The ADHEAR system is intended to treat patients of all ages with conductive hearing loss or single-sided deafness via bone conduction. The ADHEAR system is a non-invasive bone conduction hearing device which is retained on the patient's head with an elastic headband or an adhesive adapter that is placed behind the auricle.
Device Story
ADHEAR system is a non-invasive bone conduction hearing aid; consists of an audio processor (model 701) and an adhesive adapter or elastic headband. Processor contains microphones, signal processing technology, and battery; detects, processes, and amplifies sound; transmits vibrations to mastoid bone via adhesive adapter; bypasses outer/middle ear to stimulate inner ear. Used during waking hours; processor removed at night; adhesive adapter replaced every 3-7 days. Operated by patient or caregiver; user selects between 4 pre-defined settings via push button. Benefits patients with conductive hearing loss or single-sided deafness by providing improved hearing through bone conduction.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and mechanical performance testing (frequency range, gain, THD, EIN, and durability).
Technological Characteristics
Bone conduction hearing aid; electromagnetic transducer; zinc-air battery powered. Components: audio processor, adhesive adapter, headband. Materials biocompatible per ISO 10993-1. Connectivity: DAI (Direct Audio Input) for external audio sources. Software: digital signal processing for acoustic amplification. Sterilization: not applicable (non-invasive). Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-6, IEC 62366, IEC 60601-1-2, ANSI C63.19.
Indications for Use
Indicated for patients of all ages with unilateral or bilateral conductive hearing loss (chronic or temporary) with bone-conduction thresholds ≤ 25 dB HL (0.5-3 kHz), or single-sided deafness (unilateral profound sensorineural deafness) with normal contralateral air-conduction thresholds ≤ 20 dB HL (0.5-3 kHz).
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
April 27, 2018
MED-EL Elektromedizinische Geraete GmbH Elizabeth Gfoeller Corporate Director, Regulatory Affairs Fuerstenweg 77a Innsbruck, AT 6020 Tirol
Re: K172460
Trade/Device Name: ADHEAR System Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: March 23, 2018 Received: March 30, 2018
Dear Elizabeth Gfoeller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172460
Device Name ADHEAR System
Indications for Use (Describe)
The ADHEAR system is intended to treat patients of all ages with conductive hearing loss or single-sided deafness via bone conduction. The ADHEAR system is a non-invasive bone conduction hearing device which is retained on the patient's head with an elastic headband or an adhesive adapter that is placed behind the auricle.
Indications:
· Unilateral or bilateral conductive hearing loss, either chronic or temporary.
The pure tone average bone-conduction hearing threshold (measured at 0.5. 1. 2, and 3 kHz) should be better than or equal to 25 dB HL.
· Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing on the contralateral side. Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C) </div> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K172460
### 510(K) Summary
#### l. SUBMITTER
Applicant:
MED-EL Elektromedizinische Geräte GmbH Fuerstenweg 77a 6020 Innsbruck AUSTRIA Phone: +43 577 88 5614 FAX: +43 577 88 5690
Contact Person:
Elizabeth Gfoeller Corporate Director, Regulatory Affairs MED-EL Elektromedizinische Geräte GmbH Fuerstenweg 77a 6020 Innsbruck AUSTRIA Phone: +43 577 88 5614 FAX: +43 577 88 5690
Date Prepared: March 23, 2018
#### II. DEVICE
Trade Name: ADHEAR System
Generic/Common Name: Bone Conduction Hearing Prosthesis
Classification: 21 CFR§874.3300, Hearing Aid
Product Code: LXB, Hearing Aid, Bone Conduction
#### PREDICATE DEVICES III.
- Contact Mini, audifon USA Inc. (K121793) Primary Predicate ●
- . BAHA 5, Cochlear Americas (K142907) Secondary Predicate
No reference devices were used in this submission.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. A stylized, circular graphic is placed between the "D" and "E" in "MED-EL", adding a unique visual element to the logo.
#### IV. Device Description
The ADHEAR system includes a bone conduction audio processor that can be retained on the head with an adhesive adapter or by the headband situated over the mastoid behind the auricle. The ADHEAR system is intended to be used during waking hours. While the adhesive adapter is attached to the skin for 3 to 7 days and then replaced, the audio processor is removed at night. Both the audio processor and the adhesive adapter are necessary components for the system to work. Together the components work as one system to deliver vibrations to the mastoid bone in order to conductively transmit sound to the inner ear for patients with conductive hearing loss or single sided deafness via bone conduction.
The ADHEAR system consists of an adhesive adapter sitting behind the auricle and an audio processor mounted on the attachment. The audio processor contains microphones and signal processing technology as well as a battery. It detects, processes, amplifies and finally transmits sound to the adhesive adapter which transmits vibrations to the mastoid which conducts sounds to the inner ear. The audio processor is equipped with a push button that allows the user to switch between 4 pre-defined settings. A colour sleeve is provided for pediatric use. It features a tamper-proof battery door and dampens potential drops of the processor unit.
