CMN Capillary Function with Oncology Virtual Expert
K260316 · Cercare Medical A/S · QIH · Jun 22, 2026 · Radiology
Device Facts
Record ID
K260316
Device Name
CMN Capillary Function with Oncology Virtual Expert
Applicant
Cercare Medical A/S
Product Code
QIH · Radiology
Decision Date
Jun 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI and CT. The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. Oncology Virtual Expert is intended for semi-automatic labeling, visualization, and volumetric quantification of brain glioma (WHO Grade II through IV), including glioblastoma (WHO Grade IV), astrocytoma (WHO Grade II, III, and IV), and oligodendroglioma (WHO Grade III), from a set of standard MRI images of adult patients who are known to have pathologically proven glioma. Volumetric measurements may be automatically compared to prior measurements when available; to minimize measurement variability, such longitudinal comparisons should be performed using the same scanner model and consistent acquisition parameters. Oncology Virtual Expert is not to be used for primary diagnosis and is intended to be used by trained professionals as an additional source of information for decision support.
Device Story
Software-only device (CMN) for medical image processing; runs on standard PC or virtual platform (VMware). Inputs: DICOM-compliant MRI/CT images (T1 pre/post-contrast, T2, FLAIR). Oncology Virtual Expert module performs semi-automatic labeling, visualization, and volumetric quantification of brain glioma lesions. Extracts perfusion data from Capillary Function module; enables longitudinal comparison of tumor volumes. Used by physicians/technicians in clinical settings for decision support; not for primary diagnosis. Output: segmented images, volumetric reports, and perfusion metrics. Benefits: assists clinicians in assessing pathological tissue volume and monitoring progression; provides standardized quantitative data to support clinical decision-making.
Clinical Evidence
Bench testing only. Comparative performance study using 100-patient subset (425 studies) from MU-Glioma Post dataset. Ground truth established via STAPLE-based consensus algorithm with three-tiered expert review (two board-certified neuroradiologists). Primary endpoints: Dice similarity coefficient for T1 enhancing and T2/FLAIR hyperintensity regions. Results: Subject device mean Dice 0.706 (T1) and 0.822 (T2/FLAIR), demonstrating equivalence to predicate. Subgroup analyses confirmed consistency across scanner manufacturers, field strengths, and tumor types.
Technological Characteristics
Software-only image processing platform. Operates on standard PC/virtualized environments. DICOM-compliant. Algorithms perform semi-automatic segmentation and volumetric quantification of MRI data. No hardware components. Sterilization/materials not applicable. Connectivity via standard hospital network/PACS. Developed per ISO 13485, ISO 14971, IEC 62304, and IEC 62366-1.
Indications for Use
Indicated for adult patients with pathologically proven brain glioma (WHO Grade II-IV), including glioblastoma, astrocytoma, and oligodendroglioma. Used for semi-automatic labeling, visualization, and volumetric quantification of lesions from standard MRI images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
CMN Capillary Function with Virtual Expert (K241989)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 22, 2026
Cercare Medical A/S
% Jennifer Willner
President
JW Regulatory Consulting, LLC
406 Wacouta St., Unit 417
Saint Paul, Minnesota 55101
Re: K260316
Trade/Device Name: CMN Capillary Function with Oncology Virtual Expert
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: May 14, 2026
Received: May 14, 2026
Dear Jennifer Willner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260316 - Jennifer Willner
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260316 - Jennifer Willner
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260316
Device Name
CMN Capillary Function with Oncology Virtual Expert
Indications for Use (Describe)
Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians.
The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI and CT.
The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data.
