K260646 · Smith & Nephew Medical, Ltd. · OMP · May 27, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260646
Device Name
RENASYS EDGE (66803126)
Applicant
Smith & Nephew Medical, Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
May 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RENASYSTM EDGE Pump is indicated for patients who would benefit from a suction pump (NPWT), as it allows for wound management via removal of fluids, including irrigation fluids and body fluids, wound exudate and infectious materials. Appropriate wound types include: - Chronic - Acute - Traumatic - Sub-acute and dehisced wounds - Ulcers (such as pressure or diabetic) - Partial-thickness burns - Flaps - Grafts The RENASYSTM EDGE Pump is compatible with the following: - RENASYSTM Foam Dressing Kits - RENASYSTM Foam Dressing Kit - XL - RENASYSTM Gauze Dressing Kits - RENASYSTM Abdominal Dressing Kit - RENASYSTM Y Connector Kit - RENASYSTM White Foam Wound Dressing - RENASYSTM WOUND+ Dressing Kit with AIRLOCK Technology When used with the RENASYSTM WOUND+ Dressing Kit without filler, the RENASYSTM EDGE pump is indicated for the following additional wound type: - Closed surgical incisions Important information when using the RENASYSTM AB Abdominal Kit with Soft Port. When used with the RENASYSTM AB Abdominal Kit with Soft Port, the RENASYSTM EDGE Pump is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYSTM AB Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Device Story
Software-controlled, electric motor-driven diaphragm vacuum pump; provides negative pressure wound therapy (25-200mmHg) to one or two wounds; removes exudate/fluids into disposable canisters. Operated via mains power or internal battery; features tactile buttons/soft keys; includes Bluetooth connectivity. Used in clinical settings; managed by healthcare providers. System includes various compatible dressing kits (foam, gauze, abdominal, AIRLOCK technology). Auto-restart function re-initiates therapy after >1 hour pause. Alarms notify users of therapy loss. Benefits include effective wound management via fluid removal and maintenance of closed wound environment.
Clinical Evidence
No clinical data was necessary to support the demonstration of substantial equivalence. Evidence relies on bench testing and simulated wound model performance verification.
Technological Characteristics
Software-controlled, electric motor-driven diaphragm vacuum pump. Pressure range: 25-200mmHg. Power: Mains or internal Lithium-Ion battery. Connectivity: Bluetooth. User interface: Tactile buttons with soft keys. Reusable pump with disposable canisters (300ml/800ml). Non-sterile. Includes auto-restart functionality and alarm system for therapy loss.
Indications for Use
Indicated for patients requiring negative pressure wound therapy (NPWT) for chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, flaps, and grafts. Additional indication for closed surgical incisions when used with RENASYSTM WOUND+ Dressing Kit without filler. Indicated for temporary bridging of abdominal wall openings with exposed viscera (e.g., abdominal compartment syndrome) when used with RENASYSTM AB Abdominal Kit.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 27, 2026
Smith & Nephew Medical, Ltd.
Chelsea Baker
Regulatory Affairs Specialist
Contact Address
Re: K260646
Trade/Device Name: RENASYSTM EDGE
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered Suction Pump
Regulatory Class: Class II
Product Code: OMP
Dated: February 27, 2026
Received: February 27, 2026
Dear Chelsea Baker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K260646 - Chelsea Baker
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260646 - Chelsea Baker
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yu-chieh Chiu -S
Yu-Chieh Chiu, Ph.D.
Assistant Director
DHT4B: Division of Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260646
Device Name
RENASYSTM EDGE
Indications for Use (Describe)
The RENASYSTM EDGE Pump is indicated for patients who would benefit from a suction pump (NPWT), as it allows for wound management via removal of fluids, including irrigation fluids and body fluids, wound exudate and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Sub-acute and dehisced wounds
- Ulcers (such as pressure or diabetic)
- Partial-thickness burns
- Flaps
- Grafts
The RENASYSTM EDGE Pump is compatible with the following:
- RENASYSTM Foam Dressing Kits
- RENASYSTM Foam Dressing Kit - XL
- RENASYSTM Gauze Dressing Kits
- RENASYSTM Abdominal Dressing Kit
- RENASYSTM Y Connector Kit
- RENASYSTM White Foam Wound Dressing
- RENASYSTM WOUND+ Dressing Kit with AIRLOCK Technology
When used with the RENASYSTM WOUND+ Dressing Kit without filler, the RENASYSTM EDGE pump is indicated for the following additional wound type:
- Closed surgical incisions
Important information when using the RENASYSTM AB Abdominal Kit with Soft Port.
