Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Hydroureter
Deep learning neural networks
—
AUC: 0.984 [0.972-0.996]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Retroperitoneal hematoma
Deep learning neural networks
—
Balanced: Se 87.9% [80.8-92.7%], Sp 92.3% [90.4-93.9%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Large volume ascites
Deep learning neural networks
—
AUC: 0.995 [0.991-0.998]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Acute appendicitis
Deep learning neural networks
—
High Sensitivity: Se 97.4% [90.9-99.3%], Sp 86.6% [84.2-88.7%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Intraabdominopelvic abscess
Deep learning neural networks
—
Sensitivity Biased: Se 90.8% [83.9-94.9%], Sp 87.5% [85.1-89.5%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Hemoperitoneum or complex free fluid
Deep learning neural networks
—
High Specificity: Se 86.6% [80.7-90.9%], Sp 83.1% [80.3-85.5%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Splenic traumatic injury
Deep learning neural networks
—
Sensitivity Biased: Se 97.2% [90.4-99.2%], Sp 85.2% [82.7-87.4%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Hepatic traumatic injury
Deep learning neural networks
—
Balanced: Se 91.5% [81.6-96.3%], Sp 89.5% [87.3-91.3%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Renal traumatic injury
Deep learning neural networks
—
Sensitivity Biased: Se 95.7% [88.1-98.5%], Sp 85.1% [82.7-87.3%]
—
—
Independent abdominopelvic clinical test cohort: 965 studies
>1 (US board-certified radiologists)
Indications for Use
a2z-Abdo-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominopelvic CT images acquired with or without intravenous contrast in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 10 specified findings:Obstructing ureteral stoneHydroureterRetroperitoneal hematomaLarge volume ascitesAcute appendicitisIntraabdominopelvic abscessHemoperitoneum or complex free fluidSplenic traumatic injuryHepatic traumatic injuryRenal traumatic injuryThese findings are intended to be used together as one device.a2z-Abdo-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.The results of a2z-Abdo-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease.
Device Story
Software-based radiological triage tool; processes abdominopelvic CT images; utilizes deep learning AI to detect 10 specific abdominopelvic conditions; operates in parallel to standard-of-care image interpretation; deployed on cloud or on-premises server hardware; integrates with clinical PACS/worklist systems; flags suspected positive cases for prioritized review; provides DICOM instance UIDs for key images; does not alter original images; not for diagnostic use; intended for hospital/clinical settings; operated by trained medical specialists; assists in workflow prioritization to potentially reduce turnaround time for critical findings.
Clinical Evidence
Standalone performance assessment using a held-out cohort of 965 abdominopelvic CT studies (914 unique patients). Reference standard established by 2+1 U.S. board-certified radiologist review (tie-breaker used for disagreements). Primary endpoints: AUC, sensitivity, and specificity for 10 findings. Results: QFM AUCs 0.984–0.995; QAS AUCs 0.945–0.985. All findings met pre-specified acceptance criteria. No clinical diagnostic outcomes measured; bench-only performance validation.
Technological Characteristics
SaMD; non-adaptive deep learning neural networks; processes DICOM abdominopelvic CT images; cloud or on-premises server deployment; standard server hardware; output includes notifications and DICOM instance UIDs; no hardware components; software-only triage.
Indications for Use
Indicated for adults aged 22+ undergoing abdominopelvic CT (with/without contrast) to assist hospital networks/specialists in workflow triage by flagging suspected cases of 10 conditions: obstructing ureteral stone, hydroureter, retroperitoneal hematoma, large volume ascites, acute appendicitis, intraabdominopelvic abscess, hemoperitoneum/complex free fluid, and splenic, hepatic, or renal traumatic injury.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 16, 2026
A2z Radiology Ai, Inc.