Oncology Virtual Expert is intended for semi-automatic labeling, visualization, and volumetric quantification of brain glioma (WHO Grade II through IV), including glioblastoma (WHO Grade IV), astrocytoma (WHO Grade II, III, and IV), and oligodendroglioma (WHO Grade III), from a set of standard MRI images of adult patients who are known to have pathologically proven glioma. Volumetric measurements may be automatically compared to prior measurements when available; to minimize measurement variability, such longitudinal comparisons should be performed using the same scanner model and consistent acquisition parameters. Oncology Virtual Expert is not to be used for primary diagnosis and is intended to be used by trained professionals as an additional source of information for decision support.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Cercare Medical
# **510(K) SUMMARY**
510(k) Number: K260316
Date Prepared: June 17, 2026
**Table 1: Submitter Information**
| Manufacturer: Cercare Medical A/S Inge Lehmanns Gade 10 DK-8000 Aarhus C Denmark US FDA ERN: 3019844085 | Manufacturer's Contact Person: Mikkel Bo Hansen Chief Scientific Officer Phone: +45 22890125 Email: mbh@cercare-medical.com |
| --- | --- |
**Table 2: Device Information**
| **Trade Name** | CMN Capillary Function with Oncology Virtual Expert |
| --- | --- |
| **Common Name** | CMN Capillary Function with Oncology Virtual Expert |
| **Classification Name** | Medical Image Management and Processing System |
| **Regulation** | 21 CFR 892.2050 |
| **Product Code** | QIH & LLZ |
| **Regulatory Classification:** | Class II |
| **Device Panel:** | Radiology (OHT8) |
CMN Capillary Function with Oncology Virtual Expert is substantially equivalent to the NS-HGlio Predicate Device (**Table 3**) and its base platform (CMN Capillary Function) is identical to the base platform of the CMN Capillary Function with Virtual Expert Reference Device (**Table 4**). Neither of these have been subject to a design-related recall.
**Table 3: Predicate Devices**
| Predicate Device | Manufacturer | FDA 510(k) |
| --- | --- | --- |
| NS-HGlio | Neosoma, Inc. | K221738 |
**Table 4: Reference Device**
| Predicate Device | Manufacturer | FDA 510(k) |
| --- | --- | --- |
| CMN Capillary Function with Virtual Expert | Cercare Medical A/S | K241989 |
# **Device Description**
Cercare Medical Neurosuite is a software-only device designed to streamline medical image processing by providing for the visualization and study of medical images. CMN can be installed on a customer PC, or it can be accessed remotely using remote desktop technologies. CMN provides viewing, quantification, analysis and reporting capabilities. CMN is not intended as a dedicated PACS system for long term persistent storage of patient data.
Cercare Medical A/S
June 2026
{5}
Cercare Medical
CMN is software that is intended for use by trained professionals, including physicians and medical technicians. The software provides cerebral image processing capabilities for providing important clinical markers in time-critical management of suspected acute ischemic stroke patients. CMN is intended to be used as decision support software only and the clinician continues to provide all treatment decisions.
The software is intended to visualize and study neuroimaging by image viewing and registration of medical images. CMN works with MRI (Magnetic Resonance Image) and CT (computed tomography) technologies.
CMN accepts and produces data sets in the DICOM format. DICOM is a standard format for storing and transmitting medical image data in vendor neutral format and is managed by the DICOM Standards Committee. CMN also supports DICOM Structured Reporting (DICOM SR) for documenting quantitative results of study analyses.
CMN is a platform that allows for the addition of certain modules for further analysis, including Capillary Function and Oncology Virtual Expert modules. The CMN platform and Capillary Function module are identical to the Reference Device (K241989) while the Oncology Virtual Expert module is the subject of this premarket application.
#### Oncology Virtual Expert
Oncology Virtual Expert, when added to the installed CMN Capillary Function, provides further functionalities for reading, writing, visualizing and other functionalities for studying medical images.
Oncology Virtual Expert provides post-processing technologies that allow for delineation and volumetric quantification of perceived pathological tissue that may be related to neuro oncology, based on primary acquired MRI data including T1 images acquired pre- and post-contrast and T2FLAIR images. The output provides information that can be used to assess the volume of pathological tissue for a patient.
Oncology Virtual Expert also allows for extraction of information from other images within the segmented tumor lesions, such as the mean value and standard deviation for, e.g., perfusion images obtained through the Capillary Function module. This information, as well as volumetric information, can be generated as aggregated views across multiple imaging studies for a given patient.
