EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System
Applicant
Philips Ultrasound, LLC
Product Code
IYN · Radiology
Decision Date
Feb 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K243794 · Feb 6, 2025
EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System
Philips Ultrasound, LLC
Hospital-based transthoracic echocardiography (TTE) clinical records
The sponsor used a retrospective dataset of clinical TTE clips and associated Standard of Care (SOC) measurements to validate the performance and agreement of the AutoMeasure V3 AI algorithm against human-derived clinical measurements.
AI/ML algorithm validation; Retrospective clinical data; Standard of Care measurements; Echocardiography
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
AutoMeasure V3 Performance Validation Study; Retrospective performance validation study; Follow-up/Duration: Data collected between 2015 and June 2024; Study Period: 2015-2024
Adult subjects (18+ years) undergoing clinical TTE, including healthy volunteers and patients suspected of disease.; Sample Size: 3,964 subjects (7,127 TTE clips); Number of Sites: Multiple institutions (USA, Asia, Europe)
Standard of Care (SOC) measurements obtained by qualified healthcare professionals during initial subject visits
Agreement of AutoMeasure V3 software measurements with Standard of Care (ground truth) measurements.
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Adult echocardiography 2D, Doppler, and M-mode measurements
AI-based machine learning algorithm
Agreement with Standard of Care measurements based on inter-observer data
High agreement with Standard of Care measurements; confidence intervals for limits of agreement met pre-defined acceptance criteria
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode . Power Doppler and Harmonic Imaging. The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo quidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure. The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use. Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode , Power Doppler and Harmonic Imaging. The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Device Story
EPIQ and Affiniti Series Diagnostic Ultrasound Systems utilize AutoMeasure V3 software to provide semi-automated 2D, Doppler, and M-mode measurements during adult echocardiography. Input data consists of transthoracic echocardiography (TTE) clips acquired with an adult cardiology transthoracic transducer and ECG signal. The device uses an AI-based machine learning algorithm to detect anatomical structures and initialize measurement calipers. Output is presented to the clinician as semi-automated, editable measurements. Used in clinics, hospitals, and point-of-care settings by sonographers, physicians, and biomedical engineers. The clinician reviews, edits, accepts, or rejects the automated measurements, facilitating clinical decision-making for cardiovascular disease diagnosis and treatment guidance.
Clinical Evidence
Performance validation study using 7,127 TTE cardiac clips from 3,964 subjects (mean age 59.0, 51.4% female). Study compared AutoMeasure V3 parameters against Standard of Care (SOC) ground truth measurements. Results demonstrated high agreement across demographic variables (age, BMI, gender, race) and clinical status, meeting pre-defined acceptance criteria based on inter-observer variability. No new clinical investigation was required for this submission.
Technological Characteristics
Diagnostic ultrasound system with pulsed Doppler, pulsed echo, and image management capabilities. Software-based AutoMeasure V3 feature uses AI/ML algorithms for semi-automated measurement initialization. Complies with IEC 62304 (software life cycle) and ISO 14971 (risk management). Operates on EPIQ/Affiniti systems running software version 13.0 or higher.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, including abdominal, cardiac (adult/pediatric/fetal), vascular, cephalic, fetal/obstetric, gynecological, intraoperative, musculoskeletal, ophthalmic, urological, small organ, and lung applications. Used in clinics, hospitals, and point-of-care settings by trained healthcare professionals.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips EPIQ Series Diagnostic Ultrasound System and Philips Affiniti Series Diagnostic Ultrasound System (K240850)
Submission Summary (Full Text)
{0}------------------------------------------------
February 6, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Ultrasound LLC Deval Patel Principal Regulatory Affairs Specialist 22100 Bothell Everett Hwy Bothell, Washington 98021
Re: K243794
Trade/Device Name: EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OBJ, OIH Dated: December 10, 2024 Received: December 10, 2024
Dear Deval Patel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
k243794
Device Name
EPIQ Series Diagnostic Ultrasound System Affiniti Series Diagnostic Ultrasound System
Indications for Use (Describe)
Eblo:
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode . Power Doppler and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo quidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Affiniti:
The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use.
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode , Power Doppler and Harmonic Imaging.
{4}------------------------------------------------
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92.
