Morph is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. Volumetric measurements may be compared to reference percentile data.
Device Story
Morph is AI-powered brain MRI analysis software; provides quantitative assessments of brain structure volumes; assists medical practitioners in clinical decision-making. Input: 2D or 3D T1-weighted MR images (DICOM/NIfTI). Processing: Deep learning-based segmentation engine; calculates volumes of 14 brain regions (CSF areas, lateral ventricles, temporal horns); compares results to normative percentile database. Output: DICOM secondary capture images with segmented color overlays; PDF reports. Deployment: Web-based client; cloud (AWS) or on-premises server. Usage: Adjunctive tool for radiologists, neurologists, and neuroradiologists in routine patient care; not for primary diagnosis of neurodegenerative disease or atrophy. Benefits: Standardized, reproducible volumetric data to support assessment of neurodegeneration.
Clinical Evidence
Bench testing only. Validation used 64 independent 3D T1-weighted MRI scans from the ADNI dataset (diverse demographics, clinical diagnoses, and scanner manufacturers). Primary endpoint: segmentation accuracy (Dice Similarity Coefficient, DSC) and volumetric agreement (Bland-Altman). Results: Mean DSC > 0.85 (lateral ventricles > 0.97, temporal horns > 0.89). Volumetric bias and 95% limits of agreement were within predefined clinically acceptable ranges. Subgroup analyses across age, sex, diagnosis, and scanner parameters showed consistent performance.
Technological Characteristics
SaMD; deep learning-based segmentation algorithm; T1-weighted MRI input; cloud (AWS) or on-premises server deployment; DICOM/NIfTI compatibility; web-based interface; automated quality control (scan protocol verification).
Indications for Use
Indicated for adult patients (50-99 years) including cognitively normal individuals, those with mild cognitive impairment, and patients with Alzheimer's disease, to provide automated labeling, visualization, and volumetric quantification of brain structures (CSF spaces, lateral ventricles, hippocampal regions) from T1-weighted MR images for clinical assessment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 13, 2026
BeauBrain Healthcare, Inc.
% Priscilla Chung
Regulatory Affairs Consultant
LK Consulting Group USA, Inc.
2552 Walnut Ave. Suite 230
Tustin, California 92780
Re: K260032
Trade/Device Name: Morph
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: April 13, 2026
Received: April 13, 2026
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260032 - Priscilla Chung
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260032 - Priscilla Chung
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260032 | ? |
| Please provide the device trade name(s). | | ? |
| Morph | | |
| Please provide your Indications for Use below. | | ? |
| Morph is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. Volumetric measurements may be compared to reference percentile data. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Page 1 of 10
510(k) Summary
(K260032)
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
1. Date: 5/12/2026
2. Applicant / Submitter
BeauBrain Healthcare, Inc.
2F, 314, Hakdong-ro, Gangnam-gu, Seoul
South Korea 06098
Tel: +82-02-6925-0179
3. U.S. Designated Agent
Priscilla Chung
LK Consulting Group USA, Inc.
2552 Walnut Ave Ste 230, Tustin CA 92780
Tel: 714.202.5789 Fax: 714.409.3357
Email: juhee.c@LKconsultingGroup.com
4. Device Information:
- Trade/Device Name: Morph
- Common Name: Automated Radiological Image Processing Software
- Regulation Name: Medical image management and processing system
- Regulation Number: 21 CFR 892.2050
- Regulatory Class: II
- Product Code: QIH, LLZ
5. Predicate Device:
NeuroQuant (K170981) by CorTechs Labs, Inc.
6. Device Description:
Morph is an AI-powered brain MRI analysis software that provides brain quantitative assessments based on segmented quantitative information from brain MRIs. This helps medical practitioners make clinical decisions. It is an adjunctive tool and should not be used as the sole basis for final decisions or treatment.
{5}
The software automatically performs quantitative analysis of cerebrospinal fluid (CSF) volumes in cerebral areas and the lateral ventricles (LVs) to assess brain neurodegeneration, such as volumetric changes in specific brain regions. This is often associated with an increased susceptibility to neurodegenerative diseases like Alzheimer's disease (AD).
