VITA Multi-Function Head Brush (TB-2343F, TB-2442F)
K260179 · Touchbeauty Beauty & Health (Shenzhen) Co., Ltd. · NFO · Jun 3, 2026 · Neurology
Device Facts
Record ID
K260179
Device Name
VITA Multi-Function Head Brush (TB-2343F, TB-2442F)
Applicant
Touchbeauty Beauty & Health (Shenzhen) Co., Ltd.
Product Code
NFO · Neurology
Decision Date
Jun 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device. Micro-current function: The micro current stimulation mode is intended for facial stimulation and is indicated for over-the counter cosmetic use.
Device Story
Hand-held device; consists of host, removable treatment head with stainless steel bristles, and charging cable. Device generates micro-currents via stainless steel bristles applied to facial skin. User operates via button for power and mode switching; device features automatic 10-minute shut-off. Powered by internal rechargeable lithium battery. Output is transcutaneous micro-current electrical stimulation for aesthetic purposes. Healthcare provider does not operate; intended for patient self-use in home environment. Output provides cosmetic facial stimulation; intended to mimic effects of predicate aesthetic stimulators.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-11, -2-10), EMC (IEC 60601-1-2), and battery safety (IEC 62133-2). Software verification and validation confirmed mitigation of hazards to acceptable levels.
Technological Characteristics
Hand-held, battery-powered stimulator. Enclosure: ABS; Bristles: Stainless steel. Output: Monophasic pulsed micro-current (1.6kHz). Safety: Type BF protection. Biocompatibility: ISO 10993 compliant. Connectivity: None. Sterilization: Not applicable. Software: Microprocessor-controlled with 10-minute timer.
Indications for Use
Indicated for over-the-counter cosmetic facial stimulation in adult users.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 3, 2026
Touchbeauty Beauty & Health (Shenzhen) Co., Ltd.
% Bing Huang
RA Engineer
Feiying Drug & Medical Consulting Technical Service Group
Rm.2401 Zhenye International Business Center, # 3101-90
Qianhai Rd.
Shenzhen, Guangdong 518052
China
Re: K260179
Trade/Device Name: VITA Multi-Function Head Brush (TB-2343F, TB-2442F)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: May 8, 2026
Received: May 8, 2026
Dear Bing Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20903
www.fda.gov
{1}
K260179 - Bing Huang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260179 - Bing Huang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260179 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| VITA Multi-Function Head Brush (TB-2343F, TB-2442F) | | |
| Please provide your Indications for Use below. | | ? |
| VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device. Micro-current function: The micro current stimulation mode is intended for facial stimulation and is indicated for over-the counter cosmetic use. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
# 510(k) Summary
**Date: 2026-06-03**
### I. Submitter
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD. 29/F, Block B, Tiley Central Plaza, No.195 Haide 3rd Road, Nanshan District, Shenzhen, China.
Post code: 518064
Tel.: +86 400-1191-180
Zhou Elin
Title: CEO
Tel.: +86 755 3366 2222 ext.666
Email: chow@touchbeauty.com
### II. Device
Name of Device: VITA Multi-Function Head Brush
Model(s): TB-2343F, TB-2442F
Common or Usual Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes
Classification Name: Transcutaneous electrical nerve stimulator for pain relief
Regulatory Class: II
Product Code: NFO
Regulation Number: 21CRF 882.5890
**Note:** The device function-under-review in this 510(k) is transcutaneous microcurrent electrical stimulation for aesthetic purposes, which falls under the NFO product code under the regulation 21 CRF 882.5890. Only this specific device function was cleared in this 510(k).
### III. Predicate Device
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| XTREEM PULSE LLC | PureLift Pro Plus | K221443 | October 21, 2022 |
| BELEGA Co., Ltd. | Beagank 4T Plus | K233010 | November 21, 2023 |
| Carol Cole Company | NuFACE Mini Device | K133823 | Apr 17, 2014 |
### IV. Device Description
The VITA Multi-Function Head Brush is a hand-held device. It consists of a host, a removable treatment head, and a charging cable. The host has a button, indicator lights, and the device's power-
Page 1 of 11
{5}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
on/off, treatment mode switch can be achieved by pressing the button. The removable treatment head has lights row, stainless steel bristles.
The device generates a micro-currents on the stainless steel bristles which will applied on the facial skin for facial stimulation.
The device will automatically shut down after each mode completes treatment within 10 minutes. The device is powered by a built-in rechargeable lithium battery.
## V. Indications for Use
VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device.
Micro-current function: The micro current stimulation mode is intended for facial stimulation and is indicated for over-the counter cosmetic use.