510(k) Number: K243794
Date Prepared: October 10, 2024
Date of Revision: February 5, 2025
| I. | Submitter | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Manufacturer Name and<br>Address | Philips Ultrasound LLC<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 USA | |
| Contact Information | Deval Patel<br>Principal Regulatory Affairs Specialist<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 USA<br>+1 (949) 502-1823 | |
| Secondary Contact | Erdit Gremi<br>Director, Regulatory Affairs<br>Philips Ultrasound LLC<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 USA<br>+1 (617) 798-8092 | |
II. Device
| Proprietary Name | EPIQ Series Diagnostic Ultrasound System<br>Affiniti Series Diagnostic Ultrasound System |
|------------------|------------------------------------------------------------------------------------------|
| Common Name | Diagnostic Ultrasound System and Transducers |
| Regulation Description | Classification Description | 21 CFR § | Product Code |
|------------------------|---------------------------------------------------|----------|--------------|
| | Primary | | |
| | System, imaging, pulsed doppler,<br>ultrasonic | 892.1550 | IYN |
| | Secondary | | |
| | System, imaging, pulsed echo,<br>ultrasonic | 892.1560 | IYO |
| | Transducer, ultrasonic, diagnostic | 892.1570 | ITX |
| | Medical image management and<br>processing system | 892.2050 | QIH |
| | Diagnostic Intravascular Catheter | 870.1200 | OBJ |
| Device Class | Class II | | |
| Review Panel | Radiology |
|--------------|-----------|
|--------------|-----------|
Predicate Device K240850; Philips EPIQ Series Diagnostic Ultrasound System and Philips Affiniti Series Diagnostic Ultrasound System
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in blue, block letters. The letters are bold and sans-serif. The background is a light gray or white color. The word is centered in the image.
| Predicate Regulation<br>Description | Classification Description | 21 CFR § | Product Code |
|-------------------------------------|---------------------------------------------------|------------------------------------------------|--------------|
| | Primary | System, imaging, pulsed doppler,<br>ultrasonic | 892.1550 |
| Secondary | System, imaging, pulsed echo,<br>ultrasonic | 892.1560 | IYO |
| | Transducer, ultrasonic, diagnostic | 892.1570 | ITX |
| | Medical image management and<br>processing system | 892.2050 | QIH |
| | Diagnostic Intravascular Catheter | 870.1200 | OBJ |
#### III. Device Description
The purpose of this Traditional 510(k) Pre-Market Notification is to introduce the change in AutoMeasure V3 software application onto the EPIQ Series Diagnostic Ultrasound Systems and Affiniti Series Diagnostic Ultrasound Systems.
Philips AutoMeasure feature provides the end user semi-automated 2D, Doppler and M-mode measurements with an adult cardiology transthoracic transducer and acquisitions that includes an electrocardiogram (ECG). The Auto Measure feature is designed to provide semi-automated and editable measures during an echocardiography. When Auto Measure is enabled, the healthcare professional performs an echocardiography with a workflow that provides the user with a semiautomated measurement that can be edited, accepted, or rejected. AutoMeasure is available in the Adult Echo analysis applications with ECG. Availability is also controlled by transducer and preset via the EPM.
The software applications are supported by all EPIQ and Affiniti models running software version 13.0 or higher.
#### IV. Intended Use and Indications for Use
### EPIQ Intended Use
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body.
### EPIQ Indications for Use:
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
{7}------------------------------------------------
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
## Affiniti Intended Use
The intended use of Affiniti Ultrasound Diagnostic ultrasound imaging and fluid flow analysis of the human body.
## Affiniti Indications for Use:
The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaqinal, Lunq.
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler and Harmonic Imaging.
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information.
Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Note: There are no changes to the EPIQ and Affiniti Ultrasound System Indications for Use due to the introduction of the AutoMeasure software applications
#### V. Comparison of Technological Characteristics with the Predicate
The purpose of the submission is to introduce the change to AutoMeasure V3 software applications to the EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound System. The subject device is substantially equivalent to the predicate device (K240850).
The following tables provide an overview of the comparison of similarities and differences between the proposed device and the predicates.
