Retrospective clinical imaging datasets (35 datasets from 16 clinical sites)
A retrospective study was conducted to evaluate the accuracy of the 2D/3D co-registration software by comparing device performance against a gold standard defined by medical experts using routine clinical imaging data.
Retrospective analysis of clinical imaging datasets
Clinical patients (35 datasets from 16 different clinical sites); Sample Size: 35 datasets; Number of Sites: 16
Gold standard fusions defined by medical experts
Targeting accuracy of 2D/3D fusions (Euclidean distance to gold standard)
Indications for Use
Image Fusion Angio 1.0 is a software application that is intended to be used for the co-registration of cerebrovascular image data.
Device Story
Software application for co-registration of digital subtraction angiographies (2D) with volumetric medical image data (3D). Used by neurosurgeons, neuroradiologists, and radiotherapy planners in hospital/surgical environments. Inputs include 2D and 3D vascular image data. Device performs image fusion to align anatomical structures; utilizes fusion algorithms and GUI tools for manual fusion, region of interest selection, and pair editing. Output is a co-registered image set used for surgical planning or preplanning. Benefits include improved visualization and accuracy for planning cranial lesion treatments.
Clinical Evidence
Bench testing and retrospective study. Phantom bench test (n=3) compared 2D/3D fusion accuracy (CTA, 3D-DSA to 2D-DSA) against reference device; subject device accuracy 0.5mm ± 0.2mm vs 0.8mm ± 0.3mm. MRA bench test (n=3) showed 0.3mm ± 0.1mm vs 3.2mm ± 0.3mm. Retrospective study (35 datasets, 16 sites) confirmed targeting accuracy of 0.36mm ± 0.17mm, consistent with phantom results and literature.
Technological Characteristics
Software-based image processing application. Performs 2D/3D co-registration of vascular image data. Connectivity via standard hospital network/PACS. Algorithm class: image registration/fusion. No physical materials or energy sources.
Indications for Use
Indicated for use in clinical workflows benefiting from co-registration of vascular image data as a planning or preplanning step. No demographic, regional, or cultural limitations. Used in hospital office environments or surgical intervention rooms by neurosurgeons, neuroradiologists, or radiotherapy planners.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
April 25, 2019
Brainlab AG % Mr. Alexander Schwiersch Regulatory Affairs Manager Olof-Palme-Str. 9 München, 81829 GERMANY
Re: K190042
Trade/Device Name: Brainlab Elements Image Fusion Angio Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, JAK Dated: March 22, 2019 Received: March 28, 2019
Dear Mr. Schwiersch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K190042
Device Name Brainlab Elements Image Fusion Angio
Indications for Use (Describe)
The application can be used in clinical workflows that benefit from the co-registration of vascular image data as a planning or preplanning step.
There are no demographic, regional or cultural limitations for patients.
The system shall be used in a hospital office environment or rooms appropriate for surgical interventions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS FOR IMAGE FUSION ANGIO
In accordance with requirements of 21 CFR Part 807.92
| Manufacturer: | Brainlab AG<br>Olof-Palme-Str. 9<br>81829 Munich<br>Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 5033 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Rainer Birkenbach |
| Contact person: | Alexander Schwiersch |
| Summary date: | 12/21/2018 |
| Device: | Image Fusion Angio |
| Trade name: | Brainlab Elements Image Fusion Angio |
| Common/Classification<br>Name: | System, Image Processing, Radiological |
| Predicate Device: | Image Fusion (K170816) |
| Reference Device: | iPlan RT Image (K080886) |
| Device classification<br>name: | Picture archiving and communications system |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | LLZ |
| Intended use: | Image Fusion Angio 1.0 is a software application that is intended to be used<br>for the co-registration of cerebrovascular image data. |
510(k) summary
{4}------------------------------------------------
| Indications for use: | The application can be used in clinical workflows that benefit from the co-<br>registration of vascular image data as a planning or preplanning step. |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description: | Image Fusion Angio is intended to co-register digital subtraction<br>angiographies with volumetric medical image data. |
| Operator Profile: | The application is intended to be used by medical professionals and their<br>assistants working in the field of neurosurgery, neuroradiology or<br>radiotherapy planning. |
| Patient Population: | There are no demographic, regional or cultural limitations for patients. |
| Conditions of Use: | The system shall be used in a hospital office environment or rooms<br>appropriate for surgical interventions. |
| Reason for 510(k)<br>submission: | New device. |
| Substantial<br>equivalence: | Image Fusion Angio 1.0 (Subject Device) is part of a new software<br>generation at Brainlab. The functionality of the Subject Device has been<br>previously included in the Predicate Device. The Subject Device specifically<br>takes over the fusion algorithm and the GUI design from the predicate<br>device (Image Fusion, K170816). It takes over the ability to load 2D images<br>and fuse them to 3D images from the reference device (iPlan RT Image,<br>K080886).<br>All devices provide the functionality co-register anatomical structures of<br>image data and can be used in planning workflows to treat cranial lesions. |