The output maps generated can be viewed by standard DICOM image viewers. The Oncology Virtual Expert works with the following technologies:
- MRI (Magnetic Resonance Image)
Cercare Medical A/S
June 2026
{6}
Cercare Medical
## Indications for Use
Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians.
The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI and CT.
The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data.
Oncology Virtual Expert is intended for semi-automatic labeling, visualization, and volumetric quantification of brain glioma (WHO Grade II through IV), including glioblastoma (WHO Grade IV), astrocytoma (WHO Grade II, III, and IV), and oligodendroglioma (WHO Grade III), from a set of standard MRI images of adult patients who are known to have pathologically proven glioma. Volumetric measurements may be automatically compared to prior measurements when available; to minimize measurement variability, such longitudinal comparisons should be performed using the same scanner model and consistent acquisition parameters. Oncology Virtual Expert is not to be used for primary diagnosis and is intended to be used by trained professionals as an additional source of information for decision support.
## Technological Characteristics
CMN Capillary Function with Oncology Virtual Expert is image processing software for cerebral image processing, capable of generating regions of interest for tissue potentially affected by neuro oncology. The image processing software generates maps of assessed neuro oncology, intended to be used during management of brain tumors as decision support.
It is emphasized that the output from the software should not be used without close scrutiny of the relevance to the patient in question, including all available evidence and quality checking of the appropriateness of the output.
All result images generated by any module (Capillary Function or Oncology Virtual Expert) are handled and stored by the CMN, which can expose the data to the end user via the associated graphical user interface or via network communication to other hospital IT systems such as local picture archive and communication systems (PACS).
Cercare Medical A/S
June 2026
{7}
Cercare Medical
The Oncology Virtual Expert constitutes the following list of primary functionalities:
- Delineation of neuro oncology lesions, including imaging-based tissue classifications: T2/FLAIR hyperintensity and T1 enhancing tissue (US market); T2/FLAIR hyperintensity, T1 enhancing tissue, and necrotic tissue (non-US markets)
- Volumetric analysis of the individual tissue classifications
- Longitudinal comparison of tumor volumes if more than one MRI study is available
- Automatic extraction of perfusion data from each lesion type
- Automatic reporting of results in the DICOM format
The Predicate Device provides equivalent estimates of the primary clinically relevant features offered by the Oncology Virtual Expert, concerning
- Delineation and volumetric analysis of contrast enhancing region
- Delineation and volumetric analysis of edematous tissue
- Collection and reporting of volumetric data across time if more studies exist for the same patient
- Computation of perfusion metrics within the different tissue classes
Note that the first three items are all part of both the Subject Device and Predicate Device, whereas the latter is a unique data collection and reporting feature for the Subject Device.
*Differences in Technical Characteristics/Performance with Respect to the Predicate Device*
CMN Capillary Function with Oncology Virtual Expert and the Predicate Device (K221738) both apply to the field of neuro oncology (brain cancer) and are indicated for a subset of primary brain tumors, specifically brain glioma (WHO Grade II through IV).
The Subject Device is identical to the Predicate Device with the exception of the following features:
- The Subject Device does not rely on off-premises resources to perform operations, which is in contrast to the cloud-based mode of operation required by the Predicate Device
The Subject and Predicate devices are based on the same technological elements of viewing, processing and analyzing DICOM image data to assist the clinician during diagnostic procedures. These differences do not impact the intended use or raise new questions with the safety and performance of the device.
**Performance Standards**
CMN Capillary Function with Oncology Virtual Expert has been developed in conformance with the following standards and FDA guidance, as applicable:
- ISO 13485:2016, Quality management systems – Requirements for regulatory purposes
- ISO 14971:2019, Medical devices – Application of risk management to medical devices
Cercare Medical A/S
June 2026
{8}
Cercare Medical
- IEC 62366-1:2015, Medical devices – Part 1: Application of usability engineering to medical devices
- IEC 62304:2015, Medical device software – Software lifecycle processes
- IEC 82304-1:2016, Health Software Part 1: General Requirements for Product Safety
- ISO 15223-1:2021, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements
- ISO 20417:2021, Medical devices – Information to be supplied by the manufacturer
- NEMA PS 3.1-3.20:2024, Digital Imaging and Communications in Medicine (DICOM)
- Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data – Premarket Notification [510(k)] Submissions, September 2022
Performance Data
CMN Capillary Function with Oncology Virtual Expert complies with DICOM (Digital Imaging and Communications in Medicine), developed by the American College of Radiology and the National Electrical Manufacturers Association – NEMA PS 3.1-3.20.