{8}------------------------------------------------
## DHILIPS
## Table 1: Comparison to Predicate for modification of AutoMeasure V3 onto EPIQ
| Feature | EPIQ Series Diagnostic Ultrasound System<br>Feature: AutoMeasure V3<br>Proposed Device | EPIQ Series Diagnostic Ultrasound System<br>K240850<br>Predicate Device | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral Vascular,<br>Cephalic (Adult), Cephalic (Neonatal),<br>Fetal/Obstetric, Gynecological, Intraoperative<br>(Vascular), Intraoperative (Cardiac), intra-<br>luminal, intra-cardiac echo, Musculoskeletal<br>(Conventional), Musculoskeletal (Superficial)<br>Ophthalmic, Other: Urology, Pediatric,<br>Peripheral Vessel, Small Organ (Breast,<br>Thyroid, Testicle), Transesophageal<br>(Cardiac), Transrectal, Transvaginal, Lung. | Abdominal, Cardiac Adult, Cardiac other (Fetal),<br>Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult),<br>Cephalic (Neonatal), Fetal/Obstetric, Gynecological,<br>Intraoperative (Vascular), Intraoperative (Cardiac),<br>intra-luminal, intra-cardiac echo, Musculoskeletal<br>(Conventional), Musculoskeletal (Superficial),<br>Ophthalmic, Other: Urology, Pediatric, Peripheral<br>Vessel, Small Organ (Breast, Thyroid, Testicle),<br>Transesophageal (Cardiac), Transrectal, Transvaginal,<br>Lung. | Identical to predicate |
| Intended Users | Trained healthcare professionals<br>Intended for sonographers, physicians, and<br>biomedical engineers who operate and<br>maintain your product.<br>Before use of the system and user<br>information, the user must be familiar with<br>ultrasound techniques. Sonography training<br>and clinical procedures are not included in<br>the User Manual or with the EPIQ Series<br>Diagnostic Ultrasound System. | Trained healthcare professionals<br>Intended for sonographers, physicians, and biomedical<br>engineers who operate and maintain your product.<br>Before use of the system and user information, the user<br>must be familiar with ultrasound techniques.<br>Sonography training and clinical procedures are not<br>included in the User Manual or with the EPIQ Series<br>Diagnostic Ultrasound System. | Identical to predicate |
| Intended User<br>Environment | Clinics, hospitals, and clinical point-of-care<br>for diagnosis of patients. | Clinics, hospitals, and clinical point-of-care for<br>diagnosis of patients. | Identical to predicate |
| USA FDA<br>Classification | Class II | Class II | Identical to predicate |
| Primary Product<br>Code | IYN | IYN | Identical to predicate |
| Primary<br>Regulation Name | System, Imaging, Pulsed Doppler, Ultrasonic | System, Imaging, Pulsed Doppler, Ultrasonic | Identical to predicate |
| Feature | EPIQ Series Diagnostic Ultrasound System<br>Feature: AutoMeasure V3<br>Proposed Device | EPIQ Series Diagnostic Ultrasound System<br>K240850<br>Predicate Device | Comparison |
| Primary<br>Regulation<br>Number | 21 CFR 892.1550 | 21 CFR 892.1550 | Identical to predicate |
| Secondary<br>Product Codes | ITX<br>IYO<br>OBJ<br>QIH | ITX<br>IYO<br>OBJ<br>QIH | Identical to predicate |
| Secondary<br>Regulation Name | Diagnostic ultrasonic transducer<br>Ultrasonic pulsed echo imaging<br>system<br>Diagnostic intravascular catheter<br>Automated Radiological Image<br>Processing Software | Diagnostic ultrasonic transducer<br>Ultrasonic pulsed echo imaging<br>system<br>Diagnostic intravascular catheter<br>Automated Radiological Image<br>Processing Software | Identical to predicate |
| Secondary<br>Regulation<br>Number | 21 CFR 892.1570<br>21 CFR 892.1560<br>21 CFR 870.1200<br>21 CFR 892.2050 | 21 CFR 892.1570<br>21 CFR 892.1560<br>21 CFR 870.1200<br>21 CFR 892.2050 | Identical to predicate |
| Reusable-<br>Systems and<br>Transducers | Yes | Yes | Identical to predicate |
| Duration of use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Identical to predicate |
