VivaChek Blood Glucose and ß-Ketone Dual BLE Monitoring System;VivaChek Blood Glucose and ß-Ketone Dual Monitoring System.
Applicant
Vivachek Biotech (Hangzhou) Co., Ltd.
Product Code
NBW · Clinical Chemistry
Decision Date
Oct 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System is comprised of VivaChek Blood Glucose and β-Ketone Dual BLE Meter and VivaChek Blood Glucose and β-Ketone Dual Test Strips. The VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System is intended to quantitatively measure the glucose concentration and β-Ketone concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use) and should only be used by a single person and should not be shared. VivaChek Blood Glucose and β-Ketone Dual Monitoring System is comprised of VivaChek Blood Glucose and β-Ketone Dual Meter and VivaChek Blood Glucose and β-Ketone Dual Test Strips. The VivaChek Blood Glucose and β-Ketone Dual Monitoring System is intended to quantitatively measure the glucose concentration and β-Ketone concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use) and should only be used by a single person and should not be shared.
Device Story
System measures glucose and β-hydroxybutyrate in capillary whole blood via amperometric detection; uses glucose dehydrogenase/FAD and β-hydroxybutyrate dehydrogenase/NAD enzymes on test strips; capillary action draws blood into strip; meter displays both analyte concentrations simultaneously. Used by patients with diabetes at home for self-monitoring; Dual BLE model includes Bluetooth for wireless data transmission to mobile app. Healthcare providers use results to assess diabetes control effectiveness. System includes control solutions for quality checks; meter automatically recognizes control solution and excludes results from patient averages. Device is for single-patient use only to prevent bloodborne pathogen transmission.
Clinical Evidence
Lay user performance study (n=351) compared candidate device to YSI 2900C (glucose) and Randox RANBUT (ketone) comparators. Glucose results: 100% within ±15% of comparator. Ketone results: 100% within ±0.30 mmol/L for concentrations <1.5 mmol/L and 100% within ±15% for concentrations ≥1.5 mmol/L. Supplemental study confirmed accuracy at extreme analyte concentrations. Bench studies validated performance across hematocrit range (20-65%) and operating conditions.
Technological Characteristics
Amperometric detection; glucose dehydrogenase/FAD and β-hydroxybutyrate dehydrogenase/NAD enzymes. Measuring range: 20-600 mg/dL (glucose), 0.1-8.0 mmol/L (ketone). Connectivity: Bluetooth Low Energy (Dual BLE model). Power: Battery-operated. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61010-1. Single-patient use design; no calibration/coding required.
Indications for Use
Indicated for persons with diabetes for home, in vitro, quantitative measurement of glucose and β-ketone in fresh capillary whole blood from fingertips to monitor diabetes control. Not for neonatal use, diagnosis, or screening. Not for alternative site testing or healthcare/assisted-use settings.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
FORA ADVANCED GD-40 Blood Glucose and β-Ketone Monitoring System (K161738)
Submission Summary (Full Text)
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
## DECISION SUMMARY
ASSAY AND INSTRUMENT
### I Background Information:
#### A 510(k) Number
K260687
#### B Applicant
VivaChek Biotech (Hangzhou) Co., Ltd.
#### C Proprietary and Established Names
VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System
VivaChek Blood Glucose and β-Ketone Dual Monitoring System
#### D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
| JIN | Class I, meets the limitation of exemption 21 CFR 862.9(c)(5) | 21 CFR 862.1435 - Ketones (nonquantitative) test system | CH - Clinical Chemistry |
### II Submission/Device Overview:
#### A Purpose for Submission:
New devices
#### B Measurand:
Glucose and β-hydroxybutyrate in capillary whole blood obtained from the fingertip
#### C Type of Test:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
Quantitative amperometric detection of glucose using glucose dehydrogenase/FAD and β-hydroxybutyrate using 3-hydroxybutyrate dehydrogenase/NAD
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System is comprised of VivaChek Blood Glucose and β-Ketone Dual BLE Meter and VivaChek Blood Glucose and β-Ketone Dual Test Strips. The VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System is intended to quantitatively measure the glucose concentration and β-Ketone concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use) and should only be used by a single person and should not be shared.
VivaChek Blood Glucose and β-Ketone Dual Monitoring System is comprised of VivaChek Blood Glucose and β-Ketone Dual Meter and VivaChek Blood Glucose and β-Ketone Dual Test Strips. The VivaChek Blood Glucose and β-Ketone Dual Monitoring System is intended to quantitatively measure the glucose concentration and β-Ketone concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use) and should only be used by a single person and should not be shared.
#### C Special Conditions for Use Statement(s):
OTC - Over The Counter
- For in vitro diagnostic use only
- For single patient use only
- Not for neonatal use.
- Not for screening for or diagnosis of diabetes mellitus.
- Not for Alternative Site Testing (AST).
- Do not use the system above 10413 ft (3174 meters) in altitude.
