Profoject™ Safety Needle; Profoject™ Syringe with Safety Needle; Profoject™ Low Dead Space Syringe with Safety Needle
K260514 · CMT Health PTE., Ltd. · FMF · Jun 12, 2026 · General Hospital
Device Facts
Record ID
K260514
Device Name
Profoject™ Safety Needle; Profoject™ Syringe with Safety Needle; Profoject™ Low Dead Space Syringe with Safety Needle
Applicant
CMT Health PTE., Ltd.
Product Code
FMF · General Hospital
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Profoject™ Safety Needle is intended to be used with a Luer slip or Luer lock syringe for withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. The Profoject™ Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. The Profoject™ Low Dead Space Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
Device Story
Sterile, single-use manual devices for fluid withdrawal/injection; includes safety needle, syringe with safety needle, and low dead space syringe with safety needle. Components: stainless steel (SUS 304) needle tube; polypropylene (PP) hub, cap, and safety shield; polyisoprene rubber plunger stopper. Safety shield manually activated post-withdrawal to cover needle, reducing needlestick risk. User assembles syringe and needle at point-of-care. Output is fluid delivery or aspiration. Benefits include reduced fluid waste (low dead space model) and enhanced sharps injury protection. Used in clinical settings by healthcare professionals.
Clinical Evidence
Bench testing only. No clinical data. Simulated clinical use trial conducted with 520 devices to evaluate usability and reliability of the sharps injury prevention mechanism; zero functional failures reported.
Technological Characteristics
Materials: SUS 304 stainless steel (needle), polypropylene (hub/shield/barrel/plunger), polyisoprene (stopper). Standards: ISO 7864, ISO 9626, ISO 7886-1, ISO 80369-7, ISO 6009, ISO 23908. Sterilization: Ethylene oxide (SAL 10^-6). Connectivity: Luer lock. Form factor: Manual piston syringe with integrated safety shield.
Indications for Use
Indicated for withdrawal and injection of fluids for medical purposes in patients requiring hypodermic administration. Intended for use by healthcare professionals. No specific contraindications listed.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Safety Syringe with Needle for Single Use (K212920)
Sterile Safety Hypodermic Needle for Single Use (K212920)
Hypodermic Needle-Pro® EDGE™ Safety Device with Low Dead Space Syringe (K232943)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
CMT Health PTE., Ltd.
Monica Ma
Primary Correspondent
150 Beach Rd., #28-05, Gateway W.
Singapore, 189720
Singapore
Re: K260514
Trade/Device Name: Profoject™ Safety Needle; Profoject™ Syringe with Safety Needle; Profoject™ Low Dead Space Syringe with Safety Needle
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF, MEG, QNQ, FMI
Dated: May 19, 2026
Received: May 19, 2026
Dear Monica Ma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260514 - Monica Ma
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260514 - Monica Ma
Page 3
Sincerely,
KYRAN R. GIBSON -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General
Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260514
?
Please provide the device trade name(s).
?
Profoject™ Safety Needle;
Profoject™ Syringe with Safety Needle;
Profoject™ Low Dead Space Syringe with Safety Needle
Please provide your Indications for Use below.
?
The Profoject™ Safety Needle is intended to be used with a Luer slip or Luer lock syringe for withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Profoject™ Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Profoject™ Low Dead Space Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
# 510(k) Summary
1. Date of Preparation: February 14, 2026
2. Sponsor Identification
CMT HEALTH PTE. LTD.