The ADHEAR system contains the following components:
- . ADHEAR Audio Processor (model 701)
- Package of ADHEAR Adhesive Adapters
- . ADHEAR Positioning Tool
- ADHEAR Headband
- . Retention clip
- ADHEAR Sleeves
### V. Indications for Use
The ADHEAR system is intended to treat patients of all ages with conductive hearing loss or single-sided deafness via bone conduction. The ADHEAR system is a non-invasive bone conduction hearing device which is retained on the patient's head with an elastic headband or an adhesive adapter that is placed behind the auricle.
Indications:
- Unilateral or bilateral conductive hearing loss, either chronic or temporary. . The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 25 dB HL.
- Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing ● on the contralateral side.
Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL
### VI. Comparison of Technological Characteristics with The Predicate Device
The ADHEAR System and the predicate devices are bone conduction devices intended to provide hearing improvement via bone conduction. They do so by inducing vibration using the same operating principle of receiving, and transferring the sound via the bone of the skull to the inner ear of a unilaterally or bilaterally hearing impaired person.
At a high level the subject and predicate devices are based on the same technological elements:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MED-EL, a medical device company specializing in hearing implants. The logo is in red and features the company name in bold, sans-serif font. The "O" in MED-EL is replaced by a stylized graphic of a cochlea, which is a spiral-shaped part of the inner ear. The logo is simple, modern, and easily recognizable.
- -Audio processing unit with an electromagnetic transducer, generating vibration audible via bone conduction
- -Coupling of the transducer via an elastic band pressing against the skull bone
- -Battery powered using zinc-air batteries
- -Digital audio signal processing to achieve acoustic amplification
- -User controls
- -Configurable via software
- -Possibility to connect audio sources and ALDs (Assistive Listening Devices)
The following technological differences exist between the subject and predicate devices:
- -Single-unit audio processing unit with snap coupling (identical to BAHA 5 but different to contact mini)
- -Adhesive coupling option from the transducer to the skull bone
- -No coupling option to a surgically implanted abutment (as found in the BAHA 5 predicate only)
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### BIOCOMPATIBILITY TESTING
The performed biocompatibility evaluation concludes that all identified biological risks have been adequately addressed for the ADHEAR System and its accessories according to ISO 10993-1:2009/AC:2010 and the FDA ISO 10993-1 guidance (FDA-2013-D-0350). The performed evaluation provides objective evidence to support the conclusion that the ADHEAR System and its accessories can be considered biocompatible for its intended use.
## ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMCL
Electrical safety, EMC and EMI testing were conducted on the ADHEAR System. The ADHEAR system complies with the applicable sections of the following standards: IEC 60601-1:2005 +A1:2012, IEC 60601-1-11:2015, IEC 60601-1-6:2010 +A1:2013, IEC 62366:2007 +A1:2014, IEC 60601-1-2:2014, and ANSI C63.19:2011.
## MECHANICAL AND OTHER TESTING
Bench testing was successfully performed on the ADHEAR system in respect to the defined design specification requirements such as: Frequency range, OFL90, Maximum gain, Total harmonic distortion (THD), Equivalent input noise (EIN), Operating voltage, Current consumption, DAI connection, Signal processing features, Equivalent output noise (EON), Coupling distortion, Basic safety (general, EMC, usability, home healthcare environment), Removal force, Processor unit outer dimensions, Processor unit weight, Processor unit colors, Adhesive adapter colors, Batter door (battery size), Vibrator suspension, Tamper proof battery door, safety line, Push button operating force, Device marking, Drop test, Light exposure, Substance resistance, Power supply stability, Corrosion resistance, Change Battery, Switch device on/off, Adjust volume, change program, Change adhesive adapter, Place/Remove processor unit onto/from adhesive adapter, Push button longevity, DAI connector cycles, Battery door opening/closing cycles, and Protection during shipping
The collective results of the non-clinical testing demonstrate that the ADHEAR System meets the established specifications to ensure consistent and safe performance for its intended use.
{6}------------------------------------------------
### SOFTWARE VERIFICATION AND VALIDATION TESTING
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "minor" level of concern, since failures or latent flaws in the (optional) software are unlikely to result in any harm to the patient or operator.
### clinical TESTING
No clinical testing was needed to support the safety, performance, and substantial equivalence of the ADHEAR System to the predicate devices. All devices are bone conduction hearing aids and all aspects to ensure consistent and safe performance can be shown through non-clinical bench testing.
#### VIII. CONCLUSION
The non-clinical data provided support the safety of the device and the hardware verification and validation demonstrate that the ADHEAR System should perform as intended in the specified use conditions. The performance data has confirmed that the ADHEAR System is at least as safe and effective as the predicate devices, the Contact Mini and BAHA 5, and that the ADHEAR System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.