Non-inferiority in standalone cancer detection performance
AUCROC Mammogram level: 0.911 (supplementary views), 0.920 (duplicated views), 0.951 (DBT+FFDM with prior DBT), 0.937 (DBT with prior FFDM), 0.864 (FFDM with prior DBT)
—
—
Standalone performance testing: 9,789 studies
—
Indications for Use
MammoScreen 5 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen 5.
Device Story
MammoScreen 5 is a software-only AI device for radiologists interpreting screening mammograms; processes FFDM and DBT images, optionally with prior examinations. Uses deep learning to detect and characterize findings (mass/asymmetry, distortion, calcifications); assigns a 'MammoScreen Score' (1-10) indicating suspicion level at finding, breast, and case levels. Outputs graphical marks and location data (quadrant, depth, distance to nipple) to assist in reporting. Used in clinical settings by qualified physicians; acts as a concurrent reading aid. Does not replace clinical judgment; patient management decisions require physician review. Benefits include adjunctive diagnostic support to improve interpretation accuracy.
Clinical Evidence
No new clinical studies conducted; relies on data from K240301. Standalone performance testing (n=9,789) evaluated non-inferiority of new scenarios against cleared scenarios. Primary endpoints (AUCROC) for mammogram, breast, and finding levels showed non-inferiority (p<0.001 for most comparisons). Subgroups included density, lesion type, age, and race. Truthing based on biopsy-proven cancer or imaging follow-up for negative/benign cases.
Technological Characteristics
Software-only device; AI/ML-based detection using deep learning modules. Compatible with FFDM and DBT systems. Standards: IEC 62304 (software lifecycle), IEC 62366-1 (usability). No hardware components. Connectivity: processes digital images from compatible systems.
Indications for Use
Indicated for adult women undergoing screening mammography to assist physicians in interpreting full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) images. Not for use as a sole basis for patient management decisions.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
Therapixel
Shalyna Long Bansropun
Head of Quality Assurance and Regulatory Affairs
455 Promenade des Anglais,
06200 Nice
France
June 26, 2026
Re: K260714
Trade/Device Name: MammoScreen® (5)
Regulation Number: 21 CFR 892.2090
Regulation Name: Radiological Computer-Assisted Detection And Diagnosis Software
Regulatory Class: Class II
Product Code: QDQ, QIH
Dated: June 17, 2026
Received: June 17, 2026
Dear Shalyna Long Bansropun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260714 - Shalyna Long Bansropun
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260714 - Shalyna Long Bansropun
Page 3
Sincerely,
Digitally signed by Michael D. O'hara -S
Date: 2026.06.26 15:08:43 -04'00'
For
Yanna Kang
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260714 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| MammoScreen® (5) | | | |
| Please provide your Indications for Use below. | | | ? |
| MammoScreen 5 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen 5. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
**THERAPIXEL**
TECHNOLOGY. FOR LIFE. FOR ALL.
## 510(k) Summary | K260714
### MammoScreen®
This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR § 807.92.
#### Applicant Information:
Therapixel
455 Promenade des Anglais,
06200 Nice
France
Phone: +33 9 72 55 20 39
#### Company Representative:
Pierre Fillard
Chief Scientific Officer
Email: pfillard@therapixel.com
Phone: +33 6 83 71 28 09
#### Primary Correspondent:
Shalyna Bansropun
Head of Quality Assurance & Regulatory Affairs
Email: sbansropun@therapixel.com
Phone: + 33 6 20 15 11 13
**Date Summary Prepared:** March 02, 2026
#### Device Information:
| Trade Name: | MammoScreen® |
| --- | --- |
| Model: | 5 |
| Common Name: | Computer-Assisted Detection Device |
| Device Classification Name: | Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer |
| Regulation Number: | 892.2090 |
| Regulation Class: | Class II |
| Product Code: | QDQ |
| Associated Product Code: | QIH |
| Submission type: | Special 510(k) |
| 510(k) number: | K260714 |
1
{5}
**THERAPIXEL**
TECHNOLOGY FOR LIFE FOR ALL
### **Predicate Device:**
The predicate device is MammoScreen 4, cleared under K243679 (Product code QDQ).
