Retrospective clinical ultrasound images from 150 subjects
The retrospective clinical image dataset was used to validate the performance of the AI Auto Measure Abdomen algorithm by comparing AI-generated measurements against manual measurements performed by clinical experts.
AI algorithm validation; Retrospective clinical images; Abdominal ultrasound; Performance validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Auto Measure Abdomen AI Performance Validation; Retrospective performance validation study
150 subjects (adults ≥18 years) including patients referred for abdominal/renal exams and healthy volunteers; Sample Size: 150 subjects (292 kidney longitudinal images, 271 kidney transverse images, 145 spleen sagittal images); Number of Sites: 3
Manual measurements performed by 3 clinical experts (ground truth)
Bland-Altman Limits of Agreement (LoA) between AI-generated measurements and manual ground truth
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Kidney sagittal length measurement
—
Bland-Altman LoA within [-14.3%, 14.3%]
95% LoA (-6.41%, 7.33%)
—
—
Performance validation study: 150 subjects
3 (clinical experts)
Kidney transverse width measurement
—
Bland-Altman LoA within [-33.7%, 33.7%]
95% LoA (-16.97%, 17.49%)
—
—
Performance validation study: 150 subjects
3 (clinical experts)
Kidney transverse height measurement
—
Bland-Altman LoA within [-30.1%, 30.1%]
95% LoA (-11.92%, 13.00%)
—
—
Performance validation study: 150 subjects
3 (clinical experts)
Spleen length measurement
—
Bland-Altman LoA within [-15.9%, 15.9%]
95% LoA (-7.26%, 11.94%)
—
—
Performance validation study: 150 subjects
3 (clinical experts)
Indications for Use
The Alturion Series Diagnostic Ultrasound System is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body in the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. Modes of Operation include B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging. The clinical environments where the Alturion Series Diagnostic Ultrasound System can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Device Story
Alturion Series Diagnostic Ultrasound System is a software-controlled, cart-based diagnostic ultrasound system; acquires and displays ultrasound data via transducers. System includes console with user interface, display, and electronics; supports B-mode (3D/4D), M-mode, PW/CW/Color Doppler, and Harmonic Imaging. Operated by trained healthcare professionals in clinics, hospitals, and point-of-care settings. Features 'AI Auto Measure Abdomen' software for semi-automated measurement of kidney and spleen dimensions; provides editable measurements to improve workflow efficiency. Clinicians review and accept/edit AI-generated measurements to support clinical decision-making. New eV14-2v endocavitary transducer provides improved field of view and resolution. System shares common software architecture and acoustic capabilities with predicate Affiniti series.
Clinical Evidence
No clinical trials performed. Performance validation study for AI Auto Measure Abdomen used 150 retrospective ultrasound exams (78% normal, 22% abnormal kidney; 98.67% normal, 1.33% abnormal spleen). AI-generated measurements (kidney length/width/height, spleen length) compared against ground truth (average of 3 clinical experts). Results showed strong concordance; all 95% CIs for Limits of Agreement met predefined acceptance criteria. Demographic range: age 22-85, BMI 17.97-48.40 kg/m².
Technological Characteristics
Cart-based ultrasound system; software-controlled. Transducers: eV14-2v (curved linear array, 6 MHz center frequency, 436 elements). Connectivity: Standard ultrasound console interface. Biocompatibility: ISO 10993-1. Safety standards: IEC 60601-1, IEC 60601-2-37, IEC 62304 (software lifecycle), IEC 62366-1 (usability). AI feature: Semi-automated measurement algorithm.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in adult and pediatric patients across various clinical applications including abdominal, cardiac, vascular, obstetric, gynecological, musculoskeletal, ophthalmic, urological, and small organ imaging. Contraindications: None stated.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Affiniti Series Diagnostic Ultrasound System (K243794)
Reference Devices
Epiq Series Diagnostic Ultrasound System (K231190)
Epiq Series Diagnostic Ultrasound System and Affiniti System Diagnostic Ultrasound System (K253595)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 29, 2026
