K220709 · Aidoc Medical , Ltd. · QAS · Oct 7, 2022 · Radiology
Device Facts
Record ID
K220709
Device Name
BriefCase
Applicant
Aidoc Medical , Ltd.
Product Code
QAS · Radiology
Decision Date
Oct 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K220709 · Oct 7, 2022
BriefCase
Aidoc Medical , Ltd.
Retrospective clinical head CTA images from 5 US-based clinical sites
The retrospective clinical dataset was used to evaluate the sensitivity, specificity, and time-to-notification performance of the BriefCase software in a real-world clinical setting.
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of complete Large Vessel Occlusion (LVO) - MCA-M1, PCA-P1, ACA-A1, ICA, Basilar; and Medium Vessel Occlusions (MeVO) - MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, and Vertebral-V4. BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Device Story
BriefCase is a radiological triage software; processes head CTA DICOM images; utilizes deep-learning AI to detect suspected Large Vessel Occlusions (LVO) and Medium Vessel Occlusions (MeVO). System architecture: Aidoc Hospital Server (AHS) for acquisition; Aidoc Cloud Server (ACS) for processing; Aidoc Desktop Application for user notification. Operates in parallel to standard-of-care (SOC) workflow; does not alter original images; does not remove or deprioritize cases in PACS. Upon detection, sends notification to desktop application; provides unannotated, compressed, low-quality preview images for informational purposes only. Used by radiologists and medical specialists in hospital networks to prioritize worklists. Benefits patient by reducing turnaround time for critical findings; clinicians must perform primary diagnosis using full-resolution images in PACS.
Clinical Evidence
Retrospective, blinded, multicenter study (n=342 cases) from 5 US sites. Compared software performance against ground truth (3 board-certified neurologists). Primary endpoints: sensitivity 91.3% (95% CI: 83.6%-96.2%) and specificity 85.6% (95% CI: 80.6%-89.7%). Secondary metrics: NPV 98.9%, PPV 41.3%, PLR 6.34, NLR 0.10. Mean time-to-notification was 2.23 minutes, demonstrating comparability to predicate. Additional operating point (AOP1) achieved sensitivity 91.3% and specificity 85.2%.
Technological Characteristics
Software-only device; runs on Windows Server 2012 64-bit. Components: PostgreSQL, DICOM module, Image Processing Application. Architecture: AHS (orchestrator/acquisition), ACS (cloud processing), Desktop Application (integration). Deep-learning AI algorithm. Connectivity: DICOM 3.0 compliant. Input: Head CTA (slice thickness 0.5-1.25mm). Output: Notification and compressed preview images.
Indications for Use
Indicated for adults and transitional adolescents (18+) undergoing head CTA. Assists hospital networks and specialists in triage by flagging suspected complete Large Vessel Occlusions (MCA-M1, PCA-P1, ACA-A1, ICA, Basilar) and Medium Vessel Occlusions (MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, Vertebral-V4). Contraindicated for technically inadequate scans (motion/metal artifacts, suboptimal bolus, inadequate FOV).
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Aidoc Medical, Ltd. % John J. Smith Partner Hogan Lovells US LLP 555 Thirteenth Street, N.W. WASHINGTON DC 20004
# Re: K220709
October 7, 2022
Trade/Device Name: BriefCase Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: September 8, 2022 Received: September 8, 2022
Dear John J .Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K220709
Device Name
#### BriefCase
Indications for Use (Describe)
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of complete Large Vessel Occlusion (LVO) - MCA-M1, PCA-P1, ACA-A1, ICA, Basilar; and Medium Vessel Occlusions (MeVO) - MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, and Vertebral-V4.
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "aidoc.". The word "aidoc" is written in a sans-serif font, with the letters in blue. The period at the end of the word is orange. The logo is simple and modern.