### **Device Description**
**MammoScreen® 5** is a concurrent reading medical software device using artificial intelligence to assist radiologists in the interpretation of mammograms.
**MammoScreen® 5** processes the mammogram(s) and detects findings suspicious for breast cancer. Each detected finding gets a score called the **MammoScreen Score™**. The score was designed such that findings with a low score have a very low level of suspicion. As the score increases, so does the level of suspicion. For each mammogram, **MammoScreen® 5** outputs detected findings with their associated score, a score per breast, driven by the highest finding score for each breast, and a score per case, driven by the highest finding score overall. The **MammoScreen Score™** goes from one to ten.
**MammoScreen® 5** is available for 2D (FFDM images) and 3D processing (FFDM & DBT or DBT) for GE® and Hologic® devices. Optionally, **MammoScreen® 5** can use prior examinations in the analysis.
**MammoScreen® 5** can also aid in the reporting process by populating an initial report with chosen findings, including lesion type and position (quadrant, depth and distance to nipple).
Note that the **MammoScreen® 5** outputs should be used as complementary information by radiologists while interpreting mammograms. For all cases, the medical professional interpreting the mammogram remains the sole decision-maker.
### **Indication for Use**
MammoScreen® 5 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis.
Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram.
The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting.
Patient management decisions should not be made solely based on analysis by MammoScreen® 5.
### **Intended user population**
Intended users of MammoScreen are physicians qualified to read mammograms.
2
{6}
**THERAPIXEL**
TECHNOLOGY FOR LIFE FOR ALL
#### **Intended patient population**
The device is intended to be used in the population of women undergoing mammography.
#### **Warnings and precautions**
Patient management decisions should not be made solely based on analysis by MammoScreen.
3
{7}
**THERAPIXEL**
TECHNOLOGY FOR LIFE FOR ALL
# **Predicate device comparison**
| | PREDICATE DEVICE | SUBJECT DEVICE | Comparison |
| --- | --- | --- | --- |
| | MammoScreen 4 | MammoScreen 5 | |
| **Manufacturer** | Therapixel | Therapixel | Identical |
| **Classification Regulation** | 21 CFR 892.2090 Radiological Computer Assisted Detection And Diagnosis Software | 21 CFR 892.2090 Radiological Computer Assisted Detection And Diagnosis Software | Identical |
| **Medical Device Classification** | Class II | Class II | Identical |
| **Product Code** | QDQ | QDQ | Identical |
| **Intended Use** | MammoScreen® 4 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. | MammoScreen® 5 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. | Identical |
1
{8}
**THERAPIXEL**
TECHNOLOGY FOR LIFE FOR ALL
| | PREDICATE DEVICE | SUBJECT DEVICE | Comparison |
| --- | --- | --- | --- |
| | MammoScreen 4 | MammoScreen 5 | |
| | The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen® 4. | The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen® 5. | |
| **Intended user population** | Physicians qualified to read mammograms | Physicians qualified to read mammograms | Identical |
| **Intended patient population** | Women undergoing mammography. | Women undergoing mammography. | Identical |
| **Anatomical Location** | Breast | Breast | Identical |
| **Design** | Software-only device | Software-only device | Identical |
| **Type of artificial intelligence** | MammoScreen 4 is powered by artificial intelligence/machine learning-based software algorithm | MammoScreen 5 is powered by artificial intelligence/machine learning-based software algorithm | Identical |
| **Level of suspicion** | MammoScreen 4 outputs a level of suspicion at the finding, breast and case level. | MammoScreen 5 outputs a level of suspicion at the finding, breast and case level. | Identical |
| **Lesion type** | For each detected finding MammoScreen 4 classifies them as mass/asymmetry, distortion or calcifications. | For each detected finding MammoScreen 5 classifies them as mass/asymmetry, distortion or calcifications. | Identical |
2
{9}
THERAPIXEL
TECHNOLOGY FOR LIFE FOR ALL
| | PREDICATE DEVICE | SUBJECT DEVICE | Comparison |
| --- | --- | --- | --- |
| | **MammoScreen 4** | **MammoScreen 5** | |
| **Location of findings** | Marks and description including quadrant, depth, and distance from the nipple. | Marks and description including quadrant, depth, and distance from the nipple. | Identical |