Cercare conducted extensive performance validation testing and software verification and validation testing of the CMN Capillary Function with Oncology Virtual Expert. This performance validation testing demonstrated that the CMN system provides accurate representation of key processing parameters under a range of clinically relevant parameters associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the CMN system met all design requirements and specifications.
Comparative performance testing was performed for the Subject Device with respect to the Predicate Device using a 100-patient subset (425 studies) of the MU-Glioma Post dataset (University of Missouri). Both devices were independently compared against the same expert manual segmentation ground truth, established through a STAPLE-based consensus algorithm with three-tiered expert review including final consensus approval by two board-certified neuroradiologists. The comparison study population comprised: glioblastoma (n=73), astrocytoma (n=16), diffuse glioma (n=8), oligodendroglioma (n=2), and pilocytic astrocytoma (n=1), acquired predominantly on Siemens MRI systems at 1.5T and 3.0T field strengths. Testing the Predicate and Subject devices demonstrated equivalent estimates of the primary clinically relevant features offered by Oncology Virtual Expert, concerning
- Delineation and volumetric analysis of T1 enhancing region (mean Dice similarity coefficient: Subject Device 0.706, Predicate Device comparable)
- Delineation and volumetric analysis of T2/FLAIR hyperintensity region (mean Dice similarity coefficient: Subject Device 0.822, Predicate Device comparable)
Cercare Medical A/S
June 2026
{9}
Cercare Medical
- Collection and reporting of volumetric data across time if more studies exist for the same patient and quantification of imaging-based progression assessment
Subgroup analyses demonstrated consistent performance across scanner manufacturers (Siemens, GE, Philips), field strengths (1.5T and 3.0T), tumor types, and lesion size categories. Performance increased with lesion volume, as expected from the properties of the Dice similarity metric. Device labeling instructs users that longitudinal volumetric comparisons should be performed using the same scanner model and consistent acquisition parameters to minimize measurement variability.
Together with software verification and validation, the performance validation demonstrated that CMN Capillary Function with Oncology Virtual Expert satisfies all design requirements and device specifications and is substantially equivalent to the Predicate Device.
### **Substantial Equivalence**
In comparison with the Predicate NS-HGlio device (K221738) and Reference CMN Capillary Function with Virtual Expert device (K241989), CMN Capillary Function with Oncology Virtual Expert has the same intended use and similar indications, technological characteristics and principles of operation as described in the comparison table below.
