K233010 · Belega, Inc. · NFO · Nov 21, 2023 · Neurology
Device Facts
Record ID
K233010
Device Name
Beagank 4T Plus
Applicant
Belega, Inc.
Product Code
NFO · Neurology
Decision Date
Nov 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.
Device Story
Handheld portable aesthetic device; uses rechargeable lithium-ion battery; features four metal electrode pads on head. Operates by delivering transcutaneous electrical nerve stimulation (TENS) to face, neck, and body skin. Requires conductive media (wet cotton pad with purified water or cosmetic liquid) for pulse generation. Includes scalp and small attachments. User-operated for cosmetic purposes. Output consists of four selectable operational modes with varying electrical pulse parameters. Healthcare providers do not interact with output; device is for consumer self-use. Benefits include aesthetic skin stimulation.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via electrical safety, electromagnetic compatibility, and biocompatibility testing.
Technological Characteristics
Handheld aesthetic TENS device; ABS plastic housing; internal rechargeable lithium-ion battery; Type BF patient leakage isolation. Four metal electrode pads. Connectivity: standalone. Sterilization: non-sterile. Electrical safety: IEC 60601-1, IEC 60601-1-2. Biocompatibility: ISO 10993-5, 10993-10, 10993-23.
Indications for Use
Indicated for over-the-counter aesthetic use including facial, neck, and body skin stimulation for general users.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI LIFT Device (K191951)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 21, 2023
Belega Inc. % Nana Nagashita Regulatory Consultant Ken Block Consulting Co., Ltd. 1-9-9 Tsukiji Chuo-ku, Tokyo 104-0045 Japan
Re: K233010
Trade/Device Name: Beagank 4T Plus Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: September 15, 2023 Received: September 22, 2023
Dear Nana Nagashita:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233010
Device Name BEAGANK 4T PLUS
Indications for Use (Describe)
The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
| Owner/Submitter: | Belega Co., Ltd.<br>1-4-20 Sonezaki Shinchi, Kita-ku<br>Osaka-shi, Osaka, Japan 530-0002 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ken Block Consulting Co., Ltd.<br>Ms. Nana Nagashita<br>Regulatory consultant<br><br>1-9-9 Tsukiji, Chuo-ku<br>Tokyo, Japan 104-0045<br>TEL: +81-3-6260-6688<br>FAX: +81-3-6260-6689<br>nagashita@kenblockconsulting.com |
| Date Prepared: | September 20, 2023 |
| Submission Type: | Traditional 510(k) |
| Proposed Device | Manufacturer: BELEGA Co., Ltd.<br>Trade Name: BEAGANK 4T PLUS<br>Common Name: Aesthetic TENS Device<br>Classification Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief<br>Product Code: NFO<br>Regulatory Standard: 21 CFR 882.5890 |
| Predicate Device: | 510(k) Number: K191951<br>Manufacturer: Avazzia, Inc.<br>Trade Name: Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI LIFT Device<br>Common Name: Aesthetic TENS Device<br>Classification Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief<br>Product Code: NFO<br>Regulatory Standard: 21 CFR 882.5890 |
| Reference Device: | 510(k) Number: K011935<br>Manufacturer: Salton Inc.<br>Trade Name: Rejuvenique® system, Model RJV-10<br>Common Name: Aesthetic TENS Device<br>Classification Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief<br>Product Code: NFO, GYB (Media, Conductive)<br>Regulatory Standard: 21 CFR 882.5890, 21 CFR 882.1275 (Media, Conductive) |
| Reference Device: | 510(k) Number: K181008<br>Manufacturer: Carol Cole Company dba NuFACE<br>Trade Name: NuFACE Trinity<br>Common Name: Aesthetic TENS Device<br>Classification Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief<br>Product Code: NFO |
{5}------------------------------------------------
BEAGANK 4T PLUS is a cosmetic device which is intended for use on the face and body Device Description: skin. The device contains a rechargeable lithium ion battery and is provided with a charger. The device has four electrode metal pads on the head that deliver electrical impulses to the face and body skin and provide four different operational modes.