150 BEACH ROAD, #28-05, GATEWAY WEST, SINGAPORE, 189720
Establishment Registration Number: 3036722184
Contact Person: Monica Ma
Position: Management Representative
Tel.: +65 6846 1379
Email: Ra@cmthealth.com
3. Designated Submission Correspondent
Name: Monica Ma
Email: Ra@cmthealth.com
Tel.: +65 6846 1379
4. Information of Proposed Device
Trade Name: Profoject™ Safety Needle
Profoject™ Syringe with Safety Needle
Profoject™ Low Dead Space Syringe with Safety Needle
Common Name: Antistick Needle
Syringe with Antistick Needle
Low Dead Space Syringe with Antistick Needle
Classification Name: Piston Syringe
Syringe Antistick
Classification: II
Product Code: FMF, MEG, QNQ
Regulation Number: 21 CFR 880.5860
Classification Name: Hypodermic Single Lumen Needle
1 / 13
{5}
Classification: II
Product Code: FMI
Regulation Number: 21 CFR 880.5570
5. Identification of Predicate Device
5.1 Safety Needle and Syringe with Safety Needle
510(k) Number: K212920
Trade Name: Sterile Safety Syringe with Needle for Single Use (Used as predicate device)
Sterile Safety Hypodermic Needle for Single Use (Used as predicate device)
Sterile Auto-Disable Syringe with Needle for Single Use
Common Name: Piston syringe and antistick needle
Classification Name: Piston Syringe
Syringe Antistick
Hypodermic Single Lumen Needle
Classification: II
Product Code: FMF
MEG
FMI
Regulation Number: 21 CFR 880.5860
21 CFR 880.5570
5.2 Low Dead Space Syringe with Safety Needle
510(k) Number: K232943
Trade Name: Hypodermic Needle-Pro® EDGE™ Safety Device with Low Dead Space Syringe
Common Name: Sterile Low Dead Space Piston Syringe, Needle, Hypodermic
Classification: II
Product Code: QNQ
FMI
MEG
Regulation Number: 21 CFR 880.5860
21 CFR 880.5570
6. Indications for Use Statement:
The Profoject™ Safety Needle is intended to be used with a Luer slip or Luer lock syringe for withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental
2 / 13
{6}
needlesticks.
The Profoject™ Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Profoject™ Low Dead Space Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
# 7. Device Description
The Profoject™ Safety Needle is a sterile, single-use device, which consists of a needle tube, a needle hub, a needle cap, and a safety shield attached to the needle hub. The needle tube is made of stainless steel (SUS 304). The needle hub and the safety shield are made of polypropylene, and their colors comply with ISO 6009. The needle cap is also made of polypropylene and does not come into contact with the patient. This device is available in a variety of configurations, including different needle gauges, lengths, and wall thicknesses. It is compatible for use with a Luer slip or Luer lock syringe. The safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Profoject™ Syringe with Safety Needle is a sterile, single-use device, which includes a Profoject™ Safety Needle and a Luer lock syringe (Model M or Model C) comprising a polypropylene barrel, a polypropylene plunger, and a polyisoprene rubber plunger stopper. This device is available in various combination of syringe volume and needle size. The syringe and the safety needle are not provided pre-assembled; instead, they are to be assembled by the user at the time of use. The needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Profoject™ Low Dead Space Syringe with Safety Needle is a sterile, single-use device, which includes a Profoject™ Safety Needle and a Luer lock low dead space syringe comprising a polypropylene barrel, a polypropylene plunger, and a polyisoprene rubber plunger stopper. The proposed device offers 1mL syringes configurable with a range of needle sizes. The low dead space syringe and the safety needle are not provided pre-assembled; instead, they are to be assembled by the user at the time of use. The needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. This device is specifically designed to minimize dead space and reduce fluid waste.
The Profoject™ Safety Needle, the Profoject™ Syringe with Safety Needle and the Profoject™ Low Dead Space Syringe with Safety Needle are sterilized by ethylene oxide to achieve a Sterility Assurance Level (SAL) of 10⁻⁶ and are supplied in a sterility maintenance package that can maintain the sterility of the device during its 5-year shelf life.
The specifications of the proposed devices are provided in following table.