- This meter is not intended for use in healthcare or assisted-use settings, such as hospitals, physician offices, or long-term care facilities. It has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures.
- Use of this meter on multiple patients may lead to transmission of Human Immunodeficiency Virus, Hepatitis C Virus, Hepatitis B Virus, or other bloodborne pathogens.
K260687 - Page 2 of 18
{2}
• Not for use on the specific patients listed in table below:
○ Critically ill patients
○ Patients in shock
○ Patients in a hyperglycemic- hyperosmolar state with or without ketosis
○ Patients with severe dehydration
○ Patients with severe hypotension
# D Special Instrument Requirements:
VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring Meter
VivaChek Blood Glucose and β-Ketone Dual Monitoring Meter
# IV Device/System Characteristics:
# A Device Description:
VivaChek Blood Glucose and β-Ketone Dual and Dual BLE Monitoring Systems are comprised of VivaChek Blood Glucose and β-Ketone Dual or Dual BLE Meter and VivaChek Blood Glucose and β-Ketone Dual Test Strips. This system is designed to quantitatively measure both glucose and β-ketone (β-hydroxybutyrate) concentrations in fresh capillary whole blood samples drawn from fingertips. The test strip measures both analytes simultaneously from a single blood sample. Both glucose and β-ketone results are displayed on the meter screen simultaneously after each test. The Dual and Dual BLE meters differ only in the addition of Bluetooth Low Energy connectivity to the Dual BLE meter for wireless transmission of test results to the optional VivaChek GlucoKet App.
# B Principle of Operation:
VivaChek Blood Glucose and β-Ketone Dual and Dual BLE Monitoring Systems are designed to quantitatively measure both glucose and β-Ketone concentrations in a single sample of fresh capillary whole blood. The glucose measurement and β-Ketone measurements are achieved by using an amperometric detection method. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose dehydrogenase and β-Ketone (β-hydroxybutyrate) in the sample reacts with β-hydroxybutyrate dehydrogenase. The magnitude of the generated electrical current is proportional to the glucose and β-Ketone amounts in the sample. After the reaction time, the glucose and β-Ketone concentration values are reported in plasma equivalents and are displayed on meter.
# C Instrument Description Information:
1. Instrument Name:
VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring Meter
VivaChek Blood Glucose and β-Ketone Dual Monitoring Meter
2. Specimen Identification:
K260687 - Page 3 of 18
{3}
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected
3. Specimen Sampling and Handling:
Samples are to be tested immediately upon collection.
4. Calibration:
The meters do not require calibration or coding by the user.
5. Quality Control:
The sponsor manufactures VivaChek Glucose and β-Ketone Dual Control Solutions available in three (3) levels (1, 2, and 3). Each control solution level contains both glucose and β-ketone at varying concentrations: Control Solution 1 represents low concentrations of both analytes, Control Solution 2 represents mid-range concentrations, and Control Solution 3 represents high concentrations. When the user runs each control solution level, they obtain both glucose and β-ketone results simultaneously. Each test strip vial includes the expected control ranges for both analytes at each level. The user is instructed to compare the control test results with the expected control ranges printed on the test strip vial label. Users are given instructions on when to conduct control solution testing and to call customer service if repeat testing with control material is out of range. The meter automatically recognizes control solution and marks control solution results accordingly; control solution results are not included in the patient result averages.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
FORA ADVANCED GD-40 Blood Glucose and β-Ketone Monitoring System
B Predicate 510(k) Number(s):
K161738
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K260687 | K161738 |
| --- | --- | --- |
| Device Trade Name | VivaChek Blood Glucose and β -Ketone Dual BLE Monitoring System VivaChek Blood Glucose and β -Ketone | FORA ADVANCED GD-40 Blood Glucose and β-Ketone Monitoring System |
K260687 - Page 4 of 18
{4}
| | Dual Monitoring System | |
| --- | --- | --- |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | It is intended for the quantitative measurement of glucose in mg/dL and β-ketone in mmole/L in fresh capillary whole blood from the finger. For single-patient use. | Same |
| Glucose Measuring Range | 20 – 600 mg/dL | Same |
| β-Ketone Measuring Range | 0.1 - 8.0 mmol/L | Same |
| General Device Characteristic Differences | | |
| Operating Condition | 45 – 113 °F (7 - 45 °C), 10 - 90% RH | 50 – 104 °F(10 – 40 °C), 10 - 85% RH |
| Hematocrit Range | 20 - 65% | 20 - 70% |
| Sample Volume | 1.6 μL | Blood glucose: 0.9 μL; Blood β-ketone: 1.0 μL |
| Test Time | Glucose: 5 seconds, β-ketone: 7 seconds | Glucose: 5 seconds β-ketone: 10 seconds |
| Meter Memory | 300 records | 1000 records |
## VI Standards/Guidance Documents Referenced:
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)]
FDA Guidance Document: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, issued September 29, 2020
K260687 - Page 5 of 18
{5}
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
The sponsor conducted the analytical and clinical studies discussed below using the VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System to represent both candidate systems (with and without Bluetooth Low Energy capability). The differences between the two systems is limited to the presence or absence of BLE functionality. Therefore, the studies using the BLE meter are representative of the performance of both device systems in this 510(k) submission.
