K252366 · A2z Radiology Ai, Inc. · QAS · Nov 24, 2025 · Radiology
Device Facts
Record ID
K252366
Device Name
a2z-Unified-Triage
Applicant
A2z Radiology Ai, Inc.
Product Code
QAS · Radiology
Decision Date
Nov 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252366 · Nov 24, 2025
a2z-Unified-Triage
A2z Radiology Ai, Inc.
Retrospective clinical abdominal/pelvic CT studies from 12 clinical sites
The device performance was validated using a retrospective cohort of 675 clinical CT cases collected from multiple hospital sites to demonstrate accuracy across diverse patient demographics, imaging equipment, and clinical presentations.
a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. These findings are intended to be used together as one device. The device supports both cloud-based and on-premises deployment, with integration either directly with healthcare facility systems or through third-party healthcare technology platforms. a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for primary diagnosis beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Device Story
a2z-Unified-Triage is a radiological triage software that processes abdominal/pelvic CT images to detect 7 specific conditions: Acute Cholecystitis, Acute Pancreatitis, Unruptured AAA, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. Operating in parallel to standard clinical workflows, the device uses deep learning algorithms to analyze CT studies. When a suspected finding is detected, the system sends a notification to the client system (e.g., PACS/worklist) to prioritize the case for radiologist review. The output includes DICOM instance UIDs for key image slices, provided for informational purposes only. The device is used in hospital or clinical settings by trained medical specialists. It does not alter original images or provide primary diagnosis; it serves to expedite the review of urgent cases, potentially improving patient outcomes through faster clinical intervention. Deployment is supported via cloud or on-premises server hardware.
Clinical Evidence
Bench testing only. Performance validated on 675 cases (643 unique patients) from 12 clinical sites. Ground truth established by two U.S. board-certified radiologists with third-party adjudication for discordance. Results: QAS findings (Free Air, SBO) achieved >80% sensitivity/specificity; QFM findings (Cholecystitis, Pancreatitis, AAA, Diverticulitis, Hydronephrosis) achieved AUC >0.95. Mean triage turnaround time was 58.39 seconds. Stratified analysis confirmed performance consistency across demographics, scanner manufacturers, contrast status, and slice thicknesses.
Technological Characteristics
SaMD; deep learning neural networks; non-adaptive; cloud or on-premises deployment on standard server hardware; DICOM-compliant; input: abdominal/pelvic CT (contrast/non-contrast); output: JSON classification and DICOM instance UIDs.
Indications for Use
Indicated for adults aged 22+ undergoing abdominal/pelvic CT imaging. Assists clinicians in triage/prioritization by flagging suspected Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 24, 2025
a2z Radiology AI Inc.
Samir Rajpurkar
Chief Executive Officer
292 Newbury Street
Unit 235
Boston, MA 02115
Re: K252366
Trade/Device Name: a2z-Unified-Triage
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS, QFM,
Dated: November 1, 2025
Received: November 3, 2025
Dear Samir Rajpurkar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252366 – Samir Rajpurkar
Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K252366 – Samir Rajpurkar
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252366 | ? |
| Please provide the device trade name(s). | | ? |
| a2z-Unified-Triage | | |
| Please provide your Indications for Use below. | | ? |
| a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the
analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist
hospital networks and appropriately trained medical specialists in workflow triage by flagging and
communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis,
Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis,
and Small Bowel Obstruction. These findings are intended to be used together as one device. The device
supports both cloud-based and on-premises deployment, with integration either directly with healthcare
facility systems or through third-party healthcare technology platforms.
a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected
findings in parallel to the ongoing standard of care image interpretation. The device provides analysis
results that enable client systems to generate notifications for cases with suspected findings. These results
can include DICOM instance UIDs for key images, which are meant for informational purposes only and not
intended for primary diagnosis beyond notification. The device does not alter the original medical image and
is not intended to be used as a diagnostic device.
The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and
based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified
clinicians are responsible for viewing full images per the standard of care.
