23 (MQSA qualified and ABR certified radiologists)
Indications for Use
MammoScreen 3 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen 3.
Device Story
MammoScreen 3 is a software-only AI device for concurrent reading/reporting of screening mammograms. It processes current FFDM, DBT, or 2DSM images, optionally incorporating prior examinations. Using deep learning modules trained on biopsy-proven breast cancer databases, it detects suspicious findings, assigns a 'MammoScreen Score' (1-10) per finding, breast, and case, and characterizes lesion types (mass/asymmetry, distortion, calcifications). It provides spatial localization (quadrant, depth, distance to nipple). Used in clinical settings by radiologists, the output integrates into DICOM viewers and reporting systems. It serves as an adjunctive tool; radiologists remain the sole decision-makers. By highlighting suspicious areas and providing suspicion scores, it aims to improve diagnostic accuracy and efficiency in breast cancer detection.
Clinical Evidence
Multi-reader multi-case (MRMC) study with 240 combined DBT/2D mammograms and 23 radiologists. Primary endpoint: diagnostic performance (AUROC). Results: Radiologist AUC improved from 0.797 (unaided) to 0.871 (aided) (p<0.001). Sensitivity improved from 0.706 to 0.793 (p<0.001). Standalone performance AUC was 0.883, superior to unaided radiologists (p<0.0001). Standalone sensitivity 0.833, specificity 0.793. Subgroup analyses performed for density, lesion type, age, and modality.
Technological Characteristics
Software-only device. Uses deep learning modules for detection and classification. Inputs: FFDM, DBT, 2DSM, and prior exams. Outputs: graphical marks, suspicion scores (1-10), lesion characterization, and spatial localization. Complies with IEC 62304 (software lifecycle) and IEC 62366-1 (usability).
Indications for Use
Indicated for women undergoing screening mammography. Intended for use by physicians qualified to read mammograms as a concurrent reading and reporting aid to assist in the detection and characterization of suspicious findings (mass/asymmetry, distortion, calcifications) on full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) images.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{0}------------------------------------------------
August 1, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
Therapixel Quentin De Snoeck RA/QA Director/CISO 455 Promenade des Anglais NICE. 06200 FRANCE
Re: K240301
Trade/Device Name: MammoScreen® (3) Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological Computer Assisted Detection And Diagnosis Software Regulatory Class: Class II Product Code: QDQ Dated: July 2, 2024 Received: July 2, 2024
Dear Quentin De Snoeck:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Submission Number (if known)
K240301
Device Name
MammoScreen® (3)
Indications for Use (Describe)
MammoScreen® 3 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis.
Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on
mammograms along with their level of suspicion scores. The lesion type is characterized as mass/ asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram.
The location of findings including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting.
Patient management decisions should not be made solely based on the analysis by MammoScreen 3.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{3}------------------------------------------------
### 510(k) Summary
This 510(k) summary is prepared in accordance with the requirements of 21 CFR § 807.92.
Applicant Information: Therapixel 455 Promenade des Anglais, 06200 Nice France Phone: +33 9 72 55 20 39 Submission Correspondent: Quentin de Snoeck RAQA Director/CISO Email: qdesnoeck@therapixel.com Phone: +33 9 72 55 20 39 Date Summary Prepared: August 01, 2024 Device Information: Trade Name: MammoScreen® Model: 3 Common Name: Computer-Assisted Detection Device Device Classification Name: Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer Regulation Number: 892.2090 Class II Regulation Class:
#### Predicate Device:
Product Code:
Submission type 510(k) number:
The predicate device is MammoScreen, cleared under K211541 (Product code QDQ).
Traditional 510(k)
QDQ
K240301
{4}------------------------------------------------
## Device Description
MammoScreen is a concurrent reading medical software device using artificial intelligence to assist radiologists in the interpretation of mammograms.
MammoScreen processes the mammogram(s) and detects findings suspicious for breast cancer. Each detected finding gets a score called the MammoScreen Score™. The score was designed such that findings with a low score have a very low level of suspicion. As the score increases, so does the level of suspicion. For each mammogram, MammoScreen outputs detected findings with their associated score, a score per breast, driven by the highest finding score for each breast, and a score per case, driven by the highest finding score overall. The MammoScreen Score goes from one to ten.
