The BlueDop Vascular Expert (BVE) is intended to non-invasively detect the presence of blood flow in peripheral arteries using Doppler ultrasound technology. The device provides Doppler waveform information to support clinical evaluation of peripheral vascular status. The BVE is intended to be used as an adjunct to standard clinical assessment and does not provide a standalone diagnostic determination.
Device Story
Portable Doppler ultrasound system; consists of handheld probe and tablet computer. Probe transmits continuous-wave Doppler waveform data via Bluetooth to tablet. Clinician manually enters brachial blood pressure readings (obtained via external cuff) into tablet. Tablet processes, displays, and stores Doppler waveforms; calculates Ankle-Brachial Index (ABI). Used in hospitals and clinics by qualified healthcare personnel. Provides visual waveform and numeric ABI output to support clinical evaluation of peripheral vascular status. Does not provide autonomous diagnostic determinations or deliver therapy. Benefits include improved documentation and workflow efficiency via digital storage and touchscreen interface.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), ultrasound acoustic output (IEC 60601-2-37, IEC 62127-1), Doppler sensitivity, Doppler velocity measurement accuracy, temperature rise assessment, software verification/validation (IEC 62304), cybersecurity, risk management (ISO 14971), biocompatibility, battery safety (IEC 62133-2:2017), and cleaning/disinfection validation (ISO 17664-2:2021).
Technological Characteristics
Continuous-wave Doppler ultrasound system; 5 MHz operating frequency. Components: handheld probe, tablet computer. Connectivity: Bluetooth (probe to tablet). Power: rechargeable battery with inductive wireless charging. Applied part: Type BF. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62133-2:2017, ISO 14971, ISO 17664-2:2021.
Indications for Use
Indicated for non-invasive detection of blood flow in peripheral arteries in adults with intact skin. Used as an adjunct to standard clinical assessment for peripheral vascular status evaluation.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 8, 2026
BlueDop Medical, Ltd.
Jorge Millan
Regulatory Consultant
Sigma Biomedical
490 Sawgrass Corporate Pkwy.
Suite 130
Sunrise, Florida 33325
Re: K260700
Trade/Device Name: BlueDop Vascular Expert (BVE)
Regulation Number: 21 CFR 870.2100
Regulation Name: Cardiovascular Blood Flowmeter
Regulatory Class: Class II
Product Code: DPW
Dated: June 8, 2026
Received: June 9, 2026
Dear Jorge Millan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260700 - Jorge Millan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260700 - Jorge Millan
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# STEPHEN C. BROWNING -S
CDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260700
Device Name
BlueDop Vascular Expert (BVE)
Indications for Use (Describe)
The BlueDop Vascular Expert (BVE) is intended to non-invasively detect the presence of blood flow in peripheral arteries using Doppler ultrasound technology. The device provides Doppler waveform information to support clinical evaluation of peripheral vascular status.
The BVE is intended to be used as an adjunct to standard clinical assessment and does not provide a standalone diagnostic determination.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
BLUEDOP
BlueDop 510K Summary – July 6, 2026.
K260700
### 510(K) Summary
Submitter Information
| Submitter | BlueDop Medical Ltd. |
| --- | --- |
| Submitter Address | 2 Leman Street, London, UK E1W 9US |
| Contact: | Jorge Millan, PhD |
| Telephone number | +1 786 416 5587 |
| E-mail | jmillan@sigmabiomedical.com |
| Date prepared: | July 1, 2026 |
Device Name
| Trade/Proprietary Name: | BlueDop Vascular Expert (BVE) |
| --- | --- |
| 510(k) | K260700 |
| Regulation Number: | 21 CFR 870.2100 |
| Regulation Name: | Cardiovascular Blood Flow Meter |
| Product Code: | DPW |
| Class | II |
| Panel | Circulatory System Device |
Predicate Device
| Predicate Device: | VascuChek Clinical Device |
| --- | --- |
| Sponsor | Remington Medical, Inc. |
| 510(K) | K212065 |
| Regulation Number: | 21 CFR 870.2100 |
| Regulation Name: | Cardiovascular Blood Flow Meter |
| Product Code: | DPW |
| Class | II |
| Panel | Circulatory System Device |
### Indications for Use:
The BlueDop Vascular Expert (BVE) is intended to non-invasively detect the presence of blood flow in peripheral arteries using Doppler ultrasound technology. The device provides Doppler waveform information to support clinical evaluation of peripheral vascular status.
The BVE is intended to be used as an adjunct to standard clinical assessment and does not provide a standalone diagnostic determination.
Page 1 of 6
{5}
BLUEDOP
BlueDop 510K Summary – July 6, 2026.
K260700
## Device Description:
The BlueDop Vascular Expert (BVE) is a non-invasive, portable Doppler ultrasound system intended to detect the presence of blood flow in peripheral arteries using Doppler ultrasound technology. The system consists of a handheld ultrasound probe and a paired tablet computer. The probe transmits Doppler waveform data wirelessly to the tablet via Bluetooth for display and interpretation.