> The device operates by applying Transcutaneous Electrical Nerve Stimulation to the face and body skin. The electrodes that are positioned on the skin must be covered with a wet cotton pad. Purified water or commercially available cosmetic formulation liquid can be used as a conductive media to generate the electric pulses. Two attachments are provided with the BEAGANK 4T PLUS:
- Scalp attachment -
- Small attachment
Statement of The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic Intended Use: use including facial and neck stimulation or body skin stimulation.
Summary of The major differences between the proposed and predicate devices are summarized in the Technological tables below: Characteristics:
| Table 1: Basic characteristics | | | | | |
|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| | New Device | Predicate Device | Reference Device | Reference Device | |
| Trade Name | BEAGANK 4T<br>PLUS | EZZI-LIFT Device | Rejuvenique | NuFACE Trinity | |
| 510(k) | K233010 | K191951 | K011935 | K181008 | |
| Indication for Use | The BEAGANK<br>4T PLUS is a<br>handheld portable<br>device for over-<br>the-counter<br>aesthetic use<br>including facial<br>and neck<br>stimulation or<br>body skin<br>stimulation. | The Avazzia OTC<br>TENS for<br>aesthetics, model<br>BEST-AV1™:<br>EZZI-LIFT™<br>Device is indicated<br>for over-the-<br>counter aesthetic<br>use including<br>facial and neck<br>stimulation or<br>body skin<br>stimulation. | Rejuvenique<br>system is indicated<br>for cosmetic use. | The NuFACE<br>Trinity is intended<br>for facial and neck<br>stimulation and is<br>indicated for over-<br>the-counter<br>cosmetic use. | |
| Device Class | Class II | Class II | Class II | Class II | |
| Operating Power | Internal<br>rechargeable<br>Lithium-ion<br>battery | Two 1.5 V AA<br>batteries | 9 V battery | Internal<br>rechargeable<br>Lithium-ion<br>battery | |
| Method of Line Current<br>Isolation | Type BF | Unknown | Unknown | Type BF | |
| Patient<br>Leakage<br>Current | Normal | DC: <1μΑ<br>AC: <1μΑ<br>Tested per IEC<br>60601-1 | Unknown | Unknown | N/A - Battery<br>operated |
| | Single<br>fault | DC: <1μΑ<br>AC: <1μΑ<br>Tested per IEC<br>60601-1 | Unknown | Unknown | N/A - Battery<br>operated |
| Ave. DC current<br>through electrodes<br>when device is on but<br>no pulse are being<br>applied (μΑ) | | <1μΑ | Unknown | Unknown | Unknown |
| Output<br>Channels | Method<br>of<br>Channel<br>Isolation | N/A - 1 output<br>channel | Unknown | Electrode group<br>selected by relays | N/A - 1 output<br>channel |
| Regulated Current or<br>Regulated Voltage | | Voltage | Unknown | Voltage | Both |
| Automatic Overload<br>Trip | | Not required due<br>to circuit design | Unknown | No | Not required due<br>to circuit design |
| Automatic No-Load<br>Trip | | Not required due<br>to circuit design | Unknown | No | Yes |
| Automatic Shut Off | | Yes | Yes | Yes | Yes |
| Patient Override<br>Control | | Yes | Unknown | Yes | Yes |
| Weight | | 136g | 153g<br>(5.4 ounces) | 80 g | 9 oz. without<br>charging base |
| Number of output<br>modes | | 4 modes:<br>maximum 5<br>minutes each | 4 modes: 60<br>minutes | 1 | Unknown |
| Low Battery Indicator | | Yes | Yes | Yes | Yes |
| Sterility | | Non-Sterile | Non-Sterile | Non-Sterile | Non-Sterile |