Morph is an AI-powered brain MRI analysis software that provides quantitative assessments of brain structure volumes based on segmented quantitative information from brain MRIs. This helps medical practitioners make clinical decisions. It is an adjunctive tool and should not be used as the sole basis for final decisions or treatment. The software automatically performs quantitative analysis of cerebrospinal fluid (CSF) volumes in cerebral areas and the lateral ventricles (LVs) to evaluate structural variations in specific brain regions. These quantitative measurements are compared to a reference database to support clinical interpretation.
Morph identifies the 14 specific brain regions, including the left and right CSF areas of Frontal, Occipital, Parietal, and Temporal lobes, as well as Anterior LVs, Posterior LVs, and temporal horns of LVs near the hippocampus. It uses volumetric and normative percentile data to provide standardized scores based on a large database of cognitively normal individuals.
Images in standardized DICOM format are uploaded through a web-based client software and processed by the analysis engine, which can be hosted on a cloud-based server (AWS) or an on-premises server (in-hospital). During preprocessing, the software converts 2D or 3D T1 MR images in DICOM format to NIfTI format (users can also directly upload NIfTI format images). Morph receives the MR images from either PC or PACS (or imaging modalities) through DICOM interface and then process the imaging analysis of volumetric measurements of brain structures. After the analysis is completed, the result image in the DICOM secondary capture format is saved in the PACS. Otherwise, the report can be printed out directly in PDF format.
The device is not intended to diagnose any neurodegenerative disease or brain atrophy. It is used as a quantitative tool to evaluate the patient against a reference database.
## 7. Indication for use:
Morph is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. Volumetric measurements may be compared to reference percentile data.
Page 2 of 10
{6}
8. Substantial Equivalence:
Comparison Table
| Comparison Item | Proposed Device, Morph | Predicate Device, NeuroQuant, K170981 |
| --- | --- | --- |
| 510(k) Number | K260032 | K170981 |
| Trade Name | Morph | NeuroQuant |
| Design | Software as a Medical Device (SaMD) | Software as a Medical Device (SaMD) |
| Common Name | Medical Image Processing Software | Medical Image Processing Software |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Regulation Name | Automated Radiological Image Processing Software | Picture archiving and communications system |
| Regulatory Class | Class II | Class II |
| Product Code | QIH, LLZ | LLZ |
| 510(k) Review Panel | Radiology | Radiology |
| Manufacturer | BeauBrain Healthcare Inc. | CorTechs Labs, Inc. |
| Target Anatomical Sites | Brain | Brain |
| User | Medical professionals (radiologists, neurologists and neuroradiologists) | Medical professionals (radiologists, neurologists and neuroradiologists and clinical researcher) |
| Indications for Use | Morph is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. Volumetric measurements may be compared to reference percentile data. | NeuroQuant is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric measurements may be compared to reference percentile data. |
| Design and Incorporated Technology | • Automated measurement of brain volumes and structures. • Automated segmentation and quantification of brain structures using deep learning-based algorithm, with MR | • Automated measurement of brain tissue volumes and structures and lesions. • Automatic segmentation and quantification of brain structures using a dynamic |
Page 3 of 10
{7}
| Comparison Item | Proposed Device, Morph | Predicate Device, NeuroQuant, K170981 |
| --- | --- | --- |
| | image intensity. | probabilistic neuroanatomical atlas, with age and gender specificity, based on the MR image intensity. |
| Physical Characteristics | • Operates in either on-premise or a cloud service environment | • Software package installed on User hardware • Operates on off-the-shelf hardware (multiple vendors) |
| Device Description | Morph is a fully automated MR imaging post-processing medical device software that provides automatic labeling, visualization and volumetric quantification of brain structures from a set of MR images. The resulting output is provided in a standard DICOM format as additional MR series with segmented color overlays and morphometric reports that can be displayed on third-party DICOM workstations and Picture Archive and Communications Systems (PACS). The high throughput capability makes the software suitable for use in routine patient care as a support tool for clinicians in assessment of structural MRIs. | NeuroQuant is a fully automated MR imaging post-processing medical device software that provides automatic labeling, visualization and volumetric quantification of brain structures and lesions from a set of MR images. The resulting output is provided in a standard DICOM format as additional MR series with segmented color overlays and morphometric reports that can be displayed on third-party DICOM workstations and Picture Archive and Communications Systems (PACS). The high throughput capability makes the software suitable for use in both clinical trial research and routine patient care as a support tool for clinicians in assessment of structural MRIs. |
