Medical Diode Laser Systems (GBOX-6H, GBOX-10D , GBOX-12D, GBOX-15D, VELAS Pro-30B15D, VELAS II-6H, VELAS II-15D)
K241791 · Gigaalaser Company , Ltd. · GEX · Mar 19, 2025 · General, Plastic Surgery
Device Facts
Record ID
K241791
Device Name
Medical Diode Laser Systems (GBOX-6H, GBOX-10D , GBOX-12D, GBOX-15D, VELAS Pro-30B15D, VELAS II-6H, VELAS II-15D)
Applicant
Gigaalaser Company , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medical Diode Laser Systems, model: GBOX-6H, VELAS II -6H is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-6H and VELAS II -6H is 1940nm. The Medical Diode Laser Systems, model: GBOX-10D, GBOX-12D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-10D and GBOX-12D is 1470nm. The Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELAS II -15D is further indicated for laser assisted lipolysis. The output wavelength of model GBOX-15D and VELAS II -15D is 1470nm. The Medical Diode Laser Systems, model: VELAS Pro-30B15D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. Medical Diode Laser Systems, model: VELAS Pro-30B15D is further indicated for laser assisted lipolysis. The output wavelength of model VELAS Pro-30B15D is 980nm or 1470nm.
Device Story
Medical diode laser system; uses GaAlAs diode bar to generate laser energy at 980nm, 1470nm, or 1940nm. Input: user-selected parameters via touch screen; footswitch activation. Transformation: MCU controls drive current to diode module; laser beam coupled into optical fiber. Output: laser energy delivered to soft tissue for surgical applications. Used in clinical/surgical settings by medical specialists. Output affects tissue via thermal effects (vaporization, coagulation, etc.). Benefits: precise soft tissue management, minimally invasive endovenous occlusion, and lipolysis.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance verified via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), and performance standards (IEC 60601-2-22).
Technological Characteristics
GaAlAs diode laser; wavelengths 980nm, 1470nm, 1940nm. Fiber-coupled delivery. MCU-controlled drive board and temperature system. Class 4 laser. Power supply 100-240 VAC. Color LCD touch screen interface. Air cooling. Software-controlled parameters.
Indications for Use
Indicated for soft tissue surgical procedures (vaporization, incision, excision, ablation, cutting, hemostasis, coagulation) across multiple specialties including urology, thoracic, plastic, dermatology, general, ophthalmology, orthopedics, podiatry, arthroscopy, spinal, gynecology, pulmonary, neurosurgery, gastroenterology, ENT, radiology, oral, and dental. Also indicated for endovenous occlusion of greater saphenous vein in superficial vein reflux patients and laser-assisted lipolysis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 19, 2025
Gigaalaser Company Ltd.
Xinxing Nie
Regulations Control Manager
B15-8F, Wuhan Hi-Tech Medical device park,
#818 Gaoxin Road
Wu han, 430206
China
Re: K241791
Trade/Device Name: Medical Diode Laser Systems (GBOX-6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS Pro-30B15D, VELAS II-6H, VELAS II-15D)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: March 14, 2025
Received: March 14, 2025
Dear Xinxing Nie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241791 - Xinxing Nie
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K241791 - Xinxing Nie
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU -S
Digitally signed by YAN FU -S
Date: 2025.03.19 07:21:04
-04'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
## Submission Number (if known)
K241791
## Device Name
Medical diode laser systems (GBOX-10D, GBOX-12D, GBOX-15D, GBOX-6H, VELAS Pro-30B15D, VELAS II -6H and VELAS II -15D)
## Indications for Use (Describe)
(1) The Medical Diode Laser Systems, model: GBOX-6H, VELAS II -6H is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-6H and VELAS II -6H is 1940nm.
(2) The Medical Diode Laser Systems, model: GBOX-10D, GBOX-12D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-10D and GBOX-12D is 1470nm.
(3) The Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELAS II -15D is further indicated for laser assisted lipolysis. The output wavelength of model GBOX-15D and VELAS II -15D is 1470nm.
(4) The Medical Diode Laser Systems, model: VELAS Pro-30B15D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. Medical Diode Laser Systems, model: VELAS Pro-30B15D is further indicated for laser assisted lipolysis. The output wavelength of model VELAS Pro-30B15D is 980nm or 1470nm.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
510(k) Summary
K241791
Prepared on:2025-03-18
# I. SUBMITTER
Device submitter: Gigaalaser Company Ltd.
