K260750 · Varian Medical Systems, Inc. · MUJ · Jul 8, 2026 · Radiology
Device Facts
Record ID
K260750
Device Name
Virtual Cone (1.0)
Applicant
Varian Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Jul 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The Virtual Cone plan insertion tool is an ESAPI script intended to facilitate the noninvasive treatment of small, spherical intracranial targets by automating part of the planning process within the Eclipse Treatment Planning System. It enables the generation, by clinical users, of precise, conformal dose distributions comparable to those achieved with physical stereotactic cones, streamlining the treatment planning and delivery process. To achieve this, the script automates the creation of a fixed, small MLC-defined aperture by generating a standardized control point sequence with sinusoidal dose rate modulation.
Device Story
Virtual Cone is an ESAPI script accessory for Eclipse Treatment Planning System (TPS); automates planning for small, spherical intracranial targets. Input: user-defined staging plan (machine, energy, isocenter). Transformation: script automatically inserts predefined, validated templates for arc geometry, MLC-defined apertures, and delivery parameters; utilizes sinusoidal dose rate modulation to mimic physical stereotactic cones. Output: automated treatment plan ready for standard Eclipse dose calculation, review, and approval. Used in professional healthcare facilities by physicists and dosimetrists. Benefits: streamlines SRS planning workflow; provides conformal dose distributions without physical cone hardware. Operates within Eclipse environment; compatible with TrueBeam/Edge accelerators using 6X/10X High Intensity mode beams and High Definition MLCs.
Clinical Evidence
Bench testing only. No clinical data. Validation included end-to-end clinical workflow simulation (CT simulation, planning, CBCT, delivery, film/scintillator verification) using a cranial phantom. Absolute dose to isocenter agreed within 3%; film dosimetry showed >99% gamma pass rate (5%/1mm criteria).
Technological Characteristics
Software-based ESAPI script; operates within Eclipse TPS. Uses fixed MLC-based apertures and standardized control-point sequences. Compatible with TrueBeam/Edge accelerators, 6X/10X FFF beams, and HD120 MLC. Standards: ISO 13485, ISO 14971, IEC 62304, IEC 62366-1, IEC 82304-1, IEC 62083, UL ANSI 2900-1/2-1, IEC 81001-5-1.
Indications for Use
Indicated for adolescents (12-22 years) and adults (22+ years) requiring non-invasive treatment of small, spherical intracranial targets via stereotactic radiosurgery (SRS). No contraindications.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 8, 2026
Varian Medical Systems, Inc.
Lynn Allman
Senior Director, Regulatory Affairs
3100 Hansen Way
Palo Alto, California 94304
Re: K260750
Trade/Device Name: Virtual Cone (1.0)
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: MUJ
Dated: March 6, 2026
Received: March 6, 2026
Dear Lynn Allman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260750 - Lynn Allman
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260750 - Lynn Allman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260750 | ? |
| Please provide the device trade name(s). | | ? |
| Virtual Cone (1.0) | | |
| Please provide your Indications for Use below. | | ? |
| The Virtual Cone plan insertion tool is an ESAPI script intended to facilitate the noninvasive treatment of small, spherical intracranial targets by automating part of the planning process within the Eclipse Treatment Planning System. It enables the generation, by clinical users, of precise, conformal dose distributions comparable to those achieved with physical stereotactic cones, streamlining the treatment planning and delivery process. To achieve this, the script automates the creation of a fixed, small MLC-defined aperture by generating a standardized control point sequence with sinusoidal dose rate modulation. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
varian
A Siemens Healthineers Company
# Premarket Notification - 510(k) Summary
(K260750)
Traditional 510(k) Submission for Virtual Cone (1.0)
# I. Submitter's Name
Varian Medical Systems, Inc.
3100 Hansen Way
Palo Alto, CA 94304
Contact Name: Lynn, Allman, PhD., Senior Director Regulatory Affairs
Phone: (650) 424-5369
E-mail: submissions.support@varian.com
Date Prepared: July 7th, 2026
# II. Device Information
Proprietary Name: Virtual Cone (1.0)
Classification Name: Medical charged-particle radiation therapy system
Regulation Number: §892.5050
Product Code: MUJ
# III. Predicate Device
Eclipse Treatment Planning System v18.1 (K242378)
# IV. Device Description
Virtual Cone is software that is based on the technique described by Popple (Popple et al., 2018, see References section) and inserts candidate SRS photon treatment plans that allow for the treatment of small spherical, intracranial targets without the use of physical cones.
The Virtual Cone plan insertion tool enables the generation, by clinical users, of precise, conformal dose distributions comparable to those achieved with physical stereotactic cones, streamlining the treatment planning and delivery process. Virtual Cone qualifies as a medical device accessory because it is intended for use in the treatment of disease when used with the Eclipse Treatment Planning System. It is not intended to be used by itself but assists in the treatment planning process of the parent device (Eclipse Treatment Planning System).