Philips Ultrasound, LLC
Irma Sandoval-Watt
Senior Regulatory Manager
22100 Bothell Everett Hwy.
Bothell, Washington 98021
Re: K260667
Trade/Device Name: Alturion Series Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, QIH
Dated: May 29, 2026
Received: June 1, 2026
Dear Irma Sandoval-Watt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260667 - Irma Sandoval-Watt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260667 - Irma Sandoval-Watt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260667
Device Name
Alturion Series Diagnostic Ultrasound System
Indications for Use (Describe)
The Alturion Series Diagnostic Ultrasound System is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body in the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of Operation include B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where the Alturion Series Diagnostic Ultrasound System can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
K260667
PHILIPS
This summary of safety and effectiveness information is submitted in accordance with 21 CFR §807.92.
510(k) Number: K260667
**1. Submitter's name, address, telephone number, contact information**
Manufacturer: Philips Ultrasound LLC
22100 Bothell Everett Hwy
Bothell, WA 98021-8431
Contact Person: Irma Sandoval-Watt
(Primary) Senior Regulatory Manager
22100 Bothell Everett Hwy
Bothell, WA 98021-8431
Email: irma.sandoval-watt@philips.com
Phone: 1-303-453-3421
Secondary Contact: Benny Lam
(Secondary) Director of Regulatory, Software and AI
22100 Bothell Everett Hwy
Bothell, WA 98021-8431
Phone: 304-266-8208
Email: benny.lam@philips.com
Phone: 1-240-463-2613
Date Prepared: 27 Jun 2026
**2. Name of the device, including the trade of proprietary name if applicable, the common or usual name, and the classification name, if known:**
Proprietary Name: Alturion Series Diagnostic Ultrasound System
Common Name: Diagnostic Ultrasound System and Transducers
Regulatory Description:
| Classification Description | 21 CFR § | Product Code |
| --- | --- | --- |
| **Primary** | | |
| Ultrasonic pulsed doppler imaging system | 892.1550 | IYN |
| **Secondary** | | |
| Ultrasonic pulsed echo imaging system | 892.1560 | IYO |
| Diagnostic ultrasonic transducer | 892.1570 | ITX |
| Medical image management and processing system | 892.2050 | QIH |
Device Class: Class II
Review Panel: Radiology
Predicate Device: K243794, Affiniti Series Diagnostic Ultrasound System
Page 1 of 9
{5}
PHILIPS
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
Reference Device:
K231190, Epiq Series Diagnostic Ultrasound System
K253595, Epiq Series Diagnostic Ultrasound System and Affiniti System Diagnostic Ultrasound System
K231966, LOGIC E10 (GE Medical Systems Ultrasound and Primary Care Diagnostic, LLC),
### 3. Device Description
The Philips Alturion Series Diagnostic Ultrasound System is a general purpose, software controlled, diagnostic ultrasound system. Its functions are to acquire ultrasound data and to display the data in various modes of operation. The system consists of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics, and the transducer connection points. In addition to the physical knobs and buttons of the main control panel, the user interface consists of a touch screen with soft key controls. The transducers connect to the system console and facilitate ultrasound scanning of patients.
The Alturion Series Diagnostic Ultrasound System is considered a high-end refresh of the current Affiniti Series Diagnostic Ultrasound System and will be available in two configurations. At its core, the Alturion Series Diagnostic Ultrasound System is based on the Affiniti Series. Thus, enabling Alturion Series to use the same acoustic capabilities, and common software architecture as the Affiniti Series. The Alturion Series Diagnostic Ultrasound System is similar to other Philips ultrasound systems, most notably Affiniti Series Diagnostic Ultrasound System (K243794) in terms of indications for use, design, and fundamental technological characteristics.
The AI Auto Measure Abdomen feature in the Alturion Series aims to improve workflow efficiency by automating selected measurements required for routine abdominal exams. The Auto Measure Abdomen feature is designed to provide semi-automated and editable measures of abdominal organs such as kidney and spleen. The software feature offers to semi-automatically perform measurements for the user, who can always adjust caliper positions or make additional manual measurements. The Auto Measure Abdomen feature is available in C5-1 and C9-2 transducers only. This AI software feature is supported in all configurations of Alturion Series Diagnostic Ultrasound Systems running software version 14.0 or higher. The eV14-2v is a broadband, endocavitary Pure Wave transducer with elevation focus. The transducer eV14-2v includes a curved linear array with 6 MHz center frequency and 10.5 mm azimuth radius of curvature. The array has a total of 436 elements arranged in 3 rows of 218 elements. This transducer is supported in all configurations of Alturion Series Diagnostic Ultrasound System.
