The device is a wearable, non-invasive, and non-implanted bone conduction hearing aid intended to treat patients of patients aged 12 years and older with conductive hearing loss or single-sided deafness. It utilizes an ear-hook design to ensure stable mechanical coupling between the transducer and the temporal bone. • Unilateral or bilateral conductive hearing loss, either chronic or temporary. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 25 dB HL. • Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing on the contralateral side. Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
Device Story
Wearable, non-invasive, head-worn bone conduction hearing aid; utilizes ear-hook design for temporal bone coupling. Input transducer captures sound; signal conditioning unit processes/amplifies signals; output transducer (bone vibrator) transmits vibrations to temporal bone for inner ear conduction. User-operated via mode button to toggle between General, Denoise, and Deep Denoise modes. Powered by lithium battery. Used in clinical or home settings to improve auditory perception in patients with conductive hearing loss or single-sided deafness. Healthcare providers use device output to provide acoustic compensation; patient benefits from restored or improved hearing.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -23), electrical safety/EMC (IEC 60601 series), and performance verification (OFL90, AMSL, EIN, THD, frequency response, and denoise function). Software verification/validation and cybersecurity assessments were performed per FDA guidance.
Technological Characteristics
Head-worn, flexible, bendable construction. Bone conduction transducer; lithium battery powered. Features manual program selection (General, Denoise, Deep Denoise). Complies with IEC 60601-1, 60601-1-11, 60601-1-2, and ISO 10993 standards. Includes Bluetooth communication. Software-controlled signal processing.
Indications for Use
Indicated for patients aged 12+ with conductive hearing loss (BC threshold ≤25 dB HL) or single-sided deafness (contralateral AC threshold ≤20 dB HL).
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
June 15, 2026
Zdeer Technology (Henan) Co., Ltd.
Wenlin Xiang
Manager
East Of Wei Wu Rd. And S. Of Qin Gong E. Rd.,
Toubao Village Committee, Laotown Town
Changge, Xuchang City, Henan 461500
China
Re: K260452
Trade/Device Name: Bone Conduction Hearing Aid (G41157, G41158, G41159, G41160, G42151, G42152, G42153, G42154, G42155, G42156, G42157, G42158, G42159, G42160, G43151, G43152, G43153, G43154, G43155, G43156, G45151, G45152, G45153, G45154, G45155, G45156, G45157, G45158, G45159, G45160, G46151, G46152, G46153, G46154, G46155, G46156, G46157, G46158)
Regulation Number: 21 CFR 874.3302
Regulation Name: Bone-Conduction Hearing Aid
Regulatory Class: Class II
Product Code: LXB
Dated: May 20, 2026
Received: May 20, 2026
Dear Wenlin Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260452 - Wenlin Xiang
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K260452 - Wenlin Xiang
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SHUCHEN PENG -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260452 | ? |
| Please provide the device trade name(s). | | ? |
| Bone Conduction Hearing Aid (G41157, G41158, G41159, G41160, G42151, G42152, G42153, G42154, G42155, G42156, G42157, G42158, G42159, G42160, G43151, G43152, G43153, G43154, G43155, G43156, G45151, G45152, G45153, G45154, G45155, G45156, G45157, G45158, G45159, G45160, G46151, G46152, G46153, G46154, G46155, G46156, G46157, G46158) | | |
| Please provide your Indications for Use below. | | ? |
| The device is a wearable, non-invasive, and non-implanted bone conduction hearing aid intended to treat patients of patients aged 12 years and older with conductive hearing loss or single-sided deafness. It utilizes an ear-hook design to ensure stable mechanical coupling between the transducer and the temporal bone. • Unilateral or bilateral conductive hearing loss, either chronic or temporary. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 25 dB HL. • Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing on the contralateral side. Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
510(k) #: K260452
510(k) Summary
Prepared on: 2026-06-12
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Zdeer Technology (Henan) Co., Ltd |
| --- | --- |
| Applicant Address | East of Wei Wu Road and south of Qin Gong East Road, Toubao Village Committee, Laotown Town Changge, Xuchang City Henan 461500 China |
| Applicant Contact Telephone | +86-15527272371 |
| Applicant Contact | Mr. Wenlin Xiang |
| Applicant Contact Email | 1132075798@qq.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Bone Conduction Hearing Aid (G41157, G41158, G41159, G41160, G42151, G42152, G42153, G42154, G42155, G42156, G42157, G42158, G42159, G42160, G43151, G43152, G43153, G43154, G43155, G43156, G45151, G45152, G45153, G45154, G45155, G45156, G45157, G45158, G45159, G45160, G46151, G46152, G46153, G46154, G46155, G46156, G46157, G46158) |
| --- | --- |
| Common Name | Bone-conduction hearing aid |
| Classification Name | Hearing Aid, Bone Conduction |
| Regulation Number | 874.3302 |
| Product Code(s) | LXB |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K172460 | Adhere system | LXB |
| K121793 | bone conduction hearing aid | LXB |
# Device Description Summary
21 CFR 807.92(a)(4)
This Bone Conduction Hearing Aid is a head-worn device featuring flexible, bendable construction with instant rebound for a secure and comfortable fit. Integrated with an input transducer, a signal conditioning unit, an output transducer (bone vibrator), and a lithium battery, the device captures sound signals, processes and amplifies them, then transmits the audio to the temporal bone via its bone vibrator—where sound is further conducted to the inner ear. Designed for patients with conductive hearing loss or single-sided deafness, this hearing aid leverages bone conduction technology to deliver targeted vibrations to the temporal bone, enabling effective sound transmission to the inner ear for auditory perception.