# 1. Precision/Reproducibility:
# Glucose:
A glucose within-run precision (repeatability) study using 3 lots of VivaChek Blood Glucose and β-Ketone Dual Test Strips was performed. Venous blood was spiked to 5 concentrations (30-50, 51-110, 111-150, 151-250, 251-400 mg/dL). For each sample concentration, 10 meters were used, with 10 measurements taken by each meter (i.e., 100 measurements per concentration per lot). Results are summarized below:
| Venous Blood Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | CV |
| --- | --- | --- | --- | --- | --- |
| Level 1 (30-50) | Lot 1 | 100 | 45.3 | 1.7 | 3.8% |
| | Lot 2 | 100 | 45.4 | 1.8 | 3.9% |
| | Lot 3 | 100 | 45.6 | 1.7 | 3.7% |
| | Combined | 300 | 45.4 | 1.7 | 3.8% |
| Level 2 (51-110) | Lot 1 | 100 | 70.3 | 2 | 2.8% |
| | Lot 2 | 100 | 70.3 | 2.1 | 3.0% |
| | Lot 3 | 100 | 70.4 | 2.1 | 3.0% |
| | Combined | 300 | 70.3 | 2.1 | 2.9% |
| Level 3 (111-150) | Lot 1 | 100 | 135.8 | 3.6 | 2.7% |
| | Lot 2 | 100 | 135.8 | 3.5 | 2.6% |
| | Lot 3 | 100 | 136.6 | 3.3 | 2.4% |
| | Combined | 300 | 136 | 3.5 | 2.6% |
| Level 4 (151-250) | Lot 1 | 100 | 207.7 | 5.3 | 2.6% |
| | Lot 2 | 100 | 205.5 | 5.6 | 2.7% |
| | Lot 3 | 100 | 205.7 | 5.7 | 2.8% |
| | Combined | 300 | 206.3 | 5.6 | 2.7% |
| Level 5 (251-400) | Lot 1 | 100 | 326.4 | 8.3 | 2.5% |
| | Lot 2 | 100 | 324.9 | 8.3 | 2.6% |
| | Lot 3 | 100 | 325.3 | 7.9 | 2.4% |
| | Combined | 300 | 325.5 | 8.2 | 2.5% |
K260687 - Page 6 of 18
{6}
Glucose intermediate precision of the VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System was evaluated using 5 levels of glucose control solutions (30-50, 51-110, 111-150, 151-250, 251-400 mg/dL) using 3 lots of VivaChek Blood Glucose and β-Ketone Dual Test Strips. The tests were conducted with 10 meters, one measurement per meter per day at the five different glucose ranges, for ten days for a total of 300 glucose results per each glucose level. Results are summarized below:
| Glucose Control Levels (mg/dL) | Strip lot | N | Mean (mg/dL) | SD (mg/dL) | CV |
| --- | --- | --- | --- | --- | --- |
| L1 (30-50) | Lot 1 | 100 | 38.6 | 1.4 | 3.8% |
| | Lot 2 | 100 | 38.4 | 1.4 | 3.7% |
| | Lot 3 | 100 | 38.4 | 1.5 | 3.8% |
| | Combined | 300 | 38.5 | 1.4 | 3.8% |
| L2 (51-110) | Lot 1 | 100 | 72.8 | 1.9 | 2.7% |
| | Lot 2 | 100 | 72.5 | 2.1 | 2.9% |
| | Lot 3 | 100 | 72.6 | 2 | 2.8% |
| | Combined | 300 | 72.6 | 2 | 2.8% |
| L3 (111-150) | Lot 1 | 100 | 135.3 | 3.1 | 2.3% |
| | Lot 2 | 100 | 134.9 | 3.4 | 2.5% |
| | Lot 3 | 100 | 134.7 | 3.2 | 2.4% |
| | Combined | 300 | 135 | 3.2 | 2.4% |
| L4 (151-250) | Lot 1 | 100 | 215.4 | 5.4 | 2.5% |
| | Lot 2 | 100 | 214.8 | 5.5 | 2.6% |
| | Lot 3 | 100 | 215.5 | 5.2 | 2.4% |
| | Combined | 300 | 215.3 | 5.4 | 2.5% |
| L5 (251-400) | Lot 1 | 100 | 364.5 | 9.4 | 2.6% |
| | Lot 2 | 100 | 366.3 | 9.5 | 2.6% |
| | Lot 3 | 100 | 365.3 | 9.1 | 2.5% |
| | Combined | 300 | 365.5 | 9.3 | 2.6% |
