The IMDS NHanceIT Guide Wire Support Catheter with hydrophilic coating is a guidewire exchange and infusion device designed for use in the coronary and/or peripheral vasculature. The catheter is intended to support a guidewire during access of coronary and/or peripheral vasculature and allows for exchange of guidewires and provide a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.
Device Story
NHanceIT is an over-the-wire (OTW) single-lumen guide wire support catheter; used in coronary/peripheral vasculature via femoral or brachial approach. Device features 0.014" guidewire compatibility; radiopaque tip for X-ray visualization; deformable atraumatic tip for vessel centering. Physician operates device under fluoroscopic guidance to navigate to distal lesions; provides conduit for contrast agent delivery and guidewire exchange. Benefits include improved access to target lesions and procedural efficiency. No AI/ML components.
Clinical Evidence
Bench testing only. Verification tests included kink resistance, flexibility, tensile strength, torque robustness, leak testing, catheter burst, infusion flow rate, surface coating lubricity, coating integrity, and bond strength. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, and pyrogenicity. All tests met acceptance criteria.
Technological Characteristics
OTW single-lumen catheter; 135 cm length; 0.014" guidewire compatibility; 5 Fr minimum guiding catheter. Materials: Pebax/Polyethylene shaft, stainless steel reinforcement, TPE/Tungsten radiopaque marker, PUR/PVP hydrophilic coating, MDX inner lumen coating. Straight tip design. Sterilization method not specified. No software or energy source.
Indications for Use
Indicated for patients requiring guidewire support, guidewire exchange, or delivery of diagnostic contrast agents within the coronary or peripheral vasculature. Contraindicated for use in the neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 14, 2026
IMDS Operation B.V.
Edwin Schulting
CEO
Ceintuurbaan Noord 150
Roden, NL9301NZ
Netherlands
Re: K260559
Trade/Device Name: NHanceIT (NX31413525)
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY
Dated: July 15, 2026
Received: July 15, 2026
Dear Edwin Schulting:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260559 - Edwin Schulting
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260559 - Edwin Schulting
Page 3
Sincerely,
JENNY R.
KATSNELSON -S
Digitally signed by JENNY
R. KATSNELSON -S
Date: 2026.08.14 16:50:51
-04'00'
for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260559
Device Name
NHanceIT
Indications for Use (Describe)
The IMDS NHanceIT Guide Wire Support Catheter with hydrophilic coating is a guidewire exchange and infusion device designed for use in the coronary and/or peripheral vasculature. The catheter is intended to support a guidewire during access of coronary and/or peripheral vasculature and allows for exchange of guidewires and provide a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K262396
# 510(k) Summary – K262396
[As required by 21 CFR 807.92]
Date Prepared: December 15, 2025
**Submitter's Name / Contact Person**
**Manufacturer**
IMDS Operations BV
Ceintuurbaan Noord 150
9301 NZ Roden, The Netherlands
Establishment Registration #3007740583
**Contact Person**
Aljosja Kot
Director Quality Assurance and
Regulatory Affairs
Tel: 0031651453880
Fax: 0031508200231
**General Information**
**Trade Name**
**Common/ Usual Name**
**Classification Name**
**Predicate Device**
**Reference Device**
NHanceIT
Catheter
Catheter, percutaneous
K191560, Turnpike LP Catheter, Teleflex Medical LLC
K200324, NHancer Rx guide wire support catheter, IMDS Operations B.V.
**Device Description**
The NHanceIT is a guide wire support catheter intended to be used in the coronary and/ or peripheral vasculature. The NHanceIT is a 0.014" OTW guide wire support catheter. The target lesion can be reached by using the femoral or brachial approach. The catheter has a distal port, which accesses the guide wire lumen. The guide wire lumen begins at the proximal end and terminates at the distal tip. The catheter has a radiopaque tip section to indicate the distal end of the support catheter. The tip is visible under X-ray. The length of the catheter is designed that the distal end of the catheter may be used to navigate to lesions in more distal vessels. The distal part carries a deformable atraumatic tip to help center the wire in the vessel.