510(k) Summary Aidoc Medical. Ltd.'s BriefCase K220709
K220709
#### Submitter:
| Aidoc Medical, Ltd.<br>92 Yigal Alon St.<br>Tel-Aviv, Israel<br>Phone: | +972-73-7946870 |
|------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Contact Person: | Amalia Schreier, LLM |
| Date Prepared: | September 8, 2022 |
| Name of Device: | BriefCase |
| Classification Name: | Radiological computer-assisted triage and notification software<br>device |
| Requlatory Class: | Class II |
| Product Code: | QAS (21 C.F.R. 892.2080) |
| Primary Predicate Device: | BriefCase (LVO triage, K203508) |
#### Device Description
BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition: (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.
DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the desktop application, thereby facilitating triage and prioritization by the user. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.g., a chest radiologist may choose to filter out alerts on VO cases, and a neuro-radiologist would opt to divert PE alerts. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out alerts by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc desktop application.
The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed
{4}------------------------------------------------
preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.
Presenting the users with worklist prioritization facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
### Intended Use / Indications for Use
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of complete Large Vessel Occlusion (LVQ) - MCA-M1, PCA-P1, ACA-A1, ICA, Basilar; and Medium Vessel Occlusions (MeVO) - MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, and Vertebral-V4.
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
# Comparison of Technological Characteristics
The subject BriefCase for VO triage and primary predicate BriefCase for LVO triage (K203508) are identical in all aspects and defer only with respect to the training of the algorithm on VO vs LVO (M1) images.
Both devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated software packages for use with DICOM 3.0 compliant CT scanners, PACS, and radiology workstations.
Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. They differ only with regard to the training of the algorithm - the predicate device is indicated for Large Vessel Occlusions (on M1) triage and the current VO device is indicated for all brain vessel occlusions. In addition, they also differ in the image slice thickness they process (the predicate is indicated with image slice thickness of 0.5 mm -1.0 mm, and the current VO device is indicated for use with slice thickness of 0.5 mm - 1.25 mm). Both devices are intended to provide the specialists with notifications and unannotated low-quality preview images of suspect studies for the purpose of preemptive triage.
The subject and predicate BriefCase devices raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading
{5}------------------------------------------------
queue nor deprioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases. A table comparing the key features of the subject and the primary predicate devices is provided below.
| | Predicate Device<br>Aidoc Briefcase (K203508) | Subject Device<br>Aidoc Briefcase (K220709) |
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | BriefCase is a radiological computer<br>aided triage and notification software<br>indicated for use in the analysis of head<br>CTA images. The device is intended to<br>assist hospital networks and<br>appropriately trained medical<br>specialists in workflow triage<br>by flagging and communication of<br>suspected positive findings of Large<br>Vessel Occlusion (LVO) pathologies.<br><br>BriefCase uses an artificial intelligence<br>algorithm to analyze images and<br>highlight cases with detected findings<br>on a standalone desktop application in<br>parallel to the ongoing standard of care<br>image interpretation. The user is<br>presented with notifications for cases<br>with suspected findings. Notifications<br>include compressed preview images<br>that are meant for informational<br>purposes only and not intended for<br>diagnostic use beyond notification. The<br>device does not alter the original<br>medical image and is not intended to be<br>used as a diagnostic device.<br><br>The results of BriefCase are intended<br>to be used in conjunction with other<br>patient information and based on their<br>professional judgment, to assist with<br>triage/prioritization of medical images.<br>Notified clinicians are responsible for<br>viewing full images per the standard of<br>care. | BriefCase is a radiological computer<br>aided triage and notification software<br>indicated for use in the analysis of head<br>CTA images in adults or transitional<br>adolescents aged 18 and older.