| Electrical safety | | IEC 60601-1<br>Medical<br>electrical<br>equipment -<br>Part 1 General<br>requirements<br>for basic<br>safety and<br>essential<br>performance<br>IEC 60601-1-2<br>Medical<br>electrical | IEC 60601-1<br>Medical<br>electrical<br>equipment -<br>Part 1 General<br>requirements<br>for safety<br>IEC 60601-1-2<br>Medical<br>electrical<br>equipment -<br>Part 1-2:<br>General | Unknown | IEC 60601-1<br>IEC 60601-1-2 |
| | Part 1-2:<br>General<br>requirements<br>for basic<br>safety and<br>essential<br>performance –<br>Collateral<br>Standard:<br>Electromagnet<br>ic disturbances<br>–<br>Requirements<br>and tests | for basic<br>safety and<br>essential<br>performance –<br>Collateral<br>Standard:<br>Electromagnet<br>ic disturbances<br>–<br>Requirements<br>and tests<br>IEC 60601-2-<br>10 medical<br>electrical<br>equipment –<br>Part 2:10;<br>Particular<br>requirements<br>for the safety<br>of nerve and<br>muscle<br>stimulators | | | |
| Housing Materials and<br>Construction | ABS | PCBs inside<br>plastic case<br>housing | ABS Plastic, snap<br>latch assembly | Thermoplastic | |
| Accessories | Small and Scalp<br>attachment | Built-in, Y, Brush,<br>Pencil | Unknown | Unknown | |
{6}------------------------------------------------
{7}------------------------------------------------
### Table 2: Technological Characteristics
| Output Specifications: Mode 1 | | | | |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| | New Device | Predicate Device | Reference Device | Reference Device |
| Trade Name | BEAGANK 4T<br>PLUS | EZZI-LIFT Device | Rejuvenique | NuFACE Trinity |
| Indications for Use | The BEAGANK 4T<br>PLUS is a handheld<br>portable device for<br>over-the-counter<br>aesthetic<br>use<br>including facial and<br>neck stimulation or<br>body<br>skin<br>stimulation. | The Avazzia OTC<br>TENS for aesthetics,<br>model<br>BEST-<br>AV1TM:<br>EZZI-<br>LIFTTM Device is<br>indicated for over-<br>the-counter aesthetic<br>use including facial<br>and neck stimulation<br>or body<br>skin<br>stimulation. | Rejuvenique system<br>is indicated<br>for<br>cosmetic use. | The NuFACE<br>Trinity is intended<br>for facial and neck<br>stimulation and is<br>indicated for over-<br>the-counter cosmetic<br>use. |
{8}------------------------------------------------
| Waveform Shape | Rectangle, biphasic asymmetric | Positive square wave followed by a damped sinusoidal waveform of variable duration depending on damping and body loading | Pulsed biphasic, Rectangular (+phase), Spike (- phase) | Pulsed biphasic modulated square wave |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Max output voltage | (+/-15%) | (+/-20%) | (+/-10%) | 28VDC |
| @ 500Ω | 127mV (0.12V) | -42V | 18.8V | not publicly available |
| @2kΩ | 515mV (0.51V) | -122V | 24.8V | not publicly available |
| @ 10kΩ | 1.83V | -348V | 28.0V | not publicly available |
| Max output current | (+/-15%) | (+/-20%) | (+/-10%) | 400μA (0.4mA) |
| @ 500Ω | 0.19mA | 363μΑ(0.363 mA) | 37.6mA | not publicly available |
| @2kΩ | 0.18mA | 117μΑ (0.117 mA) | 12.4mA | not publicly available |
| @ 10kΩ | 0.16mA | 38μΑ (0.038 mA) | 2.8mA | not publicly available |
| Duration of primary pulse at 500Ω | Positive: 130µs (0.13msec)<br>Negative: 130µs (0.13msec) | 0.5msec | not publicly available | 60msec |
| Pulse Duration (Cycle) at 500Ω | 265µs (0.265msec) | 1.1msec | 300msec (+phase)<br>124.7msec (-phase, exponential) | 60msec |
| Frequency | 3.80kHz | 15-121Hz | 8Hz fixed | 8.3Hz |
| Net Charge per pulse at 500Ω | 0.025µC | 4µC | 0µC | not publicly available |
| Max phase charge | 0.059µC | 10µC | 11.3µC | not publicly available |