| **Localization** | For each finding MammoScreen 4 provides a quadrant, a depth and a distance to the nipple. | For each finding MammoScreen 5 provides a quadrant, a depth and a distance to the nipple. | Identical |
| **Inputs** | FFDM or 2DSM & DBT or FFDM & DBT, with an optional prior (FFDM or 2DSM & DBT) expect for the former | FFDM and/or DBT, with an optional prior (FFDM or DBT). | Since a tomosynthesis acquisition consistently produces both DBT images and a synthetic 2D mammography image (2DSM), terminology used for these modalities has been simplified |
| **Supported Scenarios** | - Cross-modality/prior comparison scenarios: | - Cross-modality/prior comparison scenarios: | New supported scenarios that do not raise different questions |
| | | Current | | |
| --- | --- | --- | --- | --- |
| | | FFDM | DBT | DBT+FFDM |
| No Prior | FFDM | Supported** | Supported* | Supported |
| | DBT+FFDM | Supported in V5 | Supported in V5 | Supported |
| Prior | DBT | | Supported | Supported in V5 |
*Only combination supported with Hologic Envision™
3
{10}
THERAPIXEL
TECHNOLOGY. FOR LIFE. FOR ALL.
| | | PREDICATE DEVICE | | | | | | SUBJECT DEVICE | | Comparison |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | MammoScreen 4 | | | | | | MammoScreen 5 | | |
| | | Current | | | | | | **Not Supported for GE - Supplementary views (XCCM, XCCL, ML) - Duplicated views | about the safety and effectiveness of the device as compared to the predicate device. The devices have the same intended use. | |
| | | FFDM | DBT | 2DSM | DBT+FFDM | DBT+2DSM | | | | |
| | | No Prior | / | Supported | Unsupported | Unsupported | Supported | | | Supported* |
| | | FFDM | Unsupported | Unsupported | Unsupported | Supported | Unsupported | | | |
| | | DBT | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported | | | |
| | | 2DSM | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported | | | |
| | | DBT + FFDM | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported | | | |
| | | DBT + 2DSM | Unsupported | Unsupported | Unsupported | Unsupported | Supported | | | |
| *Only combination supported with Hologic Envision™ | | | | | | | | | | |
| **Inclusion of PCCP** | | Included. The PCCP in the predicate device includes proposed modifications related to extending supported image acquisition systems | | | | | | No new PCCP included | | - |
4
{11}
THERAPIXEL
TECHNOLOGY. FOR LIFE. FOR ALL.
The indication for the use of MammoScreen 5 is similar to that of the predicate device. Both devices are intended for concurrent use by physicians interpreting breast images to help them with localizing and characterizing findings. The devices are not intended as a replacement for the review of a physician or their clinical judgment.
The predicate device and the subject device are two software versions of MammoScreen. They both rely on the same fundamental scientific technology. The design changes of this new version of MammoScreen have been assessed at the software design level and do not raise different questions of safety and effectiveness than the previous version. For both devices, a choice of medical image processing and machine learning techniques are implemented. The system includes 'deep learning' modules for the detection of suspicious calcifications and soft tissue lesions. These modules are trained with very large databases of biopsy-proven examples of breast cancer and normal tissue.
The overall design of MammoScreen 5 is the same as the design of the predicate device. Both versions detect and characterize findings in radiological breast images and provide information about the presence, location, and characteristics of the findings to the user in a similar manner. While MammoScreen 5, corrects some bugs found in the predicate device, and aims to demonstrate via a standalone performance analysis that MammoScreen can be validated on several scenarios, these modifications do not raise different questions about the safety and effectiveness of the device as compared to the predicate device. The devices have the same intended use. The safety and effectiveness of the device have been evaluated with a similar methodology as for the predicate device.
### Non clinical Testing
MammoScreen is a software-only device.
Tests have been performed in compliance with the following recognized consensus standards:
- IEC 62304:2006/A1:2016- Medical device software - Software life-cycle processes
- IEC 62366-1:2015+AMD1:2020- Medical devices - Application of usability engineering to medical devices.