<br>The device is intended to<br>assist hospital networks and<br>appropriately trained medical<br>specialists in workflow triage<br>by flagging and communication of<br>suspected positive findings of complete<br>Large Vessel Occlusion (LVO) - MCA-<br>M1, PCA-P1, ACA-A1, ICA, Basilar;<br>and Medium Vessel Occlusions<br>(MeVO) - MCA-M2, MCA-proximal M3,<br>PCA-P2, PCA-proximal P3, ACA-A2,<br>ACA-proximal A3, and Vertebral-V4.<br><br>BriefCase uses an artificial intelligence<br>algorithm to analyze images and<br>highlight cases with detected findings<br>on a standalone desktop application in<br>parallel to the ongoing standard of care<br>image interpretation. The user is<br>presented with for cases with<br>suspected findings. Notifications<br>include compressed preview images<br>that are meant for informational<br>purposes only and not intended for<br>diagnostic use beyond notification. The<br>device does not alter the original medical image<br>and is not intended to be used as a<br>diagnostic device.<br><br>The results of BriefCase are intended<br>to be used in conjunction with other<br>patient information and based on their<br>professional judgment, to assist with<br>triage/prioritization of medical images. |
| | Predicate Device<br>Aidoc Briefcase (K203508) | Subject Device<br>Aidoc Briefcase (K220709) |
| | | Notified clinicians are responsible for<br>viewing full images per the standard of care. |
| User population | Hospital networks and appropriately trained medical specialists | Hospital networks and appropriately trained medical specialists |
| Anatomical region of interest | Head | Head |
| Data acquisition protocol | Head CTA | Head CTA |
| Notification-only<br>(/notification alerts), parallel workflow tool | Yes | Yes |
| Images format | DICOM | DICOM |
| Interference with standard workflow | No. No cases are removed from Worklist or deprioritized. | No. No cases are removed from desktop app or deprioritized |
| Inclusion/Exclusion criteria | Inclusion criteria Head CTA protocol with a 64-slice scanner or higher. Scans performed on adults/transitional adults ≥ 18 years of age. Slice thickness 0.5 mm – 1.0 mm. Exclusion Criteria All scans that are technically inadequate, including motion artifacts, severe metal artifacts, suboptimal bolus timing or an inadequate field of view. | Inclusion criteria Head CTA protocol with a 64-slice scanner or higher. Scans performed on adults/transitional adults ≥ 18 years of age. Slice thickness 0.5 mm – 1.25 mm. Exclusion Criteria All scans that are technically inadequate, including motion artifacts, severe metal artifacts, suboptimal bolus timing or an inadequate field of view. |
| Algorithm | Artificial intelligence algorithm with database of images. | Artificial intelligence algorithm with database of images. |
| Structure | AHS module (image acquisition); ACS module (image processing)); | AHS module (orchestrator, image acquisition); ACS module (image processing)); |
| Predicate Device<br>Aidoc Briefcase (K203508) | Subject Device<br>Aidoc Briefcase (K220709) | |
| - Aidoc Worklist application for<br>workflow integration (worklist and<br>non-diagnostic Image Viewer). | - Aidoc Desktop application for<br>workflow integration (feed and non-<br>diagnostic Image Viewer). | |
## Table 1. Key feature comparison
{6}------------------------------------------------
{7}------------------------------------------------
### Performance Data
## Pivotal Study Summary
Aidoc conducted a retrospective, blinded, multicenter, study with the BriefCase software to evaluate the software's performance in identifying head CTA images containing Vessel Occlusion (VO) in 342 cases from 5 US-based clinical sites. The study compared the software's performance to the ground truth, as determined by 3 expert US board certified Neurologists reviewers, using majority voting. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm.
Primary endpoints were sensitivity and specificity with an 80% performance qoal.
Secondary endpoints were BriefCase time-to-notification compared to the predicate device, Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed.
#### Primary Endpoint
Sensitivity and specificity exceeded the 80% performance goal. Sensitivity was 91.3% (95% Cl: 83.6%, 96.2%) and specificity was 85.6% (95% CI: 80.6%, 89.7%).
#### Secondary Endpoint
In addition, the time-to-notification metric observed for the BriefCase software in the three medical centers was compared to the equivalent metric of the predicate devices.
- -The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.
The BriefCase time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate BriefCase LVO.
The time-to-notification results obtained for the subject BriefCase device show comparability with the primary predicate with regard to the standard of care review. The BriefCase mean time-to-notification for VO was 2.23 minutes (95% Cl: 2.22-2.23). The time-to-notification for the predicate BriefCase LVO was 3.80 minutes (95% Cl: 3.60-4.00).