# Ketone:
The sponsor performed a ketone repeatability (within-run) precision study using 3 lots of VivaChek Blood Glucose and β-Ketone Dual Test Strips. Venous blood samples that were spiked to 5 ketone concentrations (0.1-1.0, 1.1-2.0, 2.1-3.0, 4.1-5.0, 6.1-7.0 mmol/L) were used. For each sample concentration, 10 meters were used, with 10 measurements taken by each meter (i.e., 100 measurements per concentration per lot). Results are summarized below:
K260687 - Page 7 of 18
{7}
| Venous Blood Ketone Level (mmol/L) | Lot | N | Mean (mmol/L) | SD (mmol/L) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 (0.1-1.0) | Lot 1 | 100 | 0.6 | 0.06 | 9.2% |
| | Lot 2 | 100 | 0.58 | 0.05 | 9.3% |
| | Lot 3 | 100 | 0.59 | 0.06 | 9.5% |
| | Combined | 300 | 0.59 | 0.06 | 9.4% |
| Level 2 (1.1-2.0) | Lot 1 | 100 | 1.43 | 0.06 | 3.9% |
| | Lot 2 | 100 | 1.42 | 0.06 | 4.1% |
| | Lot 3 | 100 | 1.44 | 0.06 | 3.9% |
| | Combined | 300 | 1.43 | 0.06 | 4.0% |
| Level 3 (2.1-3.0) | Lot 1 | 100 | 2.45 | 0.08 | 3.3% |
| | Lot 2 | 100 | 2.47 | 0.08 | 3.2% |
| | Lot 3 | 100 | 2.45 | 0.08 | 3.4% |
| | Combined | 300 | 2.46 | 0.08 | 3.3% |
| Level 4 (4.1-5.0) | Lot 1 | 100 | 4.7 | 0.15 | 3.2% |
| | Lot 2 | 100 | 4.68 | 0.14 | 3.0% |
| | Lot 3 | 100 | 4.67 | 0.15 | 3.2% |
| | Combined | 300 | 4.68 | 0.15 | 3.2% |
| Level 5 (6.1-7.0) | Lot 1 | 100 | 6.57 | 0.21 | 3.2% |
| | Lot 2 | 100 | 6.58 | 0.22 | 3.4% |
| | Lot 3 | 100 | 6.63 | 0.21 | 3.2% |
| | Combined | 300 | 6.59 | 0.22 | 3.3% |
Ketone intermediate precision was evaluated using five levels of β-ketone control solutions (0.1-1.0, 1.1-2.0, 2.1-3.0, 4.1-5.0, and 6.1-7.0 mmol/L). Testing was conducted over 10 days with three test strip lots and 10 meters. For each control solution 10 replicates were taken each day resulting in 100 individual measurements per level:
| Venous Blood Ketone Level (mmol/L) | Lot | N | Mean (mmol/L) | SD (mmol/L) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 (0.1-1.0) | Lot 1 | 100 | 0.6 | 0.05 | 8.5% |
| | Lot 2 | 100 | 0.61 | 0.05 | 8.4% |
| | Lot 3 | 100 | 0.61 | 0.05 | 8.6% |
| | Combined | 300 | 0.6 | 0.05 | 8.5% |
| Level 2 (1.1-2.0) | Lot 1 | 100 | 1.49 | 0.06 | 3.8% |
| | Lot 2 | 100 | 1.5 | 0.05 | 3.6% |
| | Lot 3 | 100 | 1.49 | 0.06 | 3.8% |
| | Combined | 300 | 1.49 | 0.06 | 3.7% |
| Level 3 (2.1-3.0) | Lot 1 | 100 | 2.51 | 0.07 | 2.7% |
| | Lot 2 | 100 | 2.49 | 0.07 | 2.9% |
| | Lot 3 | 100 | 2.51 | 0.07 | 2.8% |
| | Combined | 300 | 2.5 | 0.07 | 2.8% |
| Level 4 (4.1-5.0) | Lot 1 | 100 | 4.49 | 0.13 | 2.9% |
| | Lot 2 | 100 | 4.49 | 0.14 | 3.1% |
K260687 - Page 8 of 18
{8}
| Venous Blood Ketone Level (mmol/L) | Lot | N | Mean (mmol/L) | SD (mmol/L) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| | Lot 3 | 100 | 4.52 | 0.13 | 2.8% |
| | Combined | 300 | 4.5 | 0.13 | 2.9% |
| Level 5 (6.1-7.0) | Lot 1 | 100 | 6.53 | 0.18 | 2.8% |
| | Lot 2 | 100 | 6.49 | 0.17 | 2.6% |
| | Lot 3 | 100 | 6.5 | 0.18 | 2.7% |
| | Combined | 300 | 6.5 | 0.18 | 2.7% |
Additional bench studies were conducted to assess the potential impact of various combinations of extreme glucose and ketone concentrations, under different conditions and hematocrit concentrations on both glucose and ketone results. Refer to Section E below for additional detail.
