K260729 · GuideAI Health · QAS · Jun 18, 2026 · Radiology
Device Facts
Record ID
K260729
Device Name
Vascular Assist Occlusion Triage (VAOT) (v1.0)
Applicant
GuideAI Health
Product Code
QAS · Radiology
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260729 · Jun 18, 2026
Vascular Assist Occlusion Triage (VAOT) (v1.0)
GuideAI Health
Retrospective clinical CTA studies from US-based hospital sites
A retrospective, multicenter study was conducted to evaluate the performance of the VAOT software in identifying peripheral arterial occlusions. The software's performance was compared against a reference standard established by expert radiologists.
VAOT Pivotal Study; Retrospective, blinded, multicenter study; Follow-up/Duration: Not applicable
Adult patients undergoing CTA studies of the lower extremities; Sample Size: 317 cases (165 positive, 152 negative); Number of Sites: Not specified (US-based)
Not applicable for this study
Sensitivity, specificity, and time-to-notification
VAOT is radiological computer-aided triage and notification software indicated for use in the analysis of CTA studies of adult patients. The device is intended to assist hospital networks and trained medical specialists in workflow prioritization by identifying studies with imaging patterns consistent with possible total arterial occlusion within the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries. VAOT analyzes CTA images in parallel with the on-going standard-of-care interpretation workflow. When predefined criteria are met, VAOT generates a notification to alert the user that the case may warrant prioritization. Notifications may include compressed preview images provided for informational purposes only and are not intended for diagnostic use. The results of VAOT are not intended for primary diagnosis, screening, disease detection, or patient management decisions. The interpreting clinician remains responsible for viewing the full imaging study according to standard clinical practice. VAOT does not detect, localize, characterize, measure, or diagnose vascular disease and does not alter the original medical images.
Device Story
VAOT is a cloud-based radiological triage software that analyzes CTA images to identify potential total arterial occlusions in the lower extremities (aorta to popliteal). A 'Gateway' component listens to the PACS, de-identifies CTA studies, and transmits them to the cloud. A deep neural network pipeline processes the images; if an occlusion is suspected, the system generates a notification and a compressed preview image. This preview is sent back to the Gateway, embedded into a DICOM Secondary Capture, and routed to the PACS. The device operates in parallel with standard-of-care workflows; it does not remove cases from worklists or alter original images. Radiologists use the notification to prioritize their reading queue. The device is intended for use in hospital networks by trained medical specialists to improve workflow efficiency; it is not a diagnostic tool and does not replace standard clinical interpretation.
Clinical Evidence
Retrospective, blinded, multicenter study of 317 cases (165 positive, 152 negative) from US clinical sites. Ground truth established by 3 board-certified radiologists using majority voting. Primary endpoints: sensitivity 0.915 (95% CI: 0.862, 0.953) and specificity 0.901 (95% CI: 0.842, 0.944). Secondary endpoint: time-to-notification < 5 minutes. Dataset included diverse scanner manufacturers and patient demographics.
Technological Characteristics
Software-only device; utilizes deep neural networks for image analysis. Inputs: DICOM CTA images. Outputs: DICOM Secondary Capture notifications. Connectivity: PACS-integrated via Gateway listener; cloud-based processing. Sterilization: N/A (software). Cybersecurity: Risk-assessed per 2025 FDA guidance.