| Processing Architecture | Automated internal pipeline that performs: • Segmentation • Volume calculation • Report generation | Automated internal pipeline that performs: • Artifact correction • Segmentation • Lesion quantification • Volume calculation • Report generation |
| Data Source | MRI scanner: 2D or 3D T1 MRI scans acquired with specified | MRI scanner: 3D T1 MRI scans acquired with specified |
| | dB, 3D T1 MRI scans acquired with specified | dB, 3D T1 MRI scans acquired with specified |
| Data Source Type | MRI scanner: 2D or 3D T1 MRI scans acquired with specified | MRI scanner: 3D T1 MRI scans acquired with specified |
{8}
| Comparison Item | Proposed Device, Morph | Predicate Device, NeuroQuant, K170981 |
| --- | --- | --- |
| | protocols • Supports DICOM format as input | protocols • Supports DICOM format as input |
| Output | • Provides volumetric measurements of brain structures • Includes segmented color overlays and morphometric reports • Automatically compares results to reference percentile data and to prior scans when available • Supports DICOM format as output of results that can be displayed on DICOM workstations and Picture Archive and Communications Systems | • Provides volumetric measurements of brain structures and lesions • Includes segmented color overlays and morphometric reports • Automatically compares results to reference percentile data and to prior scans when available • Supports DICOM format as output of results that can be displayed on DICOM workstations and Picture Archive and Communications Systems |
| Safety | • Automated quality control functions - Scan protocol verification • Results must be reviewed by a trained doctor (physician, neurologist, radiologist, clinician) | • Automated quality control functions - Tissue contrast check - Scan protocol verification - Atlas alignment check • Results must be reviewed by a trained physician |
# Substantial Equivalence Discussion
The proposed device, Morph is substantially equivalent to the predicate device, NeuroQuant (K170981).
Morph and the predicate device are software for automatically identifying and quantifying the volumes of brain structures, automatic labeling and visualization. The devices have the same intended use and operating principle. They take MR brain images as input and generate an electronic report with similar quantitative information. For both devices, output volumes are compared to a normative dataset of control subjects computed based on MRI data from normal control subjects.
The predicate device is classified under 21 CFR 892.2050 (Class II, Product code LLZ), while the proposed device is classified as Automated Radiological Image Processing Software (21 CFR 892.2050, Class II, Product code QIH, Radiology). The only distinction is that the QIH product
{9}
code indicates inclusion of an AI model, whereas the LLZ product code does not; however, the devices are otherwise identical.
## Indications for Use
The primary intended use of providing automatic segmentation and quantification of brain structures is equivalent. The predicate device performs volumetric quantification of brain structures including brain lesion quantification (requiring T2-FLAIR). The proposed device does not quantify brain lesions but provides volumetric analysis of anatomical brain structures from T1-weighted MR images. This difference represents a reduced functionality rather than a change in intended use. The absence of lesion quantification does not raise new questions of safety or effectiveness.
## Design and Incorporated Technology
Both devices are software applications that receive T1-weighted MR images as input and provide automated brain segmentation and volumetric quantification as output. The predicate uses an atlas-based segmentation method, whereas the proposed device applies a deep learning-based segmentation model. Although the algorithmic approach differs, both rely on established image analysis technologies. AI-based algorithm(Segmentation model) Performance testing of the proposed device (DSC and ICC) demonstrates accuracy and reproducibility equivalent to the predicate, ensuring that the technological differences do not introduce new risks.
## Segmentation Accuracy and Targeted Regions
Both devices produce segmentation maps with color overlays and volumetric reports.
However, the predicate segments whole brain regions, including gray matter, white matter, CSF, and lesions. The proposed device segments cortical CSF regions, lateral ventricles, and hippocampal areas to quantify brain regional volume. This narrower scope is a reduced functionality but uses the same mechanism of volumetric quantification with reference percentile comparison. The omission of lesion quantification does not alter the interpretation of volumetric results and does not raise questions of safety or effectiveness.
