K260576 · MeVis Medical Solutions AG · LLZ · Jun 30, 2026 · Radiology
Device Facts
Record ID
K260576
Device Name
MeVis AVM Contour
Applicant
MeVis Medical Solutions AG
Product Code
LLZ · Radiology
Decision Date
Jun 30, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260576 · Jun 30, 2026
MeVis AVM Contour
MeVis Medical Solutions AG
Retrospective clinical routine data
Retrospective clinical routine data was used to validate the accuracy of the semi-automatic image registration algorithm, demonstrating that the device performance is consistent with clinical stereotactic radiosurgery routine conditions.
MeVis AVM Contour is intended for use by qualified medical professionals working in the field of neurosurgery or radiotherapy planning to manually create target contours for regions of interest in the cranial region through semi-automatic registration of 2D digital subtraction angiography (DSA) images onto 3D computed tomography (CT) scans. The contours can be exported as DICOM RTStruct objects and are used in radiosurgery/-therapy treatment planning procedures to target specific brain regions. The created contours are intended for further review, editing and approval by radiosurgery/-therapy planning specialists in third-party radiosurgery/-therapy treatment planning systems (not part of this submission). The semi-automatic registration is restricted to/requires: • X-ray, 2D DSA and 3D CT images of the brain • X-ray images must be acquired in both frontal and sagittal direction
Device Story
Software tool for neurosurgery/radiotherapy planning; processes DICOM CT, biplanar X-ray, and DSA images. Inputs: 2D DSA and 3D CT brain images (frontal/sagittal X-ray required). Operation: semi-automatic registration of 2D DSA to 3D CT; manual contouring tools; ROI projection. Output: 3D CT-based structures in DICOM RTStruct format. Used in clinical/hospital environments by neurosurgeons and radiation oncologists. Output exported to third-party treatment planning systems for review/approval. Benefits: supports visualization and contouring for AVM treatment planning.
Clinical Evidence
No clinical data. Bench testing only. Accuracy of semi-automatic registration algorithm validated using phantom and retrospective clinical data. Achieved 3D-CT to 2D-DSA alignment accuracy of 0.44 mm (± 0.70 mm SD), compared to 0.50 mm predicate benchmark.
Technological Characteristics
Medical imaging software; runs on standard PCs. Features semi-automatic 2D/3D image registration and manual segmentation. Connectivity: DICOM RTStruct export. Software developed per IEC 62304 and ISO 14971. Cybersecurity controls include vulnerability scanning and penetration testing.
Indications for Use
Indicated for adults 18 years and older diagnosed with Arteriovenous Malformations (AVM) requiring stereotactic radiosurgery (SRS).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio (K212420)
Reference Devices
Brainlab Elements Image Fusion Angio 1.0 (K190042)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 30, 2026
MeVis Medical Solutions AG
Anne Schmidt
Regulatory Affairs Manager
Caroline-Herschel-Str. 1
Bremen, 28359
Germany
Re: K260576
Trade/Device Name: MeVis AVM Contour
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: May 28, 2026
Received: May 28, 2026
Dear Anne Schmidt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260576 - Anne Schmidt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260576 - Anne Schmidt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| # Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260576 | ? |
| Please provide the device trade name(s). | | ? |
| MeVis AVM Contour | | |
| Please provide your Indications for Use below. | | ? |
| MeVis AVM Contour is intended for use by qualified medical professionals working in the field of neurosurgery or radiotherapy planning to manually create target contours for regions of interest in the cranial region through semi-automatic registration of 2D digital subtraction angiography (DSA) images onto 3D computed tomography (CT) scans. The contours can be exported as DICOM RTStruct objects and are used in radiosurgery/-therapy treatment planning procedures to target specific brain regions. The created contours are intended for further review, editing and approval by radiosurgery/-therapy planning specialists in third-party radiosurgery/-therapy treatment planning systems (not part of this submission). The semi-automatic registration is restricted to/requires: • X-ray, 2D DSA and 3D CT images of the brain • X-ray images must be acquired in both frontal and sagittal direction | | |