{8}------------------------------------------------
| Time -to-<br>notification | Mean Estimate | 95%<br>Lower CL | 95%<br>Upper CL | Median |
|----------------------------------------------|---------------|-----------------|-----------------|--------|
| Predicate<br>K203508Time-<br>to-notification | 3.80 | 3.60 | 4.00 | 3.8 |
| BriefCase VO<br>Time-to-<br>notification | 2.23 | 2.22 | 2.23 | 2.03 |
Table 2. Time-to- Notification Comparison for BriefCase Devices (minutes)
NPV was 98.9% (95% CI: 4.66%- 99.4%) and PPV was 41.3% (95% CI: 34.1%- 49.0%).
PLR was 6.34% (95% Cl: 97.9%- 8.63%) and NLR was 0.10% (95% Cl: 0.05%- 0.20%).
Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase for VO triage the radiologists may have the same benefit in time saving as with the BriefCase for LVO triage.
As can be seen in Table 3 the mean age of patients whose scans were reviewed for VO was 64.8 years, with standard deviation of 17.1 years. Gender distribution was 42.4% male, and 56.4% female (Table 4). Scanner distribution can also be found in Table 5 below.
## Table 3. Descriptive Statistics for Age
| | Mean | Std | Min | Median | Max | N |
|-------------|------|------|-----|--------|-----|-----|
| Age (Years) | 64.8 | 17.1 | 18 | 65.0 | 90 | 342 |
| Ground<br>Truth<br>Results | Gender | | | | | | All | |
|----------------------------|--------|-------|--------|-------|-------|------|-----|-----|
| | Male | | Female | | Other | | All | |
| | N | % | N | % | N | % | N | % |
| Positive | 37 | 40.2% | 54 | 58.7% | 1 | 1.1% | 92 | 100 |
| Negative | 108 | 43.2% | 139 | 55.6% | 3 | 1.2% | 250 | 100 |
| All | 145 | 42.4% | 193 | 56.4% | 4 | 1.2% | 342 | 100 |
{9}------------------------------------------------
| Manufacturer | N | % |
|-----------------------|-----|--------|
| GE MEDICAL<br>SYSTEMS | 124 | 36% |
| TOSHIBA | 72 | 21% |
| Philips | 81 | 24% |
| SIEMENS | 65 | 19% |
| Total | 342 | 100.0% |
## Table 5. Frequency Distribution of Manufacturer
## Clinical Subgroups And Confounders:
- Pathologies present in negative cases: Vessel Occlusion (Stenosis/narrowing); Head & neck trauma; Fully negative; Vascular; Vessel Occlusion (anatomically out of product definition), Vessel Occlusion (chronic); Vessel Occlusion (partial).
- -Vessel Occlusions: MCA-M1, MCA-M2, MCA-M3, ICA, Basilar, PCA-P1, PCA-P2, PCA-P3, ACA-A1, ACA-A2, ACA-A3, and Vertebral-V4. An additional enriched dataset was collected for vessel segment-subgroup analysis on 165 positive vessel occlusions cases.
## Additional Operating Point
In addition to the default operating point that was selected to maximize both sensitivity and specificity, one additional operating point (AOP) was selected to maximize sensitivity while maintaining a lower bound 95% confidence interval of 80%:
AOP1: Sensitivity was 91.3% (95% Cl: 83.58%-96.17%) and Specificity was 85.2% (95% Cl: 80.18%-89.36%).
In summary, performance goals were achieved for the default and additional operating point. Combined with the comparison results of time-to-notification metric with the predicate device, this data establishes the achievement by the subject BriefCase of preemptive triage in the range of several minutes.
#### Conclusions
The subject BriefCase for VO triage and the predicate BriefCase for LVO triage are intended to aid in prioritization and triage of radiological images for the indications of Vessel Occlusions. Both devices are software packages with the same technological characteristics and principles of operation, incorporating deep learning AI algorithms that process images, and software to send notifications and display unannotated compressed low-quality preview images. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
{10}------------------------------------------------
Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.
The subject BriefCase device for VO triage is thus substantially equivalent to the predicate BriefCase for LVO triage.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.