MammoScreen is available for 2D (FFDM images) and 3D processing (FFDM & DBT or 2DSM & DBT). Optionally, MammoScreen can use prior examinations in the analysis. The following table describes the supported combinations:
| | | Current | | | | |
|-------|------------|-------------|-------------|-------------|-------------|-------------|
| | | FFDM | DBT | 2DSM | DBT+FFDM | DBT+2DSM |
| Prior | / | Supported | Unsupported | Unsupported | Supported | Supported |
| | FFDM | Unsupported | Unsupported | Unsupported | Supported | Unsupported |
| | DBT | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported |
| | 2DSM | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported |
| | DBT + FFDM | Unsupported | Unsupported | Unsupported | Unsupported | Unsupported |
| | DBT + 2DSM | Unsupported | Unsupported | Unsupported | Unsupported | Supported |
MammoScreen can also aid in the reporting process by populating an initial report with chosen findings, including lesion type and position (quadrant, depth and distance to nipple).
The results indicating potential breast cancer, identified by MammoScreen, are accessible via a dedicated user interface and can seamlessly integrate into DICOM viewers (using DICOM-SC and DICOM-SR). Reporting aid outputs can be incorporated into the practice's reporting system to generate a preliminary report. Additionally, certain outputs like the case score can be reported into the patient management worklist.
Note that the MammoScreen outputs should be used as complementary information by radiologists while interpreting mammograms. For all cases, the medical professional interpreting the mammogram remains the sole decision-maker.
### Indication for Use
MammoScreen 3 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis.
{5}------------------------------------------------
Output of the device includes graphical marks of findings as soft-tissue lesions on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram.
The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting.
Patient management decisions should not be made solely based on the analysis by MammoScreen 3.
| | Predicate device (MammoScreen 2) | Subject device (MammoScreen 3) | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Manufacturer | Therapixel | Therapixel | |
| Regulation number | 892.2090 | 892.2090 | |
| Product Code | QDQ | QDQ | |
| Intended Use | MammoScreen 2 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. Output of the device includes marks of findings on mammograms along with their type and level of suspicion scores. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. | MammoScreen 3 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis. Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram. The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting. Patient management decisions should not be made solely based on the analysis by MammoScreen 3. | |
| | | Predicate device (MammoScreen 2) | Subject device (MammoScreen 3) |
| Intended | user population | Physicians qualified to read mammograms. | Physicians qualified to read mammograms. |
| Intended | patient population | Women undergoing mammography. | Women undergoing mammography. |
| Anatomical Location | | Breast | Breast |
| Design | | Software-only device | Software-only device |
### Summary of Substantial Equivalence
{6}------------------------------------------------
### Predicate Device Comparison
The indication for use of MammoScreen 3 is similar to that of the predicate device. Both devices are intended for concurrent use by physicians interpreting breast images to help them with localizing and characterizing findings. The devices are not intended as a replacement for the review of a physician or their clinical judgment.
The predicate device and the subject device are two software versions of MammoScreen. They both rely on the same fundamental scientific technology. For both devices, a choice of medical image processing and machine learning techniques are implemented. The systems includes 'deep learning' modules for the detection of suspicious findings. These modules are trained with large databases of biopsy-proven examples of breast cancer and normal tissue.
Compared to the predicate device, MammoScreen 3 provides more detailed categories of finding characterization and a more detailed location of findings as a reporting aid component.
In addition, the algorithmic components have been updated to improve detection accuracy for the analysis. Moreover some additional inputs have been added to the MammoScreen 3 device compare to the predicate device. MammoScreen 3 takes into account current mammogram that can be FFDM or FFDM & DBT or 2DSM & DBT. MammoScreen 3 can also incorporate in its analysis one prior mammogram, acquired between 6 and 60 months before the current mammogram, whether it is FFDM or 2DSM & DBT.
