K261204 · Sanhe LEFIS Electronics Co., Ltd. · ONF · Jun 12, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261204
Device Name
Intense Pulsed Light Therapy Device
Applicant
Sanhe LEFIS Electronics Co., Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides and vascular lesions.
Device Story
Intense Pulsed Light (IPL) therapy device; uses high-voltage discharge to trigger avalanche ionization of gas in light source; converts electrical energy into high-intensity light pulses. Filters remove undesired spectrums to produce multi-band light output for specific dermatological treatments. System comprises main unit, handpiece, filter, and footswitch. Used in healthcare facilities/hospitals by trained professionals. Device delivers light energy to skin via sapphire treatment head; cooling provided by water, forced-air, copper, and TEC. Output parameters (wavelength, fluence, pulse sequence) adjusted based on target condition (hair, pigment, acne, vascular). Clinical benefit derived from thermal effect on target tissues. Provider monitors output to ensure safe delivery within specified energy ranges.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with ANSI/AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-57. Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-23. Software validation conducted per FDA guidance.
Indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides, and vascular lesions in patients requiring dermatological light-based therapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
Sanhe LEFIS Electronics Co., Ltd.
Dandan Wang
Registered Manager
Bldg. 11 #1-101, Phase 1, Zhongnan High Tech Yanjiao Science
Technology Innovation Smart Valley Industrial Park 1
Langfang, Hebei 065201
China
Re: K261204
Trade/Device Name: Intense Pulsed Light Therapy Device
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: ONF
Dated: March 18, 2026
Received: April 13, 2026
Dear Dandan Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261204 - Dandan Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K261204 - Dandan Wang
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.06.12
15:24:41 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261204
Device Name
Intense Pulsed Light Therapy Device
Indications for Use (Describe)
It is indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides and vascular lesions.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
## 510(k) Summary K261204
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
1. Date of Submission: 2026/3/18
2. Sponsor Identification
**Sanhe LEFIS Electronics Co., Ltd.**
Building 11 #1-101, Phase 1, Zhongnan High tech· Yanjiao Science and Technology Innovation Smart Valley Industrial Park 1, South Side of Liushan Street and West Side of Gushan West Road, Yanjiao High tech Zone, Sanhe City, 065201 Langfang City, Hebei Province, China
Contact Person: Dandan Wang
Position: Registered Manager
Tel: +86-18131561031
Email: 597782121@qq.com
3. Designated Submission Correspondent
Ms. Dandan Wang
Tel: +86-18131561031
Fax: +86-316-3096027
Email: 597782121@qq.com
4. Identification of Proposed Device
Trade Name: Intense Pulsed Light Therapy Device
Classification Name: Powered Light Based Non-Light Surgical Instrument With Thermal Effect
Regulatory Information
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Classification: II
Product Code: ONF
Regulation Number: 21CFR 878.4810
Review Panel: General & Plastic Surgery
5. Indication For Use Statement:
1
{5}
It is indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides and vascular lesions.
# 6. Environment of Use
healthcare facility/hospital
# 7. Device Description
The working principle of Intense Pulsed Light Therapy Device (IPL) is that the trigger applies a high voltage to a specific substance in the light source to trigger its ionization. After the energy storage capacitor is charged for a relatively long time, it discharges in an extremely short time, causing the avalanche ionization of the specific substance in the light source. This specific substance converts and releases the charged electrical energy in the form of high-intensity light radiation, and this discharging process is a light pulse. A filter can be used to remove the undesired emission spectrum, forming a multi-band light output to achieve different intended purposes. The device comprises a main unit, handpiece, filter, and footswitch.
# 8. Materials
| Components | Material | Category | ContactLevel | ContactDuration |
| --- | --- | --- | --- | --- |
| Treatment Head | Crystal (Sapphire) | Surface device | Intact skin | Short-term (<24h) |
| Lens hood | chrome-plated copper | Surface device | Intact skin | Short-term (<24h) |
The treatment hand piece used in the system has passed the Biocompatibility test. For details, please refer to 'Biocompatibility Discussion'.
# 9. Identification of Predicate Device(s)
510(k) Number: K122995
Product Name: Intense Pulsed Light (IPL) Systems
Manufacturer: Beijing KES Biology Technology Co., Ltd.
Reference Device : K200746
Product Name: IPL Treatment Systems
Manufacturer: Shanghai Apolo Medical Technology Co., Ltd.
2
{6}
# 10. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ANSI/AAMI ES60601-1:2005, ES60601-1:2005/AMD1:2012, ES60601-1:2005/AMD2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.1 2020-09, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-57 Edition 1.0 2011-01 Medical electrical equipment Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
The body-contacting components of this device are the treatment head. The biocompatibility evaluation for Intense Pulsed Light Therapy Device was conducted in accordance with the FDA guidance “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process’”. The treatment head is considered skin and subcutaneous tissue contacting for a duration of less than 24 hours. The biocompatible testing included In Vitro Cytotoxicity, Skin Sensitization and Intracutaneous Reactivity was conducted in compliance with:
- ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- ISO 10993- 10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO 10993-23:2021 Biological evaluation of medical devices — Part 23: Tests for irritation.