Add.:
B15-8F, Wuhan Hi-Tech Medical device park, #818 Gaoxin Road, Wuhan 430206, China
Name: Xinxing Nie
Title: Regulations Control Manager
Contact Person:
Phone: +86 18062575326
Fax: +86-27-6784 8873
E-mail: fdagigaalaser@sohu.com
# II. DEVICE
Trade Name of Device: Medical Diode Laser Systems
Model: GBOX-6H, VELAS II -6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS II -15D, VELAS Pro-30B15D
Regulation name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology
Classification name: POWERED LASER SURGICAL INSTRUMENT
Classification: Class II, 21 CFR 878.4810
Primary Product code: GEX
Review Panel: General & Plastic Surgery
510(k) number: K241791
# III. PREDICATE DEVICE
Predicate Device
Trade name: MANTA Diode Lasers
Classification name: POWERED LASER SURGICAL INSTRUMENT
Regulation number: 21 CFR 878.4810
Regulation name: GEX-Powered Laser Surgical Instrument, General and Plastic Surgery
1 / 17
{6}
2 / 17
Regulatory class: Class II
Product code: GEX
Submitter: SQUALUS MED Ltd.
510(k) number: K222701
Predicate Device
Trade name: SkyPulse Laser Platform
Classification name: POWERED LASER SURGICAL INSTRUMENT
Regulation number: 21 CFR 878.4810
Regulation name: GEX-Powered Laser Surgical Instrument, General and Plastic Surgery
Regulatory class: Class II
Product code: GEX
Submitter: Fotona d.o.o.
510(k) number: K193656
Predicate Device
Trade name: Alma Diode Tabletop Laser
Classification name: POWERED LASER SURGICAL INSTRUMENT
Regulation number: 21 CFR 878.4810
Regulation name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory class: Class II
Product code: GEX
Submitter: Alma Ltd
510(k) number: K160952
Predicate Device
Trade name: Medical Diode Laser (M2-GK)
Classification name: POWERED LASER SURGICAL INSTRUMENT
Regulation number: 21 CFR 878.4810
Regulation name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
{7}
Regulatory class: Class II
Product code: GEX
Submitter: Wuhan Pioon Technology Co., Ltd.
510(k) number: K240747
3 / 17
{8}
4 / 17
# IV. DEVICE DESCRIPTION
(1) The "Medical Diode Laser Systems", include model: GBOX-6H, VELAS II -6H, GBOX-10D, GBOX-12D, GBOX-15D, VELAS II -15D, VELAS Pro-30B15D which consist of three main components:
(2) Main device;
(3) Foot switch;
(3) Accessories.
Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.
The diode laser for this unit is GaAlAs diode bar, and the wavelength is 1940nm for GBOX-6H, VELAS II -6H; 1470nm for GBOX-10D, GBOX-12D, GBOX-15D, VELAS II -15D; 980nm or 1470nm for VELAS Pro-30B15D. It features impact structure, high efficiency and long lifetime. Generally, the beam shall be emitted as the big beam divergence of the laser from the diode. With the unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.

MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system. It supplies power for MCU, diode laser module by DC-DC conversion module and temp system by temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU has three external input sources: footswitch, remote interlock connector and information input
{9}
from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system.
V. INDICATIONS FOR USE
(1) The Medical Diode Laser Systems, model: GBOX-6H, VELAS II -6H is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-6H and VELAS II -6H is 1940nm.
(2) The Medical Diode Laser Systems, model: GBOX-10D, GBOX-12D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. The output wavelength of model GBOX-10D and GBOX-12D is 1470nm.
(3) The Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELAS II -15D is further indicated for laser assisted lipolysis. The output wavelength of model GBOX-15D and VELAS II -15D is 1470nm.
(4) The Medical Diode Laser Systems, model: VELAS Pro-30B15D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. Medical Diode Laser Systems, model: VELAS Pro-30B15D is further indicated for laser assisted lipolysis. The output wavelength of model VELAS Pro-30B15D is 980nm or 1470nm.
5 / 17
{10}
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Medical Diode Laser System has the same intended use and principal operation, the technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Medical Diode Laser System and predicate devices do not alter suitability of the proposed device for its intended use.
Table 1 Substantial equivalence discussion—GBOX-6H / VELAS II -6H
| Item | Proposed device | Predicate device (K222701) | Predicate device (K193656) | Discussion |
| --- | --- | --- | --- | --- |
| | Medical diode laser systems | MANTA Diode Lasers | SkyPulse Laser Platform | |
| Product Code | GEX | GEX | GEX | Identical |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | Medical Diode Laser Systems, model: GBOX-6H, VELAS II -6H is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, | The MANTA1940 diode laser is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, | 1940 nm Diode Laser:
• Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue. | Substantial Equivalence
Indication for Use claimed by Proposed device is the same with the Predicate device (K222701).