Please select the types of uses (select one or both, as
applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | |
{4}
K252366
a2z
RADIOLOGY AI
510(k) Summary
# 1. Administrative Information
| Submitter Name | a2z Radiology AI Inc. |
| --- | --- |
| Address | 292 Newbury Street, Unit 235, Boston, MA 02115, USA |
| Phone Number | +1 508-293-1822 |
| Fax Number | N/A |
| Company Representative | Samir Rajpurkar |
| Email | support@a2zradiology.ai |
| Date Summary Prepared | November 01, 2025 |
# 2. Subject Device Information
| Trade Name | a2z-Unified-Triage |
| --- | --- |
| Subject Device K Number | K252366 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS, QFM |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
# 3. Predicate Device Information
| Predicate Device Name | Annalise Enterprise CTB Triage Trauma |
| --- | --- |
| Predicate Device K Number | K223240 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
# 4. Device Description
a2z-Unified-Triage is a radiological computer-assisted triage and notification software device. The software consists of an algorithmic component that supports both cloud-based and on-premises
{5}
deployment on standard server hardware. The device processes abdomen/pelvis CT images from clinical imaging systems, analyzing them using artificial intelligence algorithms to detect suspected cases of 7 abdominopelvic conditions: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction.
Following the AI processing, the analysis results are returned to the client system for worklist prioritization. When a suspected case is detected, the software provides analysis results that enable the client system to generate appropriate notifications. These results can include DICOM instance UIDs for key images, which are for informational purposes only, do not contain any marking of the findings, and are not intended for primary diagnosis beyond notification.
Integration with clinical imaging systems facilitates efficient triage by enabling prioritization of suspect cases for review of the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
Algorithm Architecture: The artificial intelligence algorithms implement deep learning models that analyze CT studies and return binary outputs indicating the presence or absence of each of the 7 target abdominopelvic findings. When a finding is detected, the algorithm identifies a key image slice most likely to show the condition, provided for informational purposes only.
Training Database: The algorithms were developed on an extensive dataset of abdomen/pelvis CT studies from multiple clinical sites. The training data encompassed diverse imaging characteristics including multiple CT scanner manufacturers, both contrast-enhanced and non-contrast studies, variable slice thicknesses, and a wide range of patient demographics.
Data Independence: Strict protocols ensured complete independence between development and testing datasets, with mutually exclusive patients. No overlap existed between training/validation data and the independent test dataset used for performance evaluation.
## 5 Indications for Use
a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. These findings are intended to be used together as one device. The device supports both cloud-based and on-premises deployment, with integration either directly with healthcare facility systems or through third-party healthcare technology platforms.
a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for primary diagnosis beyond notification. The
{6}
510(k) Summary
a2z RADIOLOGY AI
device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
# Comparison of Technological Characteristics
The subject device, a2z-Unified-Triage, is substantially equivalent to the primary predicate Annalise Enterprise CTB Triage Trauma (K223240).
The subject and predicate are both radiological computer-aided triage and notification software using artificial intelligence algorithms. Both aid radiological image triage through deep learning algorithms trained on medical images, providing specialists with notifications and preview images for preemptive triage. Neither removes cases from standard care reading queues nor de-prioritizes cases. Both operate in parallel to standard care as the default option.
While the anatomical regions of interest differ (abdomen/pelvis for the subject device vs. head for the predicate), the core technology, intended use for triage, and principles of operation are similar. Both devices raise identical safety and effectiveness questions regarding accurate triage. Performance testing, summarized below, demonstrated that the subject device maintains equivalent safety and effectiveness for its intended use. A detailed comparison of key features is provided in the table below.
# 6. Comparison of Technological Characteristics with the Predicate Device
Table 1. Comparison with predicate device.