Cercare Medical A/S
June 2026
{10}
**Cercare Medical**
**Table 5: Substantial Equivalence Comparison Table**
| Description | Subject Device | Predicate Device (K221738) | Reference Device (K241989) | Conclusion |
| --- | --- | --- | --- | --- |
| **Product Name** | CMN Capillary Function with Oncology Virtual Expert | NS-HGlio | CMN Capillary Function with Virtual Expert | |
| **Manufacturer** | Cercare Medical A/S | Neosoma, Inc. | Cercare Medical A/S | |
| **Product Code / Regulation** | QIH & LLZ / 21CFR 892.2050 | QIH / 21CFR 892.2050 | LLZ / 21CFR 892.2050 | Same |
| **Indications for Use** | Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI and CT. The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This | NS-HGlio is intended for the semi-automatic labeling, visualization, and volumetric quantification of high-grade brain glioma (WHO grade 3 astrocytoma, WHO grade 4 astrocytoma and WHO grade 4 glioblastoma) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven high-grade glioma. Volumetric measurements may be compared to past measurements if available. NS-HGlio is not to be used for primary diagnosis, and is intended to be used by qualified clinical personnel as an additional source of information and is not intended to be the sole diagnostic metric. | Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI and CT. The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic | Same – SD combines indications of both the PD and RD with the exception of the Virtual Expert portion of the RD; No new questions of safety or effectiveness. |
Cercare Medical A/S
June 2026
{11}
Cercare Medical
| Description | Subject Device | Predicate Device (K221738) | Reference Device (K241989) | Conclusion |
| --- | --- | --- | --- | --- |
| | functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. Oncology Virtual Expert is intended for semi-automatic labeling, visualization, and volumetric quantification of brain glioma (WHO Grade II through IV), including glioblastoma (WHO Grade IV), astrocytoma (WHO Grade II, III, and IV), and oligodendroglioma (WHO Grade III), from a set of standard MRI images of adult patients who are known to have pathologically proven glioma. Volumetric measurements may be automatically compared to prior measurements when available; to minimize measurement variability, such longitudinal comparisons should be performed using the same scanner model and consistent acquisition parameters. Oncology Virtual Expert is not to be used for primary diagnosis and is intended to be used by trained professionals as an additional source of information for decision support. | | imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: • Lower signal intensity for CBF and CBV • Higher signal intensity for TTP, MTT, and Tmax | |
| PACS Functionality | | | | |
Cercare Medical A/S
June 2026
{12}
Cercare Medical
| Description | Subject Device | Predicate Device (K221738) | Reference Device (K241989) | Conclusion |
| --- | --- | --- | --- | --- |
| Basic PACS Functions | View, process and analyze medical images. Communication of results through service class user protocols. | Same | Same | Same |
| Computer Platform | Standard off-the-shelf PC workstation / server | Cloud-based platform only | Same | Same as RD; Difference with PD raises no new questions of safety or effectiveness |
| | Virtual platform such as VMware | | Same | |
| DICOM Compliance | Yes | Yes | Yes | Same |
| Features | | | | |
| Image Types | MRI: Acquired using four different MRI sequences either in 2D or 3D using a specified protocol: • T1 pre-contrast • T1 post-contrast • T2 (optional) • FLAIR | MRI: Acquired using four different MRI sequences either in 2D or 3D using a specified protocol: • T1 pre-contrast • T1 post-contrast • T2 • FLAIR | Same | Same |
| Intended Anatomy | Brain | Brain | Same | Same |
| Lesion Review | 2D and 3D | 2D and 3D | Same | Same |
| Segmentation | Semi-automatic and manual segmentation of brain glioma (WHO Grade II–IV) | Semi-automatic and manual segmentation of high-grade glioma | N/A | Same |
Cercare Medical A/S
June 2026
{13}
Cercare Medical
| Description | Subject Device | Predicate Device (K221738) | Reference Device (K241989) | Conclusion |
| --- | --- | --- | --- | --- |
| **Quantification** | Volumetric measurement of the sub-components of brain glioma (WHO Grade II–IV) | Volumetric measurement of the sub-components of high-grade glioma | N/A | Same |
| **Volumetric Output** | - Provides volumetric measurements of brain glioma (WHO Grade II–IV) and sub-components - Includes segmented color overlays of sub-components and reports - Automatically compares results to prior scans when available | - Provides volumetric measurements of high-grade brain glioma and sub-components - Includes segmented color overlays of sub-components and reports - Automatically compares results to prior scans when available | N/A | Same |
| **Other Output** | Summary image characteristics of other images within the glioma sub-components | Not available | N/A | Slight difference raises no new questions of safety or effectiveness |
| **Report** | Yes | Yes | Yes | Same |
Cercare Medical A/S
June 2026
{14}
Cercare Medical
## Conclusions
The CMN Capillary Function with Oncology Virtual Expert device performs as intended and presents no unacceptable risks to the intended patient population. The non-clinical bench data supports the safety of the device and demonstrates that the CMN Capillary Function with Oncology Virtual Expert device performs as intended. The CMN Capillary Function with Oncology Virtual Expert is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the legally marketed Predicate Device (K221738) with comparable characteristics to the Reference Device (K241989).
Cercare Medical A/S
June 2026
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.