Samir Rajpurkar
Chief Executive Officer
292 Newbury St.
Unit 235
Boston, Massachusetts 02115
Re: K260906
Trade/Device Name: a2z-Abdo-Triage
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological computer aided triage and notification software
Regulatory Class: Class II
Product Code: QAS, QFM
Dated: August 13, 2026
Received: August 14, 2026
Dear Samir Rajpurkar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260906 - Samir Rajpurkar
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K260906 - Samir Rajpurkar
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260906 | ? |
| Please provide the device trade name(s). | | ? |
| a2z-Abdo-Triage | | |
| Please provide your Indications for Use below. | | ? |
| a2z-Abdo-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominopelvic CT images acquired with or without intravenous contrast in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 10 specified findings:Obstructing ureteral stoneHydroureterRetroperitoneal hematomaLarge volume ascitesAcute appendicitisIntraabdominopelvic abscessHemoperitoneum or complex free fluidSplenic traumatic injuryHepatic traumatic injuryRenal traumatic injuryThese findings are intended to be used together as one device.a2z-Abdo-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.The results of a2z-Abdo-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease. | | |
| Please select the types of uses. | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{4}
K260906
a2z
RADIOLOGY AI
510(k) Summary
# Contents
1. Administrative Information ... 1
2. Subject Device Information ... 1
3. Predicate Device Information 2
4. Device Description 2
5. Indications for Use 3
6. Comparison of Technological Characteristics with the Predicate Device ... 3
7. Performance Data 5
7.1. Software Verification and Validation Testing 5
7.2. Performance Assessment 5
Patient Demographics (N=914 received patient pseudonyms) 6
Study-Level Imaging Characteristics (N=965 studies) 6
Condition Prevalence in Test Cohort 7
QFM Condition Performance 8
QAS Condition Performance 8
8. Predetermined Change Control Plan (PCCP) 9
9. Conclusion 10
# 510(k) Summary
# 1. Administrative Information
| Submitter Name | a2z Radiology AI Inc. |
| --- | --- |
| Address | 292 Newbury Street, Unit 235, Boston, MA 02115, USA |
| Phone Number | +1 508-293-1822 |
| Fax Number | N/A |
| Company Representative | Samir Rajpurkar |
| Email | support@a2zradiology.ai |
| Date Summary Prepared | September 15, 2026 |
# 2. Subject Device Information
| Trade Name | a2z-Abdo-Triage |
| --- | --- |
| Subject Device K Number | K260906 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS, QFM |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
{5}
510(k) Summary
a2z
RADIOLOGY AI
### 3. Predicate Device Information
| Predicate Device Name | a2z-Unified-Triage |
| --- | --- |
| Predicate Device K Number | K252366 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS, QFM |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
### 4. Device Description
a2z-Abdo-Triage is a radiological computer-aided triage and notification software device. The software consists of an algorithmic component that supports both cloud-based and on-premises deployment on standard server hardware. The device processes abdomen/pelvis CT images from clinical imaging systems, analyzing them using artificial intelligence algorithms to detect suspected cases of 10 abdominopelvic conditions: Obstructing ureteral stone, Hydroureter, Retroperitoneal hematoma, Large volume ascites, Acute appendicitis, Intraabdominopelvic abscess, Hemoperitoneum or complex free fluid, Splenic traumatic injury, Hepatic traumatic injury, and Renal traumatic injury.
Following the AI processing, the analysis results are returned to the client system for worklist prioritization. When a suspected case is detected, the software provides analysis results that enable the client system to generate appropriate notifications. These results can include DICOM instance UIDs for key images, which are for informational purposes only, do not contain any marking of the findings, and are not intended for diagnostic use beyond notification.
Integration with clinical imaging systems facilitates efficient triage by enabling prioritization of suspect cases for review of the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
Algorithm Architecture: The artificial intelligence algorithms implement deep learning models that analyze CT studies and identify suspected positive findings. For each detected finding, the software provides one or more image references most likely to show the condition; those references are informational only. A finding that is not reported is not a diagnostic negative result.
Training Database: The algorithms were developed using a multi-site CT dataset with diverse imaging characteristics. The filed findings were evaluated on the independent abdominopelvic clinical test cohort.
Data Independence: No received patient pseudonym appeared in both the development and test datasets, and the institutions contributing the test cohort contributed no studies to the training or tuning splits.
