K163420 · Bard Peripheral Vascular, Inc. · PNO · Jun 14, 2017 · Cardiovascular
Device Facts
Record ID
K163420
Device Name
UltraScore Focused Force PTA Balloon
Applicant
Bard Peripheral Vascular, Inc.
Product Code
PNO · Cardiovascular
Decision Date
Jun 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraScore Focused Force PTA Balloon is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expanding stents, and stent grafts in the peripheral vasculature.
Device Story
Over-the-wire (OTW) PTA catheter with semi-compliant balloon; features two scoring wires oriented 180° apart to provide focused force during dilatation. Used by physicians in clinical settings to treat peripheral arterial stenoses and AV fistulae; also used for post-dilatation of stents/stent grafts. Input: manual inflation via luer lock hub. Transformation: balloon expansion with scoring wires concentrating force on lesion. Output: mechanical dilation of vessel/fistula. Benefits: improved vessel patency; reduced risk of vessel injury through focused force. Radiopaque markers aid placement; atraumatic tip facilitates navigation. Compatible with .014" or .035" guidewires.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via in vitro testing including dimensional analysis, balloon compliance, trackability, sheath/guidewire compatibility, inflation/deflation times, rated burst pressure, fatigue testing, and tensile/torque strength. Biocompatibility testing performed per ISO 10993-1:2009, including cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, and material-mediated pyrogenicity.
Technological Characteristics
OTW catheter with semi-compliant balloon; two scoring wires (180° apart). Materials: hydrophilic coating on distal portion (.014" version). Dimensions: 2.0-8.0mm diameter; 20-300mm length. Connectivity: N/A. Sterilization: Not specified. Software: None. Mechanical principle: Focused force balloon angioplasty.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae. Also indicated for post-dilatation of balloon-expandable stents, self-expanding stents, and stent grafts in peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2017
Bard Peripheral Vascular, Inc. Mr. Timothy Wade Regulatory Affairs Project Manager 1625 West 3rd St Tempe, AZ 85281
Re: K163420
Trade/Device Name: UltraScore Focused Force PTA Balloon Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: PNO Dated: May 4, 2017 Received: May 5, 2017
Dear Mr. Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163420
Device Name UltraScore Focused Force PTA Balloon
#### Indications for Use (Describe)
The UltraScore Focused Force PTA Balloon is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expanding stents, and stent grafts in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ULTRASCORE™ Focused Force PTA Balloon 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|--------------------------------------------------------------------------------|
| Phone: | 480 638 2939 |
| Fax: | 480 449 2546 |
| Contact: | Timothy Wade |
- Date 05 Dec 2016
## Subject Device Name:
| Device Trade Name: | ULTRASCORE™ Focused Force PTA Balloon |
|-----------------------|-----------------------------------------|
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | PNO |
| Classification: | Class II |
| Classification Name: | Catheter, Percutaneous, Cutting/Scoring |
| Classification Panel: | Cardiovascular |
### Predicate Device:
- . VASCUTRAK® PTA Dilatation Catheter (K103459; cleared December 13, 2010)
### Reference Device:
- ULTRAVERSE® 035 PTA Dilatation Catheter (K142261; cleared September 24, ● 2014)
Image /page/3/Picture/13 description: The image shows the word "BARD" in a stylized, bold, sans-serif font. The letters are interconnected, with the "A" having a unique triangular shape. The overall design is clean and modern, with a focus on the geometric forms of the letters.
{4}------------------------------------------------
## Device Description:
The ULTRASCORE ™ Focused Force PTA Balloon consists of a flexible, over-the-wire (OTW) catheter shaft with a semi-compliant balloon fixed at the distal end. For all balloon lengths, radiopaque markers delineate the working length of the balloon and aid in balloon placement. For balloon lengths of 100 mm and greater, two radiopaque markers are positioned on the distal portion of the balloon and one radiopaque marker is positioned on the proximal portion of the balloon to differentiate between the distal and proximal ends of the balloon. The catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the stenosis. Two scoring wires, oriented 180° apart, provide focused force upon dilatation. The ULTRASCORE ™ Focused Force PTA Balloon is compatible with .014" or .035" quidewires, as denoted by the product labeling. The distal portion of the .014" guidewire compatible catheters is hydrophilically coated. The proximal portion of the catheter includes a female luer lock hub connected to the catheter with a guidewire lumen and an inflation lumen. Packaged with every product is a protective sheath that is positioned over the balloon and must be removed prior to use. A stylet is placed into the tip of the catheter. These products are not made with natural rubber latex.
