XPRESEN LED Light Therapy Masks are Over the Counter devices that are intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne. Red light: Treatment of full-face wrinkles. Yellow light: Repair and reduce wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; and to temporarily increase local blood circulation. Blue light: Treatment of mild to moderate inflammatory acne. Red+Blue light: Treatment of mild to moderate inflammatory acne.
Device Story
Wearable LED mask; includes mask body, controller, straps, Type-C USB charging cable. Powered by rechargeable lithium battery. Emits narrow-band light at four wavelengths (450nm blue, 580nm yellow, 630nm red, 830nm infrared) and five modes (Blue, Yellow, Red, Infrared, Red+Blue). Used in home environment by patient. Controller manages operation, mode selection, and treatment duration (10, 20, 30 minutes). Provides phototherapy for wrinkle reduction and acne treatment; infrared mode provides topical heating to increase local blood circulation. Output is light energy delivered to facial/neck skin. Benefits include non-invasive aesthetic improvement and temporary relief of local circulation issues.
Indicated for treatment of full-face wrinkles and mild to moderate inflammatory acne in adults using over-the-counter home-use LED light therapy masks.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
July 15, 2026
Dongguan Miqiang Technology Co., Ltd.
% Charles (Chengyu) Shen
Vice President
Manton Business and Technology Services
37 Winding Ridge
Oakland, New Jersey 07436
Re: K261407
Trade/Device Name: XCREEN LED Light Therapy Masks (XPRE322, XPRE326, XPRE327,
XPRE342, XPRE368, XPRE422, XPRE434, XPRE451, XPRE455, XPRE457,
XPRE448, XPRE335, XPRE350, XPRE420, XPRE436, XPRE339, XPRE445)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In
Dermatology
Regulatory Class: Class II
Product Code: OLP, OHS
Dated: April 29, 2026
Received: April 29, 2026
Dear Charles (Chengyu) Shen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above
and have determined the device is substantially equivalent (for the indications for use stated in the enclosure)
to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment
date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the
provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket
approval application (PMA). You may, therefore, market the device, subject to the general controls
provisions of the Act. Although this letter refers to your product as a device, please be aware that some
cleared products may instead be combination products. The 510(k) Premarket Notification Database
available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination
product submissions. The general controls provisions of the Act include requirements for annual registration,
listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We
remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261407 - Charles (Chengyu) Shen
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K261407 - Charles (Chengyu) Shen
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.07.15 15:28:06 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261407
Device Name
XPRESEN LED Light Therapy Masks
Indications for Use (Describe)
XPRESEN LED Light Therapy Masks are Over the Counter devices that are intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne.
Red light: Treatment of full-face wrinkles.
Yellow light: Repair and reduce wrinkles.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; and to temporarily increase local blood circulation.
Blue light: Treatment of mild to moderate inflammatory acne.
Red+Blue light: Treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) Summary: K261407
## 1.0 Submitter Information
Dongguan Miqiang Technology Co., Ltd
No. 6, Gongye 1st Road, Hengjiangsha Village, Changping Town, Dongguan City,
Guangdong Contact Person: Baotan ZHU
Tel.: +8615362851298
Email: 1069077816@qq.com
Submitter's FDA Registration Number: N/A
Submission Correspondent

Primary Contact:
Charles Shen (Ph. D)
Manton Business and Technology Services
37 Winding Ridge, Oakland, NJ 07436
Tel: 1-608-217-9358
Email: cyshen@aol.com
Secondary Contact:
Jinfeng Ning (Ph. D)
Manton Business and Technology Services
37 Winding Ridge, Oakland, NJ 07436
Tel: 1-504-256-3616
Email: jfning@gmail.com
Date of Summary: June 30, 2026
## 2.0 Device Information
| Proprietary Name: | XCREEN LED Light Therapy Masks |
| --- | --- |
| Common Name: | LED Light Therapy Masks |
| Classification Name: | Light Based Over the Counter Wrinkle Reduction |
| Device Classification: | II |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | OHS, OLP |
| Panel: | General & Plastic Surgery |
K261407
1 / 6
{5}
### 3.0 Predicate Device Information:
| | 510(K) # | Device Name | Manufacturer |
| --- | --- | --- | --- |
| Predicate Device | K253712 | Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073) | Shenzhen Rainbow Technology Co., Ltd. |
| Reference Device 1 | K223544 | LED light therapy mask (FM-01, FM-02, FM-03) | Guangdong Newdermo Biotech Co., Ltd |
| Reference Device 2 | K241293 | LED Light Therapy Device, ELIXIR MD | Yassen Wellness LLC |
### 4.0 Device Description:
The “XCREEN LED Light Therapy Masks” are home use wearable LED phototherapy devices whose purpose are to produce an even and narrow-band of light for the treatment of aesthetic indications including facial wrinkles and acnes.
The subject device consists of a mask body unit that contains light emitting diodes (LEDs), a controller, straps, and Type-C USB charging cable.
The controller is connected to the mask to control the device operation, such as turning on/off the LED lamps, switching between different LED modes. The strap is used for securing the mask to the face and neck.
The controller is powered by built-in rechargeable lithium battery. The LED generates lights with four different wavelengths and five different therapy modes:
1. Blue Light at 450 nm
2. Yellow Light at 580 nm
3. Red Lights at 630 nm
4. Infrared Light at 830 nm
5. Red and Blue Light at 450/630 nm
Blue light treatment is not available in the neck models.
Each of the “XCREEN LED Light Therapy Masks” provides three treatment options: 10 minutes, 20minutes, and 30minutes.
### 5.0 Indications for Use:
The “XCREEN LED Light Therapy Masks” are over the counter devices that are intended for use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne.