## VI. Comparison of Technological Characteristics With the Predicate Device
VITA Multi-Function Head Brush is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance:
Page 2 of 11
{6}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| **Comparison Elements** | **Subject Device** | **Primary Predicate Device** | **Predicate Device 2** | **Predicate Device 3** | **Remark** |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | Pending | K221443 | K233010 | K133823 | / |
| Trade name | VITA Multi-Function Head Brush (Model(s): TB-2343F, TB-2442F) | PureLift Pro Plus | Beagank 4T Plus | NuFACE Mini Device | / |
| Manufacturer | TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD | XTREEM PULSE LLC | BELEGA Co., Ltd. | Carol Cole Company | / |
| Regulation number | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | Same |
| Product code | NFO | NFO | NFO | NFO | Same |
| Device classification | II | II | II | II | Same |
| Prescription or OTC | OTC | OTC | OTC | OTC | Same |
| Indication for use/ Intended use | VITA Multi-Function Head Brush is an Over-the-Counter (OTC) device. Face Mode: Micro-current function: The micro current stimulation mode is intended for facial stimulation and is indicated for over-the counter cosmetic use. | Intended for facial stimulation and indicated for over-the-counter cosmetic use. | The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation. | The NuFACE Mini Facial Toning device is intended for facial stimulation and is indicated oor over-the-counter cosmetic use (21 CFR 807 Subpart C). | Same |
| Handheld or stationary | Hand-held Type | Hand-held Type | Hand-held Type | Hand-held Type | Same |
| Treatment area | Face | Face | Face and neck or body | Face | Same |
| Number of out channels | 1 output channel | 1 output channel | 1 | 1 | Same |
Page 3 of 11
{7}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| **Comparison Elements** | **Subject Device** | **Primary Predicate Device** | **Predicate Device 2** | **Predicate Device 3** | **Remark** |
| --- | --- | --- | --- | --- | --- |
| Regulated current or regulated voltage | Voltage | Regulated current | Voltage | Both | Same |
| Software/ Firmware/ Microprocessor Control | Yes | Yes | Yes | Yes | Same |
| Indicator Display | Yes | Yes | Unknown | Yes | Same |
| Timer range | 10 minutes a time | 10 minutes only | Maximum 5 minutes | 21 minutes | Same as the primary |
| Type of protection | Type BF | Type BF | Type BF | Unknown | Same |
| On/off status | Yes | Yes | Yes | Yes | Same |
| Waveform | Pulses Monophasic | Pulses Monophasic, alternating polarity | Rectangle, biphasic asymmetric | Pulsed Monophasic | Same The waveform of subject device is the same as predicate device 3. |
| Shape | Rectangular Pulses | Rectangular Pulses | Rectangle, biphasic asymmetric | Modulated Square | Same |
| Maximum output voltage | 20Vpp @500Ω 32Vpp @ 2kΩ 44Vpp @ 10kΩ | 20Vpp(@500Ω) 32Vpp(@2kΩ) 44Vpp(@10kΩ) | Mode 1: 127mV (0.12V)@500Ω 515mV (0.51V)@2KΩ 1.83V@10KΩ | Specify units: 222mV @500Ω 781mV @ 2kΩ 3.90V @ 10kΩ | Same |
Page 4 of 11
{8}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 2 | Predicate Device 3 | Remark |
| --- | --- | --- | --- | --- | --- |
| | | | Mode 2: 110mV(0.11V)@500Ω 507mV(0.50V)@2KΩ 1.87V@10KΩ Mode 3: 16.6V@500Ω 21.0V@2KΩ 22.5V@10KΩ Mode 4: 15.3V@500Ω 21.2V@2KΩ 22.9V@10KΩ | | |
| Maximum output current | 2.96mA_{r.m.s} @500Ω 1.35mA_{r.m.s} @ 2kΩ 0.515mA_{r.m.s}@10kΩ | 9mA(@500Ω) 4.4mA(@2kΩ) 1.2mA(@10kΩ) | Mode 1: 0.19mA@500Ω 0.18mA@2KΩ 0.16mA@10KΩ Mode 2: 0.45mA@500Ω 0.39mA@2KΩ 0.24mA@10KΩ Mode 3: | 396μA @500Ω 395μA@ 2kΩ 391μA@ 10kΩ | Similar Note 1 |
Page 5 of 11
{9}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 2 | Predicate Device 3 | Remark |
| --- | --- | --- | --- | --- | --- |
| | | | 30.8mA@500Ω 8.52mA@2KΩ 2.26mA@10KΩ Mode 4: 31.4mA@500Ω 8.60mA@2KΩ 2.15mA@10KΩ | | |
| Pulse width | 4μs | 4μs | Mode 1: 130μs Mode 2: 130μs Mode 3: 80μs Mode 4: 92μs | Specify unites ON Phase: 60.4ms OFF phase: 60.4ms Total Pulse Width: 120.8ms | Same |
| Frequency (Hz) | 1.6kHz | 1.37kHz ~ 1.73kHz | Mode 1: 3.80kHz Mode 2: 3.80kHz Mode 3: 3.65kHz Mode 4: 1.64kHz | 8.28Hz | **Similar Note 2** |
| Symmetrical phases | Not multiphasic | Not multiphasic | No | Not Multiphasic | Same |
| Net charge (μC per pulse train) | 0.16μC | 0μC per pulse train | Mode 1: 0.025μC@500Ω Mode 2: 0.025μC@500Ω Mode 3: -1.3μC@500Ω Mode 4: 0μC@500Ω | 1.43μC | **Different Note 3** |
Page 6 of 11
{10}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 2 | Predicate Device 3 | Remark |
| --- | --- | --- | --- | --- | --- |
