Neowise is a medical imaging software intended for use in dental and medical applications as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to elaborate (retrieve, process, render, diagnose, review, and for non-medical application as store, print and distribute) images of both adults and pediatric patients. Neowise is also software for simulating treatment options. Neowise is also intended to be used for monitoring, recording, storing and displaying mandibular jaw position related to maxilla. The software is for use by authorized healthcare professionals. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery.
Device Story
Neowise is medical imaging software for dental/medical applications; runs on standard Windows PC workstation. Inputs include 2D/3D image files, optical impressions, 3D face photos, and DICOM data from extraoral/intraoral imaging devices or TWAIN-compatible sources. Software enables acquisition, visualization, transfer, analysis, and processing of cranio-maxillofacial images. Features include brightness/contrast adjustment, image cropping, annotations, and measurement tools (segment/polyline). Supports treatment simulation and monitoring/recording of mandibular jaw position relative to maxilla. Output includes processed images, reports, and exported data for third-party software. Used by radiologists, clinicians, and referring physicians to support diagnosis and treatment planning. Benefits include enhanced visualization and standardized management of patient imaging data to assist clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Software validation conducted per IEC 62304 and FDA guidance on software validation and risk management (ISO 14971). Verification and validation testing confirmed that all functional requirements were met.
Technological Characteristics
Software-only device running on standard Windows PC. Supports DICOM, 2D/3D image formats, optical impressions, and 3D face photos. Features include image processing (brightness/contrast, crop), annotations, and measurement tools. Connectivity via TWAIN and third-party software export. Developed under ISO 13485 and 21 CFR 820; compliant with IEC 62304 for software lifecycle.
Indications for Use
Indicated for radiologists, clinicians, and referring physicians to display, visualize, and process 2D/3D dental and medical images (e.g., projection radiography, CBCT) for adult and pediatric patients. Includes treatment simulation and monitoring/recording of mandibular jaw position relative to maxilla. Requires authorized healthcare professional use; surgical planning requires maxillofacial surgery training.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 4, 2026
Cefla S.C.
% Ilenia Muccione
Regulatory Affairs Specialist
Via Selice Provinciale 23/A
Imola (Bo), BO 40026
ITALY
Re: K260716
Trade/Device Name: Neowise
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: March 5, 2026
Received: March 5, 2026
Dear Ilenia Muccione:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260716 - Ilenia Muccione
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260716 - Ilenia Muccione
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260716 | |
| Device Name Neowise | |
| Indications for Use (Describe) Neowise is a medical imaging software intended for use in dental and medical applications as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to elaborate (retrieve, process, render, diagnose, review, and for non-medical application as store, print and distribute) images of both adults and pediatric patients. Neowise is also software for simulating treatment options. Neowise is also intended to be used for monitoring, recording, storing and displaying mandibular jaw position related to maxilla. The software is for use by authorized healthcare professionals. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
CEFLA S.C.
Neowise
510(k) Premarket Notification
# 510(k) SUMMARY, AS REQUIRED BY CFR 807.92 K260716
| Submitter's Name: | CEFLA S.C. |
| --- | --- |
| Address: | Via Selice Provinciale 23/a Imola, BO 40026 ITALY Tel. +39 0542 653111 Fax +39 0542 653444 |
| Establishment Registration Number: | 3006610845 |
| Summary Preparation Date: | March 05th,2026 |
| Contact Person: | Ilenia Muccione, Regulatory Affairs |
| Telephone Number: | +39 0542 653459 |
| Email: | regulatory@cefla.it |
| Trade/Device name: | Neowise |
| Classification Name: | Device: System, Image Processing, Radiological Medical image management and processing system Device Class: II Product Code: LLZ Regulation Number: 21 CFR §892.2050 |
| Description: | Neowise is a medical imaging software intended for use in dental and medical applications as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to elaborate (retrieve, process, render, diagnose, review, and for non-medical application as store, print and distribute) images of both adults and pediatric patients. |
| | Here reported below the key functionalities of Neowise software: • acquisition • visualization • transfer • processing • analysis |
| | of dental and cranio-maxillofacial image information. |
| | Image-generating systems that can be used with the software include extraoral and intraoral devices and TWAIN compatible image sources. Neowise can also forwarding images and additional data to external software (third-party software). |
| | Neowise runs on a generic computer (Windows PC), which is considered a Workstation. All functionalities (acquisition, visualization, transfer, analysis and processing) are active and available when Neowise is installed on the Workstation. |
Page 1 of 6
{5}
CEFLA S.C.