Indications for Use
Indicated for adult patients (22+ years) undergoing CTA studies to assist hospital networks and medical specialists in workflow prioritization by identifying imaging patterns consistent with possible total arterial occlusion in the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries. Not for pediatric use.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
GuideAI
% J. David Giese
Partner
Innolitics, LLC
1101 W. 34th St. #550
Austin, TX 78705
Re: K260729
Trade/Device Name: Vascular Assist Occlusion Triage (VAOT)
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: May 22, 2026
Received: May 22, 2026
Dear J. David Giese :
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260729 - J. David Giese
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260729 - J. David Giese
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and
Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260729 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Vascular Assist Occlusion Triage (VAOT) (v1.0) | | | |
| Please provide your Indications for Use below. | | | ? |
| VAOT is radiological computer-aided triage and notification software indicated for use in the analysis of CTA studies of adult patients. The device is intended to assist hospital networks and trained medical specialists in workflow prioritization by identifying studies with imaging patterns consistent with possible total arterial occlusion within the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries. | | | |
| VAOT analyzes CTA images in parallel with the on-going standard-of-care interpretation workflow. When predefined criteria are met, VAOT generates a notification to alert the user that the case may warrant prioritization. | | | |
| Notifications may include compressed preview images provided for informational purposes only and are not intended for diagnostic use. | | | |
| The results of VAOT are not intended for primary diagnosis, screening, disease detection, or patient management decisions. The interpreting clinician remains responsible for viewing the full imaging study according to standard clinical practice. | | | |
| VAOT does not detect, localize, characterize, measure, or diagnose vascular disease and does not alter the original medical images. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
K260729

510(k) Summary
## 1. ADMINISTRATIVE INFORMATION
| Parameter | Value |
| --- | --- |
| Company Name | GuideAI Health, Inc. |
| Address | GuideAI Health, Inc. 45 Temple Street Unit 106 Boston, MA 02114 |
| Phone Number | (857) 276-8823 |
| Fax Number | N/A |
| Company Representative | Raj Shah MD, MBA, CEO |
| Email | rajshah@guideaihealth.com |
| Primary Correspondent | J. David Giese, Partner |
| Primary Correspondent Email | jdgiese@innolitics.com |
| Date Summary Prepared | February 24, 2026 |
## 2. SUBJECT DEVICE INFORMATION
| Parameter | Value |
| --- | --- |
| Trade Name | Vascular Assist Occlusion Triage (VAOT) v1.0 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS |
| Regulation Number | 21 CFR §892.2080 |
| Class | Class II |
| Panel | Radiology |
510(k) Summary
Page 1 of 9
{5}
GuideAI Health
510(k) Summary
### 3. PREDICATE DEVICE INFORMATION
| Parameter | Value |
| --- | --- |
| Predicate Device Name | Briefcase LVO Triage |
| Predicate Device K Number | K220709 |
| Product Code | QAS |
| Regulation Number | 21 CFR §892.2080 |
| Class | Class II |
| Panel | Radiology |
### 4. DEVICE DESCRIPTION
VAOT is radiological computer-assisted triage and notification software indicated for use in the analysis of CTA studies of adult patients. The software uses machine-learning techniques to automatically analyze CTA studies and to alert the PACS/RIS workstation once images with features suggestive of arterial occlusion are identified.
Using VAOT, a radiologist is able to review studies with features suggestive of arterial occlusion earlier than in the standard of care workflow.
CTA studies are captured by a CT imaging system and deposited on the PACS. The Gateway instance, a component of VAOT that runs acting as a PACS listener, receives a copy of the CTA DICOM study, checks the DICOM meta data to confirm it can be processed, de-identifies it, and securely transmits it to the VAOT cloud for processing.
The VAOT cloud receives the DICOM image(s) and determines whether a possible occlusion should be flagged using an image-processing pipeline consisting of several deep neural networks. If one or more possible occlusions are flagged, then an axial slice from the original study is selected as a “preview image” for the radiologist, and is sent back to the Gateway instance.
The Gateway instance re-associates the preview image with the patient, embeds it into a DICOM Secondary Capture, and routes it to the DICOM Destination (e.g., the PACS).
The PACS receives the DICOM Secondary Capture containing the preview image and associates it with the original study. The PACS can then use the Secondary Capture to notify radiologists of the suspected occlusion.
510(k) Summary
Page 2 of 9
{6}
GuideAI Health
510(k) Summary
In summary, the VAOT device is intended to enable a passive notification through the PACS to the radiologist indicating the existence of a case that may potentially benefit from that radiologist's prioritization. The device aims to aid in prioritization and triage of radiological medical images only.
## 5. INTENDED USE / INDICATIONS FOR USE
VAOT is radiological computer-aided triage and notification software indicated for use in the analysis of CTA studies of adult patients. The device is intended to assist hospital networks and trained medical specialists in workflow prioritization by identifying studies with imaging patterns consistent with possible total arterial occlusion within the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries.
VAOT analyzes CTA images in parallel with the on-going standard-of-care interpretation workflow. When predefined criteria are met, VAOT generates a notification to alert the user that the case may warrant prioritization.
Notifications may include compressed preview images provided for informational purposes only and are not intended for diagnostic use.
The results of VAOT are not intended for primary diagnosis, screening, disease detection, or patient management decisions. The interpreting clinician remains responsible for viewing the full imaging study according to standard clinical practice.
VAOT does not detect, localize, characterize, measure, or diagnose vascular disease and does not alter the original medical images.
## 6. CONTRAINDICATIONS FOR USE
The VAOT device was not tested on pediatric patients.