# 2. Linearity:
# Glucose Linearity
The linearity of glucose measurements was evaluated using venous whole blood samples adjusted to eleven glucose concentrations ranging from 22.6 to 583.3 mg/dL (22.6, 78.3, 136.5, 198.5, 250.0, 309.0, 365.5, 433.0, 484.5, 539.0, and 583.5 mg/dL as measured by the comparator method). Four replicates with three test strip lots per glucose concentration were assessed. The summary of the linear regression analysis for each lot was as follows:
| Lot # | Slope | y-intercept | R² value |
| --- | --- | --- | --- |
| Lot 1 | 0.995 | 0.931 | 0.998 |
| Lot 2 | 0.9903 | 2.1825 | 0.9978 |
| Lot 3 | 0.9938 | 1.4004 | 0.9976 |
| All 3 Strip Lots | 0.993 | 1.5046 | 0.9978 |
The results of the study support the sponsor's claimed glucose measuring range of 20-600 mg/dL. If a sample result is less than 20 mg/dL glucose, the result is flagged by the meter as "LO". If a sample result exceeds 600 mg/dL glucose, the result is flagged by the meter as "HI". The "LO" and "HI" functions were validated by the sponsor and were demonstrated to function as intended.
# Ketone Linearity
Linearity of ketone measurements was evaluated using venous whole blood samples adjusted to eleven ketone concentrations ranging from 0.12 to 7.76 mmol/L (0.12, 0.35, 0.65, 1.16, 1.83, 2.52, 3.32, 4.31, 5.23, 6.53, and 7.76 mmol/L as measured by the comparator method). Three test strip lots were used for the linearity evaluation study. ). Four replicates with three test strip lots per ketone concentration were assessed. The summary of the linear regression analysis for each lot was as follows:
| Lot # | Slope | y-intercept | R² value |
| --- | --- | --- | --- |
| Lot 1 | 0.9984 | 0.0076 | 0.9972 |
| Lot 2 | 0.9895 | 0.0349 | 0.9973 |
K260687 - Page 9 of 18
{9}
| Lot # | Slope | y-intercept | R² value |
| --- | --- | --- | --- |
| Lot 3 | 0.9921 | 0.0088 | 0.9978 |
| All 3 Strip Lots | 0.9951 | 0.0151 | 0.9975 |
The results of the study support the sponsor's claimed β-ketone measuring range of 0.1-8.0 mmol/L. If a sample result is less than 0.1 mmol/L ketone, the result is flagged by the meter as "< 0.1 mmol/L". If a sample result exceeds 8.0 mmol/L ketone, the result is flagged by the meter as "> 8.0 mmol/L". The "< 0.1 mmol/L" and "> 8.0 mmol/L" functions were validated by the sponsor and were demonstrated to function as intended.
### 3. Analytical Specificity/Interference:
To evaluate the potential interference exogenous and endogenous substances on both glucose and β-ketone measurement functions, venous whole blood samples at three concentration levels of glucose (50-70 mg/dL, 110-130 mg/dL, and 225-270 mg/dL) or β-ketone (0.1-1.0 mmol/L, 1.1-2.0 mmol/L, and 3.1-4.0 mmol/L) were assessed.
Each sample was divided into a control sample (with no added interferent) and a test sample (with added potential interferent). Results from the test samples obtained using the candidate system were compared to the results obtained with the control samples using the candidate system. The highest tested concentration for each substance at which no significant interference was observed (defined by the sponsor as bias within ±4 mg/dL for glucose concentrations < 75 mg/dL and ±8% for glucose concentrations ≥ 75 mg/dL, and bias within ±0.20 mmol/L for β-ketone concentrations < 1.5 mmol/L and ±10% for β-ketone concentrations ≥ 1.5 mmol/L) for either glucose or ketone measurement is summarized in the following table:
| Substance | Highest concentration at which no significant interference is observed |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Acetoacetate | 21 mg/dL |
| Acetone | 70 mg/dL |
| Acetylsalicylic Acid | 50 mg/dL |
| Acyclovir | 3.1 mg/dL |
| Allopurinol | 5 mg/dL |
| Amitriptylline | 0.27 mg/dL |
| Amoxicillin | 12.5 mg/dL |
| Ampicillin | 5 mg/dL |
| Ascorbic Acid | 6 mg/dL |
| Aspirin | 60 mg/dL |
| Atenolol | 10 mg/dL |
| Bicarbonate | 336 mg/dL |
| Bilirubin (Conjugated) | 50 mg/dL |
| Bilirubin (Unconjugated) | 40 mg/dL |
| Caffeine | 10 mg/dL |
K260687 - Page 10 of 18