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
| --- | --- | --- |
| Manufacturer | a2z Radiology AI Inc. | Annalise-AI Pty Ltd |
| Regulation Number | 892.2080 | 892.2080 |
| Regulatory Class | Class II | Class II |
| Product Code | QAS, QFM | QAS |
| Regulation Name | Radiological computer aided triage and notification software | Radiological computer aided triage and notification software |
| Device Property | SaMD (Software as a Medical Device) | SaMD (Software as a Medical Device) |
{7}
510(k) Summary
a2z
RADIOLOGY AI
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
| --- | --- | --- |
| Indications for Use | Radiological computer-aided triage and notification software for analysis of abdominal/pelvic CT images in adults aged 22+. Assists in workflow triage by flagging 7 abdominopelvic findings. | Device to aid in triage and prioritization of studies with features suggestive of 4 intracranial hemorrhage findings in patients 22+. |
| Technical Characteristics | | |
| Input | Abdomen/Pelvis CT images (with or without contrast) | Non-contrast brain CT images |
| Output | JSON object with binary classification for 7 findings and DICOM instance UIDs for positive cases | Notification to image and order management system for worklist prioritization |
| Algorithm Type | Non-adaptive machine learning (Deep Learning - Neural Networks) | Non-adaptive machine learning (Deep Learning - Neural Networks) |
| Intended Users | Appropriately trained medical specialists | Trained clinicians qualified to interpret brain CT studies |
| Target Population | Adults (22 years and older) | Adults (22 years and older) |
| Location of anatomical structures | Abdomen and Pelvis | Head (Brain) |
| Imaging Modality | Computed Tomography (CT) | Computed Tomography (CT) |
| Intended Use Environment | Hospital environment or other clinical settings with DICOM-compliant CT imaging and IT infrastructure | Medical care environment |
| Performance | 7 findings met targets: QAS findings >80% sensitivity/specificity; QFM findings >0.95 AUC | Met performance targets with sensitivity and specificity >80% across findings |
| Triage Turn-around Time | Mean: 58.39 seconds (95% CI: 56.11-60.68), Median: 55.02 seconds, 95th percentile: 90.36 seconds | Mean: 81.6 seconds (95% CI: 80.3-82.9) |
{8}
510(k) Summary
a2z
RADIOLOGY AI
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
| --- | --- | --- |
| Software device that operates on off-the-shelf hardware | Yes. Software supports both cloud-based and on-premises deployment on standard server hardware | Yes. Interfaces with image and order management systems |
# 7. Performance Data
# 7.1. Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Device Software Functions".
# 7.2. Performance Assessment
A standalone performance assessment was performed for a2z-Unified-Triage to validate the accuracy of detecting the 7 findings against a reference standard established by U.S. board-certified radiologists.
7.2.1. Ground Truth The ground truth for the 678-case Test Data was established using a $2 + 1$ methodology exclusively with U.S. board-certified radiologists. Specifically, each case was independently reviewed by two U.S. board-certified radiologists. For cases where the two readers disagreed (discordant cases), a third U.S. board-certified expert adjudicator, fellowship-trained in body imaging, provided the tiebreaking determination to establish the final ground truth. This ensures that all ground truth determinations are made by qualified U.S. board-certified radiologists using clinical standards consistent with U.S. practice patterns.
7.2.2. Development and Test Dataset a2z-Unified-Triage was trained and tested on an extensive and diverse set of abdominal/pelvic CT studies from multiple clinical sites. The final test cohort consisted of 678 cases, with 3 cases excluded due to quality control failures during model inference, resulting in an analytic cohort of 675 cases from 643 unique patients. Development and test cohorts had mutually exclusive patients. The test dataset was constructed to ensure adequate representation for each of the findings and included a wide range of patient demographics and imaging characteristics, as summarized below.
Test Cohort Demographics and Imaging Characteristics (N=675)
| Sex | N | Percentage |
| --- | --- | --- |
| Female | 345 | 51.1 |
{9}
510(k) Summary
a2z RADIOLOGY AI
| Sex | N | Percentage |
| --- | --- | --- |
| Male | 330 | 48.9 |
Age (years): Median (IQR) = 61 (44-75); Mean (SD) = 59.9 (18.2); Range = 22-90
Site and State Distribution
| Site | State | Number of Cases |
| --- | --- | --- |
| Site 1 | New York | 223 (33.0%) |
| Site 2 | Kansas | 120 (17.8%) |
| Site 3 | New York | 82 (12.1%) |
| Site 4 | Missouri | 66 (9.8%) |
| Site 5 | Texas | 61 (9.0%) |
| Other (7 Sites) | Various | 123 (18.2%) |
| Total | | 675 |
Sites in 'Other' category were located in Missouri, Texas, Kansas, and Nebraska. Total unique sites: 12