The design changes of this new version of MammoScreen have been assessed and do not raise different questions of safety and effectiveness than the previous version.
| Technological<br>characteristic | Predicate<br>(MammoScreen 2)<br>device | Subject<br>(MammoScreen 3)<br>device |
|------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Type of artificial<br>intelligence | MammoScreen 2 is powered by artificial<br>intelligence/machine learning-based software algorithm | Same. |
| Level of suspicion | MammoScreen 2 outputs a level of<br>suspicion at the finding, breast and<br>case level. | Same. |
| Lesion type | For each detected finding<br>MammoScreen 2 classifies them as<br>soft tissue lesion or calcifications. | For each detected finding<br>MammoScreen 3 classifies them as<br>mass/asymmetry, distortion or<br>calcifications. |
{7}------------------------------------------------
| Technological<br>characteristic | Predicate<br>(MammoScreen 2) | Subject<br>(MammoScreen 3) |
|---------------------------------|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Localization | MammoScreen 2 doesn't provide a<br>localization. | For each finding MammoScreen 3<br>provides a quadrant, a depth and a<br>distance to the nipple. |
| Inputs | FFMD or DBT | FFDM or 2DSM & DBT or FFDM<br>& DBT, with an optional prior<br>(FFDM or 2DSM & DBT) expect<br>for the former. |
#### Non-Clinical Testing
MammoScreen is a software-only device.
Tests have been performed in compliance with the following recognized consensus standards:
- IEC 62304:2006/A1:2016- Medical device software Software life-cycle processes ●
- IEC 62366-1:2015+AMD1:2020- Medical devices Application of usability engineering to medical ● devices.
MammoScreen 3 has successfully completed integration testing and beta validation. In addition, potential hazards have been evaluated and mitigated, and have acceptable levels.
### Clinical Performance Data
A multi-reader multi-case study (MRMC) has been conducted to establish the clinical performance of MammoScreen 3. The study applied a fully crossed design, so that each case was red by each reader both with and without the aid of MammoScreen 3.
The study aimed to determine:
- . Whether the performance of radiologists when using MammoScreen 3 is superior to unaided radiologist for interpretation of screening mammograms (primary objective).
- . Whether the performance of MammoScreen 3 standalone is superior to unaided radiologist performance.
- . Whether the performance of MammoScreen 3 standalone is non-inferior to aided radiologist performance.
The dataset included 240 combined DBT/2D mammograms (DBT+FFDM or DBT+2DSM) with a prior. Readers were all MQSA qualified and ABR certified radiologists (18 breast specialists and 5 general
{8}------------------------------------------------
radiologists) All performances were evaluated at mammogram, breast and finding level and assessed in terms of Area Under the Receiver Operating Characteristic Curve (AUROC).
Results:
- . Radiologists improved their diagnostic performance in breast cancer detection when using MammoScreen.
- o The average value of AUC went from 0.797 [0.752 - 0.843] in unaided reading conditions to 0.871 [0.829 - 0.912] in reading conditions assisted by MammoScreen, leading to a statistically significant difference of +0.074 [0.047 - 0.101] (p-value < 0.001).
- The average sensitivity went from 0.706 [0.633 0.780] in unaided reading conditions to O 0.793 [0.725 - 0.860] in reading conditions assisted by MammoScreen, leading to a statistically significant difference of +0.086 [0.040 - 0.133] (p-value < 0.001).
- The average specificity went from 0.815 [0.782 0.848] in unaided reading conditions to o 0.836 [0.805 - 0.867] in reading conditions assisted by MammoScreen, leading to a statistically significant difference of +0.021 [0.006 - 0.036] (p-value 0.007).
- . The AUC value at the mammogram level of MammoScreen as a standalone system was 0.883 [0.837 - 0.929] and was found to be superior to radiologists in unaided reading conditions (AAUC=+0.085 [0.044 - 0.127], p-value <0.0001) and non-inferior to radiologists in aided reading conditions (AAUC=+0.012 [-0.015 - 0.039], p-value <0.0001). The sensitivity and specificity values at the mammogram level of MammoScreen as a standalone system were respectively 0.833 [0.756 – 0.911] and 0.793 [0.728 – 0.858].