Conventional arm blood pressure cuff such as Summit Doppler ABI Vascular Cuff Package or other FDA-cleared equivalent (not included) are used to read the brachial blood pressure (BP) at Brachial, Posterior Tibial and Dorsalis Pedis levels bilaterally. The blood pressure readings are manually entered into the device. The probe incorporates diagnostic-level ultrasound emission and wireless communication capability. The tablet processes and displays the Doppler waveform and calculated parameters. The device is battery-powered and uses inductive wireless charging. The system does not deliver therapy and does not make autonomous diagnostic decisions.
## Non-Clinical Data:
The following verification and validation testing were performed to support substantial equivalence:
- Electrical safety testing per IEC 60601-1
- Electromagnetic compatibility testing per IEC 60601-1-2, including verification of essential performance under electromagnetic disturbance conditions
- Ultrasound acoustic output measurements in accordance with IEC 60601-2-37 and IEC 62127-1
- Doppler sensitivity measurements
- Doppler velocity measurement accuracy
- Temperature increase assessment
- Software verification and validation consistent with IEC 62304
- Cybersecurity verification
- Risk management per ISO 14971
- Biocompatibility evaluation of patient-contact materials
- Battery safety testing per IEC 62133-2:2017
- Cleaning and disinfection validation per ISO 17664-2:2021
The results of the non-clinical testing, including electrical safety, electromagnetic compatibility, ultrasound performance, software verification and validation, battery safety, cybersecurity, biocompatibility, and cleaning and disinfection validation, demonstrate that the BlueDop Vascular Expert meets all applicable safety and performance requirements and supports substantial equivalence to the predicate device.
Page 2 of 6
{6}
BLUEDOP
BlueDop 510K Summary – July 6, 2026.
K260700
## Predicate Device
The BlueDop Vascular Expert is compared to the predicate VascuChek Clinical Device (K212065) because both devices have the same intended use and employ the same continuous-wave Doppler ultrasound technology.
## Technical Characteristics Comparison:
The subject device has the same fundamental scientific technology and similar technological characteristics as the predicate device. The technological differences identified below were evaluated through appropriate verification and validation testing and demonstrated not to raise different questions of safety and effectiveness.
## Feature Comparison:
The subject device has similar features and functionality to the predicate device, as summarized in the following comparison table.
| Category | Primary Predicate: VascuChek Clinical Device (K212065) | Subject: BlueDop Vascular Expert (BVE) (K260700) | Comparison |
| --- | --- | --- | --- |
| Regulation / Classification | 21 CFR 870.2100, DPW, Class II | 21 CFR 870.2100, DPW, Class II | Same |
| Intended Use | Device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature. | The BlueDop Vascular Expert (BVE) is intended to non-invasively detect the presence of blood flow in peripheral arteries using Doppler ultrasound technology. The device provides Doppler waveform information to support clinical evaluation of peripheral vascular status. The BVE is intended to be used as an adjunct to standard clinical assessment and does not provide a standalone diagnostic determination. | Same |
| Prescription Status | Rx Only | Rx Only | Same |
| User Population | Qualified healthcare personnel | Qualified healthcare personnel | Same |
| Patient Population | Adults only | Adults, intact skin | Comparable intended patient population |
| Environment of Use | Hospitals, clinics | Hospitals, clinics | Same |
| Doppler Mode | Continuous-wave Doppler | Continuous-wave Doppler | Same |
| Operating Frequency | 8 MHz | 5 MHz | Comparable; no new risks |
Page 3 of 6
{7}
BLUEDOP
BlueDop 510K Summary – July 6, 2026.
K260700
| ISPTA.3 (mW/cm²) | 360 | 391 | Comparable acoustic output intensity values within accepted diagnostic ultrasound exposure ranges |
| --- | --- | --- | --- |
| Principle of Operation | Continuous-wave ultrasonic Doppler | Continuous-wave ultrasonic Doppler | Same fundamental Doppler ultrasound principle of operation |
| Doppler Connectivity to Base | Wireless Probe | Wireless Probe | Same |
| ABI Workflow | None | Manual entry of clinician-obtained systolic pressures; ABI calculation | Different |
| Waveform Storage | None | Waveforms displayed/stored digitally/printed | Different |
| Numeric Outputs | None | ABI L/R | Different |
| Display | None | Tablet touchscreen | Different |
| Audio Output | Internal speaker | Digital audio tablet speaker | Same function |
| Connectivity | None | Bluetooth probe + Printer | Different |
| Data Storage | None | Numeric Patient ID, Printed record, Database encrypted at rest | Different |
| Power Source | Transceiver: Permanent two cell lithium iron secondary battery. Speaker Dock includes a permanent single cell lithium ion secondary battery and an external power source, A/C to D/C power supply. | Rechargeable tablet + probe | Similar |
| Applied Part Classification | IPX1 | Type BF; IPX1 | Similar |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Ultrasound Output | Within 60601-2-37 | Within 60601-2-37 | Same limits |
Page 4 of 6
{8}
**BLUEDOP**
BlueDop 510K Summary – July 6, 2026.