| Intended Patient Population MeVis AVM Contour is used for patients diagnosed with Arteriovenous Malformations (AVM) requiring stereotactic radiosurgery (SRS). These patients are adults, 18 years and older. | | |
| Intended User Profile The intended users are qualified medical professionals, working in the field of neurosurgery or radiotherapy planning (Neurosurgeons and Radiation Oncologists). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{4}
# MeVis
## 510(k) Summary
K260576
### 1 Submitter
**Submitted Name:** MeVis Medical Solutions AG
Caroline-Herschel-Straße 1
28359 Bremen
Germany**Establishment Name:** MeVis Medical Solutions AG**Establishment
Registration Number:** 3010601176**Date Prepared:***Contact Person:** Anne Schmidt
Manager Regulatory Affairs**Telephone:** +49 (0)421 22495-120**Fax:** +49 (0)421 22495-999
### 2 Device
**Device Trade Name:** MeVis AVM Contour**Regulation:** CFR 21 892.2050**Classification Name:** Medical image management and processing system**Product Code:** LLZ -system, image processing, radiological**Class:** Class 2
---MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 1
{5}
MeVis
# 3 Predicate Device
| 510(k) Number | Predicate Device | Product Code |
| --- | --- | --- |
| K212420 | Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio | LLZ, JAK |
# 4 Reference Device¹
| 510(k) Number | Reference Device | Product Code |
| --- | --- | --- |
| K190042 | Brainlab Elements Image Fusion Angio 1.0 | LLZ JAK, LHN |
# 5 Device Description
The MeVis AVM Contour device is a specialized medical imaging software tool for DICOM CT, biplanar X-Ray and DSA images designed for qualified medical professionals working in the field of neurosurgery or radiotherapy planning.
The device integrates manual contouring tools, semi-automatic image registration, and ROI projection functions to support visualization, contouring, and export of 3D CT-based structures in DICOM RT Structure Set (RTStruct) format for subsequent treatment planning of arteriovenous malformation (AVM) diseases. The created contours are intended for further review, modification, and approval by radiosurgery or radiotherapy planning specialists using third-party treatment planning systems (not included in this submission).
# 6 Indications for Use
MeVis AVM Contour is intended for use by qualified medical professionals working in the field of neurosurgery or radiotherapy planning to manually create target contours for regions of interest in the cranial region through semi-automatic registration of 2D digital subtraction angiography (DSA) images onto 3D computed tomography (CT) scans.
The contours can be exported as DICOM RTStruct objects and are used in radiosurgery/-therapy treatment planning procedures to target specific brain regions.
The created contours are intended for further review, editing and approval by radiosurgery/-therapy planning specialists in third-party radiosurgery/-therapy treatment planning systems (not part of this submission).
The semi-automatic registration is restricted to/requires:
- X-ray, 2D DSA and 3D CT images of the brain
- X-ray images must be acquired in both frontal and sagittal direction
¹ In order to justify the acceptance criteria for the performance of the subject device's registration algorithm.
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 2
{6}
# MeVis
## Intended Patient Population
MeVis AVM Contour is used for patients diagnosed with Arteriovenous Malformations (AVM) requiring stereotactic radiosurgery (SRS). These patients are adults, 18 years and older.
## Intended User Profile
The intended users are qualified medical professionals, working in the field of neurosurgery or radiotherapy planning (Neurosurgeons and Radiation Oncologists).
## 7 Technological Characteristics Comparison
The technological characteristics of MeVis AVM Contour are similar to the identified legally marketed predicate device (see Table 2). The MeVis AVM Contour software uses fewer functionalities than the predicate device as it has a narrower indication. It only includes those functionalities necessary to fulfill its intended use.
Safety and effectiveness have been demonstrated by performance testing. Therefore, this difference in functionality is not substantial.