MammoScreen 5 has successfully completed integration and verification testing. In addition, potential hazards have been evaluated and mitigated and have acceptable levels.
The algorithm behind MammoScreen 5 did not change/evolve compared to its predecessor, accordingly a standalone performance testing was deemed sufficient to validate and claim compatibility with the newly introduced scenarios.
The standalone performance testing is summarized in what follows:
1
{12}
**THERAPIXEL**
TECHNOLOGY. FOR LIFE. FOR ALL.
| **Statistics tests for primary objective** | Non-inferiority in standalone cancer detection performance on newly introduced scenarios compared to scenarios already supported | | | | |
| --- | --- | --- | --- | --- | --- |
| **Primary endpoint** | **Supplementary views** | | | | |
| | **Metric/Level** | **Including EV** | **Excluding EV (ref.)** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.911 (0.894, 0.928) | 0.894 (0.874, 0.914) | 0.017 (0.007, 0.027) | 0.000 |
| | AUCROC Breast level | 0.926 (0.910, 0.941) | 0.912 (0.894, 0.929) | 0.014 (0.004, 0.023) | 0.000 |
| | AUCLROC Finding level | 0.909 (0.888, 0.930) | 0.895 (0.872, 0.917) | 0.014 (0.005, 0.023) | 0.000 |
| | **Duplicated views** | | | | |
| | **Metric/Level** | **Including duplicates** | **Excluding duplicates (ref.)** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.920 (0.907, 0.932) | 0.907 (0.893, 0.920) | 0.013 (0.006, 0.020) | 0.000 |
| | AUCROC Breast level | 0.938 (0.927, 0.948) | 0.927 (0.916, 0.939) | 0.010 (0.005, 0.016) | 0.000 |
| | AUCLROC Finding level | 0.908 (0.892, 0.924) | 0.900 (0.884, 0.916) | 0.007 (-0.000, 0.015) | 0.000 |
| | **Current/Prior scenario: DBT+FFDM with prior DBT** | | | | |
| | **Metric/Level** | **DBT+FFDM with prior DBT** | **DBT+FFDM with prior FFDM (ref.)** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.951 (0.937, 0.965) | 0.955 (0.943, 0.968) | -0.004 (-0.011, 0.002) | 0.000 |
| | AUCROC Breast level | 0.955 (0.942, 0.968) | 0.957 (0.943, 0.971) | -0.002 (-0.007, 0.003) | 0.000 |
| | AUCLROC Finding level | 0.936 (0.917, 0.956) | 0.935 (0.914, 0.956) | 0.001 (-0.007, 0.009) | 0.000 |
| | **Current/Prior scenario: DBT with prior FFDM** | | | | |
| | **Metric/Level** | **DBT+FFDM with prior DBT** | **DBT+FFDM with prior FFDM (ref.)** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.951 (0.937, 0.965) | 0.955 (0.943, 0.968) | -0.004 (-0.011, 0.002) | 0.000 |
| | AUCROC Breast level | 0.955 (0.942, 0.968) | 0.957 (0.943, 0.971) | -0.002 (-0.007, 0.003) | 0.000 |
| | AUCLROC Finding level | 0.936 (0.917, 0.956) | 0.935 (0.914, 0.956) | 0.001 (-0.007, 0.009) | 0.000 |
| | **Current/Prior scenario: FFDM with prior FFDM** | | | | |
| | **Metric/Level** | **DBT with prior FFDM** | **DBT with prior DBT* (ref.)** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.937 (0.922, 0.952) | 0.932 (0.916, 0.948) | 0.005 (-0.002, 0.012) | 0.000 |
| | AUCROC Breast level | 0.941 (0.924, 0.957) | 0.932 (0.913, 0.950) | 0.009 (0.000, 0.018) | 0.000 |
| | AUCLROC Finding level | 0.913 (0.891, 0.936) | 0.902 (0.878, 0.927) | 0.011 (0.001, 0.021) | 0.007 |
| | **Current/Prior scenario: FFDM with prior DBT** | | | | |
| | **Metric/Level** | **FFDM with prior DBT** | **FFDM no prior** | **Δ** | **p-value** |
| | AUCROC Mammogram level | 0.864 (0.838, 0.890) | 0.869 (0.844, 0.894) | -0.005 (-0.013, 0.004) | 0.000 |
2
{13}
**THERAPIXEL**
TECHNOLOGY. FOR LIFE. FOR ALL.
