Renasys-F XL Foam Dressing Kit With Soft Port, Renasys XL Transparent Film Dressing
K202783 · Smith & Nephew Medical Limited · OMP · Jan 14, 2022 · General, Plastic Surgery
Device Facts
Record ID
K202783
Device Name
Renasys-F XL Foam Dressing Kit With Soft Port, Renasys XL Transparent Film Dressing
Applicant
Smith & Nephew Medical Limited
Product Code
OMP · General, Plastic Surgery
Decision Date
Jan 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RENASYS-F Foam dressing kits with Soft Port and RENASYS Films are intended to be used in conjunction with Smith & Nephew RENASYS Negative Pressure Wound Therapy (NPWT) Systems. The Smith & Nephew RENASYS system is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include: Chronic, Acute, Traumatic, Sub-Acute and dehisced wounds, Ulcers (such as pressure or diabetic), Partial-thickness burns, Flaps, Grafts.
Device Story
Device consists of sterile, single-use polyurethane foam wound filler sheets and transparent film dressings; used with Smith & Nephew RENASYS NPWT pumps (e.g., RENASYS TOUCH, RENASYS GO). Foam filler placed in wound bed; secured with transparent film; connected to suction pump via soft port. System applies negative pressure to wound site to remove exudate, infectious materials, and fluids; promotes healing. Used in clinical settings by healthcare providers. Output is physical wound fluid removal; enables clinicians to manage complex wounds; benefits patients by facilitating healing environment.
Clinical Evidence
Bench testing only. Simulated wound model tests confirmed system performance (fluid removal, leak/blockage alarm functionality) with RENASYS GO and TOUCH pumps. Physical testing verified material properties (tensile strength, moisture vapor permeability, adhesion). Biocompatibility testing conducted per ISO 10993-1 (cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, implantation) and endotoxin testing per ANSI/AAMI ST72:2019.
Technological Characteristics
Polyurethane foam filler; transparent film dressing. Single-use, sterile. Connects to external NPWT pump via soft port. Materials identical to predicate. Testing per ISO 10993-1 and ANSI/AAMI ST72:2019. No electronic components; purely mechanical dressing system.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing via fluid removal. Applicable to chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, flaps, and grafts.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
RENASYS-F Negative Pressure Wound Therapy Foam Dressing Kits with Soft Port (K142979)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
January 14, 2022
Smith & Nephew Medical Limited Steeve Lamvohee RA Director 101 Hessle Road Hull, Yorkshire HU3 2BN United Kingdom
Re: K202783
Trade/Device Name: Renasys-F XL Foam Dressing Kit With Soft Port, Renasys XL Transparent Film Dressing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 23, 2021 Received: September 27, 2021
Dear Steeve Lamvohee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-device-advice-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K202783
### Device Name
RENASYS-F XL Negative Pressure Wound Therapy Foam Dressing Kit with Soft Port RENASYS XL Transparent Film Dressing
### Indications for Use (Describe)
RENASYS-F Foam dressing kits with Soft Port and RENASYS Films are intended to be used in conjunction with Smith & Nephew RENASYS Negative Pressure Wound Therapy (NPWT) Systems.
The Smith & Nephew RENASYS system is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- Sub-Acute and dehisced wounds
- · Ulcers (such as pressure or diabetic)
- · Partial-thickness burns
- Flaps
- · Grafts
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> | <input type="checkbox"/> |
|| | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
RENASYS™-F XL Negative Pressure Wound Therapy Foam Dressing Kit with Soft Port RENASYS™ XL Transparent Film Dressing Traditional 510(k) Premarket Notification : K202783
## 510(k) Summary K202783
| 21 CFR 807.92 (a)(1): Submitter's Information | |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner<br>Name | Smith & Nephew Medical Ltd |
| Address | 101 Hessle Road, Hull, HU3 2BN, United Kingdom |
| Establishment<br>Registration<br>Number | 8043484 |
| Contact Name | Dr Steeve Lamvohee, Regulatory Affairs Director |
| Telephone Number | +44 7583 048727 |
| Date Prepared | 13th January 2022 |
| 21 CFR 807.92 (a)(2): Device Information | |
| Device Name<br>(Trade/Proprietary<br>Name) | RENASYS-F XL Negative Pressure Wound Therapy Foam<br>Dressing Kit with Soft Port<br>RENASYS XL Transparent Film Dressing |
| Common Name | Negative Pressure Wound Therapy foam kit<br>Negative Pressure Wound Therapy films |
| Review Panel | General and Plastic Surgery |
| Regulation<br>Number | 21 CFR 878.4780 |
| Regulatory Class | Class II |
| Product Code | OMP |
| 21 CFR 807.92<br>(a)(3): Legally<br>marketed device to<br>which equivalence<br>is claimed | 510(k) Number: K142979<br>Device Name: RENASYS-F Negative Pressure Wound Therapy<br>Foam Dressing Kits with Soft Port |
| 21 CFR 807.92 (a)(4): Device Description | |
The RENASYS-F XL Negative Pressure Wound Therapy (NPWT) Foam Dressing Kit with Soft Port is a sterile, single use device containing one extra-large polyurethane foam wound filler sheet (18.9in x 16.1in x 0.6in / 48cm x 41cm x 1.5cm), one soft port and six transparent films (12in x 8in / 30cm x 20cm).