| Application<br>Description | Auto Measure is an optional software feature<br>on the EPIQ Series Diagnostic Ultrasound<br>System that provides the end user with<br>semiautomated adult echocardiography 2D,<br>Doppler or M-mode measurements through<br>an Al-algorithm, training via machine-learning<br>techniques.<br>It is intended to be used with an Adult<br>Cardiology Transthoracic transducer and<br>acquisitions that include an ECG. These<br>measurements are routinely collected during<br>a transthoracic ECG.<br>2D modes include the following:<br>IVSd<br>LVIDd<br>LVPWd<br>LVIDs<br>Asc Ao Diam<br>LA Vol, A2Cs<br>LA Vol, A4Cs<br>LA Diameter a.p. (PLAX)<br>RA Volume | Auto Measure is an optional software feature on the<br>EPIQ Series Diagnostic Ultrasound System that<br>provides the end user with semiautomated adult<br>echocardiography 2D and Doppler measurements<br>through an Al-algorithm, training via machine-learning<br>techniques.<br>It is intended to be used with an Adult Cardiology<br>Transthoracic transducer and acquisitions that include<br>an ECG. These measurements are routinely collected<br>during a transthoracic ECG.<br>2D modes include the following:<br>IVSd<br>LVIDd<br>LVPWd<br>LVIDS<br>Asc Ao Diam<br>Doppler modes include the following:<br>MV Peak E Vel,<br>MV Peak A Vel MV Inflow (MV Dec Time, MV Peak E<br>Vel, MV Peak A Vel) LVOT VTI, LVOT Vmax AV VTI,<br>AV Vmax PV VTI, PV Vmax TR Vmax Lat E' Vel, Lat A' | Similar to the reference device features.<br>The functionality and workflow of the<br>AutoMeasure V3 software is Identical to<br>predicate where Auto Measure quantifies<br>image data acquired on the Philips EPIQ<br>Ultrasound System through an Al based<br>algorithm that used a trained machine-<br>learning model.<br>Compared to predicate only new<br>measurements and detectors are added to<br>the existing measurements. Clinical<br>performance studies were completed, and<br>the results showed safety and effectiveness |
| Feature | EPIQ Series Diagnostic Ultrasound System<br>Feature: AutoMeasure V3<br>Proposed Device | EPIQ Series Diagnostic Ultrasound System<br>K240850<br>Predicate Device | Comparison |
| | AoR Diam(2D) = Ao Annlus diam<br>Doppler modes include the following:<br>MV Peak E Vel,<br>MV Peak A Vel MV Inflow (MV Dec Time, MV<br>Peak E Vel, MV Peak A Vel) LVOT VTI,<br>LVOT Vmax AV VTI, AV Vmax PV VTI, PV<br>Vmax TR Vmax Lat E' Vel, Lat A' Vel Lat Vel<br>(Lat E' Vel, Lat A' Vel) Med E' Vel, Med A'<br>Vel Med Vel (Med E' Vel, Med A' Vel) RV S<br>Vel, MR VTI<br>TR VTI<br>M-Modes include the following: TAPSE | Vel Lat Vel (Lat E' Vel, Lat A' Vel) Med E' Vel, Med A'<br>Vel Med Vel (Med E' Vel, Med A' Vel) RV S Vel | |
| Semi-Automation<br>Technology | AutoMeasure V3 is created semi-<br>automatically using machine learning<br>algorithm without user interaction. Semi-<br>automated adult echocardiography 2D,<br>Doppler and M-mode measurements are<br>generated using an artificial intelligence (AI)<br>detection algorithm without user<br>interaction. After measurement is generated,<br>the user can edit (manually adjust the caliper<br>positions), accept, or reject the<br>measurements. | AutoMeasure V3 is created semi-automatically using<br>machine learning algorithm without user interaction.<br>Semi-automated adult echocardiography 2D and<br>Doppler measurements are generated using an artificial<br>intelligence (AI) detection algorithm without user<br>interaction. After measurement is generated, the user<br>can edit (manually adjust the caliper<br>positions), accept, or reject the measurements. | Subject device uses identical method for<br>contour generation as the reference device<br>K240850. The only difference is the<br>algorithm is new M-mode measurement is<br>added, new detectors are added, and<br>existing detectors are re-trained based on<br>new literature available. |