| | AUCROC Breas level | 0.882 (0.858, 0.905) | 0.884 (0.860, 0.908) | -0.002 (-0.008, 0.004) | 0.000 |
| --- | --- | --- | --- | --- | --- |
| | AUCLROC Finding level | 0.803 (0.767, 0.838) | 0.807 (0.771, 0.844) | -0.005 (-0.015, 0.006) | 0.000 |
| **Acceptance criteria** | Positive lower bound of the 95% CI of the difference in endpoints between the scenario under evaluation (introduced with MammoScreen 5) and the reference scenario (already cleared with MammoScreen 4). | | | | |
| **Number of included patients** | 9,789 | | | | |
| **Number of included studies** | 9,789 | | | | |
| **Age distribution** | *Age <= 50: 2,383* *50 < Age <= 65: 3,091* *65 > Age: 1,614* | | | | |
| **Race and Ethnicity distribution** | *Asian: 652* *White: 2,295* *Black: 961* *Other (including American Indian, Alaska Native, Native Hawaiian or Other Pacific Islander): 467* *Hispanic: 328* *Not reported: 9,461* | | | | |
| **Considered subgroups** | Density, Lesion type (mass/asymmetries, calcifications, distortion), Age, Lesion size, Lesion severity, Race, Ethnicity, Data provenance, Reference standard for negative cases, Current image combination, Prior image type, Acquisition year, Presence of a non-standard view | | | | |
| **Truthing process** | Positive cases: biopsy-proven presence of cancer Benign cases: - For non-biopsied cases: verified by imaging follow-up. - For biopsied cases: confirmed by biopsy result AND imaging follow-up. Negative cases: verified by imaging follow-up. | | | | |
| **Independence of tests data from training data** | Data sources are separated into the training/tuning group and the test group. Sources in the training/tuning group may only be used for model training and tuning. Sources in the test group may only be used for external validation of the model’s performances on unseen data (i.e., from sources entirely left out during training and tuning). Data used for the standalone performance testing only belongs to the test group. | | | | |
Please note that no new usability engineering tests have been performed for MammoScreen 5 as those already submitted for the previously cleared device MammoScreen 3 (K240301) are still applicable.
The above testing confirmed that MammoScreen 5 performs in accordance with the stated intended use. All data fell within pre-determined product specifications and external standard requirements. Results of testing confirmed the substantial equivalence of the MammoScreen 5 to the predicate device.
3
{14}
**THERAPIXEL**
TECHNOLOGY FOR LIFE FOR ALL
### Clinical Testing
Therapixel has provided documentation concerning the clinical testing to support the premarket submission for the previously cleared device MammoScreen 3 (K240301).
The clinical study results from K240301 remain fully applicable to the subject device (MammoScreen 5). No additional clinical studies were conducted to support substantial equivalence to the predicate device (MammoScreen 4). Aside from the data included in the previous submissions, no new or additional clinical testing documentation is included in this submission, as the existing data continues to demonstrate the safety and effectiveness of the device.
### Conclusions
Standalone performance tests on FFDM and DBT (with and without prior) demonstrate that MammoScreen 5 achieves non-inferior performance compared to the predicate device.
Previous MRMC studies and standalone tests have demonstrated that the device is safe and effective.
Therapixel has applied a risk management process following FDA-recognized standards to identify, evaluate, and mitigate all known hazards related to MammoScreen 5. These hazards may occur when the accuracy of diagnosis is potentially affected, causing either false positives or false negatives. All identified risks are effectively mitigated, and it can be concluded that the residual risk is outweighed by the benefits. Considering all data in this submission, the data provided in these 510(k) supports the safe and effective use of MammoScreen 5 for its indications for use and substantial equivalence to the predicate device.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.