| Attribute | ULTRASCORE™ Focused Force PTA Balloon |
|---------------------------------------------------|----------------------------------------------|
| Balloon Diameter (mm) | 2.0, 2.5, 3.0, 3.5, 4.0, 5.0, 6.0, 7.0, 8.0. |
| Balloon Length (mm) | 20, 40, 60, 80, 100, 120, 150, 200, 250, 300 |
| Catheter Shaft Length (cm) | 40, 75, 100, 130, 150 |
| .014" Platform Introducer Sheath<br>Compatibility | 5F |
| .035" Platform Introducer Sheath<br>Compatibility | 6F |
{5}------------------------------------------------
### Indications for Use of Device:
The ULTRASCORE™ Focused Force PTA Balloon is intended to dilate stenoses in the illac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expandable stents, self-expanding stents, and stent grafts in the peripheral vasculature.
## Comparison of Indications for Use to Predicate Device:
The indications for use statement for the ULTRASCORE™ Focused Force PTA Balloon does not raise any new issues of safety and effectiveness based on the proposed indications for use statement as compared to the predicate device. Therefore, the subject device, the ULTRASCORE™ Focused Force PTA Balloon, is substantially equivalent to the predicate device.
## Technological Comparison to Predicate Devices:
The ULTRASCORE™ Focused Force PTA Balloon has the following similarities to the predicate device, the VASCUTRAK® PTA Dilatation Catheter (clearance to market via K103459 on December 13, 2010):
- . Same intended use
- . Same indications for use
- Same target population .
- Similar operating principle 0
- o Same fundamental scientific technology
- Same sterility assurance level and method of sterilization .
The ULTRASCORE™ Focused Force PTA Balloon has the following differences when compared to the VASCUTRAK® PTA Dilatation Catheter predicate device:
- Over-the-wire platform o
- Additional 8.0mm balloon diameter offered o
- New 0.035" quidewire compatibility offered o
- Modifications to marker bands o
- Material differences .
- Addition of packaging stylet o
{6}------------------------------------------------
### Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- . Dimensional
- Balloon Outer Diameter o
- Balloon Working Length o
- o Catheter Shaft Length
- Balloon Compliance ●
- GEOALIGN® Gradient Marking o Position
- GEOALIGN® Gradient Marking . Legibility
- GEOALIGN® Gradient Marking ● Durability
- Trackability .
- Sheath Compatibility ●
- Guidewire Compatibility ●
- Flushability ●
- Reinsertion ●
- Inflation Time ●
- Deflation Time o
- Rated Burst Pressure ●
- . Balloon Burst Mode
- . Balloon Burst in Stent/Stent Graft
- Balloon Fatique .
- Balloon Fatigue in Stent/Stent . Graft
- Balloon Removal from . Stent/Stent Graft
- Tip Taper .
- Device Compatibility ●
- Hub Stress ●
- Hub Torque ●
- Balloon to Shaft Tensile ●
- Tip to Wire Tensile ●
- Catheter Elongation ●
- Marker Band Radiopacity ●
- Marker Band Alignment ●
- Proximal Assembly Tensile .
- Packaging Pouch Tensile ●
- . Packaging Visual Inspection
- Packaging Bubble Emission ●
The following in vitro biocompatibility testing was conducted in accordance with ISO 10993-1: 2009
- Cytotoxicity
- Sensitization
- o Intracutaneous Reactivity
- Acute Systemic Toxicity 0
- Hemocompatibility
Image /page/6/Picture/44 description: The image shows the word "BARD" in a stylized, outlined font. The letters are bold and geometric, with the "A" resembling a triangle without a horizontal crossbar. The overall design is clean and modern, with a focus on the shape and form of the letters.
{7}------------------------------------------------
- Material Mediated Pyrogenicity
The results from these tests demonstrate that the technological characteristics and performance criteria of the ULTRASCORE™ Focused Force PTA Balloon are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
# Conclusions:
The subject device, the ULTRASCORE™ Focused Force PTA Balloon, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The ULTRASCORE™ Focused Force PTA Balloon is substantially equivalent to the legally marketed predicate device, VASCUTRAK® PTA Dilatation Catheter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.