Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use,Sterile Auto-Disable Syringe with Needle for Single Use
K212920 · Shandong Weigao Group Medical Polymer Co., Ltd. · FMF · Mar 11, 2022 · General Hospital
Device Facts
Record ID
K212920
Device Name
Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use,Sterile Auto-Disable Syringe with Needle for Single Use
Applicant
Shandong Weigao Group Medical Polymer Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Mar 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Safety Syringe with Needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. The Sterile Safety Hypodermic Needle for Single Use is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. The Sterile Auto-Disable Syringe with Needle for Single Use is intended for aspiration and injection of fluids.
Device Story
Manual, single-use sterile syringes and hypodermic needles; includes safety-shielded and auto-disable variants. Safety-shielded devices feature a manual shield activated post-injection to cover the needle. Auto-disable syringes prevent reuse via integrated design. Used in clinical settings for fluid aspiration and injection. Healthcare providers manually operate the device, including the safety shield activation. Output is the delivery or aspiration of fluids; safety features reduce needlestick injury risk. Materials include polypropylene barrels/hubs, polyisoprene pistons, and stainless steel needles.
Clinical Evidence
No clinical data. Bench testing only, including performance testing per ISO standards, biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, hemolysis, pyrogen), sterile barrier packaging (ASTM F1886, F88, F1929), and simulated clinical study for safety mechanism evaluation per FDA guidance and ISO 23908.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub, cap, shield), Polyisoprene (piston), Stainless Steel (needle). Standards: ISO 7864, ISO 9626, ISO 80369-7, ISO 7886-1, ISO 7886-3. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: None. Form factor: Manual syringe/needle combinations with safety shields or auto-disable mechanisms.
Indications for Use
Indicated for aspiration and injection of fluids for medical purposes. Intended for single use by clinicians. Safety-shielded versions are indicated to minimize accidental needlestick risk.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Disposable Syringe with Safety Needle (K170651)
BD SoloShot Mini Syringe/ BD Auto Disable Syringe (K201234)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency responsible for regulating food and drug products in the United States.
March 11, 2022
Shandong Weigao Group Medical Polymer Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K212920
Trade/Device Name: Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use.Sterile Auto-Disable Syringe with Needle for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF. FMF. MEG Dated: February 9, 2022 Received: February 11, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212920
## Device Name
Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use, Sterile Auto-Disable Syringe With Needle for Single Use
## Indications for Use (Describe)
The Sterile Safety Syringe with Needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Sterile Safety Hypodermic Needle for Single Use is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Sterile Auto-Disable Syringe with Needle for Single Use is intended for aspiration and injection of fluids.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K212920 510(k) Summary
- 1. Date of Preparation: March 11, 2022
- 2. Sponsor Identification:
# Shandong Weigao Group Medical Polymer Co., Ltd.
No.18 Xingshan Road, Torch Hi-tech Science Park, 264210 Weihai, Shandong Province, PEOPLE'S REPUBLIC OF CHINA.
Establishment Registration Number: 3007084575
Contact Person: Lina Liu Position: QA Manager Tel: +86-631-5716818 Fax: +86-631-5620555 Email: liulina@weigaogroup.com
- 3. Designated Submission Correspondent:
Ms. Diana Hong (Primary Contact Person) Ms. Jinlei Tang (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
- Identification of Proposed Device: 4.
Trade Name: Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use Sterile Auto-Disable Syringe with Needle for Single Use Common Name: Piston syringe and antistick needle
Classification Name: Piston Syringe Syringe Antistick Classification: II Product Code: FMF; MEG Regulation Number: 21CFR 880.5860
Classification Name: Hypodermic Single Lumen Needle Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570
- 5. Identification of Predicate Devices:
510(k) Number: K170651
Product Name: Sterile Disposable Syringe with Safety Needle (used as predicate device) Sterile Disposable Syringe with Needle Sterile Disposable Syringe Sterile Disposable Safety Needle (used as predicate device) Sterile Disposable Needle
510(k) Number: K201234 Product Name: BD SoloShot Mini Syringe/ BD Auto Disable Syringe
## Indications for Use:
The Sterile Safety Syringe with Needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The Sterile Safety Hypodermic Needle for Single Use is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately
{5}------------------------------------------------
after use to minimize risk of accidental needlesticks.
The Sterile Auto-Disable Syringe with Needle for Single Use is intended for aspiration and injection of fluids.
### Device Description:
The Sterile Safety Syringe with Needle for Single Use is intended for manual and single use only, which consists of a hypodermic needle with a safety shield attached to the needle hub and a luer slip or luer lock syringe. The proposed device is available in a variety combination of syringe volume and needle size. The safety shield will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Sterile Safety Hypodermic Needle for Single Use is intended for manual and single use only, which consists of a hypodermic needle with a safety shield attached to the connector hub. The proposed device is available in variety combination of needle gauge and needle length. The proposed device is compatible for use with a luer slip or luer lock syringe. The safety sheath will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Sterile Auto-Disable Syringe with Needle for Single Use is intended for aspiration and injection of fluids. The Sterile Auto-Disable Syringe with Needle for Single Use is available in various capacities of syringes.