**Intended Use/Indications for Use**
The IMDS NHanceIT Guide Wire Support Catheter with hydrophilic coating is a guidewire exchange and infusion device designed for use in the coronary and/or peripheral vasculature. The catheter is intended to support a guidewire during access of coronary and/or peripheral vasculature and allows for exchange of guidewires and provide a conduit for the delivery of diagnostic contrast agents.
The device is contraindicated for use in the neurovasculature.
**Technological Characteristics Comparison**
The NHanceIT is similar in design to the predicate device and both are OTW single lumen guide wire support catheters intended to support a guide wire during access of coronary and/ or peripheral vasculature, allow for exchange of guide wires and provide a conduit for the delivery of diagnostic contrast agents.
| # | Item | NHanceIT | Turnpike LP | Equivalent* (Yes or Different) |
| --- | --- | --- | --- | --- |
| | Model number | NX31413525 | 5638 | |
| 1 | Type clinically based | Over the wire single lumen guide wire support catheter | Over the wire single lumen guide wire support catheter | Y |
Page 1 of 3
{5}
K262396
| 2 | Shaft material | Pebax / Polyethylene | Unknown | D |
| --- | --- | --- | --- | --- |
| 3 | Shaft reinforcement | Stainless steel | Stainless steel | Y |
| 4 | Number of guide wire distal exit ports | 1 | 1 | Y |
| 5 | Over the wire guide wire lumens | 1 | 1 | Y |
| 6 | Radiopaque marker material | TPE / Tungsten | Unknown | D |
| 7 | Effective Length or Usable Length | 135 cm | 135 cm | Y |
| 8 | Exit Marker location (from tip) | 95 and 105 cm | No markers | D |
| 9 | Guide wire compatibility | 0.014 inch | 0.014 inch | Y |
| 10 | Distal tip outer diameter | 0.58 mm | 0.57 mm | D |
| 11 | Distal shaft outer diameter | 0.73 mm | 0.73 mm | Y |
| 12 | Proximal shaft outer diameter | 1.07 mm | 0.96 mm | D |
| 13 | Minimum Guiding Catheter size | 5 Fr | 5 Fr | Y |
| 14 | Tip design / shape | Straight | Straight | Y |
| 15 | Hydrophilic coating distal shaft | Present | Present | Y |
| 16 | Hydrophilic coating material | PUR / PVP hydrophilic coating | Unknown | D |
| 17 | Inner lumen coating | MDX | Unknown | D |
The NHanceIT is similar in design and technological characteristics to the predicate device. The differences in dimensions and materials were successfully evaluated in performance tests against the predicate device.
The reference device has a material shaft design equal to the NHanceIT and is manufactured in the same cleanrooms and with similar production equipment. The material composition does not raise different questions regarding effectiveness and/or safety compared to the predicate.
### Substantial Equivalence and Summary of Studies
The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results supported substantial equivalence.
Page 2 of 3
{6}
K262396
The NHanceIT guide wire support catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests:
1) Kink resistance/ flexibility
2) Device introduction, deployment and retraction
3) Radiopacity
4) Distal tip length
5) Tensile strength
6) Effective length
7) Shaft inner diameter
8) Torque robustness
9) Outer diameter
10) Leak testing
11) Catheter burst
12) Infusion flow rate
13) Surface coating lubricity
14) Coating integrity
15) Coating particulate evaluation
16) Catheter bond strength and tip pull
17) Packaging integrity
The NHanceIT is meeting the requirements for biocompatibility in accordance with ISO 10993-1:
- Cytotoxicity
- Sensitization
- Irritation
- Systemic toxicity
- Hemocompatibility
- Pyrogenicity
The results of the verification tests met the specified acceptance criteria and supported substantial equivalence.
Therefore, the NHanceIT guide wire support catheter is substantially equivalent to the predicate device.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.