Gains in performance have been established overall and for all the considered subgroups. Subgroups analysis were performed for breast density, lesion type, patient age, lesion size, lesion severity, current image combination (DBT+2DSM, DBT+FFDM), prior image combination (current DBT+2DSM with prior DBT+2DSM, current DBT+FFDM with prior FFDM), data sources, reader experience, reader background, and prior time difference. The performance in assessing descriptive characteristics of lessions (i.e., lesion type, quadrant and depth) has also been evaluated.
{9}------------------------------------------------
## Standalone Performance Data
| Cancer status | Normal/benign: 77% |
|-----------------------|------------------------------------------|
| | Malignant: 23% |
| Breast density | A: 10% |
| | B: 44% |
| | C: 31% |
| | D: 15% |
| Age (years old) | Minimum: 32 |
| | Maximum: 93 |
| | Mean: 58 |
| | 25th percentile: 49 |
| | 75th percentile: 65 |
| Race | White: 37% |
| | Black: 16% |
| | Asian: 18% |
| | American Indian / Alaska native: 0.8% |
| | Native Hawaiian / Pacific islander: 0.2% |
| | Unspecified: 28% |
| Imaging modality | DBT (inc. FFDM or 2DSM): 64% |
| | FFDM: 36% |
| Modality manufacturer | Hologic: 100% |
| Exam dates (range) | 2005 - 2023 |
7,544 exams from 4,429 patients screened in 3 US centers were used during this study. Demographic information about the test population is provided in the table below:
- . Standalone performances in breast cancer detection at the mammogram level overall and for all subgroups evaluated are summarized in the following table:
{10}------------------------------------------------
| Subgroup | | Number of<br>positive cases | Number of<br>negative cases | AUROC<br>(95% CI) | Sensitivity<br>(95% CI) | Specificity<br>(95% CI) |
|------------------------------------------------------------------------------------|------------------------------------|-----------------------------|-----------------------------|----------------------|-------------------------|-------------------------|
| | | Overall | | 1,053 | 6,420 | 0.927 (0.911, 0.942) |
| By density | A | 71 | 658 | 0.953 (0.924, 0.976) | 0.958 (0.901, 1.000) | 0.739 (0.701, 0.778) |
| | B | 489 | 2,682 | 0.933 (0.922, 0.945) | 0.933 (0.910, 0.954) | 0.690 (0.670, 0.709) |
| | C | 415 | 1,822 | 0.921 (0.905, 0.936) | 0.896 (0.867, 0.925) | 0.727 (0.702, 0.750) |
| | D | 77 | 1,005 | 0.891 (0.846, 0.929) | 0.816 (0.727, 0.896) | 0.815 (0.787, 0.841) |
| | By race | White | 211 | 2,780 | 0.901 (0.876, 0.924) | 0.857 (0.806, 0.900) |
| Black | | 77 | 1,085 | 0.897 (0.862, 0.929) | 0.896 (0.831, 0.961) | 0.680 (0.650, 0.708) |
| Asian | | 41 | 1,406 | 0.899 (0.849, 0.940) | 0.854 (0.732, 0.951) | 0.763 (0.736, 0.789) |
| American Indian or Alaska<br>Native / Native Hawaiian or<br>other Pacific Islander | | 3 | 76 | 0.886 (0.701, 1.000) | 0.663 (0.000, 1.000) | 0.672 (0.556, 0.784) |
| By source | US 1 | 75 | 2,616 | 0.862 (0.814, 0.905) | 0.772 (0.667, 0.867) | 0.752 (0.732, 0.771) |
| | US 2 | 717 | 915 | 0.951 (0.941, 0.961) | 0.935 (0.917, 0.952) | 0.787 (0.757, 0.817) |
| | US 3 | 261 | 2,889 | 0.908 (0.888, 0.925) | 0.889 (0.851, 0.927) | 0.689 (0.670, 0.709) |