We have also conducted:
Software verification and validation test according to the requirements of the FDA “Guidance for Pre Market Submissions and for Software Contained in Medical Devices”. The software for this device was considered as a Basic Documentation level. Software validation demonstrated that the software functions as specified in the software requirement specifications.
Bench performance testing to show that the device delivers set energy parameters within specifications.
# Sterilization and Shelf-Life
The proposed device is not provided sterile and does not need to be sterilized. The handpiece and the body are cleaned with a soft cloth moistened with isopropyl alcohol or ethanol of 75% strength or higher. The proposed device is reusable and does not have a restricted shelf-life.
# 11. Technological characteristics and substantial equivalence:
3
{7}
Table 1 General Comparison
| ITEM | Proposed Device | Predicate Device | Reference Device 1 | Remark |
| --- | --- | --- | --- | --- |
| Device Name | Intense Pulsed Light Therapy Device | Intense Pulsed Light (IPL) Systems | IPL Treatment Systems | / |
| Manufacturer | Sanhe LEFIS Electronics Co., Ltd. | Beijing KES Biology Technology Co., Ltd. | Shanghai Apolo Medical Technology Co., Ltd. | / |
| Product Code | ONF | ONF | ONF | Same |
| Class | Class II | Class II | Class II | Same |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
4
{8}
| Intended Use | It is indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides and vascular lesions. | The Intense Pulsed Light (IPL) Systems (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides (freckles), and vascular lesions. | The IPL treatment system is intended for medical use in the treatment of the following dermatologic conditions: - Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen; - Moderate inflammatory acne vulgaris; - Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); - Cutaneous lesions including scars; - Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations. | Analysis 1 |
| --- | --- | --- | --- | --- |
| Indication for use | Prescription Use | Prescription Use | Prescription Use | Same |
5
{9}
Table 2 Performance Comparison
| ITEM | ProposedDevice | PredicateDevice | ReferenceDevice1 | Remark |
| --- | --- | --- | --- | --- |
| Light source | Intense pulsed light | Intense pulsed light | Intense pulsed light | Same |
| Wavelength | 420nm-1200nm, 515nm-1200nm, 560nm-1200nm, 590nm-1200nm, 640nm-1200nm, 695nm-1200nm | 430-1200nm, 530-1200nm, 640-1200nm Optional: 480-1200nm, 560-1200nm, 590-1200nm, 690-1200nm, 750-1200nm | 420 – 1200 nm | Analysis 1 |
| Deliver system | Sapphire | Sapphire | Sapphire | Same |
| Energy output | 4J/cm2- 32J/cm2 | 10-60J/cm² | 4.1-50.8 J/cm2 | Analysis 2 |
| Pulse sequence | 2 to 10. | 1-15 pulses | / | Analysis 2 |
| Pulse width | 0.5ms/1ms | 1-20ms | 5-20 ms | Analysis 3 |
| Pulse delay | 5ms to 50ms | 5-50ms | 5-50 ms | Same |
| Spot Size | 30X12mm. | MED-210: 15mm×50mm; (optional) 12mm×33mm, 15mm×35mm MED-230: A : 12mm×33mm B : 15mm×50mm (optional) 15mm×35mm | 12*35mm, 15*50mm | Analysis 4 |
| Filters | 420nm-1200nm/560nm-1200nm: Acne vulgaris; 515nm-1200nm/560nm-1200nm: Pigmented lesions; 515nm-1200nm/560nm-1200nmEphelides; 560nm-1200nm/590nm-1200nm: vascular lesions; 640nm-1200nm/695nm-1200nm: Hair removal; | moderate inflammatory acne vulgaris: 430-1200/480-1200/530-1200 pigmented lesions: 480-1200/530-1200/560-1200 ephelides (freckles): 480-1200/530-1200 vascular lesions: 530-1200/560-1200/590-1200, | 420 -1200nm: Acne; 510 -1200nm: Acne, vascular, pigment; 560 -1200nm: Acne, vascular, pigment; | Same |
6
{10}
| ITEM | ProposedDevice | PredicateDevice | ReferenceDevice1 | Remark |
| --- | --- | --- | --- | --- |
| | | permanent hair reduction: 640-1200/690-1200/750-1200 | 610-1200nm: Hair removal; 640-1200nm: Hair removal; 690-1200nm: Hair removal; | |
| Fluences | 420nm-1200nm: 4-32J/cm2 515nm-1200nm: 4-32J/cm2 560nm-1200nm: 4-30J/cm2 590nm-1200nm: 4-28J/cm2 640nm-1200nm: 4-26J/cm2 695nm-1200nm: 4-20J/cm2 | 430-1200/480-1200/530-1200 : 10-40 J/cm2 480-1200/530-1200/560-1200 : 12-44 J/cm2 530-1200/560-1200/590-1200: 10-42J/cm2 640-1200/690-1200/ 750-1200 : 10-44 J/cm2 | 420 -1200nm: 4.1-50.8J/cm2; 510 -1200nm: 3.8-47 J/cm2; 560 -1200nm: 3.7-43.3 J/cm2; 610-1200nm: 3.5-38.7 J/cm2; 640-1200nm: 3.3-37.4J/cm2; 690-1200nm: 3.1-33.4J/cm2; | Analysis 2 |
| Deliver materials | Direct sapphire Coupling | / | Direct sapphire Coupling | Same |
| Cooling method | Water cooling, forced-air cooling, copper and TEC | / | HS-650K&HS-660K: Water cooling, forced-air cooling, copper and TEC; HS620K, HS-300CK&HS-310K: Water cooling and forced-air cooling, | Same |
