The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.
Device Story
NHancer Rx is a dual-lumen, rapid-exchange (Rx) guidewire support catheter; 135 cm length; compatible with 0.014" guidewires. Features an over-the-wire (OTW) lumen and a distal rapid-exchange lumen; includes a removable stylet for loading assistance. Distal shaft features hydrophilic coating; internal lumens feature MDX (silicone) coating. Used in clinical settings (e.g., cath lab) by physicians to facilitate vascular access, guidewire exchange, and contrast agent infusion. Radiopaque markers at distal tip and 8 mm from tip aid positioning. Device provides mechanical support for guidewires, enabling navigation through coronary/peripheral vasculature; improves procedural efficiency during wire exchanges.
Clinical Evidence
Bench testing only. Verification testing included kink resistance, flexibility, guidewire insertion, radiopacity, tensile strength, burst pressure, flow rate, leak testing, and coating integrity/lubricity. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility. All tests met acceptance criteria.
Technological Characteristics
Dual-lumen percutaneous catheter; polymeric shaft with stainless steel reinforcement. 135 cm effective length; 0.014" guidewire compatibility. Hydrophilic coating (PUR/PVP) on distal 15 cm; hydrophobic MDX (silicone) inner lumen coating. Tungsten radiopaque markers. Hub/Luer lock conforms to NEN EN ISO 80369-7. Max injection pressure 300 psi. Removable stylet included. Sterilization method not specified.
Indications for Use
Indicated for patients requiring coronary or peripheral vascular access for guidewire support, guidewire exchange, or diagnostic contrast agent delivery. Contraindicated for use in neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
August 6, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
IMDS Operations B.V. Edwin Schulting CEO Ceintuurbaan Noord 150 Roden, 9301 NA NI
Re: K200324
Trade/Device Name: NHancer Rx Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 2, 2020 Received: July 6, 2020
Dear Edwin Schulting:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200324
Device Name NHancer Rx
#### Indications for Use (Describe)
The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
#### Date Prepared: August 6th, 2020
| <b>Submitter's Name / Contact Person</b> | |
|------------------------------------------|------------------------------------|
| <b>Manufacturer</b> | <b>Contact Person</b> |
| IMDS Operations BV | Florence Wagter |
| Ceintuurbaan Noord 150 | Director of Quality and Regulatory |
| 9301 NZ Roden, The Netherlands | Tel: 0031651453880 |
| Establishment Registration #3007740583 | Fax: 0031508200231 |
#### General Information
| Trade Name | NHancer Rx |
|---------------------|------------------------------------------------------------------------------|
| Common/Usual Name | Dual lumen catheter |
| Classification Name | Catheter, percutaneous |
| Predicate Device | K162467, Twin-Pass torque, dual access catheter, Vascular<br>solutions, Inc. |
| Reference Device | K121077, NHancer guidewire support catheter |
#### Device Description
The NHancer™Rx is a dual lumen catheter intended to use in the coronary and/or peripheral vasculature. The NHancer™Rx is a 0.014" Rapid Exchange (Rx) guide wire support catheter. The NHancer™Rx has an over the wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal end. The NHancer™Rx has an effective length of 135 cm and is compatible with 0.014"/0.36mm or smaller guide wires. The NHancer™Rx has two depth markings located at 95 cm and 105 cm from the distal tip. The NHancer™Rx has a radiopaque marker identifying the distal end of the catheter and a second radiopaque marker located 8 mm from the distal tip identifies the distal end of the OTW lumen.
To reduce friction in the guide catheter and vasculature, the distal 15 cm of the shaft is fitted with a hydrophilic coating. To reduce friction of the guide wire in the entire lumens of the NHancer Rx are coated with MDX (silicone) coating.
To ease catheter loading into the hemostasis device and the guiding catheter a removable stylet (9) is placed in the OTW lumen
#### Intended Use
The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.
#### Technological Characteristics Comparison
The NHancer Rx is similar in design to the predicate device and both are dual lumen, percutaneous catheters intended to access discrete regions of the coronary and peripheral vasculature, facilitate placement and exchange of guidewires and sub selectively infuse agents.
| # | Item | NHancer Rx<br>dual lumen guidewire support<br>catheter | Twin-Pass Torque dual access<br>catheter |
|---|-----------------------|--------------------------------------------------------|------------------------------------------|
| | Model number | NRX1413518 | 5201 |
| 1 | Type clinically based | Dual lumen guidewire support catheter | Dual lumen guidewire support catheter |
| 2 | Shaft material | Polymeric | Polymeric |
{4}------------------------------------------------
| 3 | Shaft reinforcement | Stainless steel | Stainless steel |
|----|---------------------------------------|----------------------------------------------------------|----------------------------------------------------------|
| 4 | OTW Guidewire lumens | 1 | 1 |
| 5 | Rapid Exchange guidewire lumens | 1 | 1 |
| 6 | Number of guidewire distal exit ports | 2 | 2 |
| 7 | Strain Relief | present | present |
| 8 | Hub and Luer Lock | Female, 6% taper and screw conform<br>NEN EN ISO 80369-7 | Female, 6% taper and screw conform<br>NEN EN ISO 80369-7 |
| 9 | Radiopaque marker material | Tungsten | Pt/Ir |
| 10 | Effective Length or Usable Length | 135 cm | 135 cm |
| 11 | Exit Marker location (from tip) | 95 and 105 cm | 95 and 105 cm |
| 12 | Guidewire compatibility | 0.014 inch | 0.014 inch |
| 13 | Maximum injection pressure | 300 psi | 300 psi |
| 14 | Minimum Guiding Catheter size | 5 Fr | 5 Fr |
| 15 | Tip design / shape | Straight | Straight |
| 16 | Tip material | Polymeric | Polymeric |
| 17 | Hydrophilic coating distal shaft | present | present |
| 18 | Hydrophilic coating material | PUR / PVP<br>hydrophilic coating | Not known |
| 19 | Inner lumen coating | Hydrophobic | Not known |
| 20 | Removable stylet in OTW lumen | Yes | No |
| 21 | Distal shaft design | Oval | Round |
With the exception of dimensional, material and package configuration differences, the NHancer Rx is similar in design and technological characteristics to the predicate device. The dimensional, material and package configuration differences were successfully evaluated in performance tests.
{5}------------------------------------------------
#### Substantial Equivalence and Summary of Studies
The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results did not raise new questions of safety or effectiveness. The NHancer Rx dual lumen catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests:
- 1) Kind resistance/ flexibility
- 2) Guide wire insertion
- 3) Radiopacity
- 4) Distal tip length
- 5) Tensile strength
- 6) Catheter body burst
- 7) Contrast medium flow rate
- 8) Leak testing
- 9) Effective length
- 10) Shaft inner diameter
- 11) Outer diameter
- 12) Surface coating lubricity
- 13) Coating Integrity
- 14) Coating Particulate Evaluation
- 15) Packaging integrity
Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1:
- Cytotoxicity ●
- Sensitization
- Irritation ●
- Systemic toxicity
- Pyrogenicity
- Hemocompatibility
The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues.
Therefore, the NHancer Rx dual lumen catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.