### 4. Indications For Use
The Alturion Series Diagnostic Ultrasound System is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body in the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of Operation include B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, Harmonic Imaging.
Page 2 of 9
[LOGO]
{6}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
**PHILIPS**
The clinical environments where the *Alturion Series Diagnostic Ultrasound System* can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
### 5. Technological Comparison to Predicate Devices
The purpose of the submission is to introduce a new system to the Philips ultrasound systems portfolio, *Alturion Series Diagnostic Ultrasound System*, an AI Auto Measure Abdomen software feature, and a new transducer, eV14-2v.
The AI Auto Measure Abdomen feature was most recently cleared in K253595. The AI Auto Measure feature is for cardiac was cleared in the predicate device (K243794). A performance study for validation was conducted for Auto Measure Abdomen that includes patients scanned by both C5-1 and C9-2.
All comparison points between the new transducer, eV14-2v and the predicate 3D9-3v transducer are either same or equivalent to predicate. See the tabular table below for more details.
Table 1: Comparison of subject device, *Alturion Series Diagnostic Ultrasound System* to the current marketed predicate to the addition of a new transducer, eV14-2v.
| Feature | Alturion Series Diagnostic Ultrasound System and eV14-2v Transducer | Affiniti Series Diagnostic Ultrasound System and 3D9-3V Transducer K243794 | Comparison |
| --- | --- | --- | --- |
| | Subject Device | Predicate Device | |
| Scientific Technology | Ultrasound imaging | Ultrasound imaging | Same as predicate |
| Users | Trained healthcare professionals, sonographers, physicians, and biomedical engineers who operate and maintain the product. | Trained healthcare professionals, sonographers, physicians, and biomedical engineers who operate and maintain the product. | Same as predicate |
| Use Environment | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Same as predicate |
| Transducer Model | Endocavitary Curved Linear | Endocavitary Curved Linear | Same as predicate |
| Indications for Use | Fetal/OB; Fetal Echo; Trans-vaginal; GYI; Urology | Fetal/OB; Fetal Echo; Trans-vaginal; GYI; Urology | Same as predicate |
| Acoustic Output | Ispta .3 ≤ 720 (mW/cm²) TI ≤ 6.0 MI ≤ 1.9 | Ispta .3 ≤ 720 (mW/cm²) TI ≤ 6.0 MI ≤ 1.9 | Same as predicate |
| Acoustic Working Frequency | Center Frequency 6.1 MHz | Center Frequency 5.7 MHz | Equivalent to predicate |
| Geometrical Configuration | Curved Linear Array | Curved Linear Array | Same as predicate |
| Total Number of Elements | 436 addressable elements connected to form dual aperture array with 218 elements per aperture | 128 | Different than predicate. The higher number of elements allows improved image quality by an increased field of view, better resolution, reduced clutter with elevation focus. |
Page 3 of 9
{7}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
**PHILIPS**
| Feature | Alturion Series Diagnostic Ultrasound System and eV14-2v Transducer | Affiniti Series Diagnostic Ultrasound System and 3D9-3V Transducer K243794 | Comparison |
| --- | --- | --- | --- |
| | Subject Device | Predicate Device | |
| Field of View (FOV) | 180 degrees | 155.5 degrees | Different from predicate. The higher FOV allows visualization of larger anatomical areas or pathologies in a single image. |
| Transducer Modes of Operation | B (2D); M-mode; PW Doppler; 3D/4D; Color; MPR | B, M PWD, Color, 3D/4D; Doppler; Combined | Equivalent to predicate |
| Tip Dimensions | Height: 25.3 mm Width: 25.3 mm Tip Diameter: 25.3 mm | Height: 26.3 mm Width: 26.3 mm Tip Diameter: 26.3 mm | Different from predicate. The differences are for improved patient comfort. |
| Connector | Cart-based system transducer connector | Cart-based system transducer connector | Same as predicate |
| Wight (including connector and cable) | < 0.9 kg | < 1.2 kg | Equivalent to predicate |
| Radiated Field Operating Controls | The transducer is controlled via the Ultrasound System's control panel. | The transducer is controlled via the Ultrasound System's control panel. | Same as predicate |
| Patient Contact Materials | Biocompatibility ISO 10993-1 | Biocompatibility ISO 10993-1 | Same as predicate |
| Transducer Maximum Surface Temperature (per IEC 60601-2-37, clause 201.11) | Still Air: 42.6°C Simulated Use 42.5°C | Still Air: 41.8°C Simulated Use: 42.5°C | Equivalent to predicate |
| Transducer Element Check Incorporated | Yes | Yes | Same as predicate |
| Transducer Provided Sterile | No | No | Same as predicate |
| Transducer Recommended use of covers | Yes | Yes | Same as predicate |
| Transducer Reusable | Yes | Yes | Same as predicate |