{10}
| Substance | Highest concentration at which no significant interference is observed |
| --- | --- |
| Calcium | 5 mmol/L |
| Captopril | 500 mg/dL |
| Ceftriaxone | 250 mg/dL |
| Chloride | 140 mmol/L |
| Cholic Acid | 6 mg/dL |
| Cholesterol | 500 mg/dL |
| Clonidine | 2 mg/dL |
| Creatinine | 30 mg/dL |
| Digoxin | 0.16 mg/dL |
| Diphenhydramine | 1 mg/dL |
| Dopamine | 1.25 mg/dL |
| EDTA | 0.1 mg/dL |
| Enalapril | 0.15 mg/dL |
| Ephedrine HCl | 60 mg/dL |
| Erythromycin | 20 mg/dL |
| Estrone | 0.1 mg/dL |
| Famotidine | 0.13 mg/dL |
| Fluoxetine | 0.8 mg/dL |
| Folic Acid | 13.3 mg/dL |
| Fructose | 1000 mg/dL |
| Furosemide | 2 mg/dL |
| Galactose | 1000 mg/dL |
| Gamma-Globulin | 12000 mg/dL |
| Gentisic Acid | 2 mg/dL |
| Glyburide | 1.07 mg/dL |
| Glycerol | 1000 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Heparin | 300 IU/dL |
| Ibuprofen | 55 mg/dL |
| Icodextrin | 2000 mg/dL |
| Isomalt | 1000 mg/dL |
| Lactitol | 1000 mg/dL |
| Lactose | 1000 mg/dL |
| Levo-Dopa (L-Dopa) | 0.75 mg/dL |
| Lidocaine | 6 mg/dL |
| Magnesium | 5 mmol/L |
| Maltitol | 1000 mg/dL |
| Maltose | 1000 mg/dL |
| Mannitol | 1800 mg/dL |
| Mannose | 1.15 mg/dL |
K260687 - Page 11 of 18
{11}
| Substance | Highest concentration at which no significant interference is observed |
| --- | --- |
| Metaproterol | 1.81 mg/dL |
| Metformin HCl | 50 mg/dL |
| Methyldopa | 2 mg/dL |
| Metoprolol | 0.3 mg/dL |
| N-Acetylcysteine | 2.59 mg/dL |
| Naproxen | 100 mg/dL |
| Nifedipine | 0.17 mg/dL |
| Nortriptyline | 0.15 mg/dL |
| Penicillin | 12 mg/dL |
| Phenytoin | 10 mg/dL |
| Piroxicam | 5 mg/dL |
| Potassium | 10 mmol/L |
| Pralidoxime Iodide | 5 mg/dL |
| Reduced Glutathione | 30 mg/dL |
| Salicylic Acid | 60 mg/dL |
| Sodium | 200 mmol/L |
| Sorbitol | 1000 mg/dL |
| Sulfamethoxazole | 120 mg/dL |
| Sulfate | 5 mmol/L |
| Tetracycline | 30 mg/dL |
| Theophylline | 25 mg/dL |
| Terfenadine | 0.45 mg/dL |
| Tolazamide | 9 mg/dL |
| Tolbutamide | 72 mg/dL |
| Triglycerides | 3000 mg/dL |
| Trimethoprim | 12.5 mg/dL |
| Urea | 600 mg/dL |
| Uric Acid | 24 mg/dL |
| Vancomycin | 25 mg/dL |
| Verapamil | 0.45 mg/dL |
| Vitamin E | 20 mg/dL |
| Warfarin | 2 mg/dL |
| Xylitol | 1000 mg/dL |
| Xylose | 1000 mg/dL |
All substances were tested at the concentrations listed above and demonstrated no significant interference with either glucose or β-ketone measurements according to the predefined acceptance criteria.
In addition, to assess the impact of pH changes on device performance, pH interference was evaluated at two levels: pH 7.07 (representing severe acidosis as observed in DKA patients)
K260687 - Page 12 of 18
{12}
and pH 7.31 (representing mild acidosis at the lower end of the normal physiological range). Testing was conducted using venous whole blood samples at three glucose and β-ketone concentration levels (50-70 mg/dL and 0.1-1.0 mmol/L; 110–130 mg/dL and 1.1–2.0 mmol/L; 225–270 mg/dL glucose and 3.1–4.0 mmol/L β-ketone) across 10 meters and three test strip lots. Results were compared results from a control sample at normal pH and demonstrated no significant impact of pH on either glucose or β-ketone measurements.
4. Detection Limit and Assay Reportable Range:
20 - 600 mg/dL for glucose
0.1 - 8.0 mmol/L for ketone
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Glucose Traceability
The system reports plasma-equivalent glucose values and is calibrated using the YSI 2900C Biochemistry Analyzer, which is traceable to NIST SRM 917d glucose reference material.
β-Ketone Traceability
The system reports plasma-equivalent ketone values that are traceable to commercially available D-3 Hydroxybutyrate calibrator material.
Stability:
Accelerated and real-time closed-vial and open-vial stability study protocols and acceptance criteria were reviewed and found acceptable to support the sponsor's claimed 24-month closed-vial shelf life and 6-month open-vial use claim for the VivaChek Blood Glucose and β-Ketone Test Strips when stored at 36–86°F (2–30°C) and 10–90% relative humidity.