| By age | ≤ 50 years old | 273 | 3,879 | 0.912 (0.894, 0.929) | 0.879 (0.839, 0.916) | 0.720 (0.704, 0.736) |
| | 50 < age ≤ 65 | 397 | 1,630 | 0.942 (0.929, 0.954) | 0.932 (0.907, 0.957) | 0.756 (0.732, 0.780) |
| | > 65 | 382 | 911 | 0.921 (0.902, 0.938) | 0.914 (0.885, 0.942) | 0.705 (0.669, 0.740) |
| By lesion type | Mass | 180 | 6,420 | 0.945 (0.927, 0.961) | 0.939 (0.902, 0.973) | 0.727 (0.715, 0.740) |
| | Calcifications | 271 | 6,420 | 0.926 (0.909, 0.942) | 0.912 (0.878, 0.941) | 0.727 (0.715, 0.740) |
| | Asymmetries | 117 | 6,420 | 0.884 (0.854, 0.911) | 0.835 (0.768, 0.897) | 0.727 (0.715, 0.740) |
| | Focal asymmetries | 262 | 6,420 | 0.921 (0.901, 0.938) | 0.902 (0.861, 0.936) | 0.727 (0.715, 0.740) |
| | Distortions | 223 | 6,420 | 0.945 (0.929, 0.960) | 0.938 (0.906, 0.969) | 0.727 (0.715, 0.740) |
| By lesion size | < 20 mm | 380 | 6,420 | 0.905 (0.887, 0.922) | 0.887 (0.852, 0.924) | 0.727 (0.715, 0.740) |
| | 20 ≤ size ≤ 30 mm | 355 | 6,420 | 0.933 (0.915, 0.948) | 0.916 (0.881, 0.946) | 0.727 (0.715, 0.740) |
| | > 30 mm | 318 | 6,420 | 0.959 (0.947, 0.971) | 0.959 (0.934, 0.983) | 0.727 (0.715, 0.740) |
| By lesion<br>severity | BI-RADS 4 | 525 | 6,420 | 0.924 (0.910, 0.936) | 0.910 (0.884, 0.935) | 0.727 (0.715, 0.740) |
| | BI-RADS 5 | 178 | 6,420 | 0.972 (0.960, 0.982) | 0.983 (0.961, 1.000) | 0.727 (0.715, 0.740) |
| By current<br>image<br>combination | FFDM | 886 | 4,303 | 0.908 (0.896, 0.919) | 0.915 (0.895, 0.933) | 0.632 (0.615, 0.647) |
| | DBT + FFDM | 823 | 3,744 | 0.929 (0.919, 0.939) | 0.925 (0.905, 0.942) | 0.710 (0.694, 0.725) |
| | DBT + 2DSM | 161 | 2,114 | 0.904 (0.878, 0.927) | 0.858 (0.806, 0.903) | 0.762 (0.739, 0.785) |
| By prior image<br>combination | FFDM | 782 | 4,216 | 0.926 (0.916, 0.936) | 0.901 (0.880, 0.921) | 0.761 (0.746, 0.776) |
| | DBT + FFDM | 534 | 1833 | 0.930 (0.918, 0.942) | 0.907 (0.882, 0.930) | 0.775 (0.755, 0.795) |
| | DBT + 2DSM | 80 | 921 | 0.903 (0.861, 0.937) | 0.838 (0.750, 0.912) | 0.767 (0.735, 0.797) |
| By current &<br>prior image<br>combination | No prior | 1,037 | 6,211 | 0.925 (0.915, 0.933) | 0.933 (0.918, 0.948) | 0.667 (0.653, 0.681) |
| | Current FFDM – Prior FFDM | 679 | 2,135 | 0.905 (0.891, 0.920) | 0.908 (0.887, 0.929) | 0.664 (0.643, 0.683) |
| | Current DBT+FFDM<br>Prior FFDM | 630 | 1,717 | 0.932 (0.920, 0.943) | 0.914 (0.894, 0.935) | 0.767 (0.746, 0.788) |
| | Subgroup | Number of<br>positive cases | Number of<br>negative cases | AUROC<br>(95% CI) | Sensitivity<br>(95% CI) | Specificity<br>(95% CI) |
| | Current DBT+FFDM | 482 | 1,586 | 0.928 (0.913, 0.941) | 0.906 (0.880, 0.932) | 0.771 (0.750, 0.791) |
| | Prior DBT+FFDM | | | | | |
| | Current DBT+FFDM<br>Prior DBT+2DSM | 13 | 3 | 0.945 (0.769, 1.000) | 1.000 (1.000, 1.000) | 0.663 (0.000, 1.000) |
| | Current DBT+2DSM<br>Prior FFDM | 120 | 1,196 | 0.908 (0.875, 0.936) | 0.875 (0.817, 0.933) | 0.764 (0.737, 0.789) |
| | Current DBT+2DSM<br>Prior DBT+FFDM | 75 | 164 | 0.939 (0.905, 0.967) | 0.935 (0.867, 0.987) | 0.755 (0.680, 0.823) |
| | Current DBT+2DSM<br>Prior DBT+2DSM | 65 | 878 | 0.890 (0.844, 0.931) | 0.817 (0.723, 0.908) | 0.770 (0.738, 0.799) |
| | Current FFDM - No prior | 878 | 4,167 | 0.899 (0.887, 0.911) | 0.911 (0.892, 0.930) | 0.611 (0.594, 0.628) |