State Distribution:
| State | N | Percentage |
| --- | --- | --- |
| New York | 305 | 45.2 |
| Kansas | 143 | 21.2 |
| Missouri | 124 | 18.4 |
| Texas | 101 | 15.0 |
| Nebraska | 2 | 0.3 |
7.3.3. Imaging Characteristics Manufacturer Distribution:
| Manufacturer | N | Percentage |
| --- | --- | --- |
| GE | 277 | 41.0 |
| SIEMENS | 217 | 32.1 |
| Canon | 107 | 15.9 |
| TOSHIBA | 52 | 7.7 |
| Other | 22 | 3.3 |
{10}
510(k) Summary
a2z RADIOLOGY AI
Manufacturers in 'Other' category: FUJI, Hitachi, Philips
Contrast Status:
| Contrast Status | N | Percentage |
| --- | --- | --- |
| With Contrast | 385 | 57 |
| Without Contrast | 290 | 43 |
Slice Thickness:
| Slice Thickness | N | Percentage |
| --- | --- | --- |
| 5mm | 464 | 69 |
| 2mm to <5mm | 211 | 31 |
Mean (SD): 4.30 (1.06) mm; Median (IQR): 5.00 (3.00-5.00) mm; Range: 2-5mm
7.2.4. Clinical Subgroups and Confounders The test dataset included diverse disease presentations across all 7 target conditions. U.S. board-certified radiologists annotated clinically relevant anatomical variations and disease subtypes on subsets of positive cases, including: calculous and acalculus acute cholecystitis, gallbladder wall thickening, and gallbladder hydrops; acute pancreatitis with and without fluid collections; sigmoid and other colonic locations of acute diverticulitis including cases with perforation; fusiform and saccular unruptured AAA morphologies across multiple size categories and anatomical locations; varying severities of hydronephrosis; small bowel obstruction from adhesive and mechanical etiologies; and free air from various etiologies and volumes. The test cohort also included cases with concurrent findings. Comprehensive stratified analyses demonstrated consistent device performance across all demographic subgroups, imaging parameters, clinical sites, and disease presentations.
7.2.5. Primary Endpoints
| Condition | Positive Cases | Total Cases | Prevalence (%) |
| --- | --- | --- | --- |
| Acute Cholecystitis | 77 | 675 | 11.4% |
| Acute Pancreatitis | 99 | 675 | 14.7% |
| Unruptured AAA | 76 | 675 | 11.3% |
| Acute Diverticulitis | 76 | 675 | 11.3% |
| Free Air | 112 | 675 | 16.6% |
| Hydronephrosis | 97 | 675 | 14.4% |
| Small Bowel Obstruction | 99 | 675 | 14.7% |
{11}
510(k) Summary
a2z
# 7.2.4. QFM Conditions
| Condition | Available Operating Points and Performance |
| --- | --- |
| Acute | High Sensitivity: Se 96.1% [89.2-98.7%], Sp 89.3% [86.6-91.5%] |
| Cholecystitis (QFM) | Sensitivity Biased: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%] |
| Total N=675, Positive=77 | Balanced: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%] |
| AUC: 0.985 [0.972-0.998] | |
| Acute | High Sensitivity: Se 98.0% [92.9-99.4%], Sp 87.8% [84.9-90.3%] |
| Pancreatitis (QFM) | Sensitivity Biased: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%] |
| Total N=675, Positive=99 | Balanced: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%] |
| AUC: 0.994 [0.985-1.000] | High Specificity: Se 92.9% [86.1-96.5%], Sp 99.8% [99.0-100.0%] |
| Acute | High Sensitivity: Se 98.7% [92.9-99.8%], Sp 89.3% [86.6-91.5%] |
| Diverticulitis (QFM) | Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%] |
| Total N=675, Positive=76 | Balanced: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%] |
| AUC: 0.995 [0.990-1.000] | High Specificity: Se 94.7% [87.2-97.9%], Sp 98.7% [97.4-99.3%] |
| Unruptured | High Sensitivity: Se 100.0% [95.2-100.0%], Sp 86.3% [83.3-88.8%] |
| AAA (QFM) | Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 95.8% [93.9-97.2%] |
| Total N=675, Positive=76 | Balanced: Se 97.4% [90.9-99.3%], Sp 97.5% [95.9-98.5%] |
| AUC: 0.995 [0.991-0.999] | |
| Hydronephrosis (QFM) | High Sensitivity: Se 89.7% [82.1-94.3%], Sp 92.9% [90.5-94.7%] |
| Total N=675, Positive=97 | |
| AUC: 0.976 [0.960-0.991] | |
# 7.2.5. QAS Conditions
{12}
510(k) Summary
a2z
| Condition | Available Operating Points and Performance |
| --- | --- |
| Small Bowel | High Sensitivity: Se 94.9% [88.7-97.8%], Sp 91.7% [89.1-93.7%] |
| Obstruction | Sensitivity Biased: Se 91.9% [84.9-95.8%], Sp 96.0% [94.1-97.3%] |
| (QAS) | Balanced: Se 88.9% [81.2-93.7%], Sp 98.1% [96.6-98.9%] |
| Total N=675, | |
| Positive=99 | |
| Free Air (QAS) | Balanced: Se 89.3% [82.2-93.8%], Sp 88.6% [85.7-91.0%] |
| Total N=675, | High Specificity: Se 88.4% [81.1-93.1%], Sp 90.8% [88.1-92.9%] |
| Positive=112 | |
7.2.6. Performance Discussion The comprehensive performance evaluation of a2z-Unified-Triage demonstrates consistently exceptional performance across all seven abdominopelvic findings, meeting or substantially exceeding FDA-specified acceptance criteria. All QAS findings (Free Air, Small Bowel Obstruction) met the acceptance criteria of sensitivity and specificity $>80\%$ , while 5 QFM findings (Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Hydronephrosis) met the acceptance criteria of AUC $>0.95$ .