| Maximum Insertion Portion Width | 25.3 mm | 27.3mm | Equivalent to predicate The difference in width enhances patient comfort |
| Working Length (Intended Insertion Length) | 150 mm | 165 mm | Equivalent to predicate The difference in length enhances patient comfort |
| Transducer cleaning and disinfection instructions provided | Yes | Yes | Same as predicate |
| Professional Use | Yes | Yes | Same as predicate |
| Transducer Packaging | Transducer placed in a cardboard box with internal foam | Transducer placed in a cardboard box with internal foam | Same as predicate |
| 510(k) Track | Track 3 | Track 3 | Same as predicate |
Page 4 of 9
{8}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
PHILIPS
Table 1 – Comparison of subject device, Alturion Series Diagnostic Ultrasound System to the current marketed predicate to the addition of a new transducer, eV14-2v.
For testing, all pre-determined acceptance criteria were met as per the supporting evidence provided as part of the verification report with this 510(k)-submission including Biocompatibility, EMC, Safety testing, Acoustic output, Design Qualification Test (DQT), Patient contact surface verification. The results of these tests show that the subject device, Alturion Series Diagnostic Ultrasound System meets the intended use.
The subject device and predicate device:
- Are indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
- Have identical intended users and use environments.
- Have the same device classification and product codes.
- Have identical software architecture.
- Have identical hardware and imaging modes.
- Have equivalent acoustic output characteristics.
- Meet biocompatibility requirements per ISO 10993-1.
- Have equivalent transducer modes of operations.
- Have acoustic output levels within the Track 3 FDA limits and in accordance with IEC 60601-2-37.
6. Safety Considerations
The proposed Alturion Series Diagnostic Ultrasound System including AI Auto Measure Abdomen feature software application, and new transducer, eV14-2v, and compatible transducers are all Track 3 devices and comply with the reference standards and with the FDA ultrasound guidance document, Guidance for Industry and FDA Staff – Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, February 21, 2023.
7. Non-Clinical Performance Data
Philips Ultrasound performed the following testing to ensure the safety and effectiveness of the proposed subject device, Alturion Series Diagnostic Ultrasound System:
- IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical equipment – Part 1: General requirements for basic and essential performance
- IEC 60601-1-2: 2014+AMD1:2020 Medical Electrical Equipment – Part 1-2: General requirements for safety; electromagnetic compatibility – requirements and tests
- IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – collateral standard: usability
- IEC 60601-2-37:2024 Medical electrical equipment – Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- IEC 60601-2-18:2009 Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 62304:2006+AMD1:2015 Medical Device Software – Software life cycle process
- IEC 62359:2010+AMD1:2017 CSV Ultrasonics – Field Characterization – Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic field
- IEC 62366-1:2015+AMD1:2020 Medical devices – Application of usability engineering to medical devices
- ISO 10993-1:2018 Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process
- ISO 14971:2019 Medical devices – Application of risk management to medical devices
Non-Clinical verification testing has been performed addressing system level requirements according to system and design specifications, and risk control measures. Design Control activities to assure the safety
Page 5 of 9
{9}
PHILIPS
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
and effectiveness of the Alturion Series Diagnostic Ultrasound System include but are not limited to the following:
- Requirements Review
- Risk Analysis and Risk Management Review
- Product Specification Review
- Design Review
Non-clinical testing also included the Performance Validation Study for the proposed Auto Measure Abdomen software application. A data analysis study was conducted to evaluate the performance of the AI Auto Measure Abdomen feature, where ultrasound images previously collected from 150 subjects were used to obtain AI Auto Measure Abdomen algorithm-generated measurements which were then compared to manual measurements performed by 3 clinical experts, with the manual measurements used as ground truth for the study. The clinical experts also reviewed the AI algorithm-generated measurements to either accept or edit the measurements. Images that contributed to the study were obtained from subjects who represented a broad range of demographics, body habitus, and the range of measurements that were representative of the intended population.