| | Current DBT+FFDM<br>No prior | 807 | 3,606 | 0.927 (0.917, 0.937) | 0.941 (0.924, 0.957) | 0.659 (0.642, 0.676) |
| | Current DBT+2DSM<br>No prior | 158 | 2,040 | 0.902 (0.874, 0.928) | 0.893 (0.842, 0.943) | 0.689 (0.665, 0.712) |
| By prior time<br>difference | 6 months ≤ prior < 18 months | 237 | 1,866 | 0.913 (0.893, 0.932) | 0.869 (0.823, 0.912) | 0.774 (0.755, 0.793) |
| | 18 ≤ prior < 30 months | 436 | 1,735 | 0.931 (0.916, 0.944) | 0.917 (0.889, 0.942) | 0.752 (0.728, 0.777) |
| | 30 months ≤ prior < 60 months | 177 | 628 | 0.935 (0.912, 0.955) | 0.910 (0.865, 0.949) | 0.754 (0.718, 0.789) |
{11}------------------------------------------------
{12}------------------------------------------------
- . Standalone performances in lesion type assessment were measured in terms of Positive Percentage Agreement (PPA) and Negative Percentage Agreement (NPA) against the reference standard. Results are summarized in the following table:
| | Counts | PPA (95% CI) | NPA (95% CI) |
|----------------|--------|-----------------------|-----------------------|
| Overall | 1,039 | 0.784, (0.758, 0.811) | 0.893, (0.880, 0.906) |
| Mass/asymmetry | 548 | 0.868, (0.838, 0.894) | 0.783, (0.752, 0.815) |
| Distortion | 221 | 0.544, (0.475, 0.611) | 0.947, (0.932, 0.962) |
| Calcifications | 270 | 0.941, (0.911, 0.967) | 0.950, (0.934, 0.964) |
The associated confusion matrix is provided for completeness:
Image /page/12/Figure/3 description: This image is a confusion matrix that visualizes the performance of a classification model. The matrix compares the true labels (rows) against the predicted labels (columns) for three classes: 'mass/asym.', 'distortion', and 'calcification'. The diagonal elements show the number of correctly classified instances for each class, with 476, 120, and 254 instances respectively. Off-diagonal elements indicate misclassifications, such as 94 instances of 'distortion' being misclassified as 'mass/asym'.
- Standalone performances in CC/MLO quadrant assessment were measured in terms of Positive Percentage Agreement (PPA) and Negative Percentage Agreement (NPA) against the reference standard. Results are provided in the following tables separately for CC and MLO quadrants:
| | CC quadrant: |
|--|--------------|
| | |
| | Counts | PPA (95% CI) | NPA (95% CI) |
|-----------------|--------|----------------------|----------------------|
| Overall | 736 | 0.765 (0.726, 0.810) | 0.963 (0.951, 0.965) |
| Inner / Medial | 149 | 0.866 (0.812, 0.919) | 0.978 (0.966, 0.988) |
| Central | 179 | 0.782 (0.715, 0.844) | 0.925 (0.903, 0.943) |
| Outer / Lateral | 372 | 0.884 (0.852, 0.914) | 0.953 (0.929, 0.973) |
| Retroareolar | 36 | 0.528 (0.361, 0.667) | 0.994 (0.987, 0.999) |
Associated confusion matrix of CC quadrant assessment:
{13}------------------------------------------------
Image /page/13/Figure/0 description: The image is a confusion matrix, which is a table that is used to evaluate the performance of a classification model. The matrix shows the counts of the true vs predicted labels. The matrix has labels such as inner/medial, central, outer/lateral, retroareolar, and unassigned. For example, the matrix shows that 129 instances were correctly classified as inner/medial.