{6}
510(k) Summary
a2z
RADIOLOGY AI
## 5. Indications for Use
a2z-Abdo-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominopelvic CT images acquired with or without intravenous contrast in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 10 specified findings:
- Obstructing ureteral stone
- Hydroureter
- Retroperitoneal hematoma
- Large volume ascites
- Acute appendicitis
- Intraabdominopelvic abscess
- Hemoperitoneum or complex free fluid
- Splenic traumatic injury
- Hepatic traumatic injury
- Renal traumatic injury
These findings are intended to be used together as one device.
a2z-Abdo-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of a2z-Abdo-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease.
## 6. Comparison of Technological Characteristics with the Predicate Device
The subject device, a2z-Abdo-Triage, is substantially equivalent to the predicate device a2z-Unified-Triage (K252366), a cleared radiological computer-aided triage and notification software for the abdominopelvic CT region.
The subject and predicate are both radiological computer-aided triage and notification software using artificial intelligence algorithms. Both provide notifications for worklist prioritization. Neither removes cases from standard-care reading queues nor de-prioritizes cases; both operate in parallel with standard care.
The technology type, intended use for triage, and principles of operation are shared with the predicate. Both provide positive notifications for worklist prioritization. The subject device reports
{7}
510(k) Summary
a2z
RADIOLOGY AI
suspected positive findings; an unreported finding is not a diagnostic negative result. The specified finding sets differ, and each subject-device finding is supported by standalone performance testing. The table below compares key features.
Table 1. Comparison with predicate device.
| Characteristic | Subject Device: a2z-Abdo-Triage | Predicate Device: a2z-Unified-Triage (K252366) |
| --- | --- | --- |
| Manufacturer | a2z Radiology AI Inc. | a2z Radiology AI Inc. |
| Regulation Number | 892.2080 | 892.2080 |
| Regulatory Class | Class II | Class II |
| Product Code | QAS, QFM | QAS, QFM |
| Regulation Name | Radiological computer aided triage and notification software | Radiological computer aided triage and notification software |
| Device Property | SaMD (Software as a Medical Device) | SaMD (Software as a Medical Device) |
| Indications for Use | Radiological computer-aided triage and notification software for analysis of abdominopelvic CT images acquired with or without intravenous contrast in adults aged 22+. Assists in workflow triage by flagging 10 specified findings. | Radiological computer-aided triage and notification software for analysis of abdominal/pelvic CT images in adults aged 22+. Assists in workflow triage by flagging 7 abdominopelvic findings. |
| Technical Characteristics | | |
| Input | Abdominopelvic CT images acquired with or without intravenous contrast | Abdomen/Pelvis CT images (with or without contrast) |
| Output | Suspected positive findings for worklist prioritization and informational DICOM instance UIDs for key images; unreported findings are not negative diagnostic results | Notification to image and order management system for worklist prioritization |
| Algorithm Type | Non-adaptive machine learning (Deep Learning - Neural Networks) | Non-adaptive machine learning (Deep Learning - Neural Networks) |
| Intended Users | Appropriately trained medical specialists qualified to interpret abdominal/pelvic CT studies | Appropriately trained medical specialists qualified to interpret abdominal/pelvic CT studies |
| Target Population | Adults (22 years and older) | Adults (22 years and older) |
| Location of anatomical structures | Abdomen and Pelvis | Abdomen and Pelvis |
| Imaging Modality | Computed Tomography (CT) | Computed Tomography (CT) |
| Intended Use Environment | Hospital environment or other clinical settings with DICOM-compliant CT imaging and IT infrastructure | Hospital environment or other clinical settings with DICOM-compliant CT imaging and IT infrastructure |
| Performance | Met pre-specified performance targets for all findings | Met pre-specified performance targets for all findings |
{8}
510(k) Summary
| Characteristic | Subject Device: a2z-Abdo-Triage | Predicate Device: a2z-Unified-Triage (K252366) |
| --- | --- | --- |
| Turnaround Time | Technical processing interval – mean 133.52 seconds (95% CI: 129.26-137.78), median 97.19 seconds, 95th percentile 276.74 seconds | Filed predicate turnaround values – mean 58.39 seconds (95% CI: 56.11-60.68), median 55.02 seconds, 95th percentile 90.36 seconds. Measurement boundaries differed, so the values are not a direct speed comparison. |
| Software device that operates on off-the-shelf hardware | Yes. Software supports both cloud-based and on-premises deployment on standard server hardware | Yes. Interfaces with image and order management systems |
Table 1 compares the subject device to the predicate. The Turnaround Time row reproduces each device's own filed turnaround-time figures as measured; the measurement boundaries differ between the two devices, and no direct speed comparison is claimed.