| Conclusion: | Functionality and features considered as substantially equivalent have been<br>verified and validated. The system Image Fusion Angio with its set of<br>functionalities is substantially equivalent to its predicate and reference<br>devices. |
| Changes to Predicate<br>Device: | The new Image Fusion Angio 1.0 Element enlarges the portfolio of already<br>cleared Brainlab Elements. Central changes compared to the main<br>predicate device Image Fusion are based in focusing on image fusion<br>functionality in general. |
{5}------------------------------------------------
# Technological Characteristics of the Subject Device in comparison to the Predicate Device:
| | Cleared device<br>feature/specifications<br>Image Fusion 3.0<br>(K170816) | Modified device<br>feature/specifications<br>Image Fusion Angio<br>1.0 |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| HW Requirements and<br>Operating Systems | Equivalent | |
| Data Inputs | Equivalent | |
| Viewing Data | Equivalent | |
| Performing an Image<br>Fusion | Equivalent as follows:<br>The predicate provides identical functions<br>compared to the subject device except for the<br>ability to load 2D images and to co-register<br>them to 3D images. This is not critical for<br>equivalence rating, since this is a known and<br>existing feature. | |
| Fusion Algorithm | Equivalent as follows:<br>The predicate provides the same algorithm<br>compared to the subject device except for the<br>ability to fuse 3D to 2D images. This is not<br>critical for equivalence rating, since this is a<br>known and existing feature. | |
| Fusion and<br>Verification Tools | | Equivalent |
| Accepting a Fusion<br>Result | | Equivalent |
| Performing a manual<br>fusion | | Equivalent |
| Choosing a Region of<br>Interest | | Equivalent |
| Edit and Select Pairs | | Equivalent |
Technological Characteristics of the Subject Device in
comparison to the Reference Device:
| | Cleared device<br>feature/specifications<br>iPlan RT Image<br>(K080886) | Modified device<br>feature/specifications<br>Image Fusion Angio<br>1.0 |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Fusion Algorithm | Equivalent as follows:<br>The reference device provides functionality for a<br>2D/3D co-registration in an equivalent context<br>but uses a differing algorithm with respect to the<br>2D/3D co-registration. This is not critical for<br>equivalence rating, since the resulting co-<br>registration accuracy is equally good, as proven<br>by verification and validation. | |
{6}------------------------------------------------
| Verification/validation<br>summary: | Verification<br>Verification has been performed according to the Verification Plan and<br>following internal processes to demonstrate that design specifications are<br>met by the device. |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Validation
Validation of the functionalities, including nonclinical performance tests (accuracy tests, see below), has been performed in accordance with the Clinical Evaluation Plan and following internal processes.
Usability tests have been performed as well to ensure that the device can be used safely and that measures against potential use related risks are effective.
The validation has been performed with software and units that are considered equivalent to the final version of the product (in terms of functionality and user interface), as demanded by 21 CFR 820.30 (q).
#### Nonclinical performance testing (Accuracy)
To evaluate the accuracy of the 2D/3D co-registration we took the following pathway (detailed results see table below):
- Phantom Bench Test to compare the 2D/3D fusion accuracy of । CTA, 3D-DSA to 2D-DSA of the Subject Device to the Reference Device.
- -MRA Bench Test - to compare the 2D/3D fusion accuracy of MRA to 2D-DSA of the Subject Device to the Reference Device.
- -Retrospective Study - to confirm similar accuracy as shown in the Phantom Bench Test.
Due to the results of the nonclinical tests we demonstrated that the device is as safe, as effective and performs as well as the reference device
| Test | Test Method Summary | Results |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phantom<br>Bench Test | We performed 2D and 3D scans using a physical vessel phantom.<br>2D and 3D images were fused with Image Fusion Angio and iPlan<br>RT Image (reference device). Targeting accuracies of both devices<br>to a gold standard fusion were compared to each other (Euclidean<br>Distances between predefined landmarks). The test was repeated 3<br>times. | Targeting accuracy, Image<br>Fusion Angio:<br>0.5 mm +/- 0.2 mm<br>Targeting accuracy, iPlan<br>RT Image:<br>0.8 mm +/- 0.3 mm |
| MRA Bench<br>Test | We performed 2D and 3D scans using a retrospective patient data<br>set. 2D and 3D images were fused with Image Fusion Angio and<br>iPlan RT Image (reference device). Targeting accuracies of both<br>devices to a gold standard fusion were compared to each other<br>(Euclidean Distances between predefined landmarks). The test was<br>repeated 3 times. | Targeting accuracy, Image<br>Fusion Angio:<br>0.3 mm +/- 0.1 mm<br>Targeting accuracy, iPlan<br>RT Image:<br>3.2 mm +/- 0.3 mm |
| Retrospective<br>Study | We used 35 datasets from 16 different clinical sites (11 different<br>scanner types). We performed 2D/3D fusions with Image Fusion<br>Angio. The gold standard fusions were defined by medical experts.<br>Targeting accuracies of predefined landmarks (Euclidean distance<br>to gold standard fusion) were calculated. All fusions were further<br>reviewed by medical experts. | Targeting accuracy: 0.36<br>mm +/- 0.17 mm<br>The result is similar to the<br>result of the phantom<br>bench test and similar to<br>the findings in the<br>literature. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.