When used with the RENASYSTM AB Abdominal Kit with Soft Port, the RENASYSTM EDGE Pump is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYSTM AB Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
{5}
K260646
# 510(k) Summary
| 21 CFR 807.92 (a)(1): Submitter’s Information | |
| --- | --- |
| Applicant | Smith & Nephew Medical Ltd |
| Address | 101 Hessle Road, Hull, HU3 2BN, United Kingdom |
| Establishment Registration Number | 8043484 |
| Telephone Number | +44 (0)7793553737 |
| Contact Person | Chelsea Baker |
| Date Prepared | 30th April 2026 |
| 21 CFR 807.92 (a)(2): Device Information | |
| Device Name | RENASYS™ EDGE |
| Common Name | Negative Pressure Wound Therapy System |
| Classification Name | General and Plastic Surgery |
| Regulation Number | 21 CFR 878.4780 |
| Regulatory Class | Class II |
| Product Code | OMP |
| 21 CFR 807.92 (a)(3): Legally marketed device to which equivalence is claimed | |
| 510(k) Number | K231939 |
| Device Name | RENASYS™ EDGE |
| 21 CFR 807.92 (a)(4): Device Description | |
| The RENASYS™ EDGE device is a software-controlled suction device consisting of an electric motor driven, diaphragm, vacuum pump. The device is designed to provide Negative Pressure Wound Therapy at a range of pressure settings between 25-200mmHg to a closed environment over one or two wounds. The device removes exudate from the wound site(s) to a disposable container, which allows for wound management via removal of fluids, including irrigation of body fluids, wound exudates and infectious materials. The RENASYS™ EDGE Negative Pressure Wound Therapy Pump can be operated by either a mains power supply or internal battery. The RENASYS™ EDGE device is compatible with Smith & Nephew RENASYS™ Dressing Kits. | |
Page 1 of 5
{6}
K260646
| 21 CFR 807.92 (a)(5): Intended Use/Indications for use |
| --- |
| The RENASYS™ EDGE Pump is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it allows for wound management via removal of fluids, including irrigation fluids and body fluids, wound exudate and infectious materials. |
| Appropriate wound types include: • Chronic • Acute • Traumatic • Sub-acute and dehisced wounds • Ulcers (such as pressure or diabetic) • Partial-thickness burns • Flaps • Grafts |
| The RENASYS™ EDGE Pump is compatible with the following: • RENASYS™ Foam Dressing Kits • RENASYS™ Foam Dressing Kit – XL • RENASYS™ Gauze Dressing Kits • RENASYS™ Abdominal Dressing Kit • RENASYS™ Y Connector Kit • RENASYS™ White Foam Wound Dressing • RENASYS™ WOUND+ Dressing Kit with AIRLOCK™ Technology |
| When used with the RENASYS™ WOUND+ Dressing Kit without filler, the RENASYS™ EDGE pump is indicated for the following additional wound type: • Closed surgical incisions When used with the RENASYS™ AB Abdominal Kit with Soft Port, the RENASYS™ EDGE pump is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS™ AB Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre |
| 21 CFR 807.92 (a)(6): Comparison of Technological Characteristics between the Subject and Predicate Devices |
| The RENASYS™ EDGE device was previously cleared under K231939, whereby the application was to add the RENASYS™ White Foam dressing kits as compatible devices. No significant changes have been made to the subject device since this clearance; a summary of the similarities and differences are included below: |
Page 2 of 5
{7}
K260646
| Characteristics | Subject Device | Predicate Device (K231939) | Comparison |
| --- | --- | --- | --- |
| Trade Name | RENASYS™ EDGE | RENASYS™ EDGE | |
| Intended Use | Intended to create a closed environment over a wound to allow negative pressure wound therapy to evacuate from the wound into a collection canister. | Intended to create a closed environment over a wound to allow negative pressure wound therapy to evacuate from the wound into a collection canister. | Same as predicate |
| Indications for Use | Appropriate wound types include: • Chronic • Acute • Traumatic • Sub-acute and dehisced wounds • Ulcers (such as pressure or diabetic) • Partial-thickness burns • Flaps • Grafts When used with the RENASYS™ WOUND+ Dressing Kit without filler; the RENASYS™ EDGE pump is indicated for the following additional wound type: • Closed surgical incisions | Appropriate wound types include: • Chronic • Acute • Traumatic • Sub-acute and dehisced wounds • Ulcers (such as pressure or diabetic) • Partial-thickness burns • Flaps • Grafts | Same as predicate with the addition of RENASYS™ WOUND+ Dressing Kit without filler indication – Indication cleared in K251826 |