| Power Supply | 110V 60Hz or230V 50/60Hz | 220V±20V 50Hz or 110V±20V 60Hz | / | Similar |
| Max. power consumption | 3300VA | MED-210: 1400 W MED-230: 2000 W | / | Analysis 5 |
7
{11}
Table 3 Safety Comparison
| ITEM | ProposedDevice | PredicateDevice | ReferenceDevice | Remark |
| --- | --- | --- | --- | --- |
| Biocompatibility | Passed the test as per ISO 10993-5, ISO 10993-10, ISO10993-23 | / | Passed the test as per ISO 10993-1 | Same |
| Patient Contact Sites | Skin | Skin | Skin | Same |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Safety | Comply with IEC 60601-2-57 | / | Comply with IEC 60601-2-57 | Same |
### Analysis 1 Wavelength
The wavelength difference between the proposed device and the predicate device is only minor deviations in the starting wavelength, while the end wavelength (1200nm) is completely consistent. The characteristic absorption spectra of the core target tissues (pigment, blood vessels, hair follicles) are fully covered, and the device complies with relevant standards. This difference does not alter the treatment principle or performance, and thus does not affect the safety and effectiveness of the product.
### Analysis 2 Energy output & Pulse sequence
The differences in energy output and pulse sequence are both differences in the breadth of parameter ranges. The effective parameter ranges required for core treatment completely overlap, and they comply with relevant standards, so they do not affect the safety and effectiveness of the product.
### Analysis 3 Pulse width
The pulse width of the test device is set to 0.5ms/1ms, while the predicate device adopts a pulse width range of 1-20ms. The 1ms mode of the test device is fully covered within the validated safety range of the predicate device. The newly added 0.5ms mode, which is narrower than the predicate's minimum 1ms pulse width, is controlled under the safety boundary of the peak power and total energy verified by the predicate device through the hardware power limiting design, with no higher energy density or
8
{12}
thermal effect introduced. The key safety parameter, pulse duration (5-50ms), remains identical between the two devices, ensuring no change in the upper limit of total energy output. This difference is a controlled contraction and low-risk adjustment of the validated safety range of the predicate device, and will not adversely affect the safety of the device.
### Analysis 4 Spot Siz
The proposed device is different in Spot Size from the predicate device. Spot size only affects the area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness.
### Analysis 5 Max. power consumption
There is a difference in Max. Power between the proposed device and the predicate device. After verification, the proposed device complies with the requirements of AAMI/ANSI/ES60601-1 and IEC 60601-1-2. The verification results show that this difference in Max. Power will not give rise to any safety or effectiveness issues; therefore, this difference will not affect the safety and effectiveness of the proposed device.
### Substantially Equivalent(SE) Conclusion
Intense Pulsed Light Treatment device has the same intended use, similar indications for use, the same technological characteristics, the same energy used, and the same operating principles as its predicates. The non-clinical data and performance testing reports in this submission demonstrate that Intense Pulsed Light Treatment device meets the expected performance requirements. Any difference between the subject and predicate device do not raise new issues of safety or effectiveness. Based on above analysis, Intense Pulsed Light Treatment device is substantial equivalent to the cited predicate device.
### 12. Clinical Test Conclusion
No clinical study is included in this submission.
### 13. Substantially Equivalent(SE) Conclusion
Intense Pulsed Light Therapy Device has the same intended use, similar indications for use, the same technological characteristics, the same energy used, and the same operating principles as its predicates. The non-clinical data and performance testing reports in this submission demonstrate that Intense Pulsed Light Therapy Device meets the expected performance requirements. Any difference between the subject and predicate device do not raise new issues of safety or
9
{13}
effectiveness. Based on above analysis, Intense Pulsed Light Therapy Device is substantial equivalent to the cited predicate device.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.