K111433 · DePuy Orthopaedics, Inc. · OIY · Aug 30, 2011 · Orthopedic
Device Facts
Record ID
K111433
Device Name
DEPUY STTUNE (TM) PS KNEE SYSTEM
Applicant
DePuy Orthopaedics, Inc.
Product Code
OIY · Orthopedic
Decision Date
Aug 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total knee replacement is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. The Attune PS Total Knee is intended to accommodate knee flexion to 145 degrees in those patients able to attain a high degree of knee flexion.
Device Story
DePuy Attune PS Knee System is a total knee replacement prosthesis consisting of posterior stabilized (PS) femoral components and PS fixed bearing tibial inserts. Femoral components feature asymmetric trochlear grooves, textured fixation surfaces, and two lugs for cement fixation. Tibial inserts utilize multi-radii topography to balance anterior constraint and posterior rollback. Components are implanted by orthopedic surgeons in a clinical setting to replace damaged knee joints, aiming to restore mobility and reduce pain. The system accommodates up to 145 degrees of knee flexion. The device is intended for cemented use.
Clinical Evidence
No clinical data provided; substantial equivalence determined via bench testing in compliance with FDA guidance for knee joint prostheses.
Indicated for cemented total knee replacement in patients with severely painful and/or disabled joints due to osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or failed previous implants.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
K111M33 #'/4
AUG 3 0 2011
### Summary of Safety and Effectiveness Attune Total Knee System – PS Femoral Components and PS Fixed Bearing Inserts DePuy Orthopaedics
ﻟﻤﺴﺎﻋ
| Submitted by: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, IN 46581<br>Phone: (574) 371-4923<br>Fax: (574) 371-4987 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Nancy Friddle, Project Manager Regulatory Affairs |
| Date Prepared: | May 16, 2011 |
| Proprietary Name: | DePuy Attune™ Knee System |
| Common Name: | Total Knee Replacement Prosthesis |
| Classification Name: | 21 CFR 888.3560 Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented prosthesis.<br>Class II |
| Product Code: | JWH<br>OIY |
| Predicate Devices: | The DePuy Attune PS Femoral Components and PS Fixed Bearing<br>Inserts, which are part of the DePuy Attune Knee System, are<br>substantially equivalent to currently marketed devices including:<br>• Attune Knee System, K101433<br>• Sigma Tibial Inserts, K033272<br>• Sigma XLK Tibial Inserts, K040166<br>• Sigma PS Femoral Components, K073529<br>• Zimmer NexGen LPS Femoral Components / Articular<br>Surfaces, K960279 |
| Device Description: | The Attune™ posterior stabilized (PS) femoral components have an<br>asymmetric trochlear groove and are available in sizes 1-10 in right<br>and left options. Sizes 3-6 are available in standard and narrow<br>options. The fixation surface is textured. It incorporates two lugs<br>(pegs) to provide additional stability and recessed cement pockets for<br>enhanced cement fixation. The Attune femoral components are<br>manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. |
| | The Attune PS fixed bearing tibial insert is designed with a<br>topography that, in the sagittal view, consists of multiple radii. The<br>sagittal anterior radii are relatively conforming to the femoral<br>component to aid in anterior constraint while the posterior radius is<br>less conforming to allow rollback. The Attune PS fixed bearing tibial<br>inserts are available in sizes 1-10 and in thicknesses of 5, 6, 7, 8, 10,<br>12, 14, 16, 18, 20, and 22mm. The inserts are manufactured from<br>AOX ultra high molecular weight polyethylene conforming to ASTM<br>F648. |
| | The Attune PS femoral components and PS fixed bearing inserts are<br>compatible with the Attune FB tibial bases (K101433) and patellae<br>(K103756). |
| | The Attune PS Total Knee is designed to accommodate knee flexion<br>to 145 degrees in those patients able to attain a high degree of knee<br>flexion. |
| Intended Use: | Total knee replacement is intended to provide increased patient<br>mobility and reduced pain by replacing the damaged knee joint<br>articulation in patients where there is evidence of sufficient sound<br>bone to seat and support the components. |
| | The Attune PS Total Knee is intended to accommodate knee flexion<br>to 145 degrees in those patients able to attain a high degree of knee<br>flexion. |
| Indications for Use: | As part of the DePuy Attune Knee System, the DePuy Attune PS<br>femoral components and tibial inserts are intended for cemented use<br>as a total knee replacement system.<br><br>Candidates for total knee replacement include patients with a severely<br>painful and/or severely disabled joint resulting from osteoarthritis,<br>post-traumatic arthritis, rheumatoid arthritis, or a failed previous<br>implant. |
| Technological<br>Characteristics: | As shown in the following tables, the technological characteristics of<br>the Attune PS femoral components and tibial inserts are similar to<br>the predicate devices including design and material. |
・
{1}------------------------------------------------
K1114.33 *2/4
| Characteristic | Attune PS Femoral<br>Component | Attune CR Femoral<br>Component (K101433) |
|----------------|-----------------------------------------------|-----------------------------------------------|
| Material | Cast Co-Cr-Mo alloy<br>conforming to ASTM F75 | Cast Co-Cr-Mo alloy<br>conforming to ASTM F75 |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K111433 *3/4". The text is written in a handwritten style. The numbers are written in a clear and legible manner.