K261407
2 / 6
{6}
Red light: Treatment of full-face wrinkles.
Yellow light: Repair and reduce wrinkles.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; and to temporarily increase local blood circulation.
Blue light: Treatment of mild to moderate inflammatory acne.
Red+Blue light: Treatment of mild to moderate inflammatory acne.
### 6.0 Comparison to Predicate Devices
The “XCREEN LED Light Therapy Masks”, marketed by “Dongguan Miqiang Technology Co., Ltd.” are compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance.
Predicate Device
(1) K253712, “LED Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073)”, manufactured by “Shenzhen Rainbow Technology Co., Ltd.”
Reference Device
(1) K223544, “LED light therapy mask (FM-01, FM-02, FM-03)” manufactured by “Guangdong Newdermo Biotech Co., Ltd.”
(2) K241293, “LED Light Therapy Device, ELIXIR MD”, manufactured by “Yassen Wellness LLC”
The following table shows similarities and differences of use, design, and material between our devices and the predicate device.
K261407
3 / 6
{7}
Table 5.1: Comparison of Intended Use, Design, and Material
| Comparison Items | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 |
| --- | --- | --- | --- | --- |
| 510(k) number | | K253712 | K223544 | K241293 |
| Trade Name | XCREEN LED Light Therapy Masks | LED Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073) | LED Light Therapy Mask | LED Light Therapy Device, ELIXIR MD |
| Manufacturer | Dongguan Miqiang Technology Co., Ltd. | Shenzhen Rainbow Technology Co., Ltd. | Guangdong Newdermo Biotech Co., Ltd. | Yassen Wellness LLC |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 21 CFR 890.5500 | 21 CFR 878.4810 |
| Product code | OHS, OLP | OHS, OLP, ILY | OHS, OLP, ILY | GEX |
| Device classification | Class II | Class II | Class II | Class II |
| Indication for Use/Intended Use | XCREEN LED Light Therapy Masks are Over the Counter devices that are intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne. Red light: Treatment of full-face wrinkles. Yellow light: Repair and reduce wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; and to temporarily increase local blood circulation. Blue light: Treatment of mild to moderate inflammatory acne. Red+Blue light: Treatment of mild to moderate inflammatory acne. | The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment. Red+Infrared light: Treatment of full-face wrinkles. Red light: Treatment of full-face wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Blue light: Treatment of mild to moderate inflammatory acne. Red+Blue light: Treatment of mild to moderate inflammatory acne. | Red: Treatment of full-face wrinkles. Blue: Treatment of mild to moderate inflammatory acne. Infrared: Topical heating for elevating tissue temperature; arthritis and muscle spasm; etc; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne. | ELIXIR MDTM use of the red, blue, yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions. The red light (633±10nm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions. The blue light (417±10 nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. The yellow light (590±10nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital |
K261407
4 / 6
{8}
| | | | | wrinkles and rhytides. The infrared light (835±15nm wavelength) is generally use for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied. |
| --- | --- | --- | --- | --- |
| Prescription or OTC | OTC | OTC | OTC | Prescription |
| Location for use | Face and body (neck) | Face | Face and body | Face |
| Power supply | Lithium battery: 3.7V, 1200-5200 mAh (4.44 – 19.24 Wh) | Lithium battery: 5V, 1A | Input: 100-240 V~, 50/60 Hz, 0,25 A Output: DC 5 V, 500 mA | AC 100-240 50/60Hz |
| Light source | LED | LED | LED | LED |
| Wavelength | 450nm (Blue), 580nm (Yellow), 630nm (Red), 830nm (Infrared) | 465nm (Blue), 620nm (Red), 850nm (Infrared) | 460 (Blue), 620nm (Red), 850nm (Infrared) Mixed: 620 nm and 850nm and 460nm | 417±10 nm (Blue), 633 ±10 nm (Red), 590±10 nm (Yellow), and 835±15 nm (Infrared) |
| LED Intensity | Red 14.43 mw/cm² Blue: 33.90 mw/cm² Yellow: 2.55 mw/cm² Infrared: 2.70 mw/cm² Red+Blue: 24.17 mw/cm² | Red: 18-26 mw/cm² Blue: 31.5-45.5 mw/cm² Infrared: 3.6-5.2 mw/cm² Red+Blue: 36-52 mw/cm² | Red light: 2.0~3.0 mW/cm² Blue light: 2.0~4.0 mW/cm² Infrared light: 2.0~4.0 mW/cm² Mixed light: 9.0~12.0 mW/cm² | Red light: 20~96 mw/cm² Blue light: 10~120 mw/cm² Yellow light: 5~35 mw/cm² Infrared: ≤ 70 mw/cm² Red/IR: 166mW/cm² Blue/IR: 190mW/cm² |
The Indications for Use, working mechanisms, design and technological characteristics of the subject device is similar to the predicate device. The minor differences between the subject device and predicate device do not raise any concerns in terms of safety and effectiveness.
K261407
5 / 6
{9}
## 7.0 Non-Clinical Study Summary
Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
1) IEC 60601-1: 2005/AMD1: 2012/AMD2: 2020, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
2) IEC 60601-1-2: 2014/ AMD 2020: Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances –Requirements and tests
3) IEC TS 60601-4-2:2024: Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
4) IEC 60601- 1- 11:2015+A1:2020: Medical Electrical Equipment– Part 1- 11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
5) IEC 60601-2-83:2019: Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
6) IEC 62471:2006: Photobiological safety of lamps and lamp systems
7) IEC 62133-2: 2017/AMD 2021: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
## 8.0 Clinical Study Summary
Clinical study is not performed for this product.
## 9.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate is that the subject device is as safe, as effective, and performs as well as the legally marketed device (K253712).
K261407
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.