## 7. SUBSTANTIAL EQUIVALENCE DISCUSSION
The subject device, Vascular Assist Occlusion Triage (VAOT), and the predicate device, Aidoc BriefCase LVO Triage (K220709), share the same intended use as radiological computer-aided triage and notification software under 21 CFR §892.2080 (product code QAS). Both are Class II, notification-only tools that analyze CTA studies in adult patients using deep-learning algorithms and generate alerts to assist hospital networks and trained clinicians with workflow prioritization. Both devices operate entirely in parallel with the standard-of-care interpretation workflow: neither removes cases from the reading queue, alters original DICOM images, directs attention to specific locations in the diagnostic images, nor provides primary diagnostic output. Notifications, which may include compressed preview images for informational purposes only, are explicitly not intended for diagnostic use. The difference between the two devices lies
510(k) Summary
Page 3 of 9
{7}
GuideAI Health
510(k) Summary
in the vascular territory evaluated—intracranial vessels for the predicate versus lower-extremity arteries (aorta through popliteal) for the subject device—which represents a difference in indications for use within the same clinical purpose of rapid identification and prioritization of suspected acute vascular occlusion, rather than a new intended use.
The subject and predicate devices share nearly identical technological characteristics. Both are software packages that:
- Utilize artificial intelligence and deep learning algorithms to analyze DICOM CTA images.
- Receive images from DICOM-compliant imaging devices (CT scanners) or PACS.
- Process images automatically to detect specific findings.
- Notify clinicians of suspected findings to facilitate triage.
- Provide compressed preview images for informational purposes only, which are not intended for diagnostic use.
- Do not alter the original medical images or remove cases from the standard worklist.
The devices differ in the anatomical region and specific pathology targeted:
- The **Predicate Device** is indicated for the analysis of Head CTA images to detect Large Vessel Occlusions (LVO) and Medium Vessel Occlusions (MeVO) associated with stroke.
- The **Subject Device** is indicated for the analysis of CTA images of the lower extremities to detect total arterial occlusions in the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries.
The primary difference between the subject and predicate devices is the anatomical location of the suspected findings (Lower Extremities vs. Head). This difference requires the subject device's AI algorithms to be trained on a different dataset specific to lower extremity vasculature. However, this difference does not raise new questions of safety and effectiveness.
Both devices serve the same triage function within their respective clinical contexts. The risks—false positives (unnecessary prioritization) or false negatives (standard of care review applies)—remain similar. Neither device replaces the radiologist's interpretation, and both rely on the clinician to view the full original study for diagnosis. Standalone performance testing, along with verification and validation, demonstrates that the subject device performs as intended for its specific anatomical region.
A table comparing the key features of the subject and the primary predicate devices is provided below.
| Parameter | VAOT Subject Device | Briefcase LVO Triage (K220709) Predicate Device | Discussion |
| --- | --- | --- | --- |
| Product Code | QAS | QAS | Same |
510(k) Summary
Page 4 of 9
{8}
**GuideAI Health**
**510(k) Summary**
| Parameter | VAOT Subject Device | Briefcase LVO Triage (K220709) Predicate Device | Discussion |
| --- | --- | --- | --- |
| Regulation | 21 CFR §892.2080 | 21 CFR §892.2080 | Same |
| Intended Use | Radiological computer aided triage and notification | Radiological computer aided triage and notification | Same |
| Indications for Use | VAOT is radiological computer-aided triage and notification software indicated for use in the analysis of CTA studies of adult patients. The device is intended to assist hospital networks and trained medical specialists in workflow prioritization by identifying studies with imaging patterns consistent with possible total arterial occlusion within the aorta, common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries. VAOT analyzes CTA images in parallel with the on-going standard-of-care interpretation workflow. When predefined criteria are met, VAOT generates a notification to alert the user that the case may warrant prioritization. Notifications may include compressed preview images provided for informational purposes only and are not intended for diagnostic use. The results of VAOT are not intended for primary diagnosis, screening, disease detect, or patient management decisions. The interpreting | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of complete Large Vessel Occlusion (LVO) - MCA-M1, PCA-P1, ACA-A1, ICA, Basilar; and Medium Vessel Occlusions (MeVO) - MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, and Vertebral-V4. BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. | Both devices are radiological computer-aided triage and notification software that analyze CTA images in parallel with the standard-of-care interpretation workflow and provide notification with compressed preview images intended for informational use only. Each device is intended to assist appropriately trained clinical users with workflow prioritization by flagging studies with imaging patterns consistent with suspected vascular occlusion in its indicated anatomy. |