Neowise
# 510(k) Premarket Notification
| Indication for Use: | Neowise is a medical imaging software intended for use in dental and medical applications as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to elaborate (retrieve, process, render, diagnose, review, and for non-medical application as store, print and distribute) images of both adults and pediatric patients. Neowise is also software for simulating treatment options. Neowise is also intended to be used for monitoring, recording, storing and displaying mandibular jaw position related to maxilla. The software is for use by authorized healthcare professionals. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery. | | | |
| --- | --- | --- | --- | --- |
| Identification of Predicate Device: | CEFLA S.C. will refer to the following predicate device: Proprietary Name: Planmeca Romexis Classification Name: System, Image Processing, Radiological, 21 CFR 892.2050 Product Code: LLZ Applicant/Manufacturer: Planmeca Oy Asentajankatu 6 Helsinki, 00880 FINLAND 510 (k): K200572 | | | |
| Identification of Reference Device: | CEFLA S.C. will refer to the following reference device: Proprietary Name: DTX Studio Clinic Classification Name: System, Image Processing, Radiological, 21 CFR 892.2050 Product Code: LLZ Applicant/Manufacturer: Nobel Biocare AB Nobel Biocare AB Vastra Hamngatan 1 Göteborg, SE-411 17, Sweden 510 (k): K203156 | | | |
| Comparison of technological characteristics with the predicate and reference devices: | Characteristic | Proposed Device | Predicate Device | Reference Device |
| | Device Name | Neowise | Planmeca Romexis | DTX Studio Clinic |
| | Manufacturer | CEFLA S.C | Planmeca Oy | Nobel Biocare AB |
| | 510(K) No. | - | K200572 | K203156 |
| | Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 |
| | Regulation Name | Picture archiving and communications system | Picture archiving and communications system | Picture archiving and communications system |
| | Regulatory Class | Class II | Class II | Class II |
| | Classification Product Code | LLZ | LLZ | LLZ |
| Indication for use | Neowise is a medical imaging software intended for use in dental and medical applications as a tool for | Planmeca Romexis is a medical imaging software intended for use in dental and medical care as tool | DTX Studio Clinic is a software program for the acquisition, management, transfer | |
Page 2 of 6
{6}
CEFLA S.C.
510(k) Premarket Notification
Neowise
| | | displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to elaborate (retrieve, process, render, diagnose, review, and for non medical application as store, print and distribute) images of both adults and pediatric patients. Neowise is also software for simulating treatment options. Neowise is also intended to be used for monitoring, recording, storing and displaying mandibular jaw position related to maxilla. The software is for use by authorized healthcare professionals. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery. | for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, diagnose, review, store, print and distribute images of both adult and pediatric patients. Planmeca Romexis is also a preoperative software used for dental implant palling. Based on the planned implant position a model of surgical guida gor a guided implant surgery can be designed. The designed bojects can be exported to manufacture a separate physical product. Planmeca Romexis is also intended to be sued for monitoring, recording, sotring and displaying mandibular jaw positions and movements relative to maxilla. Additionally, Planmeca Romexis includes monitoring features for Planmeca devices for maintenance process. The software is design to work as stand-alone or as an accessory to Planmeca imaging and Planmeca dental unit products in standard PC. The software is for use by authorized healthcare professionals. Use of the software for implant planning requires that the user has the necessary medical training | and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental restorative solutions. It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations |
| --- | --- | --- | --- | --- |
Page 3 of 6
{7}
CEFLA S.C.
Neowise
# 510(k) Premarket Notification
| | | | in implantology and surgical dentistry. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery. Indication of the dental implants do not change with guided surgery compared to conventional surgery. | |
| --- | --- | --- | --- | --- |
| | Main features | | | |
| | Input data/Image acquisition | 2D, 3D, optical impressions, 3D face photos, DICOM image data volumes | 2D, 3D, optical impressions, 3D face photos | DICOM image data volumes, 2D and 3D images |
| | Output data | Data is stored in local database. 2D and 3D image export and related reporting. | Data is stored in local database (also data can be shared by cloud management – refer to page 35 of Planmeca Romexis user manual) 2D and 3D image export and related reporting. | Data is stored locally or in remotely accessible database in the network (DTX Studio Core). 2D and 3D image export (DICOM image data volumes, 2D and 3D images such as (CB)CT scans and 2D images such as OPG/panorex images, acquisition of X-Ray images from intra-oral sensors, cephalometric images, intra-oral images and clinical pictures.), STL, PLY, NXA file export. Export implant or restorative treatment plan Diagnostic findings report export. |
| | Image Processing methodology | Image processing, brightness & contrast, crop, annotations, measurements, import/export. | Image processing, brightness & contrast, crop, annotations, | Enhancement (image filter application), annotations, measurements (distance and |
Page 4 of 6
{8}
CEFLA S.C.