**Table 1** Comparison table for indications for use.
| Device | Indications for Use |
| --- | --- |
| **Subject Device MeVis AVM Contour** | MeVis AVM Contour is intended for use by qualified medical professionals working in the field of neurosurgery or radiotherapy planning to manually create target contours for regions of interest in the cranial region through semi-automatic registration of 2D digital subtraction angiography (DSA) images onto 3D computed tomography (CT) scans. The contours can be exported as DICOM RTStruct objects and are used in radiosurgery/-therapy treatment planning procedures to target specific brain regions. The created contours are intended for further review, editing and approval by radiosurgery/-therapy planning specialists in third-party radiosurgery/-therapy treatment planning systems (not part of this submission). The semi-automatic registration is restricted to/requires: • X-ray, 2D DSA and 3D CT images of the brain • X-ray images must be acquired in both frontal and sagittal direction Intended Patient Population MeVis AVM Contour is used for patients diagnosed with Arteriovenous Malformations (AVM) requiring stereotactic radiosurgery (SRS). These patients are adults, 18 years and older. Intended User Profile The intended users are qualified medical professionals, working in the field of neurosurgery or radiotherapy planning (Neurosurgeons and Radiation Oncologists). |
| **Primary Predicate device:** **Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio (K212420)** | Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The generated 3D structures are not intended to create physical replicas used for diagnostic purposes. The device itself does not have clinical indications. Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Tensor Imaging (DTI) data for use in treatment planning procedures. The device itself does not have clinical indications. Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. The device itself does not have clinical indications. Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. The device itself does not have clinical indications. |
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 3
{7}
# MeVis
**Table 2** Comparison table for Substantial Equivalence Discussion
| Feature/ Item | MeVis AVM Contour (subject device) | Predicate device | Comparison Result |
| --- | --- | --- | --- |
| Device Name | MeVis AVM Contour | **Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio** | n/a |
| 510(k) Number | n/a | K212420 | n/a |
| Indications for Use | See Table 1 | See Table 1 | Substantially equivalent The MeVis AVM Contours indications for use covers a subset of indications of the predicate device. The MeVis AVM Contours indications for use refer to 2 out of 4 applications in Brainlab Elements: Brainlab Elements Contouring and Brainlab Elements Image Fusion Angios indications for use. The subject device is raising no new concerns regarding safety or effectiveness, as these are the only applications necessary to perform the intended use of MeVis AVM Contour. |
| Product Code | LLZ | LLZ, JAK | Same as the predicate device. |
| Intended Use | The MeVis AVM Contour device is a specialized medical imaging software tool designed for qualified medical professionals working in the field of neurosurgery or radiotherapy planning. It enables healthcare professionals to create manual 3D CT-based target contours for AVM as input for further radiosurgery/therapy treatment planning. | Brainlab Elements is a medical device for processing of medical images, that is used to support treatment planning of surgical or radio-therapeutical procedures. | Substantially equivalent The MeVis AVM Contour intended use is a subset of the intended use of the predicate device. Both devices are used to process a variety of medical images in support of radiosurgery/therapy planning. |
| Intended Users | The intended users are qualified medical professionals, working in the field of neurosurgery or radiotherapy planning (Neurosurgeons and Radiation Oncologists). | The intended users are medical professionals. Typical users as defined in the use case descriptions are: • Image Guided Surgery: Neurosurgeons, Ear-Nose-Throat (ENT) surgeons and Cranio-Maxillofacial (CMF) surgeons including their assistants • Radiotherapy: medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | Substantially equivalent The MeVis AVM Contouring intended users are a subset of the users for the predicate device. Both devices are intended for use by medical professionals. |
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 4
{8}
# MeVis
| **Intended Patient Population** | MeVis AVM Contour is used for patients diagnosed with Arteriovenous Malformations (AVM) requiring stereotactic radiosurgery (SRS). | In general, there are no demographic, regional or cultural limitations for patients. It is up to the user to decide if the system shall be used to assist a certain procedure. | Substantially equivalent This is a subset of the patients mentioned in the indications of the predicate device. |
| --- | --- | --- | --- |
| **Environment of Use** | MeVis AVM Contour is intended for use in clinical and hospital environments, particularly in radiotherapy planning departments and neurosurgical units. It is designed for flexible use, allowing it to run on standard PCs. | The device shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. | Substantially equivalent |
| **Intended Part of the Body** | cranial region | Spine, cranial region | Substantially equivalent MeVis AVM Contouring supports a subset of the regions that are supported in the predicate device. |
| **Import of medical device image series** | X-ray, 2D DSA and 3D CT images of the brain | 2D DSA, 3D DSA, 3D CT, 3D MRA, DICOM support | Substantially equivalent MeVis AVM Contouring supports a subset of image series that are supported in the predicate device. X-ray images are a technical prerequisite for 2D-3D image registration and do not raise new concerns regarding safety and effectiveness. |
| **Image registration** | 2D/3D | 2D/3D, 3D/3D | Substantially equivalent MeVis AVM Contouring supports a subset of the image registration techniques that are supported in the predicate device. |
| **Image segmentation of anatomical structures** | manual | Manual, automated | Substantially equivalent MeVis AVM Contouring supports a subset of the image segmentation techniques/tools that are supported in the predicate device. |
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 5
{9}
MeVis
# 8 Performance Data
# Non-clinical Tests Discussion:
MeVis AVM Contour is validated and verified against its requirements and intended use by the successful execution of the Validation and Verification Plan:
- Specifications have been tested to demonstrate the software performs as intended for its intended use according to IEC 62304 and ISO 14971.
