Mean DSC 0.88 (95% CI 0.86-0.90) on preoperative imaging; Mean DSC 0.80 (95% CI 0.77-0.83) on postoperative imaging; Mean ICC 0.98 (95% CI 0.97-0.99)
—
—
33 subjects and 132 MRIs
3 (US board certified neuroradiologists)
Indications for Use
NS-HGlio is intended for the semi-automatic labeling, visualization, and volumetric quantification of high-grade brain glioma (WHO grade 3 astrocytoma, WHO grade 4 astrocytoma and WHO grade 4 glioblastoma) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven high-grade glioma. Volumetric measurements may be compared to past measurements if available. NS-HGlio is not to be used for primary diagnosis, and is intended to be used by qualified clinical personnel as an additional source of information and is not intended to be the sole diagnostic metric.
Device Story
NS-HGlio is a SaMD tool for high-grade brain glioma analysis. Input: DICOM MRI images (T1 pre-contrast, T1 post-contrast, T2, FLAIR). Operation: Deep learning methodology semi-automatically labels tumor subcomponents; calculates volumetric measurements. Output: Segmented color overlays, volumetric reports, longitudinal comparison to prior scans. Usage: Clinical setting by qualified personnel; integrates with viewing software or PACS. Benefit: Provides quantitative data to assist clinicians in assessing disease condition pre- and post-operatively; serves as an additional information source for clinical decision-making.
Clinical Evidence
Bench testing using 132 MRIs from 33 subjects (diverse demographics, Siemens/GE/Toshiba scanners). Ground truth established by three board-certified neuroradiologists. Primary endpoints: Dice Similarity Coefficient (DSC) and Intraclass Correlation Coefficient (ICC). Results: Mean DSC 0.88 (pre-op) and 0.80 (post-op), exceeding expert mean DSC (0.84 pre-op, 0.74 post-op). Mean ICC 0.98 (95% CI 0.97-0.99).
Technological Characteristics
SaMD; deep learning-based image processing. Inputs: 2D/3D MRI (T1 pre/post-contrast, T2, FLAIR). Outputs: Volumetric quantification, segmented color overlays, reports. Connectivity: DICOM-compliant; integrates with PACS/viewing software. Software lifecycle: IEC 62304:2006/AC:2015 compliant.
Indications for Use
Indicated for semi-automatic labeling, visualization, and volumetric quantification of high-grade brain glioma (WHO grade 3/4 astrocytoma, WHO grade 4 glioblastoma) in patients 18+ years old with pathologically proven high-grade glioma. Not for primary diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
September 27, 2022
Neosoma, Inc. % Aly Abayazeed Co-founder and Chief Medical Officer 44 Farmers Row GROTON MA 01450
Re: K221738
Trade/Device Name: NS-HGlio Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: September 1, 2022 Received: September 2, 2022
Dear Aly Abayazeed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel M. Krainak, Ph.D.
Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221738
Device Name NS-HGlio
#### Indications for Use (Describe)
NS-HGlio is intended for the semi-automatic labeling, visualization, and volumetric quantification of high-grade brain glioma (WHO grade 3 astrocytoma, WHO grade 4 astrocytoma and WHO grade 4 glioblastoma) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven high-grade glioma. Volumetric measurements may be compared to past measurements if available. NS-HGlio is not to be used for primary diagnosis, and is intended to be used by qualified clinical personnel as an additional source of information and is not intended to be the sole diagnostic metric.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------|-----------------------------------------------------------------|
| <div> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### 1.1 General Information
## K221738
| 510(k) Sponsor | Neosoma, Inc. |
|--------------------------|---------------------------------------------------------------|
| Address | 44 Farmers Row<br>Groton, MA 01450 |
| Correspondence<br>Person | Aly H. Abayazeed, MD.<br>Co-founder and Chief Medical Officer |
| Contact Information | aly.abayazeed@neosomainc.com<br>443-804-8096 |
| Date Prepared | June 14, 2022 |
#### 1.2 Proposed Device
| Proprietary Name | NS-HGlio |
|---------------------|--------------------------------------------------|
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Regulatory Class | II |
## 1.3 Predicate Device
| Proprietary Name | NeuroQuant |
|------------------------|--------------------------------------------------|
| Premarket Notification | K170981 |
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | LLZ |
| Regulatory Class | II |
## 1.4 Device Description
NS-HGlio is a non-invasive software as a medical device (SaMD) tool intended for labeling, visualization, and volumetric quantification of high-grade brain gliomas for a population that has been pathologically diagnosed to have brain tumors. The device is used
{4}------------------------------------------------
as a tool by clinicians in determining the patient's disease conditions on pre- and post-operative MRI images. The device is not used for primary diagnosis.