Version 1.0 is limited to generation of Virtual Cone techniques with a dose distribution comparable to that delivered with a 5.0 mm physical stereotactic cone. It supports use with a 6X or 10X High Intensity mode beam (6FFF or 10FFF) on a TrueBeam or Edge accelerator equipped with a High Definition MLC. Virtual Cone is compatible with Eclipse Treatment Planning System v16.1, 18.0 and 18.1.
510(k) Summary
Traditional 510(k) Application
Virtual Cone
Page 1 of 7
{5}
varian
A Siemens Healthineers Company
# V. Indications for Use
The Virtual Cone plan insertion tool is an ESAPI script intended to facilitate the non-invasive treatment of small, spherical intracranial targets by automating part of the planning process within the Eclipse Treatment Planning System. It enables the generation, by clinical users, of precise, conformal dose distributions comparable to those achieved with physical stereotactic cones, streamlining the treatment planning and delivery process. To achieve this, the script automates the creation of a fixed, small MLC-defined aperture by generating a standardized control point sequence with sinusoidal dose rate modulation.
# VI. Comparison of Technological Characteristics with the Predicate Device
The subject device and the predicate device have the same technological characteristics with respect to SRS cone planning workflow, with differences limited to the automation of existing manual planning steps within Eclipse TPS. These differences do not introduce new functionality and do not represent a change in the fundamental technology used for treatment planning.
The predicate device supports SRS physical cone planning through standard Eclipse functionality, in which the user manually performs the following steps:
- creation of a treatment plan,
- definition of cone size and beam geometry,
- manual configuration of arc geometry and delivery parameters,
- plan evaluation and approval within Eclipse.
The subject device (Virtual Cone) performs the same overall clinical task but introduces automation for specific workflow steps:
- the user defines a staging plan containing machine, energy, and isocenter parameters;
- the device automatically inserts arc geometry, MLC-defined apertures, and delivery parameters using predefined, validated templates;
- the resulting treatment plan is then evaluated, calculated, and approved using the standard Eclipse workflow.
- Note: these subsequent treatment planning steps within Eclipse are identical to the predicate device workflow
Both the predicate device and the subject device are based on the same characteristics:
- Both the subject device and the predicate are radiotherapy treatment plan software applications. They are computer-based software devices used by trained medical professionals to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment.
- Both devices are used by the same type of clinical users.
- Virtual Cone is an accessory to the Eclipse Treatment Planning System and operates within the Eclipse environment using the Eclipse Scripting API.
Certain technological characteristics differ in that Virtual Cone 1.0 provides automation for the insertion of predefined stereotactic arc geometries and multileaf collimator apertures. These differences represent workflow automation of existing Eclipse TPS capabilities and do not change the fundamental principles of operation of the predicate device. All other planning functions performed by the clinical team, including dose prescription, dose calculation, plan review, and plan approval, remain exclusively within the standard Eclipse environment.
510(k) Summary
Traditional 510(k) Application
Virtual Cone
Page 2 of 7
{6}
Varian
A Siemens Healthineers Company
# **Comparison Table between Predicate Device and Subject Device**
| Feature/Specification | Predicate Device Eclipse TPS 18.1 (K242378) | Subject Device Virtual Cone 1.0 | Comparison |
| --- | --- | --- | --- |
| **General Details** | | | |
| Intended Use Statement | The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. | The Virtual Cone plan insertion tool is an ESAPI script intended to facilitate the non-invasive treatment of small, spherical intracranial targets by automating part of the planning process within the Eclipse Treatment Planning System. It enables the generation, by Clinical users, of precise, conformal dose distributions comparable to those achieved with physical stereotactic cones, streamlining the treatment planning and delivery process. To achieve this, the script automates the creation of a fixed, small MLC-defined aperture by generating a standardized control point sequence with sinusoidal dose rate modulation. | Virtual Cone 1.0 does not expand or modify the existing Eclipse TPS intended use/indications for use. It is an accessory to Eclipse TPS that automates part of the planning workflow already supported by Eclipse TPS. |
| Indications for Use | | | |
| Contraindications for Use | No contraindications for use | No contraindications for use | Identical |
| Environment | Professional healthcare facilities | Professional healthcare facilities | Identical |
| Type of Users | Qualified healthcare professionals | Qualified healthcare professionals | Identical |
| **Cone Planning Workflow** | | | |
| Type of Cone Planning | SRS physical cone planning | MLC-based SRS virtual (non-physical) cone planning | This pertains to the intended use of the subject device, Virtual Cone – that users are intended to plan radiation treatments that are comparable to those which include the use of physical cones. This occurs through the design and usage of virtual (as in software-based and algorithmically modeled) cones such that the physical cones are not utilized. |