Neowise
# 510(k) Premarket Notification
| | | measurements, import/export. | angular, volume and surface area for data segmentation), import/export. Airway volume segmentation. Alignment of surface scans, such as intra-oral or dental cast scans .STL/.PLY files with (CB)CT data for accurate implant planning |
| --- | --- | --- | --- |
| Software Feature: Patient Records | | | |
| List of exams associated with a selected patient | List of exams associated with a selected patient is reported on dedicated “Patient” tab section on Neowise user interface | The list of patients with dedicated list of exams for each patient is included on dedicated section “Patients” on the user interface. | The list of patients with dedicated list of exams for each patient is included on dedicated section “Patients” on the user interface. |
| Import exams | Data and related exam can be imported on the software including DICOM, Dental scans, Face scans,2D and 3D images | Data and related exam can be imported on the software including DICOM, Dental scans, Face scans,2D and 3D images | Data and related exams can be imported on the software including DICOM, Dental scans, Face scans,2D and 3D images |
| Software Feature: Scan data | | | |
| Scan Data (performing 2D examination) | It is possible to perform 2D scans (and related image capturing interfacing with proprietary devices). | It is possible to perform 2D scans (and related image capturing interfacing with proprietary devices) | It is possible to perform 2D scans (and related image capturing interfacing with proprietary devices) |
| Scan Data (performing 3D examination) | It is possible to perform 3D scans | It is possible to perform 3D scans | It is possible to perform 3D scans |
| Software Feature: Image management | | | |
| Change image palette | A cursor within the window enables the modification of the image brightness and contrast: dragging the cursor to the right/left changes the contrast and to the top/bottom the brightness. | The software allows to adjust contrast, brightness and softness | Information not available |
Page 5 of 6
{9}
CEFLA S.C.
Neowise
# 510(k) Premarket Notification
| | Measurement tools | Measurement button allows the user to draw a segment on the image to make a measurement. Continuous measurement tool button allows the user to draw several contiguous segments on the image to make a measurement. | It is possible to perform a measurement of a length (length -single which is the segment and polyline which is the continuous measurement tool) | The software allows measurement tools |
| --- | --- | --- | --- | --- |
| | Software Feature: Twain and third-party | | | |
| | Twain sources (data import) | It is possible to select and acquire images from external TWAIN devices. | It is possible to perform scanning from TWAIN sources | It is possible to perform scanning from TWAIN sources |
| | Bench testing | | | |
| | Software validation | IEC62304 + Guidance FDA on MD SW Risk management following the ISO 14971 | IEC62304 + Guidance FDA on MD SW Risk management following the ISO 14971 | IEC62304 + Guidance FDA on MD SW Risk management following the ISO 14971 |
| Non-clinical Performance Testing: | Neowise is designed and manufactured under the Quality System Regulations as outlined in 21 CFR §820 and ISO 13485:2016 Standards. This device is in conformance with the applicable parts of IEC 62304 standard. Design Control Activities, including risk management following the ISO 14971 verification/validation testing, were conducted and are included in this submission. The performance of the subject device was verified and validated following the guidance provided in FDA Guidance General Principles of Software Validation. This documentation includes testing which demonstrates that the requirements for the features have been met. V&V testing evidences as per “Content of Premarket Submissions for Device Software Functions ” issued on June 14, 2023, is also included as part of this submission. | | | |
| Clinical Testing: | Clinical performance testing was not conducted. | | | |
| Conclusion: | CEFLA S.C. considers Neowise to be substantially equivalent to the predicate device. This conclusion is based on the similarities in intended use, principle of operation, functional design, and established medical use. Differences between the devices shown in the comparison section above are minor and do not have any negative effect on substantial equivalence. | | | |
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.