6. Assay Cut-Off:
Not applicable.
# B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable.
2. Matrix Comparison:
Not applicable. The device is only intended for use with fresh capillary whole blood from a fingerstick.
# C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
K260687 - Page 13 of 18
{13}
2. Clinical Specificity:
Not applicable.
3. Clinical Cut-Off
Not applicable.
4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
## Method Comparison/Lay User Performance Study
### Glucose
A total of 351 lay-user participants obtained their own fingerstick samples and both glucose and ketone results using only the system components and instructions provided in the labeling. Three glucose test strip lots were used for the study. The glucose concentrations of the participant samples ranged from 61.7 to 371 mg/dL as measured by the comparator method (YSI 2900C Biochemistry Analyzer). The study included 12 unaltered samples with blood glucose concentrations < 80 mg/dL and 36 unaltered samples with blood glucose concentration >250 mg/dL.
The \( \beta \) -ketone concentrations obtained by the participant samples ranged from 0.07 to 1.81 mmol/L as measured by the comparator method (Randox RANBUT D-3-Hydroxybutyrate Assay run on the Randox RX Imola Chemistry Analyzer).
The glucose results obtained by the participants using the candidate device were compared to results obtained using the comparator method and summarized in the table below:
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 219/351 (62.4%) | 334/351 (95.2%) | 351/351 (100%) | 351/351 (100%) |
Regression analysis results: \( y = 0.9912x + 1.3505 \) , \( R^{2} = 0.9734 \)
### Ketone
The \( \beta \) -ketone results from the same 351 lay user participants using the candidate device were compared to the results obtained from the comparator method (Randox RANBUT D-3-Hydroxybutyrate Assay run on the Randox RX Imola Chemistry Analyzer; 0.07-1.81 mmol/L) and are summarized in the tables below:
| For all β-ketone concentrations <1.5 mmol/L | | |
| --- | --- | --- |
| Within ±0.15 mmol/L | Within ±0.225 mmol/L | Within ±0.30 mmol/L |
| 330/345 (95.7%) | 344/345 (99.7%) | 345/345 (100%) |
| For β-ketone concentrations ≥1.5 mmol/L | | |
| --- | --- | --- |
| Within ±10% | Within ±15% | Within ±20% |
| 5/6 (83.3%) | 6/6 (100%) | 6/6 (100%) |
K260687 - Page 14 of 18
{14}
Regression analysis results: y = 0.9890x + 0.0074; R² = 0.9621
## Usability Assessment:
A usability evaluation was conducted to assess the ability of intended users to understand the labeling and operate the VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System. The study demonstrated that participants were able to understand the labeling, use the VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System, and interpret testing results and error messages without assistance.
## Readability
A Flesch-Kincaid readability assessment was conducted on the user manuals and test strip insert demonstrating that the labeling was written at an 8th grade reading level or less.
## Supplemental Method Comparison Study
To address the limited number of samples with both high glucose and high ketone levels in the lay user study, the sponsor conducted a small supplemental study with native patient capillary fingertip samples. The samples contained both elevated glucose and elevated β-ketone concentrations. Glucose results were compared to results obtained from the glucose comparator and β-ketone results were compared to result obtained from the β-ketone comparator. The results demonstrated that combinations of high glucose and high ketone concentrations did not significantly impact the accuracy of either glucose or β-ketone results.
## Accuracy at extreme glucose and ketone concentrations:
An additional study was conducted to evaluate glucose and β-ketone accuracy performance with altered capillary fingerstick samples with extreme glucose and ketone concentrations and 3 test strip lots. Fifty (50) samples were glycolyzed to achieve samples with glucose concentrations less than 80 mg/dL (range: 21.7-79.5 mg/dL). Glucose results from these samples using the candidate device were compared to the glucose comparator method. Fifty (50) samples were spiked to achieve glucose concentrations greater than 250 mg/dL (258-593 mg/dL) and β-ketone concentrations ≥1.5 mmol/L (1.63-7.8 mmol/L). Glucose and β-ketone results from the candidate device were compared to results obtained using the glucose and the ketone comparator. Results are summarized in the tables below:
Glucose results
| For glucose concentration <80 mg/dL | | | |
| --- | --- | --- | --- |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 60% (30/50) | 90% (45/50) | 100% (50/50) | 100% (50/50) |
| For glucose concentration >250 mg/dL | | | |
| 66% (33/50) | 94% (47/50) | 100% (50/50) | 100% (50/50) |
Ketone results
| For β-ketone concentrations >1.5 mmol/L | | | |
| --- | --- | --- | --- |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 58% (29/50) | 88% (44/50) | 100% (50/50) | 100% (50/50) |
K260687 - Page 15 of 18
{15}
See section E below for additional bench studies conducted to assess the potential impact of various combinations of extreme glucose and ketone concentrations, operating conditions and hematocrit concentrations on both glucose and ketone results.