Table 1 Specifications of the Profoject™ Safety Needle
3 / 13
{7}
| Needle Gauge | Designated Metric Size (mm) | Needle Length (mm) | Wall Thickness |
| --- | --- | --- | --- |
| 18G | 1.20 | 22, 25, 30, 38 | RW, TW |
| 19G | 1.10 | 20, 22, 25, 30, 38 | RW, TW |
| 20G | 0.90 | 20, 22, 25, 30, 38 | RW, TW |
| 21G | 0.80 | 20, 22, 25, 30, 38 | RW, TW |
| 22G | 0.70 | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| 23G | 0.60 | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| 25G | 0.50 | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| 26G | 0.45 | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| 27G | 0.40 | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| 30G | 0.30 | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
Table 2 Specifications of the Profoject™ Syringe with Safety Needle
| Model | Syringe Volume | Needle Gauge | Connector Type | Needle Length (mm) | Wall Thickness |
| --- | --- | --- | --- | --- | --- |
| Model C / Model M | 1mL | 25G | LL | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 26G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 27G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 30G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| Model C / Model M | 3mL | 20G | LL | 20, 22, 25, 30, 38 | RW, TW |
| | | 21G | | 20, 22, 25, 30, 38 | RW, TW |
| | | 22G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 23G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 25G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| Model C / Model M | 5mL | 20G | LL | 20, 22, 25, 30, 38 | RW, TW |
| | | 21G | | 20, 22, 25, 30, 38 | RW, TW |
| | | 22G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
4 / 13
{8}
| | | 23G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| --- | --- | --- | --- | --- | --- |
| Model C | 6mL | 20G | LL | 20, 22, 25, 30, 38 | RW, TW |
| | | 21G | | 20, 22, 25, 30, 38 | RW, TW |
| | | 22G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 23G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| Model C / Model M | 10mL | 20G | LL | 20, 22, 25, 30, 38 | RW, TW |
| | | 21G | | 20, 22, 25, 30, 38 | RW, TW |
| | | 22G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 23G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| Model C | 12mL | 20G | LL | 20, 22, 25, 30, 38 | RW, TW |
| | | 21G | | 20, 22, 25, 30, 38 | RW, TW |
| | | 22G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | | 23G | | 13, 15, 20, 22, 25, 30, 38 | RW, TW |
Table 3 Specifications of the Profoject™ Low Dead Space Syringe with Safety Needle
| Syringe Volume | Needle Gauge | Connector Type | Needle Length (mm) | Wall Thickness |
| --- | --- | --- | --- | --- |
| 1mL | 25G | LL | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | 26G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | 27G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
| | 30G | | 10, 13, 15, 20, 22, 25, 30, 38 | RW, TW |
8. Substantially Equivalent (SE) Comparison
5 / 13
{9}
Table 4 General Comparison of Profoject™ Safety Needle
| Item | Proposed Device | Predicate Device K212920 | Comments |
| --- | --- | --- | --- |
| Product | Profoject™ Safety Needle | Sterile Safety Hypodermic Needle for Single Use | / |
| Product Code | FMI | FMI | Same |
| Regulation No. | 21 CFR 880.5570 | 21 CFR 880.5570 | Same |
| Indications for Use | The Profoject™ Safety Needle is intended to be used with a Luer slip or Luer lock syringe for withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | The Sterile Safety Hypodermic Needle for Single Use is intended to be used with a Luer slip or Luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | Same |
| Configuration | Needle hub Needle tube Needle cap Safety shield | Needle hub Needle tube Needle cap Safety shield | Same |
| Operation Mode | For manual use only | For manual use only | Same |
| Single Use | Single Use | Single Use | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Connector Type | Luer Lock/Luer Slip | Luer Lock/Luer Slip | Same |
| Needle Gauge | 18G, 19G, 20G, 21G, 22G, 23G, 25G, 26G, 27G, 30G | 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | Different See comment 1 |
| Needle Length | 10mm, 13mm, 15mm, 20mm, 22mm, 25mm, 30mm, 38mm | 13mm, 16mm, 19mm, 22mm, 25mm, 32mm, 38mm | Different See comment 1 |
| Needle Hub | Color-coded per ISO 6009 | Color-coded per ISO 6009 | Same |
| Needle Performance | Complied with ISO 7864 ISO 9626 ISO 80369-7 | Complied with ISO 7864 ISO 9626 ISO 80369-7 | Same |
| Safety Mechanism | Similar safety shield and same manual activated mechanism. | Similar safety shield and same manual activated mechanism. | Same |
| Material | Needle hub Polypropylene (PP) | Needle hub Polypropylene (PP) | Same |
| | Needle tube Stainless Steel | Needle tube Stainless Steel | Same |
| | Safety shield Polypropylene (PP) | Safety shield Polypropylene (PP) | Same |
| | Needle cap Polypropylene (PP) | Needle cap Polypropylene (PP) | Same |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Same |
| Sterilization Method | EO Sterilized | EO Sterilized | Same |
| SAL | 10^{-6} | 10^{-6} | Same |
6 / 13
{10}
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same |
| --- | --- | --- | --- |
Discussion:
Comment 1
The needle gauge and needle length of the proposed device are different from those of the predicate device. However, this difference is just in dimension. Different needle specifications will be selected by physicians based on patients' conditions, and this difference does not affect the intended use. In addition, bench performance testing was conducted on the proposed device according to ISO 7864, ISO 9626, and ISO 80369-7, and the results met the requirements of these standards. Therefore, the difference in needle gauge and needle length does not raise different questions of safety and effectiveness when compared to the predicate device.