Indication for Use claimed by Proposed device has been |
6 / 17
{11}
| | General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. | Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. | | covered by Predicate device (K193656). |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | 1940nm | 1940nm | 1940nm | Identical |
| Output power | 1-6W | Up to 5 W | Up to 7.5W | Similar Comment 1 |
| Pulse width | GBOX-6H: 10ms-10s
VELAS II -6H: 10ms-10s | 10 msec – 30 sec adjustable | CW or 10ms – 10s | Similar Comment 2 |
| Operation Mode | CW, single pulse, repeat pulse | CW, pulsed, single pulse | / | Identical |
| Operation interface | Color LCD touch screen | Color touch screen | Touch screen control | Identical |
| Repetition rate | 0.05Hz-50Hz | 0.02–50Hz | CW or up to 100Hz | Similar Comment 3 |
| Aiming beam | Diode laser of 650nm, power ≤3mW, adjustable brightness. Red, Class1M, CW | Red 635-650nm (<5mW) | / | Substantial Equivalence |
| Laser Class | 4 | 4 | / | Identical |
| Power supply | 100-240 VAC, 50-60Hz, 200VA | 100-240V, 47-63 Hz | / | Substantial |
7 / 17
{12}
8 / 17
| | | | | Equivalence |
| --- | --- | --- | --- | --- |
## Discussion:
**Clinical**: Indication for Use claimed by Proposed device is the same with the Predicate device (K222701). Indication for Use claimed by Proposed device has been covered by Predicate device (K193656). The indication for use depends on product performance and this difference does not affect safety and effectiveness.
**Technology**: The subject device has the same operation principle and same wavelengths of diode laser to achieve its intended use. The operation mode, Laser Classification, Operation interface are either identical to the predicate. The main differences are Output power, Pulse width and Repetition rate.
**Comment 1** Output power:
The power output range of proposed device is 1-6W, which is similar to the predicate device (K222701): 5W; Also similar to Predicate device (K193656): Up to 7.5 W. By Proposed device has been covered by predicate device (K222701) and (K193656).
**Comment 2** Pulse width:
The pulse width of the proposed device: 10ms-10s, somewhat different from the predicate device K222701 (10 msec – 30 sec), is within the scope of what predicate device claims. The pulse width of the proposed device: 10ms-10s is the same with the predicate device K193656 (10 ms – 10s).
**Comment 3** Pulse Repetition rate:
{13}
Repetition rate of the subject device: 0.05Hz-50Hz is within the scope of what predicate device claims. By Proposed device has been covered by Predicate device (K222701) and (K193656).
**Material**: The fiber in contact with the subject device has been approved in K140470.
**Conclusion**:
The subject device has the same intended use and operation principal, the technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices (K222701) and (K193656). The differences between the subject and predicate devices do not alter suitability of the proposed device for its intended use.
9 / 17
{14}
Table 2 Substantial equivalence discussion—GBOX-10D / GBOX-12D / GBOX-15D / VELAS II -15D
| Item | Proposed device | Predicate device (K222701) | Predicate device (K160952) | Discussion |
| --- | --- | --- | --- | --- |
| | Medical diode laser systems | MANTA Diode Lasers | Alma Diode Tabletop Laser | |
| Product Code | GEX | GEX | GEX | Identical |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | Medical Diode Laser Systems, model: GBOX-10D/12D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and | The MANTA1470 diode laser is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures. | The Alma 1470 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470 nm diode tabletop laser is further indicated for laser assisted lipolysis. | Substantial Equivalence Indication for Use claimed by Proposed device has been covered by Predicate device (K222701) and (K160952). Indication for Use claimed by Proposed device Medical Diode Laser Systems, model: GBOX-10D/12D is the same with the Predicate device (K222701). Indication for Use |
{15}
| | Dental procedures.
Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The GBOX-15D, VELAS II -15D is further indicated for laser assisted lipolysis. | | | claimed by Proposed device Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is the same with the Predicate device (K160952). |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | 1470nm | 1470nm | 1470nm | Identical |
| Output power | GBOX-10D: 1-10W
GBOX-12D: 1-12W
GBOX-15D: 1-15W | Up to 12 W | Up to 15 W | Similar Comment 1 |
| Pulse width | 10ms-10s | 10 msec – 30 sec adjustable | 10-990ms | Similar Comment 2 |
| Operation Mode | CW, single pulse, repeat pulse | CW, pulsed, single pulse | Continuous wave, single pulse, pulsed | Identical |
| Operation interface | Color LCD touch screen | Color touch screen | LCD touch screen | Identical |
| Repetition rate | 0.05Hz-50Hz | 0.02–50Hz | CW or up to 100Hz | Similar Comment 3 |
| Aiming beam | GBOX-10D/12D/15D: Diode laser of 650nm, power ≤3mW, adjustable brightness. Red, Class 1M, CW; | Red 635-650nm (<5mW) | 635nm | Substantial Equivalence |
11 / 17
{16}
| | VELAS II -15D: Diode laser of 650nm, power ≤2mW, adjustable brightness. | | | |
| --- | --- | --- | --- | --- |
| Laser Class | 4 | 4 | / | Identical |
| Power supply | 100-240 VAC, 50-60Hz, 200VA | 100-240V, 47-63 Hz | / | Substantial Equivalence |
## Discussion:
**Clinical:** Indication for Use claimed by Proposed device Medical Diode Laser Systems, model: GBOX-10D/12D is the same with the Predicate device (K222701). Indication for Use claimed by Proposed device Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D is the same with the Predicate device (K160952).