| User Interface | User selects an adult echocardiography 2D,<br>Doppler or M-mode measurement to perform<br>then the caliper positions are initialized based<br>on the output of the Al detection algorithm.<br>The user can edit, accept, or reject the<br>measurements. | User selects an adult echocardiography 2D or Doppler<br>measurement to perform then the caliper positions are<br>initialized based on the output of the Al detection<br>algorithm.<br>The user can edit, accept, or reject the<br>measurements. | Similar. The proposed AutoMeasure V3<br>software allows very similar semi-automate<br>measurements as the reference software<br>application AutoMeasure.<br>The proposed Auto Measure V3 software<br>adds additional M-mode measurement and<br>new detectors to the existing measurements<br>(2D and Doppler). |
| Feature | Affiniti Series Diagnostic Ultrasound System<br>Feature: AutoMeasure V3<br>Proposed Device | Affiniti Series Diagnostic Ultrasound System<br>K240850<br>Predicate Device | Comparison |
| Indications for<br>Use | Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral Vascular,<br>Cephalic (Adult), Cephalic (Neonatal),<br>Fetal/Obstetric, Gynecological, Intraoperative<br>(Vascular), Intraoperative (Cardiac), intra-<br>luminal, intra-cardiac echo, Musculoskeletal<br>(Conventional), Musculoskeletal (Superficial)<br>Ophthalmic, Other: Urology, Pediatric,<br>Peripheral Vessel, Small Organ (Breast,<br>Thyroid, Testicle), Transesophageal<br>(Cardiac), Transrectal, Transvaginal, Lung. | Abdominal, Cardiac Adult, Cardiac other (Fetal),<br>Cardiac Pediatric, Cerebral Vascular, Cephalic<br>(Adult), Cephalic (Neonatal), Fetal/Obstetric,<br>Gynecological, Intraoperative (Vascular),<br>Intraoperative (Cardiac), intra-luminal, intra-<br>cardiac echo, Musculoskeletal (Conventional),<br>Musculoskeletal (Superficial), Ophthalmic,<br>Other: Urology, Pediatric, Peripheral Vessel,<br>Small Organ (Breast, Thyroid, Testicle),<br>Transesophageal (Cardiac), Transrectal,<br>Transvaginal, Lung. | Identical to predicate |
| Intended Users | Trained healthcare professionals<br>Intended for sonographers, physicians, and<br>biomedical engineers who operate and<br>maintain your product.<br>Before use of the system and user<br>information, the user must be familiar with<br>ultrasound techniques. Sonography training<br>and clinical procedures are not included in<br>the User Manual or with the Affiniti Series<br>Diagnostic Ultrasound System. | Trained healthcare professionals<br>Intended for sonographers, physicians, and<br>biomedical engineers who operate and maintain<br>your product.<br>Before use of the system and user information,<br>the user must be familiar with ultrasound<br>techniques. Sonography training and clinical<br>procedures are not included in the User Manual<br>or with the Affiniti Series Diagnostic Ultrasound<br>System. | Identical to predicate |
| Intended User<br>Environment | Clinics, hospitals, and clinical point-of-care for<br>diagnosis of patients. | Clinics, hospitals, and clinical point-of-care for<br>diagnosis of patients. | Identical to predicate |
| USA FDA<br>Classification | Class II | Class II | Identical to predicate |
| Primary Product<br>Code | IYN | IYN | Identical to predicate |
| Primary<br>Requlation Name | System, Imaging, Pulsed Doppler, Ultrasonic | System, Imaging, Pulsed Doppler, Ultrasonic | Identical to predicate |
| Feature | Affiniti Series Diagnostic Ultrasound System<br>Feature: AutoMeasure V3<br>Proposed Device | Affiniti Series Diagnostic Ultrasound System<br>K240850<br>Predicate Device | Comparison |
| Primary<br>Regulation<br>Number | 21 CFR 892.1550 | 21 CFR 892.1550 | Identical to predicate |
| Secondary<br>Product Codes | ITX<br>IYO<br>OBJ<br>QIH | ITX<br>IYO<br>OBJ<br>QIH | Identical to predicate…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.