Table 5 General Comparison of Profoject™ Syringe with Safety Needle
| Item | Proposed Device | Predicate Device K212920 | Comments |
| --- | --- | --- | --- |
| Product | Profoject™ Syringe with Safety Needle | Sterile Safety Syringe with Needle for Single Use | / |
| Product Code | FMF FMI MEG | FMF FMI MEG | Same |
| Regulation No. | 21 CFR 880.5860 21 CFR 880.5570 | 21 CFR 880.5860 21 CFR 880.5570 | Same |
| Class | Class II | Class II | Same |
| Indications for Use | The Profoject™ Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | The Sterile Safety Syringe with Needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | Same |
| Configuration | Syringe | Barrel (Luer lock) | Syringe | Barrel (Luer lock/Luer slip) | Different See comment 2 |
| Plunger | Plunger | Same |
| Plunger stopper | Plunger stopper | Same |
| Needle | Needle hub | Needle | Needle hub | Same |
| Needle tube | Needle tube | Same |
| Needle cap | Needle cap | Same |
| Safety shield | Safety shield | Same |
| Operation Mode | For manual use only | For manual use only | Same |
| Single Use | Single Use | Single Use | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Syringe Volume | 1mL, 3mL, 5mL, 6mL, 10mL, 12mL | 1ml, 2ml, 2.5ml, 3ml, 5ml, 10ml, 20ml, | Different |
7 / 13
{11}
| | | | 25ml, 30ml, 50ml, 100ml | See comment 3 |
| --- | --- | --- | --- | --- |
| Needle | Connector Type | Luer Lock | Luer Lock/Luer Slip | Different See comment 4 |
| | Needle Gauge | 20G, 21G, 22G, 23G, 25G, 26G, 27G, 30G | 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | Different See comment 5 |
| | Needle Length | 10mm, 13mm, 15mm, 20mm, 22mm, 25mm, 30mm, 38mm | 13mm, 16mm, 19mm, 22mm, 25mm, 32mm, 38mm | Different See comment 5 |
| | Needle Hub | Color-coded per ISO 6009 | Color-coded per ISO 6009 | Same |
| Needle & Syringe Performance | Complied with ISO 7864 ISO 9626 ISO 7886-1 ISO 80369-7 | Complied with ISO 7864 ISO 9626 ISO 7886-1 ISO 80369-7 | | Same |
| Material | Barrel | Polypropylene (PP) | Barrel | Polypropylene (PP) |
| | Plunger | Polypropylene (PP) | Plunger | Polypropylene (PP) |
| | Plunger stopper | Polyisoprene | Plunger stopper | Polyisoprene |
| | Needle hub | Polypropylene (PP) | Needle hub | Polypropylene (PP) |
| | Needle tube | Stainless Steel | Needle tube | Stainless Steel |
| | Safety shield | Polypropylene (PP) | Safety shield | Polypropylene (PP) |
| | Needle cap | Polypropylene (PP) | Needle cap | Polypropylene (PP) |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | | Same |
| Sterilization Method | EO Sterilized | EO Sterilized | | Same |
| SAL | 10^{-6} | 10^{-6} | | Same |
| Endotoxin Limit | 20 EU per device | 20 EU per device | | Same |
#### Discussion:
##### Comment 2
The predicate device's syringe barrel connector is available in both Luer lock and Luer slip configurations, whereas the proposed device features only the Luer lock configuration. This design difference does not affect the device's intended use. Furthermore, the Luer lock connector on the proposed device was tested per ISO 80369-7, and the results met the requirements of the standard. Therefore, the difference in syringe barrel connector design does not raise different questions of safety and effectiveness when compared to the predicate device.
##### Comment 3
The syringe volume of the proposed device is different from that of the predicate device. However, this difference is just in dimension. Different syringe specification will be selected by physicians based on patients' conditions, and this difference does not affect the intended use. In addition, bench performance testing was conducted on the proposed device according to ISO 7886-1 and ISO 80369-7, and the results met the requirements of these standards. Therefore, the difference in syringe volume does not raise different questions of safety and effectiveness when compared to the predicate device.
8 / 13
{12}
Comment 4
The predicate device's needle hub is available in both Luer lock and Luer slip connector configurations, whereas the proposed device's needle hub incorporates only the Luer lock configuration. This design difference does not affect the device's intended use. Furthermore, the Luer lock connector on the proposed device was tested per ISO 80369-7, and the results met the requirements of the standard. Therefore, the difference in needle hub connector types does not raise different questions of safety and effectiveness when compared to the predicate device.
