JADE SCORE PTA Scoring Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Device Story
JADE SCORE PTA Scoring Balloon Dilatation Catheter is a peripheral angioplasty device; available in 14 Rx (0.014-inch guidewire) and 18 OTW (0.018-inch guidewire) versions. Device features a scoring element designed to focus dilatation force on obstructive lesions in peripheral arteries and dialysis fistulae. Operated by physicians in clinical settings; catheter is inserted over a guidewire to the target lesion site; balloon is inflated to perform angioplasty. Scoring element enhances lesion dilation; device facilitates improved blood flow; benefits patients by treating obstructive vascular disease. Output is mechanical dilatation of the vessel.
Clinical Evidence
Bench testing only. No clinical data provided. In vitro performance testing conducted per FDA guidance 'Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters' (April 2023) included visual inspection, dimensional verification, flexibility/kink, shaft burst, torque strength, bond strength, leakage, corrosion resistance, radiopacity, simulated use (prep/deployment/retraction), inflation/deflation time, coating integrity, particulate evaluation, rated burst pressure, fatigue, and compliance. Packaging, sterilization, and shelf-life validation were also performed. All results met acceptance criteria.
Technological Characteristics
Catheter-based scoring balloon system. Features include scoring elements for focused force, hydrophilic coating, and compatibility with 5F-6F guiding catheters. Available in Rapid-exchange (Rx) and Over-The-Wire (OTW) configurations. Sterilization via Ethylene Oxide (EO). Materials and dimensions are optimized for peripheral vasculature. No software or electronic components.
Indications for Use
Indicated for Percutaneous Transluminal Angioplasty in peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra-popliteal, renal arteries) and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae in adults.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
JADE PLUS PTA Balloon Dilatation Catheter (K241025)
Submission Summary (Full Text)
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
September 16, 2026
OrbusNeich Medical (Shenzhen) Co., Ltd.
Dora Zhang
Manager of Regulatory Affairs
#1 Jinkui Rd.
Futian Free Trade Zone
Shenzhen, 518038
China
Re: K261778
Trade/Device Name: JADE SCORE PTA Scoring Balloon Dilatation Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: PNO
Dated: August 13, 2026
Received: August 13, 2026
Dear Dora Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261778 - Dora Zhang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261778 - Dora Zhang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
GREGORY W.
O'CONNELL -S
Digitally signed by GREGORY
W. O'CONNELL -S
Date: 2026.09.16 11:10:08
-04'00'
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261778 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| JADE SCORE PTA Scoring Balloon Dilatation Catheter | | | |
| Please provide your Indications for Use below. | | | ? |
| JADE SCORE PTA Scoring Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
## 510(k) Summary - K261778
This 510(k) Summary is submitted in accordance with 21 CFR 807.92.
Submitter: OrbusNeich Medical (Shenzhen) Co., Ltd.
No.1 Jinkui Road
Futian Free Trade Zone
Shenzhen 518038, China
Phone: +86-755-83580181-8326
Fax: +86-755-83580169
Contact Person: Name: Dora Zhang
Job Title: Manager of Regulatory Affairs
Email: dorazhang@orbusneich.com
Date Prepared: May 20, 2026
Device: Name of Device:
JADE Score PTA Scoring Balloon Dilatation Catheter
Classification Name:
Catheter, Percutaneous, Cutting/Scoring
Regulatory Number: 21 CFR 870.1250
Regulatory Class: II
Product Code: PNO
Predicate Device: UltraScore Focused Force PTA Balloon (K163420, PNO.
Cleared June 14, 2017)
This predicate has not been subject to a design-related recall.
Reference Devices: Scoreflex PTA Scoring Balloon Catheter (K182713, PNO,
cleared May 07, 2019)
JADE PLUS PTA Balloon Dilatation Catheter (K241025,
LIT, cleared June 11, 2024)
Device Description: The Peripheral Transluminal Angioplasty (PTA) Catheters
are designed for peripheral indications.
The JADE Score PTA Scoring Balloon Dilatation Catheter
contains two device versions: JADE Score PTA Scoring
Balloon Dilatation Catheter (14 Rx version) and JADE Score
PTA Scoring Balloon Dilatation Catheter (18 OTW version).
The JADE Score 14 Rx is designed with Rapid-exchange
(Rx) structure, permits the use of standard 0.014-inch
guidewire, while the JADE Score 18 OTW is designed with
510(k) Notification-K261778
JADE Score PTA Scoring Balloon Dilatation Catheter
Page 1 of 3
{5}
Over-The-Wire (OTW) structure, permits the use of standard 0.018-inch guidewire.
Indications For Use: JADE Score PTA Scoring Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Technological Characteristics: The subject device has the following similarities to the predicate devices:
- Same indications for use
- Similar catheter design
- Similar materials of construction
- Same hydrophilic coating
- 0.014” guidewire compatibility
- 5F-6F guiding catheter compatibility
- Same method of EO sterilization
- Same Scoring Element, element direction and element shape
- Same Mechanism of Action
- Similar performance specifications
The following technological differences exist between the subject and predicate device:
- Specific materials selected
- Exact dimensions of components
Performance Data: The following performance data were provided in support of the substantial equivalence determination.
➤ In vitro performance tests were conducted on subject device in accordance with FDA guidance “Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters” issued on April 14, 2023, including:
- Visual Inspection
- Dimensional Verification
- Flexibility and Kink
- Shaft burst
- Torque Strength
- Catheter Bond Strength
- Leakage Test
510(k) Notification -K261778
JADE Score PTA Scoring Balloon Dilatation Catheter
Page 2 of 3
{6}
- Corrosion Resistance
- Radiopacity
- Balloon Preparation, Deployment, and Retraction (Simulated Use)
- Balloon Inflation and Deflation Time
- Coating Integrity
- Particulate Evaluation
- Balloon Rated Burst Pressure
- Balloon Fatigue
- Balloon Compliance
➤ Packaging and sterilization validation
➤ Shelf Life
The test results met all acceptance criteria, which are the same or similar to the predicate device and ensure that the JADE Score PTA Scoring Balloon Dilatation Catheter design and construction are suitable for their intended use.
Conclusion:
This information supports a determination of substantial equivalence between the JADE Score PTA Scoring Balloon Dilatation Catheter and the predicate device described above.
510(k) Notification-K261778
JADE Score PTA Scoring Balloon Dilatation Catheter
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.