### D Expected Values/Reference Range:
The sponsors included the following in the labeling:
#### Glucose
The ideal ranges for adults without diabetes are:
- Under 100 mg/dL before meals.
- Under 140 mg/dL after meals.
American Diabetes Association, Standard of Medical Care in Diabetes 2020, Vol. 43 (Suppl. 1).
#### Ketone
Based on published literature, the sponsor included < 0.6 mmol/L as the expected value for ketones and cites the following reference:
Evans K. Diabetic ketoacidosis: update on management. Clin Med (Lond). 2019 Sep; 19 (5): 396-398.
### E Other Supportive Instrument Performance Characteristics Data:
The sponsor conducted an additional bench study to evaluate the potential impact of various combinations of extreme glucose and ketone concentrations, under different conditions and hematocrit concentrations on both glucose and ketone results. The study assessed venous whole blood samples prepared at four glucose concentration levels (50–70, 110–130, 225–270, and 500–600 mg/dL) at each of four β-ketone concentration ranges (~0.6, ~1.5, ~3.0, and ~7.0 mmol/L). The samples were assessed in replicates of 30 under each of the conditions. Each analyte combination was tested in replicates of 30 and three test strip lots. Glucose and β-ketone results were compared to results from the respective glucose and β-ketone comparator methods. The SDs/%CVs were calculated from the replicate measurements for each analyte combination. The results demonstrated that various analyte combinations did not significantly impact either glucose or β-ketone measurement accuracy or precision.
#### Hematocrit
The effect of different hematocrit levels was evaluated using venous whole blood samples with hematocrit levels of 20-65% (20, 25, 31, 35, 42, 49, 55, 61, and 65%). The same blood samples contained both glucose and β-ketone at five concentration levels each: glucose levels of 30-50, 51-110, 111-150, 151-250, and 251-400 mg/dL, and β-ketone levels of 0.1-1.0, 1.1-2.0, 2.1-3.0, 4.1-5.0, and 6.1-7.0 mmol/L. Glucose and β-ketone results from the candidate system were compared to results obtained using the respective comparator methods and to the candidate system results from samples with normal hematocrit (~42%). The results demonstrated acceptable performance across the claimed hematocrit range of 20-65%.
#### System Operating Conditions
K260687 - Page 16 of 18
{16}
The effect of different operating conditions was evaluated using venous whole blood samples that were spiked to contain both glucose (60-70, 100-120, and 300-400 mg/dL) and β-ketone (0.1-1.0, 1.1-2.0, and 3.1-4.0 mmol/L) concentrations. Candidate system results were obtained at the following temperature and relative humidity (RH) combinations: low temperature/low humidity (43°F and 8% RH), low temperature/high humidity (43°F and 92% RH), high temperature/low humidity (115°F and 8% RH), and high temperature/high humidity (115°F and 92% RH). The results were compared to candidate system results obtained at nominal conditions (73.4°F/40% RH): The results support the operating condition claims in the labeling of 45-113°F and relative humidity of 10% to 90%.
## Sample Volume
The sponsor performed a study to support the claimed minimum sample volume of 1.6 μL. Venous whole blood samples with three levels of glucose concentrations (approximately 55, 115, and 230 mg/dL) and three levels of β-ketone concentrations (approximately 0.66, 1.46, and 3.55 mmol/L) were tested at three sample volumes (0.8, 1.2, and 1.6 μL) using 3 lots of test strips. The results of the study support the claimed sample volume of 1.6 μL and demonstrate that the insufficient sample volume error message functions as intended.
## Infection Control
The device systems are intended for single-patient use only. Disinfection efficacy studies were performed on the external materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, Clorox™ Healthcare Bleach Germicidal Wipes (EPA Registration No. 67619-12). A robustness study was also conducted by the sponsor demonstrating that there was no change in glucose or ketone performance or in the external materials of the meter after 608 cleaning and disinfection cycles using the Clorox™ Healthcare Bleach Germicidal Wipes. The robustness studies were designed to simulate 5 years of single-patient device use.
## Flex Studies
The following additional flex studies were performed with the VivaChek Blood Glucose and β-Ketone Dual BLE Monitoring System: shipping and handling simulation, low battery effect, intermittent sampling, sample perturbation, incorrect strip insertion, testing with used test strips, and test strip early removal. Both glucose and β-ketone measurements were assessed across these flex studies. The testing performed demonstrated that the device is robust under these expected use conditions.
## Electrical Safety and EMC Testing
The sponsor provided documentation certifying that acceptable electrical safety and electromagnetic compatibility (EMC) testing had been performed, and the system was found to be compliant.
## Software and Cybersecurity
Software and cybersecurity documentation was reviewed and found to be acceptable.
## Test Strip Lot Release Protocol
Glucose test strip lot release protocols and criteria were reviewed and found to be acceptable.
K260687 - Page 17 of 18
{17}
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K260687 - Page 18 of 18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.