## 7. Performance Data
### 7.1. Software Verification and Validation Testing
Software verification and validation testing was conducted.
### 7.2. Performance Assessment
A standalone performance assessment was performed for a2z-Abdo-Triage to validate the accuracy of detecting the 10 findings against a reference standard established by U.S. board-certified radiologists.
#### 7.2.1. Reader Reference Standard
The reference standard for the 989-case test dataset used a 2+1 methodology with U.S. board-certified radiologists. Each study-finding classification received two independent U.S. board-certified radiologist determinations. When those readers disagreed, an independent U.S. board-certified radiologist adjudicator provided the tiebreaking determination. Readers were blinded to device output.
#### 7.2.2. Development and Test Dataset
a2z-Abdo-Triage was evaluated on a held-out reader-reference-standard cohort of 989 abdominopelvic CT studies. After prespecified exclusions, 965 CT studies were included in the performance analysis. The enriched test design provided positive representation for each filed finding; the cohort's observed demographics and imaging characteristics are reported below.
#### 7.2.3. Test Cohort Characteristics
The 965 studies in the analytic cohort are associated with 914 received patient pseudonyms. Some pseudonyms are associated with more than one study. Age and sex are reported at the received-
{9}
a2z
RADIOLOGY AI
# 510(k) Summary
pseudonym level; imaging characteristics (site, CT scanner manufacturer, contrast status, slice thickness) are reported at the study level because acquisition parameters can vary across studies.
# Patient Demographics (N=914 received patient pseudonyms)
# Age Distribution
- Mean: 54.6 years (SD: 19.7)
- Median: 54 years
- Range: 22-90 years
| Age Group | N | Percentage |
| --- | --- | --- |
| 22-39 | 262 | 28.7% |
| 40-59 | 272 | 29.8% |
| 60-79 | 260 | 28.4% |
| 80+ | 120 | 13.1% |
Sex Distribution
| Sex | N | Percentage |
| --- | --- | --- |
| Female | 435 | 47.6% |
| Male | 477 | 52.2% |
| Not available | 2 | 0.2% |
# Study-Level Imaging Characteristics (N=965 studies)
Site Distribution
| Site | N | Percentage |
| --- | --- | --- |
| Site 1 | 548 | 56.8% |
| Site 2 | 294 | 30.5% |
| Site 3 | 101 | 10.5% |
| Not available | 22 | 2.3% |
Geographic Distribution
| State | N | Percentage |
| --- | --- | --- |
| FL | 548 | 56.8% |
| TX | 294 | 30.5% |
| OH | 101 | 10.5% |
| Not available | 22 | 2.3% |
CT Scanner Manufacturer Distribution
| Manufacturer | N | Percentage |
| --- | --- | --- |
| SIEMENS | 497 | 51.5% |
| GE | 391 | 40.5% |
{10}
a2z
RADIOLOGY AI
# 510(k) Summary
| Manufacturer | N | Percentage |
| --- | --- | --- |
| Canon | 34 | 3.5% |
| TOSHIBA | 34 | 3.5% |
| FUJI | 7 | 0.7% |
| Other | 2 | 0.2% |
# Contrast Status
| Contrast | N | Percentage |
| --- | --- | --- |
| With contrast | 558 | 57.8% |
| Without contrast | 407 | 42.2% |
# Slice Thickness Distribution
- Range: 0.6-5.0mm, Mean: 2.9mm, Median: 2.0mm
| Slice Thickness | N | Percentage |
| --- | --- | --- |
| <5mm | 707 | 73.3% |
| 5mm | 258 | 26.7% |
### 7.2.4. Clinical Subgroups and Confounders
The test dataset included positive representation for each of the 10 target findings and cases with concurrent findings. Descriptive stratified results were evaluated for the observed demographic, imaging, site, and disease-presentation groups. Sparse or non-estimable strata were identified; no separate per-stratum acceptance criterion is claimed.