**510(k) Summary**
Traditional 510(k) Application
Virtual Cone
Page 3 of 7
{7}
varian
A Siemens Healthineers Company
| Feature/Specification | Predicate Device Eclipse TPS 18.1 (K242378) | Subject Device Virtual Cone 1.0 | Comparison |
| --- | --- | --- | --- |
| Plan Setup | User manually creates an SRS treatment plan and selects physical cone configuration within Eclipse | User creates a photon staging plan with defined machine, energy, and isocenter as input to Virtual Cone | Substantially equivalent Both workflows begin with plan setup in Eclipse. Virtual Cone requires a staging plan but does not change intended use or planning scope. |
| Geometry Definition (Cone / Beam Geometry) | User manually defines cone size, arc geometry, couch/collimator angles, and beam parameters | User selects a predefined Virtual Cone template that defines arc geometry and beam parameters | Substantially equivalent Predicate supports manual definition of all geometric parameters. Virtual Cone applies predefined validated templates representing the same parameters. |
| Arc Creation / Configuration | User manually creates and configures arcs (angles, sequencing) | Automatically inserts standardized arcs using validated templates derived from staging plan | Substantially equivalent Predicate supports arc creation; Virtual Cone automates this exact workflow step without introducing new delivery mechanisms. |
| Collimation Approach | Achieved using physical stereotactic cones (fixed hardware geometry) | Achieved using fixed MLC-based aperture and standardized control-point sequence | Substantially equivalent Both produce small-field SRS dose distributions. Virtual Cone uses MLC-based implementation instead of hardware, within existing Eclipse capabilities. |
| Automation / Workflow Execution | All steps performed manually via Eclipse UI | Plan generation automated, including MLC aperture, arcs, and dose-rate modulation | Substantially equivalent Virtual Cone automates user-executable actions without introducing new planning capabilities or altering underlying algorithms. |
| Scripting Capability / Control | ESAPI scripting is available, but not required for cone planning; users perform planning through UI | Virtual Cone is a fixed ESAPI-based tool; scripts are not user-modifiable | Substantially equivalent Predicate provides scripting infrastructure; Virtual Cone uses it in a controlled, deterministic manner with no user programmability. |
| User Interaction | User manually defines all geometry and parameters | User selects staging plan and template; no direct parameter editing or scripting | Substantially equivalent Virtual Cone reduces manual input but does not introduce new user-controlled parameters. |
| MLC / Beam Parameter Configuration | Determined by cone hardware and user-defined beam settings | Programmatically applied from template and commissioning constraints | Substantially equivalent Parameters correspond to configurations achievable in Eclipse; Virtual Cone enforces validated machine/energy combinations. |
510(k) Summary
Traditional 510(k) Application
Virtual Cone
Page 4 of 7
{8}
varian
A Siemens Healthineers Company
| Feature/Specification | Predicate Device Eclipse TPS 18.1 (K242378) | Subject Device Virtual Cone 1.0 | Comparison |
| --- | --- | --- | --- |
| Dose Calculation | Performed using Eclipse algorithms (e.g., AAA/AXB) | Performed using same Eclipse algorithms after plan creation | Identical – Virtual Cone does not perform dose calculation. |
| Plan Review and Approval | User reviews and approves plan within Eclipse workflow | Identical Eclipse workflow used for review, QA, and approval | Identical – No change to clinical decision-making or QA workflow. |
| Treatment Delivery Compatibility | Plans delivered via standard Eclipse-integrated systems (e.g., TrueBeam) | Generated plans are delivered using same clinical workflow and systems, **see below limitations on delivery system support.** | Substantially equivalent Virtual Cone does not control delivery and produces plans compatible with existing systems. |
| Scope of Functionality | Full treatment planning workflow (including dose calculation, optimization, approval) | Limited to arc insertion only; all other planning remains in Eclipse | Substantially equivalent Virtual Cone does not expand functionality; it automates a subset of the existing workflow only. |
| **Compatibility** | | | |
| SRS Planning application | Yes | Yes, automated insertion of arcs into plan for SRS planning using 6X-FFF and 10X-FFF photon beams | Substantially equivalent Virtual Cone supports SRS treatments only using 6X-FFF and 10X-FFF photon beams |
| Treatment Delivery Support | TrueBeam and other delivery machines | TrueBeam only | Substantially equivalent Both devices support treatment delivery units. Virtual Cone is limited to a specific delivery system already supported by Eclipse TPS. |
| MLC Support | HD120 multi-leaf collimator and other MLCs | HD120 multi-leaf collimator only | Substantially equivalent Both devices support MLCs. Virtual Cone is limited to a specific MLC already supported by Eclipse TPS. |
510(k) Summary
Traditional 510(k) Application
Virtual Cone
Page 5 of 7
{9}
varian
A Siemens Healthineers Company
# VII. Summary of Performance Testing (Non-Clinical Testing)
The following performance data was provided in support of the substantial equivalence determination.