# Auto Measure Abdomen AI Feature Summary
a. Summary of test statistics or other test results, including acceptance criteria and any additional information supporting the appropriateness of the characterized performance
The study evaluated AI algorithm performance for four abdominal measurements: kidney sagittal length, kidney transverse width, kidney transverse height, and spleen length. Table 2 presents the comparative clinical measurement results between AI Auto Measure Abdomen and the ground truth. Table 2 provides the Bland-Altman Limits of Agreement (LoA) analysis of the AI Auto Measure Abdomen in relation to the ground truth, with reference to established acceptance criteria.
Table 2. Clinical measurement summary for AI Auto Measure Abdomen and Ground Truth.
| Measurement | AI Auto Measure (cm) Mean ± SD (Min, Max) | Ground Truth (cm) Mean ± SD (Min, Max) |
| --- | --- | --- |
| Kidney Sagittal Length | 10.55 ± 1.20 (7.54, 14.80) | 10.50 ± 1.17 (7.20, 14.13) |
| Kidney Transverse Width | 5.19 ± 0.60 (3.25, 7.26) | 5.17 ± 0.71 (3.62, 7.30) |
| Kidney Transverse Height | 5.11 ± 0.80 (2.95, 7.71) | 5.10 ± 0.83 (2.84, 7.76) |
| Spleen Length | 10.77 ± 2.41 (6.08, 20.36) | 10.53 ± 2.39 (5.52, 19.59) |
Table 2 – Clinical Measurement Summary for Auto Measure Abdomen AI and Ground Truth.
Table 3. Primary endpoint: Bland-Altman limits of agreements (in percentage) for clinical measurements between AI Auto Measure Abdomen and Ground Truth.
Page 6 of 9
[LOGO]
{10}
PHILIPS
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
| Measurement | Mean Difference ± SD | 95% LoA | 95% CI of LoA | Acceptance Criteria |
| --- | --- | --- | --- | --- |
| Kidney Sagittal Length | 0.46% ± 3.51% | (-6.41%, 7.33%) | (-7.10%, 8.02%) | [-14.3%, 14.3%] |
| Kidney Transverse Width | 0.26% ± 8.79% | (-16.97%, 17.49%) | (-18.77%, 19.29%) | [-33.7%, 33.7%] |
| Kidney Transverse Height | 0.54% ± 6.36% | (-11.92%, 13.00%) | (-13.22%, 14.30%) | [-30.1%, 30.1%] |
| Spleen Length | 2.34% ± 4.90% | (-7.26%, 11.94%) | (-8.63%, 13.32%) | [-15.9%, 15.9%] |
Table 3 – Primary endpoint: Bland-Altman agreement for clinical measurements between Auto Measure Abdomen AI and Ground Truth.
The primary endpoint analysis indicated a strong concordance between measurements generated by the AI Auto Measure Abdomen algorithm and those obtained manually by clinical experts (serving as ground truth). The confidence intervals (CI) for the LoA satisfied the predefined acceptance criteria established for the study.
# b. The number of individual patients from whom the images were collected
A total of 150 subjects (i.e., 150 ultrasound exams) from whom images were collected for the performance validation study.