## Physical / Operational Characteristics
The predicate operates on off-the-shelf hardware, while the proposed device supports both on-premise and cloud-based environments, accessible through a web browser. This deployment difference does not affect analysis performance. Verification and validation testing confirm that the proposed device performs reliably across environments, and therefore the difference does not impact safety or effectiveness.
## Quality Control Functions
The predicate device includes tissue contrast and atlas alignment checks as part of automated quality control. The proposed device, based on deep learning, does not require these checks. Instead, performance has been validated through algorithm testing, ensuring equivalent accuracy and safety. The absence of these checks does not affect the reliability of segmentation or volumetric outputs.
Page 6 of 10
{10}
Page 7 of 10
# Output and Safety Perspective
Both devices generate segmentation maps with color overlays and quantitative volumetric reports, including percentile comparisons. The only difference is that the predicate includes lesion quantification, while the proposed device does not. For its intended function—volumetric quantification of segmentable brain structures—the proposed device provides equivalent outputs. Software validation and performance testing confirms that the device meets acceptance criteria for accuracy and reproducibility. Therefore, these differences do not raise new questions of safety or effectiveness.
# Summary
The proposed device, Morph (BeauBrain Healthcare Inc.), has the same intended use and fundamental technological characteristics as the predicate device, NeuroQuant (K170981, CorTechs Labs, Inc.). Both are Software as a Medical Device (SaMD) that automatically label, visualize, and quantify brain structures from MR images, providing volumetric reports with percentile-based comparisons. Differences between the devices include the absence of lesion quantification in Morph, use of a deep learning algorithm instead of an atlas-based method, a narrower segmentation scope (CSF, ventricles, hippocampus), and the option to operate in cloud-based environments. These differences represent reduced functionality or alternative implementations rather than new intended uses. Performance testing (DSC, ICC) confirm that Morph achieves accuracy, reproducibility, and safety equivalent to the predicate. Accordingly, these differences do not raise new questions of safety or effectiveness, and Morph is therefore substantially equivalent to NeuroQuant (K170981). Accordingly, Morph is substantially equivalent to the predicate device NeuroQuant (K170981).
# 9. Performance Data:
The device employs an AI-based segmentation algorithm to automatically identify and quantify selected brain structures from structural MRI images, including cortical CSF spaces, lateral ventricles, and CSF spaces surrounding the hippocampal region. Performance characteristics were established through independent validation using datasets that were fully separated from the training data. These datasets were designed to reflect variability in scanner manufacturers, acquisition protocols, demographics, and clinical diagnoses.
a) Training Data
The AI-based segmentation algorithm was trained using a total of 1,947 3D T1-weighted MRI scans obtained from Samsung Medical Center (SMC) and several publicly available datasets including the IXI, ICBM, and OASIS-3.
The training dataset included adult subjects across a wide age range (50-99 years) and included both male and female subjects. The cohort also represented diverse racial groups, including White, Black, and Asian subjects. The clinically relevant groups were included, such as cognitively normal individuals, subjects with mild cognitive impairment, and patients diagnosed with Alzheimer's disease.
{11}
MRI scans were acquired using scanners from multiple major vendors including Siemens, Philips, and GE Healthcare, with magnetic field strengths of 1.5 T and 3 T and slice thicknesses of 1.0 mm and 1.2 mm. All scans consisted of 3D T1-weighted structural MRI images acquired using protocols compatible with volumetric brain analysis.
Training and validation datasets were strictly separated at the subject level. No MRI scans from the training datasets were reused in the validation datasets, ensuring independence of performance estimation.
### b) Performance Testing
To support transparency of the AI-based segmentation model, the validation dataset was characterized with respect to demographics, clinical subgroups, imaging equipment, and the reference standard generation. The validation dataset consisted of 64 MRI scans from unique subjects obtained from the ADNI, a U.S.-based dataset.
The validation cohort included adult subjects aged 50--99 years, with representation across multiple age groups, including 50--59 years (6.3%), 60--69 years (31.3%), 70--79 years (32.8%), 80--89 years (28.1%), and 90--99 years (1.5%).
The dataset included female (40.6%) and male (59.4%) subjects, and the racial distribution consisted of White (89.0%), Black (6.3%), and Asian (4.7%).