The device is fitted with a mode button, which allows users to toggle between three preset modes: General Mode, Denoise Mode and Deep Denoise Mode.
There are 38 different models (G41157, G41158, etc.), but that they have the same intended use and internal electrical components, and they only differ in button location, casing material, and performance parameters.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The device is a wearable, non-invasive, and non-implanted bone conduction hearing aid intended to treat patients of patients aged 12
{5}
years and older with conductive hearing loss or single-sided deafness. It utilizes an ear-hook design to ensure stable mechanical coupling between the transducer and the temporal bone.
- Unilateral or bilateral conductive hearing loss, either chronic or temporary.
The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 25 dB HL.
- Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing on the contralateral side. Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The Subject Device is substantially equivalent to the Predicate Device in terms of intended use, clinical indications, and fundamental operating principle. Both devices provide non-invasive bone conduction acoustic compensation for patients with conductive hearing loss or SSD, utilizing identical clinical thresholds (BC≤25 dB HL).
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device maintains the same technological characteristics as the predicate and reference devices regarding the indicated degree of hearing loss, target population, circuit type, and user features such as volume control and low battery alerts. Since the subject device's OFL90 performance falls within the established performance envelope defined by the Predicate and Reference devices, the differences in maximum output are considered clinically insignificant. The subject device has comparable electroacoustic performance in terms of ASML50, Equivalent Input Noise (EIN), and Total Harmonic Distortion (THD) when compared to the predicate device.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
1)BIOCOMPATIBILITY TESTING
The zdeer Bone Conduction Hearing Aid is compliant with the standard of ISO 10993-5 and ISO 10993-10, ISO 10993-23. The tests were conducted in compliance with applicable requirements in the 21 CFR Part 58 regulations.
2)ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC)
Electrical safety, EMC and EMI testing were conducted on the zdeer Bone Conduction Hearing Aid. The zdeer Bone Conduction Hearing Aid complies with the applicable sections of the following standards: IEC 60601-1:2005+A2:2020, IEC 60601-1-11:2015+A1:2020, IEC 60601-1-6:2010 +A1:2013+A2:2020, IEC 62366:2007 +A1:2014, IEC 60601-1-2:2014 +A1:2020 and IEC TS 60601-4-2 :2024.
3)MECHANICAL AND OTHER TESTING
Bench testing was successfully performed in respect to the defined technical requirements such as: Appearance and Structure, Output force level for input sound pressure level of 90 dB (OFL90), Acousto-mechanical Sensitivity Level (AMSL) at Reference Frequency, Equivalent Input Noise(EIN), Total Harmonic Distortion(THD), Frequency Response Range, Rated Power Supply Current Consumption, Manual Program Selection, Automatic Denoise Function, Number of Channels, Bluetooth Communication Function, Electrical Safety Requirements.
4)SOFTWARE VERIFICATION AND VALIDATION TESTING
Software verification and validation testing were conducted and basic level of documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions(2023)".
5)CYBERSECURITY
The Cybersecurity verification and validation was provided in accordance with the Guidance: Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions.
The non-clinical data provided support the safety of the device and the hardware verification and validation demonstrate that the zdeer Bone Conduction Hearing Aid should perform as intended in the specified use conditions. The performance data has confirmed that the product is substantially equivalent as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.