### 7.2.5. Primary Endpoints (Prevalence)
Condition Prevalence in Test Cohort
| Condition | N Total | N Positive | Prevalence |
| --- | --- | --- | --- |
| Acute appendicitis | 965 | 76 | 7.9% |
| Hemoperitoneum or complex free fluid | 965 | 172 | 17.8% |
| Hepatic traumatic injury | 965 | 59 | 6.1% |
| Hydroureter | 965 | 162 | 16.8% |
| Intraabdominopelvic abscess | 965 | 109 | 11.3% |
| Large volume ascites | 965 | 79 | 8.2% |
| Obstructing ureteral stone | 965 | 118 | 12.2% |
| Renal traumatic injury | 965 | 70 | 7.3% |
| Retroperitoneal hematoma | 965 | 116 | 12.0% |
| Splenic traumatic injury | 965 | 72 | 7.5% |
{11}
510(k) Summary
a2z
RADIOLOGY AI
#### 7.2.6. QFM Conditions
QFM Condition Performance
| Condition | Available Operating Points and Performance |
| --- | --- |
| Hydroureter (QFM)Total N=965,Positive=162AUC: 0.984[0.972-0.996] | High Sensitivity: Se 97.5% [93.8-99.0%], Sp 88.4% [86.0-90.5%]Sensitivity Biased: Se 94.4% [89.8-97.0%], Sp 94.4% [92.6-95.8%]Balanced: Se 91.4% [86.0-94.8%], Sp 96.8% [95.3-97.8%] |
| Large volume ascites (QFM)Total N=965,Positive=79AUC: 0.995[0.991-0.998] | High Specificity: Se 100.0% [95.4-100.0%], Sp 83.1% [80.5-85.4%] |
| Obstructing ureteral stone (QFM)Total N=965,Positive=118AUC: 0.994[0.989-0.998] | High Sensitivity: Se 99.2% [95.4-99.9%], Sp 85.4% [82.8-87.6%]Sensitivity Biased: Se 97.5% [92.8-99.1%], Sp 95.3% [93.6-96.5%]Balanced: Se 94.1% [88.3-97.1%], Sp 98.8% [97.8-99.4%] |
#### 7.2.7. QAS Conditions
QAS Condition Performance
| Condition | Available Operating Points and Performance |
| --- | --- |
| Acute appendicitis (QAS)Total N=965,Positive=76 | High Sensitivity: Se 97.4% [90.9-99.3%], Sp 86.6% [84.2-88.7%]Sensitivity Biased: Se 94.7% [87.2-97.9%], Sp 96.6% [95.2-97.6%]Balanced: Se 88.2% [79.0-93.6%], Sp 99.1% [98.2-99.5%] |
| Hemoperitoneum or complex free fluid (QAS)Total N=965,Positive=172 | High Specificity: Se 86.6% [80.7-90.9%], Sp 83.1% [80.3-85.5%] |
| Hepatic traumatic injury (QAS)Total N=965,Positive=59 | Balanced: Se 91.5% [81.6-96.3%], Sp 89.5% [87.3-91.3%]High Specificity: Se 81.4% [69.6-89.3%], Sp 96.0% [94.5-97.1%] |
| Intraabdominopelvic abscess (QAS)Total N=965,Positive=109 | Sensitivity Biased: Se 90.8% [83.9-94.9%], Sp 87.5% [85.1-89.5%] |
| Renal traumatic injury (QAS)Total N=965,Positive=70 | Sensitivity Biased: Se 95.7% [88.1-98.5%], Sp 85.1% [82.7-87.3%]Balanced: Se 94.3% [86.2-97.8%], Sp 95.3% [93.7-96.5%] |
{12}
a2z
RADIOLOGY AI
510(k) Summary
| Condition | Available Operating Points and Performance |
| --- | --- |
| Retroperitoneal hematoma (QAS) Total N=965, Positive=116 | Balanced: Se 87.9% [80.8-92.7%], Sp 92.3% [90.4-93.9%] |
| Splenic traumatic injury (QAS) Total N=965, Positive=72 | Sensitivity Biased: Se 97.2% [90.4-99.2%], Sp 85.2% [82.7-87.4%] Balanced: Se 88.9% [79.6-94.3%], Sp 95.0% [93.3-96.2%] |
### 7.2.8. Performance Discussion
The standalone performance assessment validated 10 abdominopelvic findings comprising 3 QFM findings (Hydroureter, Large volume ascites, Obstructing ureteral stone) and 7 QAS findings (Acute appendicitis, Hemoperitoneum or complex free fluid, Hepatic traumatic injury, Intraabdominopelvic abscess, Renal traumatic injury, Retroperitoneal hematoma, Splenic traumatic injury). All 10 findings met pre-specified acceptance criteria, consistent with the statistical methodology used in the predicate device submission.