In terms of clinical subgroup distribution, the cohort included cognitively normal subjects (CN) (29.7%), subjects with mild cognitive impairment (MCI) (37.5%), and patients diagnosed with Alzheimer's disease (AD) (32.8%). MRI data were collected from multiple clinical sites across the United States, including Eastern, Western, and Midwestern regions. Images were acquired on scanners from major manufacturers, including Siemens (48.4%), GE Healthcare (31.3%), and Philips (20.3%), using 3D T1-weighted structural MRI protocols.
Imaging parameters reflected typical variability encountered in clinical practice. The dataset included images acquired with slice thicknesses of 1.0 mm (51.6%) and 1.2 mm (48.4%), and magnetic field strengths of 1.5 T (51.6%) and 3 T (48.4%).
The reference standard for segmentation accuracy evaluation was established through manual segmentation of 64 MRI scans by two U.S. board-certified neuroradiologists using predefined anatomical criteria and the ITK-SNAP software. Inter-rater agreement was assessed using the Dice Similarity Coefficient, and cases with lower agreement were reviewed by a senior neuroradiologist to establish the final consensus reference segmentation standard used for performance evaluation.
Standalone performance testing was conducted using a validation dataset consisting of 64 independent 3D T1-weighted MRI scans.
Segmentation performance was evaluated by comparing automated segmentation results with expert manual segmentations using the DSC. The predefined acceptance criterion for segmentation performance was a mean DSC greater than or equal to 0.70 across evaluated brain regions.
{12}
Dice coefficients were highest in the lateral ventricles (>0.97) and in the temporal horn adjacent to the hippocampus (>0.89). Other CSF regions demonstrated Dice coefficient ranging from approximately 0.79 to 0.82, while occipital CSF regions achieved Dice coefficients of approximately 0.71. Across the full cohort, the overall mean DSC exceeded 0.85, satisfying the predefined acceptance criteria.
To further assess the robustness and generalizability of the algorithm, additional subgroup analyses were conducted across demographic characteristics, clinical diagnosis, and imaging acquisition parameters. Mean DSC values were consistent across demographic subgroups including age and sex (0.85 to 0.88), clinical diagnostic categories including AD, MCI, and cognitively normal subjects (0.86 to 0.87), and imaging acquisition parameters including scanner manufacturer, magnetic field strength, and slice thickness (0.86 to 0.87), with no subgroup falling below the predefined acceptance criterion.
Agreement between automated and expert reference volumetric measurements was evaluated using Bland-Altman analysis. Mean bias, standard deviation of volume differences, and 95% limits of agreement (LoA) were assessed for all evaluated brain regions. In ventricular regions, mean bias ranged from -30.8 mm³ to 138.4 mm³, with standard deviations ranging from 62.6 mm³ to 478.5 mm³ and 95% LoA ranging from -799.6 mm³ to 1076.3 mm³. In lobar regions, mean bias ranged from approximately -2,071 mm³ to 223 mm³, with standard deviations ranging from 318.1 mm³ to 5105.5 mm³ and 95% LoA ranging from -12078.2 mm³ to 7935.3 mm³. These results demonstrated low volumetric bias and agreement within predefined clinically acceptable 95% LoA ranges between automated and expert reference measurements, with no evidence of clinically significant systematic bias.
## Summary
The proposed device, Morph (BeauBrain Healthcare Inc.), has the same intended use and similar fundamental technological characteristics as the predicate device, NeuroQuant (CorTechs Labs, Inc.). Both devices are Software as a Medical Device (SaMD) that automatically segment, visualize, and quantify brain structures from MR images and generate volumetric reports with percentile-based comparisons.
Differences include the use of a deep learning-based segmentation algorithm in Morph, a narrower segmentation scope, the absence of lesion quantification, and the option for cloud-based deployment. These differences do not introduce new intended uses or raise new questions of safety or effectiveness.
Performance testing demonstrated that Morph provides accurate and reproducible volumetric measurements comparable to those of the predicate device. Therefore, Morph is substantially equivalent to the predicate device.
## 10. Conclusion:
The subject device is substantially equivalent in the areas of technical characteristics, general
Page 9 of 10
{13}
function, application, and indications for use. The test results also support the substantial equivalence to the predicate devices. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.