K260700
| Software | Embedded firmware | 62304 Class B tablet app | Different |
| --- | --- | --- | --- |
| Cleaning | Wipe-only | Wipe-only | Same |
| Biocompatibility | Intact Skin | Intact Skin | Same |
| Sterility | Non-sterile | Non-sterile | Same |
## Evaluation of similarities and differences:
### Similarities
The BlueDop Vascular Expert (BVE) is substantially equivalent to the predicate VascuChek Clinical Device (K212065) with respect to its intended use, technological characteristics, and fundamental scientific technology. Both devices are prescription-use, non-invasive continuous-wave (CW) Doppler ultrasound systems intended for use by qualified healthcare professionals to evaluate peripheral arterial blood flow in hospital and clinical environments. Both devices employ the same Doppler ultrasound principle of operation to detect blood flow and generate audible Doppler signals for vascular assessment. The devices are intended for use on adults with intact skin and comply with the same applicable medical electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and diagnostic ultrasound output requirements (IEC 60601-2-37). The BVE has an acoustic output intensity comparable to that of the predicate and well within accepted diagnostic ultrasound exposure limits. In addition, both devices have comparable cleaning methods, are supplied non-sterile, are intended for repeated external use, and present the same overall risk profile associated with non-invasive Doppler vascular assessment.
### Differences
The differences between the BlueDop Vascular Expert and the predicate device relate primarily to user interface, data management, and additional workflow functionality rather than to the fundamental technology used to detect blood flow. The BVE incorporates a tablet-based touchscreen interface, digital waveform display, electronic waveform storage, encrypted patient database, Bluetooth communication, and printing capability, whereas the predicate relies on a more limited hardware interface and does not provide digital waveform storage or integrated patient record management. These technological enhancements improve usability and documentation but do not alter the underlying Doppler measurement methodology or the intended clinical use.
The BVE also includes an Ankle-Brachial Index (ABI) calculation feature that allows the clinician to manually enter systolic blood pressure measurements obtained using standard clinical techniques and calculates the ABI value. The device does not automatically measure blood pressure or independently diagnose peripheral arterial disease; rather, it performs a mathematical calculation using clinician-acquired measurements to assist clinical assessment. This functionality serves as a workflow aid and does not introduce a new scientific technology or new intended use. The ABI calculation algorithm was verified and validated through software verification and validation testing to confirm the accurate implementation of the mathematical calculation using manually entered systolic blood pressure values. Testing demonstrated that the software performs the calculation as intended and does not introduce new questions of safety or effectiveness.
Page 5 of 6
{9}
BLUEDOP
BlueDop 510K Summary – July 6, 2026.
K260700
The BVE operates at a nominal Doppler frequency of 5 MHz compared with 8 MHz for the predicate device. Both frequencies are well established for peripheral vascular Doppler examinations and utilize the same continuous-wave Doppler technology. Performance testing, including Doppler sensitivity measurements, Doppler velocity measurement accuracy testing, and ultrasound acoustic output measurements, confirmed that operation at 5 MHz provides equivalent performance for the intended use and does not introduce new questions of safety or effectiveness. Similarly, differences in power supply architecture, software implementation, applied part classification, and system electronics reflect design evolution and modern implementation while remaining compliant with the same consensus safety standards applicable to diagnostic ultrasound devices.
The technological differences between the BlueDop Vascular Expert and the predicate device were evaluated through appropriate verification and validation testing, including electrical safety testing (IEC 60601-1), electromagnetic compatibility testing (IEC 60601-1-2), battery safety testing (IEC 62133-2:2017), ultrasound acoustic output measurements, Doppler sensitivity testing, Doppler velocity measurement accuracy testing, temperature rise assessment, software verification and validation, cybersecurity verification, risk management in accordance with ISO 14971, biocompatibility evaluation, and cleaning and disinfection validation (ISO 17664-2:2021).
## Conclusion:
Non-clinical safety and performance testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), battery safety (IEC 62133-2:2017), ultrasound acoustic performance (IEC 60601-2-37), software verification and validation, cybersecurity, biocompatibility, and cleaning and disinfection validation (ISO 17664-2:2021), demonstrated that the BlueDop Vascular Expert met all predetermined acceptance criteria. The results demonstrate that the technological differences do not raise new questions of safety and effectiveness and that the BlueDop Vascular Expert performs as intended. Therefore, based on the intended use, technological characteristics, and results of the non-clinical performance testing, the BlueDop Vascular Expert is as safe and effective as the predicate device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.