The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10th and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.
The specifications of the proposed devices are provided in following table.
| Needle Gauge | Needle Length |
|--------------|------------------------|
| 30G | 13mm, 16mm, 19mm |
| 29G | 13mm, 16mm, 19mm |
| 28G | 13mm, 16mm, 19mm |
| 27G | 13mm, 16mm, 19mm |
| 26G | 13mm, 16mm, 19mm |
| 25G | 16mm, 19mm, 25mm, 38mm |
| 24G | 16mm, 19mm, 25mm |
| 23G | 19mm, 25mm, 32mm, 38mm |
| 22G | 19mm, 25mm, 32mm, 38mm |
Table 1. Specifications of the Sterile Safety Hypodermic Needle for Single Use
{6}------------------------------------------------
| 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
|-----|------------------------------|
| 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| 19G | 25mm, 32mm, 38mm |
| 18G | 19mm, 22mm, 32mm, 38mm |
Table 2. Specifications of the Sterile Safety Syringe with Needle for Single Use
| Syringe volume | Needle Gauge | Needle Length |
|----------------|------------------------|------------------------------|
| 1mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 2mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| 18G | 19mm, 22mm, 32mm, 38mm | |
| 2.5mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| 3mL | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 5mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 10mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 20mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm, 19mm |
| | 26G | 13mm, 16mm, 19mm |
| | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 25mL | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 30mL | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 50mL | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
| 100mL | 25G | 16mm, 19mm, 25mm, 38mm |
| | 24G | 16mm, 19mm, 25mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| | 22G | 19mm, 25mm, 32mm, 38mm |
| | 21G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 20G | 19mm, 22mm, 25mm, 32mm, 38mm |
| | 19G | 25mm, 32mm, 38mm |
| | 18G | 19mm, 22mm, 32mm, 38mm |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Table 3. Specifications of the Sterile Auto-Disable Syringe with Needle for Single Use
| Syringe volume | Needle Gauge | Needle Length |
|----------------|--------------|------------------------|
| 0.2mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm |
| | 26G | 16mm, 19mm |
| | 25G | 16mm, 19mm, 32mm, 38mm |
| | 24G | 16mm, 19mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| 0.5mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm |
| | 26G | 16mm, 19mm |
| | 25G | 16mm, 19mm, 32mm, 38mm |
| | 24G | 16mm, 19mm |
| | 23G | 19mm, 25mm, 32mm, 38mm |
| 1mL | 30G | 13mm, 16mm, 19mm |
| | 29G | 13mm, 16mm, 19mm |
| | 28G | 13mm, 16mm, 19mm |
| | 27G | 13mm, 16mm |
| | 26G | 16mm, 19mm |
| | 25G | 16mm, 19mm, 32mm, 38mm |
| | 24G | 16mm, 19mm |
{10}------------------------------------------------
| | 23G | 19mm, 25mm, 32mm, 38mm |
|--|-----|------------------------|
|--|-----|------------------------|
- 6. Substantially Equivalent (SE) Comparison:
| | | Table 4. General Comparison of Sterile Safety Syringe with Needle for Single Use | |
|--|--|----------------------------------------------------------------------------------|--|
| | | | |
| ITEM | Proposed Device | Predicate Device K170651 | Comments | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------|
| Product | Sterile Safety Syringe with Needle for Single Use | Sterile Disposable Syringe with Safety Needle | / | |
| Product Code | FMF<br>FMI<br>MEG | FMF<br>FMI<br>MEG | Same | |
| Regulation Number | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same | |
| Class | Class II | Class II | Same | |
| Indication for Use | The Sterile Safety Syringe with Needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. | Same | |
| Configuration | Syringe Barrel (luer lock/luer slip)<br>Plunger<br>Piston<br>Needle Hub<br>Needle tube<br>Needle cap<br>Safety shield | Syringe Barrel (luer lock/luer slip)<br>Plunger<br>Piston<br>Needle Hub<br>Needle tube<br>Needle cap<br>Safety shield | Same | |
| Operation Mode | For manual use only | For manual use only | Same | |
| Single Use | Single Use | Single Use | Same | |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same | |
| Syringe | 1ml, 2ml, 2.5ml, 3ml, 5ml, 10ml, 20ml, 25ml, 30ml, 50ml, 100ml | 1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml, 60ml | Different<br>Comment #1 | |
| Syringe Volume | | | | |
| The Syringe volume for proposed device is different from the predicate device. This difference does not affect<br>intended use and this technological difference does not raise new questions of safety and effectiveness. | | | | |
| Syringe<br>performance | Complied with<br>ISO 7886-1 | Complied with<br>ISO 7886-1 | Same | |
| Needle Type | Connector Luer Lock/Luer Slip | Luer Lock /Luer Slip | Same | |