QFM AUC values ranged from 0.984 to 0.995, and QAS AUC values ranged from 0.945 to 0.985. Each finding met its prespecified, finding-type-specific acceptance criteria at one or more reported operating points.
Stratified subgroup analyses were conducted across patient demographics, clinical sites and regions, imaging acquisition parameters, and concurrent findings. These analyses are descriptive and supportive: estimates and confidence intervals characterize the observed data, but no formal per-stratum acceptance criterion or claim of absence of subgroup degradation is applied. Sparse or non-estimable strata were identified.
The technical processing interval had a mean of 133.52 seconds, a median of 97.19 seconds, and a 95th percentile of 276.74 seconds per study. This interval excludes deployment-specific overhead and does not represent end-to-end clinical-workflow turnaround.
In summary, a2z-Abdo-Triage met the prespecified performance criteria for the 10 validated findings. These results support the substantial-equivalence determination for the device's intended triage-and-notification use.
### 8. Predetermined Change Control Plan (PCCP)
This submission contains a Predetermined Change Control Plan (PCCP). The PCCP does not authorize adaptive algorithms that continuously learn in the field. A modified model is trained, validated, and locked before release. Labeling is updated for implemented PCCP changes as specified in the plan.
The PCCP specifies possible modifications to a2z-Abdo-Triage and the verification, validation, risk-management, and acceptance criteria that must be satisfied before an in-scope modification is implemented. The PCCP applies to the 10 conditions validated in this 510(k) submission: Obstructing ureteral stone, Hydroureter, Retroperitoneal hematoma, Large volume ascites, Acute
{13}
510(k) Summary
a2z
RADIOLOGY AI
appendicitis, Intraabdominopelvic abscess, Hemoperitoneum or complex free fluid, Splenic traumatic injury, Hepatic traumatic injury, and Renal traumatic injury. For each condition, the PCCP covers operating points that meet the pre-specified performance criteria upon revalidation.
The planned modifications include:
- Training data volume and diversity expansion
- Training data quality and annotation refinement
- Ensemble composition optimization
• Data augmentation parameter adjustments
- Training parameter optimization
• Weight initialization method selection
- Neural network architecture component modifications
- Input data preprocessing adjustments
The PCCP defines bounded specifications and prespecified validation requirements. Each modification must complete the applicable validation requirements before implementation. For substantive modifications, these requirements include independent test data, U.S. board-certified-radiologist reference standards, data sequestration, descriptive subgroup analysis, and prespecified statistical acceptance criteria.
The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices. Detailed validation activities, testing methodologies, and performance requirements have been established for each modification. Each change must complete the specified verification and validation before implementation and meet the PCCP acceptance criteria.
## 9. Conclusion
The subject device, a2z-Abdo-Triage, and the predicate, a2z-Unified-Triage, are substantially equivalent. Both are software devices intended to aid in the prioritization and triage of radiological images using AI algorithms. They share the same fundamental scientific technology and principles of operation.
While the specific clinical findings differ, the intended use for triage, the AI-based approach, the integration into clinical workflows, and the parallel nature of operation are highly similar. Both devices target the same anatomical region (abdomen/pelvis) and aim to reduce turnaround time through preemptive triage without altering the standard of care.
The performance data show that a2z-Abdo-Triage met the prespecified criteria for its intended triage-and-notification use and support the determination that a2z-Abdo-Triage is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.