NS-HGlio device takes as an input imported Digital Imaging and Communications in Medicine (DICOM) images of high-grade brain glioma acquired with standard brain tumor MRI protocols and uses a deep learning methodology to semi-automatically label the different subcomponents of the high-grade glioma. Results are displayed on a Neosoma viewing software. Optionally, the software connects to clinicians' applications (e.g., PACS).
# 1.5 Indications for Use
NS-HGlio is intended for the semi-automatic labeling, visualization, and volumetric quantification of high-grade brain glioma (WHO grade 3 astrocytoma, WHO grade 4 astrocytoma and WHO grade 4 glioblastoma) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven high-grade glioma. Volumetric measurements may be compared to past measurements if available. NS-HGlio is not to be used for primary diagnosis, and is intended to be used by qualified clinical personnel as an additional source of information and is not intended to be the sole diagnostic metric.
| Feature/<br>Function | Subject Device:<br>NS-HGlio | Predicate Device<br>NeuroQuant manufactured by<br>CorTechs Labs<br>K170981 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Scans | MRI:<br>Acquired using four different MRI<br>sequences either in 2D or 3D using<br>a specified protocol<br>(T1 pre-contrast, T1 post-contrast,<br>T2 and FLAIR) | MRI:<br>Neuroquant: 3D T1 pre-contrast scans<br>acquired with specified protocols<br>Lesionquant: 3D FLAIR scan acquired<br>with specified protocol |
| Intended<br>Anatomy | Brain | Brain |
| Lesion Review | 2D and 3D | 2D |
# 1.6 Comparison of Technological Characteristics with the Predicate Device
{5}------------------------------------------------
| Feature/<br>Function | Subject Device:<br>NS-HGlio | Predicate Device<br>NeuroQuant manufactured by<br>CorTechs Labs<br>K170981 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Segmentation | Semi-automatic and manual<br>segmentation of high-grade glioma | Semi-automatic and manual<br>segmentation of brain structures |
| Quantification | Volumetric measurement of the<br>sub-components of high-grade<br>glioma | Automated measurement of brain<br>tissue volumes and structures and<br>lesions |
| Output | - Provides volumetric<br>measurements of high-grade<br>brain glioma and<br>sub-components<br>- Includes segmented color<br>overlays of sub-components and<br>reports<br>- Automatically compares results<br>to prior scans when available | - Provides volumetric measurements<br>of brain structures and lesions<br>- Includes segmented color overlays<br>and morphometric reports<br>- Automatically compares results to<br>reference percentile data and to prior<br>scans when available |
| Image Format | DICOM | DICOM |
| Report | YES | YES |
# 1.7 Performance Data
Safety and performance of NS-HGlio have been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/AC:2015 - Medical device software -Software life cycle processes, in addition to the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
{6}------------------------------------------------
The testing dataset consisted of 33 subjects and 132 MRIs used for the evaluation of the machine learning model performance. The test dataset was acquired from medical sites that were not included in the training dataset to ensure device generalizability. The data were acquired using standard of care MRI protocols on Siemens, GE, and Toshiba scanners. Following the real world prevalence of the high grade glioma, the data consisted of more males than females within the age range of 18 to 79 and covering a diverse group of ethnic backgrounds.
The reference standard (ground truth) was established using three US board certified neuroradiologists with expertise in measuring high grade gliomas. The dataset was evaluated using the DSC (Dice Similarity Coefficient) assessing the degree of overlap between device output and the reference standard as well as the Intraclass correlation coefficient (ICC) of the device output volumes and the reference standard.
The device achieved a mean DSC of 0.88 with 95% CI of 0.86-0.90 on preoperative imaging and 0.80 with 95% CI of 0.77-0.83 on postoperative imaging which is higher than the mean DSC of the average of the three experts for the same task, which was 0.84 on preoperative imaging and 0.74 for postoperative imaging respectively. The mean ICC was 0.98 with 95% CI of 0.97-0.99.
## 1.8 Conclusion
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, and performance testing, NS-HGlio raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.