#### MLO quadrant:
| | Counts | PPA (95% CI) | NPA (95% CI) |
|------------------|--------|----------------------|----------------------|
| Overall | 749 | 0.471 (0.425, 0.523) | 0.889 (0.878, 0.902) |
| Lower / Inferior | 130 | 0.623 (0.538, 0.708) | 0.861 (0.835, 0.884) |
| Central | 151 | 0.351 (0.272, 0.430) | 0.766 (0.731, 0.796) |
| Upper / Superior | 428 | 0.458 (0.300, 0.600) | 0.941 (0.910, 0.966) |
| Retroareolar | 40 | 0.450 (0.300, 0.575) | 0.990 (0.983, 0.997) |
Associated confusion matrix of MLO quadrant assessment:
Image /page/13/Figure/4 description: This image is a confusion matrix, which is a table that is used to evaluate the performance of a classification model. The rows of the matrix represent the true labels of the data, and the columns represent the predicted labels. The diagonal elements of the matrix represent the number of correctly classified instances, while the off-diagonal elements represent the number of misclassified instances. For example, the matrix shows that 81 instances of the 'lower/inferior' class were correctly classified, while 14 instances were misclassified as 'central'.
- Standalone performances in depth assessment were measured in terms of Positive Percentage ● Agreement (PPA) and Negative Percentage Agreement (NPA) against the reference standard. Results are provided in the following table:
{14}------------------------------------------------
| | Counts | PPA (95% CI) | NPA (95% CI) |
|-----------|--------|----------------------|----------------------|
| Overall | 775 | 0.617 (0.587, 0.644) | 0.943 (0.932, 0.953) |
| Posterior | 346 | 0.907 (0.873, 0.936) | 0.891 (0.865, 0.918) |
| Middle | 355 | 0.852 (0.814, 0.887) | 0.904 (0.879, 0.931) |
| Anterior | 74 | 0.699 (0.595, 0.797) | 0.992 (0.984, 0.997) |
Associated confusion matrix:
Image /page/14/Figure/2 description: This image is a confusion matrix that visualizes the performance of a classification model. The matrix has four categories: posterior, middle, anterior, and unassigned. The matrix shows the counts of true vs predicted labels, with the diagonal elements representing correct classifications. For example, 314 posterior labels were correctly predicted, while 45 middle labels were incorrectly predicted as posterior.
{15}------------------------------------------------
### Conclusions
Standalone performance tests on FFDM and DBT demonstrated that MammoScreen 3 achieved superior performance compared to the predicate device.
Therapixel applied a risk management process following FDA-recognized standards to identify, evaluate, and mitigate all known hazards related to MammoScreen 3. These hazards may occur when the accuracy of diagnosis is potentially affected, causing either false positives or false negatives. All identified risks are effectively mitigated, and it can be concluded that the benefits outweigh the residual risk.
The indication for use of MammoScreen 3 is substantially the same as the predicate devices are intended for use by physicians in interpreting mammograms to help them identify and characterize findings. The devices are not intended as a replacement for the review of a physician or their clinical judgment. The overall design of MammoScreen 3 is the same as the predicate devices detect and characterize findings in radiological breast images and provide information about the presence, location, and characteristics of the findings to the user.
The new features introduced in MammoScreen 3 do not raise different questions of safety and effectiveness compared to the predicate device. The standalone performances, clinical validation, risk analysis, and usability studies confirm that MammoScreen 3 maintains its intended use and risk profile, and demonstrates substantial equivalence to the predicate device, MammoScreen 2 (K211541).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.