 10ms – 10s | | **Repetition Rate** | 0.05 to 50 Hz | 0.05-5Hz | 0.05-5Hz | 0.05-50Hz | 0.05-50Hz | up to 100 Hz | | **Aiming Beam** | Red 650nm (<2mW) | Red 650nm (≤10mW) | Red 650nm (<2mW) | Red 650nm (≤3mW) | Red 650nm (≤3mW) | Red 650nm (<2mW) | | **Cooling Method** | Air cooling | Air cooling | Air cooling | Air cooling | Air cooling | Air cooling | | **Delivery System** | Fiber delivery | Fiber delivery | Fiber delivery | Fiber delivery | Fiber delivery | Fiber delivery | | **User Interface** | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Touch screen control | Color LCD touch screen | | **Microprocessor Control** | Yes | Yes | Yes | Yes | Yes | Yes | Page 7 of 9 {11} | Power Supply | 100-240VAC, 50/60Hz, 2.5A max; 19V DC, 9.47A | 100-240VAC, 50/60Hz | 100-240VAC, 50/60Hz, 4A-2A | 100-240 VAC, 50-60Hz, 200VA | 100-240VAC, 50/60Hz, 200VA | 100-240VAC, 50/60Hz, 4A-2A | | --- | --- | --- | --- | --- | --- | --- | | Safety Feature | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | Complied with: IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60825-1 | From the above table, the proposed device Medical Diode Laser (Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) and the predicate devices have the same indications for use. The subject device uses similar 980nm, 1470nm and 1940nm diode lasers technology as that used by the predicates. The output parameters of the proposed device are similar to the output parameters of the predicates, and subtle differences in the output parameters do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use. Page 8 of 9 {12} # 7. Performance Data # Clinical data: Not applicable. # Non-clinical data: # Electrical Compatibility and Electrical Safety The Medical Diode Laser was tested and found to conform to the criteria of the following performance standards: - IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. - IEC 60601-1-2:2014+A1:2020 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. - AIM Standard 7351731 Rev. 3.00 2021-06-04 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers - An AIM Standard # Performance Testing – Bench The Guidance Document, Laser Products – Conformance with IEC 60825-1 and IEC 60601-2-22 (Laser Notice 56) January 19, 2018 was used. Pioon has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specifications and meets relevant consensus standards. - IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification, and requirements - IEC 60601-2-22:2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment # Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Function”. The recommended Documentation Level for software of this device was considered as a “Basic Documentation Level” and the documentation was provided accordingly. # 8. Conclusions The non-clinical data support the safety of the proposed Medical Diode Laser (Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) device, and the software verification and validation demonstrate that the device can perform as intended. The proposed device uses similar laser diode technology as that used by the predicate devices and difference in parameters do not raise new types of questions regarding safety and effectiveness for the proposed indications for use. The Medical Diode Laser (Models: S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) is considered to be substantially equivalent to the predicate devices. Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...