| Maximum phase charge (μC) | 0.2304μC | 5.81μC@500Ω | Mode1: 0.059μC@500Ω Mode 2: 0.025μC@500Ω Mode 3: 2.2μC@500Ω Mode 4: 0.29μC@500Ω | 23.7μC | Different Note 3 |
| Maximum current density (mA/cm²) | 1.62mA/cm² | 8.8mA/cm²@500Ω | With a wet cotton pad (Purified water) Mode 1: 0.15~1.36mA/cm²@500Ω Mode 2: 0.10~0.92mA/cm²@500Ω Mode 3: 4.26~38.43mA/cm²@500Ω Mode 4: 3.13~28.21mA/cm²@500Ω | 0.514mA/cm² | Similar Note 4 |
| Maximum power density | 2394μW/cm² | 39600μW/cm² | With a wet cotton pad: Mode 1: 11.39μW/cm²@500Ω Mode 2: 33.87μW/cm²@500Ω Mode 3: 1124.0μW/cm²@500Ω Mode 4: 556.96μW/cm²@500Ω | 1981μW/cm² | Different Note 4 |
Page 7 of 11
{11}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 2 | Predicate Device 3 | Remark |
| --- | --- | --- | --- | --- | --- |
| Main Materials | Enclosure: ABS; Stainless Steel Bristles Button: Stainless; | Not publicly available | ABS | Thermo Plastic | Different, but solved by biocompatibility tests |
| Electrical safety | IEC 60601-1 IEC 60601-1-11 IEC 60601-1-2 IEC 60601-2-10 IEC 62133-2 | ANSI/AAMI ES60601-1 IEC 60601-1-2 IEC 60601-2-10 IEC 60601-1-11 | IEC 60601-1 IEC 60601-1-2 | IEC 60601-1 IEC 60601-1-2 | Same |
| Biocompatibility feature | All patient contacting materials are complied with ISO 10993-5, ISO 10993-10 and ISO 10993-23 | ISO 10993-5 ISO 10993-10 | ISO 10993-5 ISO 10993-10 ISO 10993-23 | Not publicly available | Same |
# Note 1:
The maximum output voltage of the subject device is same as the primary predicate device(K221443). The maximum output current seems to be different from the predicate devices, but it is within the range of the predicate devices. Besides, the device has been tested for electrical safety as per IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 requirements, so this difference does not impact its safety and effectiveness.
# Note 2:
Though the Frequency of the subject device is a little different from the primary predicate device, it is within the rang of the primary predicate device. So this difference will not raise any safety or effectiveness issues.
# Note 3:
Though Maximum phase charge of the subject device is a little different from the primary predicate device, it is within the rang of the primary predicate device and secondary predicate device. So this difference will not raise any safety or effectiveness issues.
The waveform of the subject device is the same as the predicate device 3, both are Pulsed Monophasic. And the net charge (0.16µC) of the subject device, which is much lower than the requirement in clause 8.4.3 of IEC 60601-1 that the shored charge should not exceed 45µC, is much smaller than that of the predicate device 3 (1.43µC), as well as that of the mode 3 of the predicate device 2 (-1.3µC). Besides, the subject device has been tested for electrical safety as per IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 requirements, so this difference does not impact its safety and
Page 8 of 11
{12}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary---
effectiveness. Therefore, the waveform and net charge of the subject device will not increase the risk of charge accumulation at the electrode-skin interface.
# Note 4:
Though the Maximum current density and Maximum power density of the subject device is a little different from the primary predicate device, it is within the range of the primary predicate device and secondary predicate device, and the 1.62mA/cm² is less than the limit of 2mA/cm² specified in IEC 60601-2-10. The subject device has provided adequate electrical safety and performance testings, which were conducted according to applicable standards for nerve and muscle stimulators (including IEC 60601-1 and IEC 60601-2-10) as well as lab bench performance evaluation. So this difference will not raise any safety or effectiveness issues.
Page 9 of 11
{13}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020’”, as recommended by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation
### 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
- IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-10 Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
### 3) Software Verification and Validation
Software documentation consistent with *Basic Documentation* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
### Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.
## VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, it can be concluded that the
Page 10 of 11
{14}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary
VITA Multi-Function Head Brush is as safe, as effective, and performs as well as the legally marketed predicate devices.
Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.