Comment 5
The needle gauge and needle length of the proposed device are different from those of the predicate device. However, this difference is just in dimension. Different needle specifications will be selected by physicians based on patients' conditions, and this difference does not affect the intended use. In addition, bench performance testing was conducted on the proposed device according to ISO 7864, ISO 9626, and ISO 80369-7, and the results met the requirements of these standards. Therefore, the difference in needle gauge and needle length does not raise different questions of safety and effectiveness when compared to the predicate device.
Table 6 General Comparison of Profoject™ Low Dead Space Syringe with Safety Needle
| Item | Proposed Device | Predicate Device K232943 | Comments |
| --- | --- | --- | --- |
| Product | Profoject™ Low Dead Space Syringe with Safety Needle | Hypodermic Needle-Pro® EDGE™ Safety Device with Low Dead Space Syringe | / |
| Product Code | QNQ, FMI, MEG | QNQ, FMI, MEG | Same |
| Regulation No. | 21 CFR 880.5860 21 CFR 880.5570 | 21 CFR 880.5860 21 CFR 880.5570 | Same |
| Class | Class II | Class II | Same |
| Indications for Use | The Profoject™ Low Dead Space Syringe with Safety Needle is intended for use in the withdrawal and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | This device is intended for use to inject fluids into or withdraw fluids from the body. The needle protection device covers the needle after use to help prevent needle sticks. | Different See comment 6 |
| Configuration | Syringe | Barrel (Luer lock) | Barrel (Luer lock) |
| | | Plunger | Plunger |
| | | Plunger stopper | Plunger stopper |
| | Needle | Needle hub | Needle hub |
| | | Needle tube | Needle tube |
| | | Needle cap | Needle cap |
| | | Safety shield | Safety shield |
| Operation Mode | For manual use only | For manual use only | Same |
9 / 13
{13}
| Single Use | | Single Use | | Single Use | | Same |
| --- | --- | --- | --- | --- | --- | --- |
| Syringe Volume | | 1mL | | 1mL, 3mL | | Different See comment 7 |
| Needle | Connector Type | Luer Lock | | Luer Lock | | Same |
| | Needle Gauge | 25G, 26G, 27G, 30G | | 23G, 25G | | Different See comment 8 |
| | Needle Length | 10mm, 13mm, 15mm, 20mm, 22mm, 25mm, 30mm, 38mm | | 16mm, 25mm, 38mm | | Different See comment 8 |
| Needle & Syringe Performance | | Complied with ISO 7864 ISO 9626 ISO 7886-1 ISO 80369-7 | | Complied with ISO 7864 ISO 9626 ISO 7886-1 ISO 80369-7 | | Same |
| Material | Barrel | Polypropylene (PP) | | Barrel | Polypropylene (PP) | Same |
| | Plunger | Polypropylene (PP) | | Plunger | Polypropylene (PP) | Same |
| | Plunger stopper | Polyisoprene | | Plunger stopper | IR rubber | Same |
| | Needle tube | Stainless Steel | | Needle tube | Stainless Steel | Same |
| | Lubricant | Silicone | | Needle hub | Silicone | Same |
| Dead Space Specification | | Low Dead Space Maximum dead space: ≤ 0.035mL | | Low Dead Space Maximum dead space: ≤ 0.015mL | | Different See comment 9 |
| Biocompatibility | | Conforms with ISO 10993-1 | | Conforms with ISO 10993-1 | | Same |
| Sterilization Method | | EO Sterilized | | EO Sterilized | | Same |
## Discussion:
### Comment 6
The proposed device and the predicate device share the same core intended use: fluid withdrawal and injection, and post-use needle protection to reduce needlestick injuries. Both devices also feature a low dead space design. The minor differences in descriptive phrasing (e.g., 'manually activated,' 'immediately after use' in the proposed device) do not alter the intended use or affect safety and effectiveness. These phrases provide additional clarity to users and are consistent with the predicate device's actual use. Therefore, the difference in descriptive phrasing does not raise different questions of safety and effectiveness when compared to the predicate device.
### Comment 7
The syringe volume of the proposed device is different from that of the predicate device. However, this difference is just in dimension. Different syringe specification will be selected by physicians based on patients' conditions, and this difference does not affect the intended use. In addition, bench performance testing was conducted on the proposed device according to ISO 7886-1 and ISO 80369-7, and the results met the requirements of these standards. Therefore, the difference in syringe volume does not raise different questions of safety and effectiveness when compared to the predicate device.