| | Size | 18G, 19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G,<br>30G | 16G,18G, 19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G, 30G | Different<br>Comment #2 |
| | Length | 13mm, 16mm, 19mm, 22mm,<br>25mm, 32mm, 38mm | 8mm(5/16"), 13mm(1/2"),<br>16mm(5/8"), 20mm(3/4"), 25mm(1"),<br>32mm(1-1/4"), 38mm(1-1/2") | Different<br>Comment #2 |
| | Needle<br>hub | Color-coded per ISO 6009 | Color-coded per ISO 6009 | Same |
| Comment #2 | | | | |
| Needle Size and Length | | | | |
| The needle size and length for proposed device is different from the predicate device. However, the needle size<br>and length of the proposed device is included in the range of the needle size and length of the predicate device. | | | | |
| Therefore, this difference does not affect safety and effectiveness of the proposed device. | | | | |
| Needle<br>Performance | Syringe | Complied with<br>ISO 7886-1<br>Iso 9626-1 | Complied with<br>ISO 7886-1<br>Iso 9626-1 | Same |
| Material | | Barrel Polypropylene (PP) | Barrel Polypropylene (PP) | Same |
| | | Plunger Polypropylene (PP) | Plunger Polypropylene (PP) | Same |
| | | Piston Polyisoprene | Piston Polyisoprene | Same |
| | | Hub Polypropylene (PP) | Hub Polypropylene (PP) | Same |
| | | Needle<br>tube Stainless Steel | Needle<br>tube Stainless Steel | Same |
| | | Needle<br>cap Polypropylene (PP) | Needle<br>cap Polypropylene (PP) | Same |
| | | Safety<br>shield Polypropylene (PP) | Safety<br>shield Polypropylene (PP) | Same |
| | | | | |
| Cytotoxicity | | No cytotoxicity | No cytotoxicity | Same |
| Irritation | | No intracutaneous reactivity | No intracutaneous reactivity | Same |
| Sensitization | | No skin sensitization | No skin sensitization | Same |
| Systemic Toxicity | | No systemic toxicity | No systemic toxicity | Same |
| Hemolysis | | No Hemolysis | No Hemolysis | Same |
| Pyrogen | | No Pyrogen | No Pyrogen | Same |
| Method | | | | |
| SAL | 10-6 | 10-6 | Same | |
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same | |
{11}------------------------------------------------
{12}------------------------------------------------
# Table 5. General Comparison of Sterile Safety Hypodermic Needle for Single Use
| ITEM | Proposed Device | Predicate Device K170651 | Comment |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Product | Sterile Safety Hypodermic Needle for Single Use | Sterile Disposable Safety Needle | / |
| Product Code | FMI | FMI | Same |
| Regulation No. | 21 CRF 880.5570 | 21 CRF 880.5570 | Same |
| Class | Class II | Class II | Same |
| Indication for Use | The Sterile Safety Hypodermic Needle for Single Use is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. | The Sterile Disposable Safety Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick. | Same |
| Configuration | Hub<br>Needle tube<br>Needle cap<br>Safety shield | Hub<br>Needle tube<br>Needle cap<br>Safety shield | Same |
| Operation Mode | For manual use only | For manual use only | Same |
| Single Use | Single Use | Single Use | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Connector Type | Luer Lock/Luer Slip | Luer Lock /Luer Slip | Same |
| Needle Gauge | 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | 16G,18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | Different<br>Comment #3 |
| Needle Length | 13mm, 16mm, 19mm, 22mm, 25mm, 32mm, 38mm | 8mm(5/16"),<br>13mm(1/2"), 16mm(5/8"), 20mm(3/4"), 25mm(1"), 32mm(1-1/4"), 38mm(1-1/2") | Different<br>Comment #3 |
| Comment #3<br>Needle Gauge and length | | | |
{13}------------------------------------------------
| Needle Hub | | | | | |
|--------------------|------------------------------------------------------------|------------------------------------------------------------|---------------|--------------------|------|
| Needle Performance | Complied with<br>ISO 7864<br>ISO 9626 | Complied with<br>ISO 7864,<br>ISO 9626 | Same | | |
| Safety mechanism | Similar safety shield and same manual activated mechanism. | Similar safety shield and same manual activated mechanism. | Same | | |
| Material | Needle hub | Polypropylene (PP) | Needle hub | Polypropylene (PP) | Same |
| | Needle tube | Stainless Steel | Needle tube | Stainless Steel | Same |
| | Safety Shield | Polypropylene (PP) | Safety Shield | Polypropylene (PP) | Same |
| | Needle Cap | Polypropylene (PP) | Needle Cap | Polypropylene (PP) | Same |
| Biocompatibility | | | | | |
| Cytotoxicity | No cytotoxicity | No cytotoxicity | Same | | |
| Irritation | No intracutaneous reactivity | No intracutaneous reactivity | Same | | |
| Sensitization | No skin sensitization | No skin sensitization | Same | | |
| Systemic Toxicity | No systemic toxicity | No systemic toxicity | Same | | |
| Hemolysis | No Hemolysis | No Hemolysis | Same | | |
| Pyrogen | No Pyrogen | No Pyrogen | Same | | |
| Sterilization | | | | | |
| Method | EO Sterilized | EO Sterilized | Same | | |
| SAL | 10-6…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.