The extensive stratified analyses across patient demographics (sex, age, site, state), imaging parameters (contrast status, slice thickness, equipment manufacturer), and clinical scenarios (concurrent findings) reveal robust performance maintenance with no clinically significant degradation across any subgroup. Additionally, comprehensive stratified analyses by anatomical variations and edge cases were performed for all seven conditions, with U.S. board-certified radiologists annotating clinically relevant attributes on subsets of positive cases. These analyses—including disease subtypes, confounders, and morphological features specific to each condition—demonstrate robust device performance across all clinically relevant variations, validating that the device independently and accurately detects each indication based on its primary imaging features rather than confounding concurrent findings. This validates the algorithm's generalizability and reliability across diverse clinical presentations. The triage effectiveness study confirms rapid processing times well within clinical workflow requirements, establishing a2z-Unified-Triage as a clinically viable solution for enhancing radiological workflow efficiency.
# 8. Predetermined Change Control Plan (PCCP)
This submission contains a Predetermined Change Control Plan (PCCP). The PCCP does not include provisions for implementation of adaptive algorithms that will continuously learn in the field. All algorithm modifications will be trained, validated, and locked prior to release of the software to the field. A procedure has been established for updating device labeling to inform users about changes implemented under this PCCP, including a summary of the changes, characterization of algorithm performance, and compatibility information.
The PCCP specifies possible modifications to a2z-Unified-Triage, as well as verification and validation activities to implement changes in a controlled manner such that the modified device remains as safe and effective as the originally cleared device. The PCCP applies to the seven conditions val
{13}
510(k) Summary
a2z RACIOLOGY
0
2z
510(k) Summary
idated in this 510(k) submission: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. For each condition, the PCCP covers all operating points that meet pre-specified performance criteria upon revalidation.
The planned modifications include:
- Training data volume and diversity expansion
- Training data quality and annotation refinement
- Ensemble composition optimization
- Data augmentation parameter adjustments
- Training parameter optimization
- Weight initialization method selection
- Neural network architecture component modifications
- Input data preprocessing adjustments
- Validated slice thickness range expansion
All modifications implement bounded specifications with specific parameter ranges to ensure changes remain within validated frameworks. Comprehensive validation requirements apply to all modifications, including minimum test dataset size, ground truth determination by U.S. board-certified radiologists, complete data sequestration, stratified subgroup analysis, and acceptance criteria requiring sensitivity and specificity $>80\%$ for all findings plus AUC $>0.95$ for QFM findings.
The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices. Detailed validation activities, testing methodologies, and performance requirements have been established for each modification. All changes will undergo appropriate verification and validation testing before implementation to ensure the modified device maintains the safety and effectiveness profile of the originally cleared device.
## 9 Conclusion
The subject device, a2z-Unified-Triage, and the predicate, Annalise Enterprise CTB Triage Trauma, are substantially equivalent. Both are software devices intended to aid in the prioritization and triage of radiological images using AI algorithms. They share the same fundamental scientific technology and principles of operation.
While the specific anatomical regions and findings differ, the intended use for triage, the AI-based approach, the integration into clinical workflows, and the parallel nature of operation are highly similar. Both devices aim to reduce turnaround time through preemptive triage without altering the standard of care.
The performance data for a2z-Unified-Triage demonstrates that the device is safe and effective for its intended use and performs as intended. Therefore, a2z-Unified-Triage is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.