The RENASYS XL Transparent Film Dressing is a sterile, single use device provided in a pack containing five transparent films (15in x 24in / 38cm x 60cm). The RENASYS XL Transparent Film Dressing are provided where additional film dressings may be needed to supplement other RENASYS dressing kits.
The RENASYS-F XL Negative Pressure Wound Therapy Foam Dressing Kit with Soft Port and RENASYS XL Transparent Film Dressing are intended to be used in conjunction with
{4}------------------------------------------------
ENASYS™-F XL Negative Pressure Wound Therapy Foam Dressing Kit with Soft Port RENASYS™ XL Transparent Film Dressing Traditional 510(k) Premarket Notification : K202783
Smith & Nephew RENASYS Negative Pressure Wound Therapy (NPWT) systems and dressing kits.
These devices are compatible with the RENASYS TOUCH NPWT pump (cleared in K181822) and RENASYS GO NPWT pump (cleared in K152163).
The devices for which clearance is being sought are:
RENASYS-F XL Negative Pressure Wound Therapy Foam Dressing Kit with Soft Port Containing cleared components (soft-port and transparent film) and a larger sized foam sheet consisting of identical material to the foam sheets in the predicate device RENASYS XL Transparent Film Dressing
a larger sized transparent film of identical material to the film sheets in the predicate device
# 21 CFR 807.92 (a)(5): Intended Use / Indications for Use
RENASYS-F Foam dressing kits with Soft Port and RENASYS Films are intended to be used in conjunction with Smith & Nephew RENASYS Negative Pressure Wound Therapy (NPWT) Systems.
The Smith & Nephew RENASYS system is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include:
- . Chronic
- Acute .
- Traumatic
- . Sub-Acute and dehisced wounds
- Ulcers (such as pressure or diabetic) ●
- . Partial-thickness burns
- Flaps
- . Grafts
# 21 CFR 807.92 (a)(6): Comparison of Technological Characteristics between the Subject and Predicate Devices
The subject device and predicate device have identical indications for use, similar technological characteristics and the same principles of operation. The minor technological difference in the dimensions of the foam and transparent film does not raise any new or different questions of safety and effectiveness compared to the predicate device. The changes made to the Soft Port from the predicate device specifically relate to the release handles and the controlled leak pathway, and do not impact the safety or effectiveness.
The subject device and the predicate device are based on the following same technological elements:
- · The devices are single patients use, supplied sterile and disposable
- · The devices have the similar design and utilise the same materials of construction
{5}------------------------------------------------
NASYS™-F XL Neqative Pressure Wound Therapy Foam Dressing Kit with Soft Port RENASYS™ XL Transparent Film Dressing Traditional 510(k) Premarket Notification : K202783
- · The devices are connected to a negative pressure pump via the RENASYS Soft Port
# 21 CFR 807.92 (b)(1): Brief discussion of nonclinical tests submitted/referenced/ relied on in this submission to determine substantial equivalence
Verification activities were conducted which demonstrate the overall system performance of the RENASYS-F XL Foam Kit and RENASYS XL Film. The principal test methods used to demonstrate performance were simulated wound model tests.
- Wound model tests demonstrating performance of RENASYS-FXL Foam Kit and RENASYS XL Film when used with RENASYS GO and RENASYS TOUCH NPWT pumps at various pressures, minimum/maximum leak rates and use in combination with cleared RENASYS consumables
- · Wound model testing demonstrating no change in performance between large and extra-large wound sizes
- · Wound model tests demonstrating RENASYS GO and RENASYS TOUCH NPWT leak and blockage alarms are not impacted when used with RENASYS-F XL Foam Kit and RENASYS XL Film
- · Physical testing on the XL Foam (tensile strength, elongation at break, loss on drying)
- · Physical testing on the XL Transparent Film Dressing (Moisture Vapor permeability, Waterproofness, Adhesion to Steel, Tensile strength, Weight of Adhesive (g/m2), Weight of Base Film (g/m2), Extensibility, Permanent Set)
RENASYS-F XL Foam Kit and RENASYS XL Film meets the biocompatibility requirements of:
- · FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995.
- · International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The biological endpoints for the subject devices are in line with the requirements of ISO 10993 with consideration to: Cytotoxicity, Sensitization, Irritation, Material mediated pyrogenicity, Systemic toxicity (acute, sub-acute and sub-chronic) and Implantation
- · Evaluation and risk assessment of bacterial endotoxin contamination in accordance with ANSI/AAMI ST72:2019
# 21 CFR 807.92 (b)(3): Conclusions drawn
Based on the test results and the additional supporting information provided in this submission, S+N believe the subject device is substantially equivalent to the legally marketed predicate device. To the extent that where there are differences between the subject device and the predicate device, these differences do not raise new or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.