ATTUNE Total Knee System; ATTUNE Cementless Knee System
Applicant
DePuy Ireland UC
Product Code
JWH · Orthopedic
Decision Date
Jun 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total knee arthroplasty is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee replacement may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely disabled patients with multiple joint involvement for whom a gain in knee mobility may lead to an expectation of significant improvement in the quality of their lives.
Device Story
Total knee replacement prosthesis consisting of femoral, tibial, and patellar components; replaces natural articular surface of knee joint. Subject of 510(k) is updated surgical technique and new accessory cases/trays for existing ATTUNE implants. No changes to implant design, manufacturing, or principle of operation. Used by orthopedic surgeons in clinical settings to restore mobility and reduce pain in patients with severe joint degeneration. Output is the physical implantation of the prosthesis to facilitate joint articulation. Benefits include improved patient mobility and quality of life.
Clinical Evidence
No clinical tests were conducted to demonstrate substantial equivalence. Evidence is limited to non-clinical bench testing, including simulated wear performance, computational wear analysis, cementless fixation strength, and bacterial endotoxin testing per ANSI/AAMI ST 72:2019.
Technological Characteristics
Total knee prosthesis comprising metal femoral components, metal/polyethylene tibial components, and polyethylene patellar components. Features AFFIXIUM 3DP technology for porous-coated implants. No changes to materials or energy sources. Sterilization and biocompatibility maintained per existing standards.
Indications for Use
Indicated for adults with severely painful and/or disabled joints resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or failed previous implants. Cementless components require adequate bone for seating and support.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K202194 - ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM 3DP Technology
Reference Devices
K233980 - Bundled Knee MRI - ATTUNE, LPS, Sigma HP Uni
K232303 - ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology
Submission Summary (Full Text)
{0}

June 25, 2026
DePuy Ireland UC
% Paige Myers
Senior Regulatory Affairs Specialist
Depuy Orthopaedics, Inc.
700 Orthopaedic Dr.
Warsaw, Indiana 46582
Re: K261458
Trade/Device Name: ATTUNE Total Knee System; ATTUNE Cementless Knee System
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Codes: JWH, MBH, OIY
Dated: May 1, 2026
Received: May 1, 2026
Dear Paige Myers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261458 - Paige Myers
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K261458 - Paige Myers
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G.
Allen -S
Digitally signed by Peter
G. Allen -S
Date: 2026.06.25 17:15:21
-04'00'
for Lixin Liu, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261458 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| ATTUNE Total Knee System ; ATTUNE Cementless Knee System | | | |
| Please provide your Indications for Use below. | | | ? |
| ATTUNE™ Total Knee System Indications for Use Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant. | | | |
| ATTUNE™ Cementless Knee System Indications for Use The ATTUNE™ Cementless CR and PS Femoral Components are intended for cementless use within the ATTUNE™ Total Knee Replacement System. Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present). | | | |
| The ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless use within the ATTUNE® Total Knee Replacement System. Porous coated implants may be used with or without cement. Candidates for total knee replacement include patients with a severely painful and/or impaired knee function resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present). | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
K261458 page 1 of 6

Bundled Traditional 510(k)
ATTUNE™ Total Knee System,
ATTUNE™ Cementless Knee System
# **510(K) SUMMARY**
(As required by 21 CFR 807.92 and 21 CFR 807.93)
| **Contact Information** | |
| --- | --- |
| **Applicant Name** | DePuy Ireland UC |
| **Applicant Address** | Loughbeg, Ringaskiddy Co. Cork Munster, IRELAND |
| **Applicant Establishment Registration Number** | 3015516266 |
| **Applicant Contact Telephone** | +1 574-367-4249 |
| **Applicant Contact Email** | pmyers1@its.jnj.com |
| **Applicant Contact** | Paige Myers |
| **Correspondent Name** | DePuy Orthopaedics, Inc. |
| **Correspondent Address** | 700 Orthopaedic Drive, Warsaw, IN 46582 USA |
| **Alt. Contact Telephone** | +1 925-890-1497 |
| **Alternate Contact Email** | Mliao7@its.jnj.com |
| **Alternate Contact Name** | Michael Liao |
| **Date prepared** | 1 May 2026 |
| **Name of device** | |
| **Trade or proprietary name** | ATTUNE™ Total Knee System |
| **Common or usual name** | Total Knee Replacement Prosthesis |
| **Classification name** | 21 CFR 888.3560 – Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis |
| **Class** | II |
| **Classification panel** | 87 Orthopedics |
| **Regulation** | Class II - 21 CFR 888.3560 |
| **Product Code(s)** | **JWH:** Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer **OIY:** Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive |
| **Legally marketed device(s) to which equivalence is claimed** | Primary Predicate: **K111433** - DePuy Attune™ Knee System Secondary Predicate: **K170806** – ATTUNE Cemented Tibial Base, Fixed Bearing |
{5}
K261458 page 2 of 6