Indication for Use claimed by Proposed device has been covered by Predicate device (K222701) and (K160952) which depends on product performance, and this difference does not affect safety and effectiveness.
**Technology:** The subject device has the same operation principle and same wavelengths of diode laser to achieve its intended use. The operation mode, Laser Classification, Operation interface are either identical to the predicate. The main differences are Output power, Pulse width and Repetition rate.
**Comment 1 Output power:**
The Proposed device Medical Diode Laser Systems, model: GBOX-10D/12D has a power output range of 1-10W/1-12W, which is with the scope of what Predicate device (K222701): Max. 12W claims.
12 / 17
{17}
The Proposed device Medical Diode Laser Systems, model: GBOX-15D, VELAS II -15D has a power output range of 1-15W, which is with the scope of what Predicate device (K160952): Max. 15W claims.
Comment 3 Pulse width:
The pulse width of the subject device: 10ms~10s, somewhat different from the Predicate device (K160952), but within the scope of what Predicate device (K222701) claims.
Comment 3 Repetition rate:
Repetition rate of the subject device: 0.05Hz-50Hz is within the scope of what Predicate device claims. By Proposed device has been covered by Predicate device (K222701) and (K160952).
Material: The fiber in contact with the subject device has been approved in K140470.
**Conclusion:**
The subject device has the same intended use and operation principle, the technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices (K222701) and (K160952). The differences between the subject and predicate devices do not alter suitability of the proposed device for its intended use.
13 / 17
{18}
Table 3 Substantial equivalence discussion—VELAS Pro-30B15D
| Item | Proposed device | Predicate device (K240747) | Discussion |
| --- | --- | --- | --- |
| | Medical diode laser systems | Medical Diode Laser (M2-GK) | |
| Product Code | GEX | GEX | Identical |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Classification | Class II | Class II | Identical |
| Indication for Use | Medical Diode Laser Systems, model: VELAS Pro-30B15D is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. Medical Diode Laser Systems, | The Medical Diode Laser (Model: M2-GK) is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Medical Diode Laser (Model: M2-GK) is further indicated for laser assisted | Identical |
{19}
15 / 17
| | model: VELAS Pro-30B15D is further indicated for laser assisted lipolysis. | lipolysis. | |
| --- | --- | --- | --- |
| Laser type | diode laser | diode laser | Identical |
| Wavelength | 980nm; 1470nm | 980nm; 1470nm | Identical |
| Max Output power | 980nm: 30W
1470nm: 15W | 980nm: 30W
1470nm: 15W | Identical |
| Pulse width | 10ms-1s | 10ms-1s | Identical |
| Operation Mode | CW, single, repeat | CW, single pulse, repeat pulse | Identical |
| Aiming beam | Diode laser of 635nm power<5mW, and 520nm, power< 3.5mW; adjustable brightness. | Red 650nm (<2mW) | Substantial Equivalence |
| Cooling | Air | Air | Identical |
| Laser Class | 4 | 4 | Identical |
| Power supply | 100-240 VAC, 50-60Hz, 200VA | 100-240VAC, 50/60Hz, 4A-2A | Substantial Equivalence |
## Discussion:
**Technology**: The subject device has the same operation principle and same wavelengths of diode laser to achieve its intended use.
{20}
The operation mode, Laser Classification, Operation interface and main core technical specifications are either identical to the predicate.
Material: The fiber in contact with the subject device has been approved in K140470.
**Conclusion:**
The subject device has the same intended use and operation principle, the technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices (K240747). There are no differences between the subject and predicate devices.
16 / 17
{21}
17 / 17
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Medical Diode Laser Systems.
- Verify the conformity of the proposed devices with the requirements of 2005(R)2012 IEC60601-1 : (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
- Verify the conformity of the proposed devices with the requirements of 2020-09 IEC60601-1-2 : (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic compatibility).
- Verify the conformity of the proposed devices to 2007-03 IEC 60825-1 (Safety of laser products - Part 1: Equipment classification and requirements).
- Verify the performance of the proposed devices according to 2012-10 IEC 60601-2-22: (Medical electrical equipment Part 2: Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment).
- Verify the conformity of the proposed devices with the requirements of 2020-07 IEC 60601-1-6: (Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance Collateral standard: Usability)
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
## Mechanical and acoustic testing
It is not applicable.
## Animal Study
It is not applicable.
## Clinical Studies
It is not applicable.
# VIII. CONCLUSIONS
The Medical Diode Laser Systems is substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.