# Software Verification and Validation Testing:
Software design verification and design validation testing was performed according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System Standard, ISO 14971 Risk Management Standard, and IEC 62304 Software Life Cycle Process standard. Test results demonstrate conformance to applicable requirements specifications and assure hazard safeguards function properly. Testing also included interoperability validation of the device within the intended operating software environment alongside the Eclipse Treatment Planning software. Software verification and validation was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions."
Cybersecurity and Interoperability requirements were assessed per FDA guidance's "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (Feb 2026)" "Design Considerations and Premarket Submission Recommendations for Interoperable Medical Devices (Sept 2017)".
# Workflow and Implementation Validation Testing:
Device testing was performed to validate the accurate implementation of the Virtual Cone MLC-based SRS technique within a representative clinical case scenario. Completed alongside the validation of the full end-to-end clinical workflow was dosimetric validation of plans generated by the Virtual Cone script. Using a cranial phantom designed for SRS validation, the following steps for the clinical workflow simulation were completed by testers: CT simulation, treatment planning, image guidance (CBCT), delivery, and film/scintillator verification. Following delivery, film gamma analysis was completed.
The measured absolute dose to isocenter agreed with the Eclipse TPS calculation within 3%, and the high-resolution film dosimetry showed >99% gamma pass rates with 5%/1mm evaluation criteria (dose/distance tolerance). The implemented pre-treatment QA process demonstrated high sensitivity, with deliberate MLC aperture changes of ± 0.1 mm clearly identified.
# Human factors Validation Testing:
Usability formative testing was conducted with representative users, physicists and dosimetrists, to obtain feedback on the user interface. Participants also provided overall impressions on the device and its use with their clinical workflows. From this study, no significant safety-related concerns were identified. All identified usability issues were evaluated for impact on task-safety or existing risk controls; no residual safety issues from use error remain after validated design and labeling changes.
Test results demonstrate conformance to applicable requirements and specifications.
No animal studies or clinical tests have been included in this pre-market submission.
# Use of Consensus Standards:
The following list of FDA-recognized, voluntary consensus standards were utilized in the design and evaluation of the subject device's safety and efficacy.
ISO 14971:2019
Medical devices - Application of risk management to medical devices
510(k) Summary
Traditional 510(k) Application
Virtual Cone
Page 6 of 7
{10}
**varian**
A Siemens Healthineers Company
| IEC 62304:2006 + A1:2016 | Medical Device Software - Software Lifecycle processes |
| --- | --- |
| IEC 62366-1:2015+A1:2020 | Application of Usability Engineering to Medical Devices |
| IEC 82304-1:2016 | Health software Part 1: General requirements for product safety |
| IEC 62083:2009 | Medical electrical equipment – Requirements for the safety of radiotherapy treatment planning systems |
| UL ANSI 2900-1:2017 | Standard for Software Cybersecurity for Network-Connectable Products, Part 1: General Requirements |
| UL ANSI 2900-2-1:2017 | Software Cybersecurity for Network-Connectable Products, Part 2-1: Particular Requirements for Network Connectable Components of Healthcare and Wellness Systems |
| IEC 81001-5-1:2021 | Health Software and Health IT Systems Safety, Effectiveness and Security - Part 5-1: Security - Activities In The Product Life Cycle |
# **VIII. Determination of Substantial Equivalence to the Predicate**
Virtual Cone is an accessory to the Eclipse Treatment Planning System and does not introduce a new intended use/indication. Virtual Cone automates a portion of an existing stereotactic radiosurgery planning workflow already supported by Eclipse TPS. These do not constitute a new intended use as Virtual Cone is used to support certain existing treatment planning workflows within the Eclipse TPS.
Certain technological characteristics differ in that Virtual Cone 1.0 provides automation for the insertion of predefined stereotactic arc geometries and multileaf collimator apertures. These differences represent workflow automation of existing Eclipse TPS capabilities and do not change the fundamental principles of operation of the predicate device.
Verification and validation demonstrate that the subject device is as safe and effective as the predicate. Varian therefore believes that the subject device is substantially equivalent to the predicate device.
# **IX. References**
Popple RA, Wu X, Brezovich IA, Markert JM, Guthrie BL, Thomas EM, Bredel M, Fiveash JB. The virtual cone: A novel technique to generate spherical dose distributions using a multileaf collimator and standardized control-point sequence for small target radiation surgery. Med Phys. 2013;40(5):051707. doi:10.1118/1.4802752.
**510(k) Summary**
Traditional 510(k) Application
Virtual Cone
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.