# c. The number of samples, if different from above, and the relationship between the two
A total of 292 images in kidney longitudinal view were used for kidney length measurement.
A total of 271 images in kidney transverse view were used for kidney width and height measurement.
A total of 145 images in spleen sagittal view were used for spleen length measurement.
# d. Demographic distribution including sex, age, and ethnicity
- Age (y): 58.59 ±14.70 (Mean ± SD), range (22, 85)
- Gender: Female 50%; Male 50%
- BMI (kg/ m²): 26.20 ± 4.74 (Mean ± SD), range (17.97, 48.40)
- Ethnicity:
- American Indian 0.67%
- Asian 31.33%
- Black or African American 3.33%
- Hispanic 0.67%
- White 64.00%
# e. Information about clinical subgroups and confounders present in the dataset
Validation images were obtained from subjects who represent the intended population with a broad range of demographics, body types and organ measurement values.
Page 7 of 9
[LOGO]
{11}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
PHILIPS
More specifically, patients from all BMI categories: Underweight (<18.5), Healthy Weight (18.5 to <25), Overweight (25 to <30), and Obesity (≥30) were included to ensure balanced body habitus representation.
Additionally, the study included patients referred for abdominal or renal exams and healthy volunteers, with datasets reflecting different clinical statuses to maintain clinical relevance:
Kidney Clinical Status
- Abnormal 22.00%
- Normal 78.00%
Spleen Clinical Status
- Abnormal 1.33%
- Normal 98.67%
f. Information about equipment and protocols used to collect images
Adults (≥18 years) were enrolled at three clinical sites, including patients referred for abdominal or renal ultrasound and healthy volunteers. Imaging was performed using Philips EPIQ and Affiniti ultrasound systems with C5-1 and C9-2 transducers as data deemed equivalent between Epiq, Affiniti and Alturion products given its shared codebase used for Auto Measure Abdomen. Each site followed a standardized study protocol for data collection.
g. Information about how reference standard was derived from the dataset (i.e., the “truthing” process)
Three clinical experts independently carried out manual measurements during the performance assessment. The average values obtained from their measurements served as the ground truth and were used to evaluate AI performance. All three experts are registered clinical sonographers with ten or more years of experience in general imaging and abdominal imaging, and each holds active certification by American Registry for Diagnostic Medical Sonography (ARDMS).
h. Description of how independence of test data from training data was ensured
The datasets used in the validation study and for regulatory clearance were distinct from those employed during algorithm training. Data independence between performance validation and model development was ensured by sourcing data from different clinical sites, different time periods and different ultrasound systems.
Since this is a software-only change and no new hardware was added, no acoustic output, cleaning and disinfectants, thermal, electrical, electromagnetic, and mechanical safety testing were required. Biocompatibility testing is not needed for the proposed Alturion Series Diagnostic Ultrasound System with Auto Measure Abdomen AI feature. The transducer patient contact materials and manufacturing processes are not impacted by the release of the proposed Alturion Series Diagnostic Ultrasound System with Auto Measure Abdomen.
8. Clinical Data
The proposed Alturion Series Diagnostic Ultrasound System did not require clinical data for determination of substantial equivalence, with substantial equivalence being demonstrated based on the following attributes:
- Design features
- Indications for use
Page 8 of 9
{12}
Traditional 510(k)
Alturion Series Diagnostic Ultrasound System
510(k) Summary
# PHILIPS
- Fundamental scientific technology
- Non-clinical performance testing
- Safety and effectiveness
9. Sterilization
The proposed *Alturion Series Diagnostic Ultrasound System* is not provided sterile; therefore, this is not applicable.
10. Conclusion
The subject device, *Alturion Series Diagnostic Ultrasound System*, has predicate devices which are legally marketed. Both subject device and predicate devices have the same intended use. The technological differences between the subject device and the predicate devices do not raise new or different questions of safety and effectiveness.
Both the subject device and the predicates were tested with the same types of non-clinical testing methods and follow the same set of standards.
Therefore, the subject device *Alturion Series Diagnostic Ultrasound System* is substantially equivalent to the predicate devices in terms of indications for use, design, technological characteristics, modes of operations, safety, and effectiveness.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.