| Compatible Devices | • RENASYS™ Foam Dressing Kits K142979 | • RENASYS™ Foam Dressing Kits K142979 | Same as predicate with the addition of RENASYS™ WOUND+ Dressing Kit |
Page 3 of 5
{8}
K260646
| | • RENASYS™ Foam Dressing Kit – XL K202783 • RENASYS™ Gauze Dressing Kits K142979 • RENASYS™ Abdominal Dressing Kit K14133 • RENASYS™ Y Connector Kit K181822 • RENASYS™ White Foam Wound Dressing K213853 • RENASYS™ WOUND+ Dressing Kit with AIRLOCK™ Technology K243576 | • RENASYS™ Foam Dressing Kit – XL K202783 • RENASYS™ Gauze Dressing Kits K142979 • RENASYS™ Abdominal Dressing Kit K14133 • RENASYS™ Y Connector Kit K181822 • RENASYS™ White Foam Wound Dressing K213853 | with AIRLOCK™ Technology cleared under K243576 |
| --- | --- | --- | --- |
| Mode of Action | Negative Pressure Wound Therapy | Negative Pressure Wound Therapy | Same as predicate |
| Device Operation | Software | Software | Same as predicate |
| User Interface | Utilizes tactile buttons with soft key function | Utilizes tactile buttons with soft key function | Same as predicate |
| Single-use or Reusable | Reusable pump unit, Disposable canister | Reusable pump unit, Disposable canister | Same as predicate |
| Method of Sterilization | Non-sterile | Non-sterile | Same as predicate |
| O-Ring | Incorporates O-ring on the canisters allowing for a new O-ring with each therapy | Incorporates O-ring on the canisters allowing for a new O-ring with each therapy | Same as predicate |
| Battery type | Lithium Ion | Lithium Ion | Same as predicate |
| Negative Pressure | 25-200mmHg | 25-200mmHg | Same as predicate |
| Therapy Mode | Continuous & Intermittent | Continuous & Intermittent | Same as predicate |
| Alarms | Device utilizes alarms to notify the user of any loss of therapy | Device utilizes alarms to notify the user of any loss of therapy | Same as predicate |
| Auto Restart | Incorporates an auto-restart function to re-initiate therapy if the device has been left on pause for more than 1 hour | Incorporates an auto-restart function to re-initiate therapy if the device has been left on pause for more than 1 hour | Same as predicate |
| Wireless Technology | Incorporates wireless technology to allow for Bluetooth connectivity | Incorporates wireless technology to allow for Bluetooth connectivity | Same as predicate |
Page 4 of 5
{9}
K260646
| Configurations | 300ml & 800ml canisters available | 300ml & 800ml canisters available | Same as predicate |
| --- | --- | --- | --- |
| **Accessories** | | | |
| Carry Bag | Carry bag for use with device and 300ml Canister | Carry bag for use with device and 300ml Canister | Same as predicate |
| Carry Strap | Carry strap fixes to carry handle | Carry strap fixes to carry handle | Same as predicate |
| **21 CFR 807.92 (b)(1): Brief discussion of nonclinical tests submitted/referenced/relied on in the submission to determine substantial equivalence** | | | |
| Verification activities were conducted which demonstrate the overall system performance of the RENASYS™ EDGE Negative Pressure Wound Therapy System. Performance testing for the use of RENASYS™ WOUND + is covered within the K243576 and K251826 submissions. Verification activities were conducted which demonstrate the overall system performance of the RENASYS™ EDGE device is not impacted by the cumulative updates to the RENASYS™ EDGE NPWT System. The principal test methods used to demonstrate performance were simulated wound model tests to ensure satisfactory performance and substantial equivalence to the predicate device. | | | |
| **21 CFR 807.92 (b)(2): Brief discussion of clinical tests submitted/references/relied on in this submission to determine substantial equivalence** | | | |
| No Clinical performance data was necessary to support the demonstration of substantial equivalence to the predicate device. | | | |
| **21CFR 807.92 (b) (3): Conclusions drawn from the nonclinical tests** | | | |
| Based on the non-clinical performance testing provided in this submission, the subject device is substantially equivalent to the legally marketed predicate device (K231939) and there are no new questions of safety or effectiveness. | | | |
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.