| Max current density at 500Ω | - Standard: 0.15mA/cm²<br>- Scalp: 0.53mA/cm²<br>- Small: | Built-in, Y, Brush: 800μA/cm²(0.80mA /cm²) | 46.4mA/cm² | not publicly available |
| | 1.36mA/cm2 | Pencil: | | |
| | | 19,000μA/cm2<br>(19mA/cm2) | | |
| | With a wet cotton<br>pad (Purified water) | | | |
| | - Standard:0.15<br>mA/cm2 | | | |
| | - Scalp:<br>0.53mA/cm2<br>Small: 1.36mA/cm2 | | | |
| Max power density<br>at 500Ω (Smallest<br>electrode) | 8.51μW/cm2 | Built-in, Y, Brush:<br>Average<br>500μW/cm2 | 2.31mW/cm2<br>(2,310μW/cm2) | not publicly<br>available |
| | - With a wet<br>cotton pad<br>11.39μW/cm2 | Pencil: Average<br>3,500μW/cm2 | | |
| Max. Average<br>Current (mA) | 0.105 | not publicly<br>available | not publicly<br>available | not publicly<br>available |
| Burst mode | | | | |
| Pulse per burst | N/A | N/A | 160 | 20 |
| Burst per second | N/A | N/A | 1/240 (per electrode<br>group) | 8.3 |
| Burst duration | N/A | N/A | 20sec | 2.4 |
| Duty cycle | N/A | N/A | 1/12 | 20.2sec |
| On Time (seconds) | N/A | N/A | 20sec/electrode<br>group | not publicly<br>available |
| Off Time (seconds) | N/A | N/A | not publicly<br>available | not publicly<br>available |
| Output Specifications: Mode 2 | | | | |
| | New Device | Predicate Device | Reference Device | Reference Device |
| Trade Name | BEAGANK 4T<br>PLUS | EZZI-LIFT Device | Rejuvenique | NuFACE Trinity |
| Indications for Use | The BEAGANK 4T<br>PLUS is a handheld<br>portable device for<br>over-the-counter<br>aesthetic use<br>including facial and | The Avazzia OTC<br>TENS for aesthetics,<br>model BEST-<br>AV1TM: EZZI-<br>LIFTTM Device is<br>indicated for over- | Rejuvenique system<br>is indicated for<br>cosmetic use. | The NuFACE<br>Trinity is intended<br>for facial and neck<br>stimulation and is<br>indicated for over- |
{9}------------------------------------------------
{10}------------------------------------------------
| | neck stimulation or<br>body skin stimulation. | the-counter aesthetic<br>use including facial<br>and neck stimulation<br>or body skin stimulation. | | the-counter cosmetic<br>use. |
|------------------------------------------------|----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------|
| Waveform Shape | Rectangle, biphasic<br>asymmetric | Positive square<br>wave followed by a<br>damped sinusoidal<br>waveform of<br>variable duration<br>depending on<br>damping and body<br>loading | Pulsed biphasic,<br>Rectangular<br>(+phase), Spike (-<br>phase) | Pulsed biphasic<br>modulated square<br>wave |
| Max output voltage | (+/-15%) | (+/-20%) | (+/-10%) | 28VDC |
| @ 500Ω | 110mV (0.11V) | -42V | 18.8V | not publicly<br>available |
| @2kΩ | 507mV (0.50V) | -122V | 24.8V | not publicly<br>available |
| @ 10kΩ | 1.87V | -348V | 28.0V | not publicly<br>available |
| Max output current | (+/-15%) | (+/-20%) | (+/-10%) | 400μA (0.4mA) |
| @ 500Ω | 0.45mA | 363μA (0.363mA) | 37.6mA | not publicly<br>available |
| @2kΩ | 0.39mA | 117μA (0.117mA) | 12.4mA | not publicly<br>available |
| @ 10kΩ | 0.24mA | 38μA (0.038mA) | 2.8mA | not publicly<br>available |
| Duration of primary<br>pulse at 500Ω | Positive: 130µs<br>(0.13msec)<br>Negative: 130µs<br>(0.13msec) | 0.5msec | not publicly<br>available | 60msec |
| Pulse Duration<br>(Cycle) at 500Ω | 265µs (0.265msec) | 1.1msec | 300msec (+phase)<br>124.7msec (-phase,<br>exponential) | 60msec |
| Frequency | 3.80kHz | 15-121Hz | 8Hz fixed | 8.3Hz |
| Net Charge per pulse<br>at 500Ω | 0.025μC | 4μC | 0μC | not publicly<br>available |