- Usability tests have been performed to demonstrate that usability requirements are met.
- Security tests have been performed covering vulnerability scanning, penetration testing, and verification of implemented cybersecurity measures.
- Bench tests for semi-automatic image registration were performed to demonstrate that the product is substantially equivalent.
A summary of the software verification and validation activities is provided below:
# Software verification and validation
| Test | Objective | Acceptance Criteria | Method | Result |
| --- | --- | --- | --- | --- |
| Unit Testing | Verify correct implementation of individual software components | Corresponding automated tests pass | Automated unit testing framework | Passed |
| Functional Testing | Verify correct implementation of system/software requirements | All system/software requirements are fulfilled; Pass/Fail criteria were based on the requirements | Functional test plan in simulated use environment | Passed |
| Acceptance Testing | Validate that user needs are fulfilled; Confirm the completeness of the implemented system requirements from an end-user perspective. | System operates as expected across all major areas; all system/software requirements are fulfilled; Pass/Fail criteria were based on the requirements and intended use | System test plan in simulated use environment, performed by subject matter experts | Passed |
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 6
{10}
# MeVis
## Usability Testing
| Test | Objective | Acceptance Criteria | Method | Result |
| --- | --- | --- | --- | --- |
| Usability Testing | Confirm safe and effective use | No critical use errors | Simulated use covering all tasks with representative users of intended user group | Passed |
| | | Usability design principles are fulfilled | Heuristic evaluation (Nielsen usability heuristics) with in-house usability specialist | Passed |
| | | High user acceptance | Qualitative assessment using a 5-point Likert score (system usability scale) with representative users of intended user group | Passed |
## Security Testing
| Test | Objective | Acceptance Criteria | Method | Result |
| --- | --- | --- | --- | --- |
| Security Testing | Verify correct implementation of security requirements and security controls | Security requirements are fulfilled | Security requirements test, vulnerability scanning/dynamic code analysis | Passed |
| Penetration Testing | Avoid exploitable vulnerabilities | No critical security issues | Grey-box test | Passed |
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 7
{11}
MeVis
# Bench Testing
For the semi-automatic image registration validation tests were performed to demonstrate that the product is substantially equivalent:
- The objective accuracy of the automated image registration algorithm (without manual refinement by the user) was evaluated using phantom data and retrospective clinical routine data.
- The expected alignment accuracy was set as ≤1 mm.
- The datasets used are suitable for demonstrating the performance of the device is consistent with U.S. clinical use conditions, demographically and technically representative of typical clinical imaging conditions encountered in clinical stereotactic radiosurgery routine.
- The objectively achieved 3D-CT to 2D-DSA alignment accuracy is 0.44 mm (± 0.70 mm sd)
The accuracy testing confirms the device meets or exceeds the accuracy of 0.50 mm of the predicate device (K212420) and reference device (K190042).
# Clinical Tests Discussion:
N/A - No clinical testing has been conducted to demonstrate substantial equivalence.
# 9 Conclusion
The 510(k) Pre-Market Notification for MeVis AVM Contour contains adequate information, data, and non-clinical test results to enable FDA – CDRH to determine substantial equivalence to the predicate device. MeVis Medical Solutions AG has determined that its device, MeVis AVM Contour, is substantially equivalent to the identified predicate device listed above.
Non-clinical performance testing has demonstrated that MeVis AVM Contour performs at least as safely and effectively as the predicate device in the context of its narrower indications for use.
MeVis AVM Contour – K260576
Rev 3 – 06/30/2026
Page 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.