510(k) Summary
Page 5 of 9
{9}
GuideAI Health
510(k) Summary
| Parameter | VAOT Subject Device | Briefcase LVO Triage (K220709) Predicate Device | Discussion |
| --- | --- | --- | --- |
| | clinician remains responsible for viewing the full imaging study according to standard clinical practice. VAOT does not detect, localize, characterize, measure, or diagnose vascular disease and does not alter the original medical images. | The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. | |
| User Population | Hospital networks and appropriately trained medical specialists | Hospital networks and appropriately trained medical specialists | Same |
| Anatomical Site | Lower Extremities | Head | The devices are indicated for different anatomical sites. |
| Disease/Condition | Suspected Peripheral Arterial Occlusion | Suspected Stroke/LVO | Both devices attempt to identify occlusions to the large vessels in their indicated anatomical sites. |
| Input Data | CTA | CTA | Same |
| Data Acquisition | Acquires medical image data from DICOM-compliant imaging devices and modalities | Acquires medical image data from DICOM-compliant imaging devices and modalities | Same |
| Preview Images | Preview images provided for informational purposes only and not intended for diagnostic use beyond notification. | Preview images provided for informational purposes only and not intended for diagnostic use beyond notification. | Same |
| Algorithm | Artificial intelligence algorithm with database of images | Artificial intelligence algorithm with database of images | Same |
| Patient Population | Adult | Adult | Same |
510(k) Summary
Page 6 of 9
{10}
GuideAI Health
510(k) Summary
| Parameter | VAOT Subject Device | Briefcase LVO Triage (K220709) Predicate Device | Discussion |
| --- | --- | --- | --- |
| Images format | DICOM | DICOM | Same |
| Notification only, parallel workflow tool | Yes | Yes | Same |
| Notification delivery | Using DICOM Secondary Captures | Using a Standalone Desktop Application | Both devices provide notifications to support workflow prioritization; however, the mechanisms differ (DICOM secondary captures versus a standalone desktop application). Despite different delivery approaches, both present compressed preview images for informational purposes to facilitate timely user awareness. |
| Interference with standard workflow | No. No cases are removed from Worklist or deprioritized. | No. No cases are removed from Worklist or deprioritized | Same |
### 8. PERFORMANCE TESTING
### 8.1. Pivotal Study Summary
GuideAI conducted a retrospective, blinded, multicenter study with the VAOT software to evaluate the software's performance in identifying peripheral arterial occlusions in 317 cases from US-based clinical sites. The study compared the software's performance to the ground truth, as determined by 3 expert US board certified Radiologists, using majority voting. The cases collected for the pivotal dataset were all distinct from the cases used to train the algorithm.
The study included 317 cases (165 positive for occlusion, 152 negative). The dataset included cases from multiple scanner manufacturers and covered a broad range of patient ages. The gender distribution was 180 Male and 137 Female. The study sample was enriched for cases with aortic occlusions.
510(k) Summary
Page 7 of 9
{11}
GuideAI Health
510(k) Summary
### 8.2. Primary Endpoint
VAOT met the primary endpoints of sensitivity and specificity with an 80% performance goal.
Sensitivity was 0.915 (95% CI: 0.862, 0.953) and specificity was 0.901 (95% CI: 0.842, 0.944).
### 8.3. Secondary Endpoint
VAOT met the secondary endpoint of time-to-notification less than 5 minutes.
The time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.
### 8.4. Clinical Subgroups and Confounders
Frequency Distribution of Manufacturer:
| Manufacturer | N | % |
| --- | --- | --- |
| GE MEDICAL SYSTEMS | 192 | 60.6% |
| SIEMENS | 73 | 23.0% |
| Philips | 14 | 4.4% |
| Canon/Toshiba | 37 | 11.7% |
| Other | 1 | 0.3% |
| Total | 317 | 100.0% |
# Clinical Subgroups And Confounders:
- Pathologies present in negative cases: Non-occlusive arterial disease (mimics); Post-operative and device confounders; Out-of-scope vessels/locations; Other vascular pathologies; Acquisition/technical quality issues.
- Vessel Occlusions: Aorta, Common Iliac, External Iliac, Common Femoral, Superficial Femoral, and Popliteal. The set was enriched with 21 positive vessel aortic occlusions cases.
510(k) Summary
Page 8 of 9
{12}
GuideAI Health
510(k) Summary
### 8.5. Software Validation
Software verification and validation testing were conducted, and documentation was provided as recommended by 2023 FDA Guidance “Content of Premarket Submissions for Device Software Functions”.
The software verification and validation testing verified that the design requirements were successfully met. The Intended use and user needs were successfully validated.
### 8.6. Cybersecurity
GuideAI conducted a cybersecurity risk assessment and testing in accordance 2025 FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”.
### 9. CONCLUSIONS
VAOT shares the same intended use, technological characteristics, and principles of operation as the predicate device. The differences in anatomical indication are addressed through standalone performance testing. We believe the introduction of VAOT into the U.S. market should not raise any safety or efficacy concerns. The device demonstrates substantial equivalence to the predicate device.
510(k) Summary
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.