| Characteristic | Attune PS Femoral<br>Component | Attune CR Femoral<br>Component (K101433) |
|---------------------|-------------------------------------------|-------------------------------------------|
| Sizes | Sizes 1 to 10 standard, Left<br>and Right | Sizes 1 to 10 standard, Left<br>and Right |
| | Sizes 3-6 narrow, Left and<br>Right | Sizes 3-6 narrow, Left and<br>Right |
| | Proportional Sizing | Proportional Sizing |
| Fixation<br>Surface | Cemented | Cemented |
| Characteristic | Attune PS Fixed Bearing<br>Tibial Inserts | Attune CR Fixed Bearing<br>Tibial Inserts (K101433) and<br>Sigma Tibial Inserts<br>(K033272) |
|----------------|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | AOX UHMWPE conforming<br>to ASTM F648 | Attune CR: AOX UHMWPE<br>conforming to ASTM F648 |
| Sizes | Proportional sizing<br>PS inserts, sizes 1 to 10<br>Actual thicknesses 5, 6, 7, 8,<br>10, 12, 14, 16, 18, 20, and<br>22mm. | Attune CR:<br>Proportional sizing<br>CR inserts, sizes 1 to 10<br>Actual thicknesses 5, 6, 7, 8,<br>10, 12, 14, 16, and 18mm.<br><br>Sigma Stabilized:<br>Constant sizing<br>Stabilized inserts, sizes 1.5 to 6<br>Composite thicknesses 8, 10,<br>12.5, 15, 17.5, 20, 22.5 and<br>25mm. |
Summary of Substantial Equivalence:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The DePuy Attune Total Knee System is substantially equivalent to currently marketed devices as demonstrated with preclinical data.
{3}------------------------------------------------
K111433*4/4
### Non-Clinical Testing:
Functional testing was conducted in compliance with FDA guidance, Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses: Guidance for Industry and FDA, to verify that the implant performance would be substantially equivalent to predicate devices for anticipated in vivo loading via various constraint, contact, wear, and fatigue tests.
#### Clinical Testing:
None provided as it was not necessary to determine substantial equivalence between the Attune Knee System's PS femoral and PS fixed bearing insert components and the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DePuy Orthopaedics, Inc. % Ms. Nancy Friddle 700 Orthopaedic Drive Warsaw, Indiana 46581
AUG 3 0 2011
Re: K111433 Trade/Device Name: DePuy Attune™ PS Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: OIY, JWH Dated: August 10, 2011 Received: August 11, 2011
Dear Ms. Friddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Ms. Nancy Friddle
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
urs,
For Pati
Der
Cli
B.N
Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 2. INDICATIONS FOR USE
KU11433 510(k) Number (if known): _
Device Name: DePuy Attune™ PS Knee System
Indications for Use:
The DePuy Attune™ Knee System is intended for cemented use as a total knee replacement system.
Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteo.arthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
for M.Mellerson
(Division Sign-Oft) (Division Sign-on) )
Division of Surgical, Orthopedic, Division of Surgices
510(k) Number K111433.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.