| Max phase charge | $0.025\mu C$ | $10\mu C$ | $11.3\mu C$ | not publicly available |
| Max current density at 500Ω | Standard:<br>$0.36mA/cm^2$<br>Scalp:<br>$1.25mA/cm^2$<br>Small:<br>$3.21mA/cm^2$ | Built-in, Y, Brush:<br>$800\mu A/cm^2$<br>( $0.8mA/cm^2$ )<br>Pencil:<br>$19,000\mu A/cm^2$<br>( $19mA/cm^2$ ) | $46.4mA/cm^2$ | not publicly available |
| With a wet cotton pad (Purified water) | Standard:<br>$0.10mA/cm^2$<br>Scalp: $0.36 mA/cm^2$<br>Small: $0.92 mA/cm^2$ | | | |
| Max power density at 500Ω (Smallest electrode) | $17.46\mu W/cm^2$ | Built-in, Y, Brush:<br>Average $500\mu W/cm^2$ | $2.31mW/cm^2$<br>( $2,310\mu W/cm^2$ ) | not publicly available |
| With a wet cotton pad | $33.87\mu W/cm^2$ | Pencil: Average $3,500\mu W/cm^2$ | | |
| Max. Average Current (mA) | 0.105 | not publicly available | not publicly available | not publicly available |
| Burst mode | | | | |
| Pulse per burst | N/A | N/A | 160 | 20 |
| Burst per second | N/A | N/A | 1/240 (per electrode group) | 8.3 |
| Burst duration | N/A | N/A | 20sec | 2.4 |
| Duty cycle | N/A | N/A | 1/12 | 20.2sec |
| On Time (seconds) | N/A | N/A | 20sec/electrode group | not publicly available |
| Off Time (seconds) | N/A | N/A | not publicly available | not publicly available |
| Output Specifications: Mode 3 | | | | |
| | New Device | Predicate Device | Reference Device | Reference Device |
{11}------------------------------------------------
{12}------------------------------------------------
| Trade Name | BEAGANK 4T<br>PLUS | EZZI-LIFT Device | Rejuvenique | NuFACE Trinity |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The BEAGANK 4T<br>PLUS is a handheld<br>portable device for<br>over-the-counter<br>aesthetic<br>use<br>including facial and<br>neck stimulation or<br>body<br>skin<br>stimulation. | The Avazzia OTC<br>TENS for aesthetics,<br>model BEST-<br>AV1TM. EZZI-<br>LIFT™ Device is<br>indicated for over-<br>the-counter aesthetic<br>use including facial<br>and neck stimulation<br>or body skin<br>stimulation. | Rejuvenique system<br>is indicated for<br>cosmetic use. | The NuFACE<br>Trinity is intended<br>for facial and neck<br>stimulation and is<br>indicated for over-<br>the-counter cosmetic<br>use. |
| Waveform Shape | Rectangle, biphasic<br>symmetrical | Positive square<br>wave followed by a<br>damped sinusoidal<br>waveform of<br>variable duration<br>depending on<br>damping and body<br>loading | Pulsed biphasic,<br>Rectangular<br>(+phase), Spike (-<br>phase) | Pulsed biphasic<br>modulated square<br>wave |
| Max output voltage | (+/-15%) | (+/-20%) | (+/-10%) | 28VDC |
| @ 500Ω | 16.6V | -42 | 18.8V | not publicly<br>available |
| @2kΩ | 21.0V | -122 | 24.8V | not publicly<br>available |
| @10kΩ | 22.5V | -348 | 28.0V | not publicly<br>available |
| Max output current | (+/-15%) | (+/-20%) | (+/-10%) | 400μA |
| @ 500Ω | 30.8mA | 363μA(0.363 mA) | 37.6mA | not publicly<br>available |
| @2kΩ | 8.52mA | 117μA (0.117 mA) | 12.4mA | not publicly<br>available |
| @ 10kΩ | 2.26mA | 38μA (0.038 mA) | 2.8mA | not publicly<br>available |
| Duration of primary<br>pulse at 500Ω | Positive: 80μs<br>(0.08msec)<br>Negative: 80μs<br>(0.08msec) | 0.5msec | not publicly<br>available | 60msec |
| Pulse Duration<br>(Cycle) at 500Ω | 274μs (0.247msec) | 1.1msec | 300msec (+phase)<br>124.7msec (-phase,<br>exponential) | 60msec |
| Frequency | 3.65kHz | 15-121Hz | 8Hz fixed | 8.3Hz…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.