**Bundled Traditional 510(k)**
**ATTUNE™ Total Knee System,**
**ATTUNE™ Cementless Knee System**
| | Reference Device: **K233980** – Bundled Knee MRI - ATTUNE, LPS, Sigma HP Uni |
| --- | --- |
| **Reason for 510(k) submission** | In accordance with Section 510(k) of the Medical Device Amendments of 1976 and Subpart E of Part 807, Title 21 of the Code of Federal Regulations, and as per the FDA Guidance, Bundling Multiple Devices or Multiple Indications in a Single Submission: Guidance for Industry and FDA Staff (June 2007), DePuy Ireland UC has compiled a Bundled Traditional 510(k) Premarket Notification to modify labeling to include an updated surgical technique for DePuy ATTUNE™ Total Knee System (K111433, K170806). Additionally, cases and trays that are associated with this new surgical technique are being introduced as accessories to the existing implant devices. |
| **Device description** | A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design. |
| **Intended use of the device** | Total knee arthroplasty is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee replacement may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely disabled patients with multiple joint involvement for whom a gain in knee mobility may lead to an expectation of significant improvement in the quality of their lives. |
| **Indications for use** | Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant. |
| **Substantial Equivalence** | There are no changes in design, manufacturing, principle of operation, indication, or intended use. The only change is the creation of an additional surgical technique for the existing implants and the addition of new case and trays that are now to be considered as device accessories. |
{6}
K261458 page 3 of 6

Bundled Traditional 510(k)
ATTUNE™ Total Knee System,
ATTUNE™ Cementless Knee System
# SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
# PERFORMANCE DATA
# SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The following tests were performed to address:
- Simulated Wear Performance
- Computation Wear Analysis
The proposed devices also meet the requirement of bacterial endotoxin testing as specified in ANSI/AAMI ST 72:2019.
# SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
# CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject device, DePuy ATTUNE Total Knee System, is substantially equivalent to the predicate ATTUNE Total Knee System (K11433, K170806).
# Contact Information
| **Applicant Name** | DePuy Ireland UC |
| --- | --- |
| **Applicant Address** | Loughbeg, Ringaskiddy Co. Cork Munster, IRELAND |
| **Applicant Establishment Registration Number** | 3015516266 |
| **Applicant Contact Telephone** | +1 574-367-4249 |
| **Applicant Contact Email** | pmyers1@its.jnj.com |
| **Applicant Contact** | Paige Myers |
| **Correspondent Name** | DePuy Orthopaedics, Inc. |
| **Correspondent Address** | 700 Orthopaedic Drive, Warsaw, IN 46582 USA |
| **Alt. Contact Telephone** | +1 925-890-1497 |
| **Alternate Contact Email** | Mliao7@its.jnj.com |
| **Alternate Contact Name** | Michael Liao |
| **Date prepared** | 1 May 2026 |
{7}
K261458 page 4 of 6

**Bundled Traditional 510(k)**^{}[] ATTUNE™ Total Knee System,^{}[] ^{}[] ATTUNE™ Cementless Knee System
| Name of device | |
| --- | --- |
| **Trade or proprietary name** | ATTUNE™ Cementless Knee System |
| **Common or usual name** | Total Knee Prosthesis |
| **Classification name** | 21 CFR 888.3565 - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis |
| **Class** | II |
| **Classification panel** | 87 Orthopedics |
| **Regulation** | Class II - 21 CFR 888.3565 |
| **Product Code(s)** | **MBH:** Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer **JWH:** Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer |
| **Legally marketed device(s) to which equivalence is claimed** | Primary Predicate: **K140881** - ATTUNE™ Knee System- Cementless CR and PS Femoral Components Secondary Predicate: **K202194** – ATTUNE™ Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology Reference Devices: **K233980** – Bundled Knee MRI - ATTUNE, LPS, Sigma HP Uni, **K232303** - ATTUNE™ Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology |
| **Reason for 510(k) submission** | In accordance with Section 510(k) of the Medical Device Amendments of 1976 and Subpart E of Part 807, Title 21 of the Code of Federal Regulations, and as per the FDA Guidance, Bundling Multiple Devices or Multiple Indications in a Single Submission: Guidance for Industry and FDA Staff (June 2007), DePuy Ireland UC has compiled a Bundled Traditional 510(k) Premarket Notification to modify labeling to include an updated surgical technique for DePuy, ATTUNETM Cementless Knee System (K140881). Additionally, cases and trays that are associated with this new surgical technique are being introduced as accessories to the existing implant devices |
| **Device description** | A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design. |
{8}
K261458 page 5 of 6

**Bundled Traditional 510(k)**
**ATTUNE™ Total Knee System,**
**ATTUNE™ Cementless Knee System**
| **Intended use of the device** | Total knee arthroplasty is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee replacement may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely crippled disabled patients with multiple joint involvement for whom a gain in knee mobility may lead to an expectation of significant improvement in the quality of their lives. |
| --- | --- |
| **Indications for use** | The ATTUNE™ Cementless CR and PS Femoral Components are intended for cementless use within the ATTUNE™ Total Knee Replacement System. Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present). The ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless use within the ATTUNE® Total Knee Replacement System. Porous coated implants may be used with or without cement. Candidates for total knee replacement include patients with a severely painful and/or impaired knee function resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present). |
| **Substantial Equivalence** | There are no changes in design, manufacturing, principle of operation, indication, or intended use. The only change is the creation of an additional surgical technique for the existing implants and the addition of new case and trays that are now to be considered as device accessories. |
| **SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE** |
| --- |
| **PERFORMANCE DATA** |
| **SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE** |
| The following tests were performed: - Computational Wear Analysis - Cemented Fixation Strength |
{9}
K261458 page 6 of 6
DePuy Synthes
Bundled Traditional 510(k)
ATTUNE™ Total Knee System,
ATTUNE™ Cementless Knee System
The proposed devices also meet the requirement of bacterial endotoxin testing as specified in ANSI/AAMI ST 72:2019.
# SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
# CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject DePuy ATTUNE Cementless Knee System is substantially equivalent to the predicate ATTUNE Knee System – Cementless (K140881) and the ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology (K202194).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.