10 / 13
{14}
# Comment 8
The needle gauge and needle length of the proposed device are different from those of the predicate device. However, this difference is just in dimension. Different needle specifications will be selected by physicians based on patients' conditions, and this difference does not affect the intended use. In addition, bench performance testing was conducted on the proposed device according to ISO 7864, ISO 9626, and ISO 80369-7, and the results met the requirements of these standards. Therefore, the difference in needle gauge and needle length does not raise different questions of safety and effectiveness when compared to the predicate device.
# Comment 9
While the dead space acceptance criteria for the proposed device differ from the predicate device, the proposed device fully complies with the requirement specified in ISO 7886-1 that mandates a dead space of ≤ 0.070 mL for syringes with a nominal volume (V) < 2 mL. Therefore, the difference in dead space acceptance criteria does not raise different questions of safety and effectiveness when compared to the predicate device.
# 9. Performance data
To establish substantial equivalence to the identified predicate device, the tests noted below were performed on the proposed device. The testing results proved that the proposed device complied with the applicable standards requirements and is substantially equivalent to the predicate device.
# **Biocompatibility testing**
In accordance with ISO 10993-1 the proposed device is classified as an externally communicating device, in contact with "Blood path, indirect" with a contact duration of less than 24 hours. A biocompatibility evaluation for the proposed device was conducted per ISO 10993-1:2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," and the FDA Guidance document (2023), "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The evaluation results indicated that proposed device's biocompatibility met the requirements in the standards.
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
# **Performance testing**
The bench performance testing performed verified that the performance of the proposed device was substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
ISO 7864:2016 Sterile hypodermic needles for single use — Requirements and test methods
ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods
ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use
ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications
ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methods
11 / 13
{15}
ISO 6009:2016 Hypodermic needles for single use — Colour coding for identification
ISO 23908:2024 Sharps injury protection — Sharps protection mechanisms for single-use needles, introducers for catheters and needles used for blood testing, monitoring, sampling and medical substance administration — Requirements and test methods
### **Sterilization, packaging, and shelf-life testing**
The sterilization process was validated in accordance with ISO 11135:2014, establishing the routine control and monitoring parameters. Simulated distribution testing and package integrity testing were conducted to demonstrate that the packaging effectively protects the product and maintains sterility. To support the claimed 5-year shelf-life, accelerated aging study was performed, followed by package integrity testing, sterility testing, and bench performance testing, confirming that the device remains functional and sterile throughout its shelf-life.
| Sterilization validation | ISO 11135: 2014 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices |
| --- | --- |
| EO and ECH residue test | ISO 10993-7:2008+AMD1:2019 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — AMENDMENT 1: Applicability of allowable limits for neonates and infants |
| Bacterial endotoxins test | USP <85> Bacterial Endotoxins Test |
| Package integrity testing | Seal strength ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials |
| Shelf-life evaluation | Dye penetration ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration Visual inspection ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection Internal pressurization ASTM F1140/F1140M-13 (Reapproved 2020)e1 Standard Test Methods for Internal Pressurization Failure Resistance of Packages Sterility test USP <71> Sterility tests Physical, mechanical, chemical, and package tests were performed on aging samples to verify the claimed shelf-life of the device |
### **Simulated clinical study**
Beyond performance testing, a dedicated human factors study was conducted to evaluate the usability of the investigational device. In alignment with FDA's Medical Devices with Sharps Injury Prevention Features guidance, we carried out a simulated clinical use trial to examine both the usability and operational reliability of the safety needle's sharps injury prevention mechanism. Healthcare professionals assessed the safety needle's performance against predefined pass/fail criteria and provided qualitative feedback on their experience with the device's functional design. A total of 520 devices were tested, and no functional failures were identified throughout the trial.
12 / 13
{16}
#### 10. Clinical Test Conclusion
Clinical studies are not required to demonstrate substantial equivalence to the predicate device.
#### 11. Conclusion
The Profoject™ Safety Needle, Profoject™ Syringe with Safety Needle and Profoject™ Low Dead Space Syringe with Safety Needle are substantially equivalent to their predicate device. The non-clinical testing demonstrates that the devices are as safe, as effective and perform as well as the legally marketed device.
13 / 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.