The Ossiclear system is a non- invasive two-point bone conduction hearing aid solution. The Ossiclear system, attaches to the skin via Ossiclear Stick Pads, uses bone conduction to transmit sounds to the cochlea (inner ear) with the purpose of enhancing hearing. Ossiclear Pro Fitting Suite is a fitting software used to program Ossiclear BHD and modify hearing profiles to provide comfortable and usable gain for recipients. Ossiclear SmartApp is a mobile application that provides digital support for patients in controlling and monitoring the Ossiclear BHD. Ossiclear System is indicated for the following patients aged 12 years and older: - Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) should be better than or equal to 55 dB HL. - Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally moderate to profound conductive or mixed hearing loss. Symmetrical bone conduction thresholds are defined as less than a 10 dB HL average difference between ears (measured at 0.5 1 2 and 3 kHz), or less than a 15 dB HL difference at individual frequencies. - Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single sided deafness SSD). Normal hearing is defined as a pure tone average air conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) of better than or equal to 55 dB HL. - Any patient who is indicated for an air conduction contralateral routing of signals (AC CROS) hearing aid.
Device Story
Non-invasive bone conduction hearing aid; transmits sound vibrations to cochlea via skin-adhered processor. Input: dual microphones capture ambient sound. Processing: signal processing technology adjusts gain/frequency response based on audiometric profile. Output: two axisymmetric electromagnetic transducers convert signals to mechanical vibrations. Used in home/clinical settings; operated by patient or hearing care professional. Professional fitting via Pro Fitting Suite; patient control via SmartApp. Benefits: improved hearing for conductive/mixed loss or SSD patients. Powered by rechargeable Li-ion battery; secured by single-use adhesive pads.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electroacoustic performance (gain, frequency response, noise, distortion), biocompatibility, electrical/thermal safety (IEC 60601-1 series), EMC (IEC 60601-1-2), and software/cybersecurity verification.
Technological Characteristics
Non-invasive bone conduction hearing aid. Materials: biocompatible. Sensing: dual microphones. Actuation: two axisymmetric electromagnetic transducers. Energy: 1454 Li-ion rechargeable battery. Connectivity: Bluetooth (SmartApp). Fixation: single-use adhesive pads (ASTM D3330 peel strength validated). Software: Class A (IEC 62304), Minor level of concern. Standards: IEC 60118-9:2019 (electroacoustic), IEC 60601-1 (safety), ISO 10993 (biocompatibility).
Indications for Use
Indicated for patients aged 12+ with conductive or mixed hearing loss (bone conduction threshold ≤55 dB HL), bilateral moderate to profound conductive/mixed hearing loss, unilateral sensorineural deafness (SSD) with normal hearing in the contralateral ear (air conduction threshold ≤20 dB HL), or patients indicated for AC CROS hearing aids.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 22, 2026
Sounduct
Camille Bonnassieux
Quality and Regulatory Manager
2, Rue Marie Hamm
Strasbourg, 67000
France
Re: K254296
Trade/Device Name: Ossiclear (Ossiclear System)
Regulation Number: 21 CFR 874.3302
Regulation Name: Bone-Conduction Hearing Aid
Regulatory Class: Class II
Product Code: LXB
Dated: December 31, 2025
Received: December 31, 2025
Dear Camille Bonnassieux:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254296 - Camille Bonnassieux
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254296 - Camille Bonnassieux
Page 3
Sincerely,
JOYCE C.
LIN-S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254296
Device Name
Ossiclear (Ossiclear System)
Indications for Use (Describe)
Ossiclear System is indicated for the following patients aged 12 years and older:
- Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) should be better than or equal to 55 dB HL.
- Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally moderate to profound conductive or mixed hearing loss. Symmetrical bone conduction thresholds are defined as less than a 10 dB HL average difference between ears (measured at 0.5 1 2 and 3 kHz), or less than a 15 dB HL difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single sided deafness SSD). Normal hearing is defined as a pure tone average air conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) of better than or equal to 20 dB HL.
- Any patient who is indicated for an air conduction contralateral routing of signals (AC CROS) hearing aid.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
1 / 8
# 510(k) Summary
# A. Submitter information
Submitted by: SOUNDUCT
2 rue Marie HAMM
6700 Strasbourg – FRANCE
Contact: Camille BONNASSIEUX
Quality and Regulatory Affairs Manager
SOUNDUCT
+33 (0) 9 67 41 18 46
camille@sounduct.com ; regulatory@sounduct.com
B. Date Prepared July 22nd, 2026
# C. Device Name and Classification
Device Name: Ossiclear System
Device Trade Name: Ossiclear
Classification Name: Hearing aid, bone conduction - 21 CFR 874.3302, Class II
Classification Panel Ear Nose & Throat
Device Product Code: LXB
# D. Predicate and Reference devices
Predicate device Trade Name: Cochlear™ Baha® System, Cochlear™ Baha® Fitting Software 6, Cochlear™ Baha® Baha Smart App
Classification Name: Hearing Aid, Bone Conduction - 21 CFR 874.3302, Class II
Classification Panel Ear Nose & Throat
510(k) Number: K212136
Product Code: LXB
ADHEAR System
This summary includes only information that is publicly available and does not contain any confidential information.
{5}
| | 510(k) Summary - Ossiclear System [As Required by 21 CFR 807.92] |
| --- | --- |
| | 2 / 8 |
# **Reference device Trade Name:**
| Classification Name: | Hearing Aid, Bone Conduction - 21 CFR 874.3302, Class |
| --- | --- |
| Classification Panel | II Ear Nose & Throat |
| 510(k) Number: | K172460 |
| Product Code: | LXB |
# **E. Device Description**
The Ossiclear System is a non-invasive bone conduction hearing aid intended to improve hearing by transmitting sound vibrations through bone to the cochlea. The system includes the Ossiclear Bone Hearing Device (BHD), which contains microphones, signal processing technology, and a rechargeable battery. The BHD is secured to the skin using single-use adhesive (Ossiclear Stick Pads) specifically designed for compatibility with the Ossiclear System.
The Ossiclear DOCK is a portable charging station and storage case for one BHD, allowing safe and convenient charging. The Ossiclear Pro Fitting Suite is a professional software application intended for qualified hearing care professionals to configure, program, and fit the Ossiclear BHD according to each patient's audiometric profile. The Ossiclear SmartApp is an optional mobile application that enables patients to monitor and control their device.
Together, these components work as one system for patients with conductive or mixed hearing loss or single-sided deafness.
# **F. Intended / Indication for use statement**
The Ossiclear system is a non- invasive two-point bone conduction hearing aid solution. The Ossiclear system, attaches to the skin via Ossiclear Stick Pads, uses bone conduction to transmit sounds to the cochlea (inner ear) with the purpose of enhancing hearing. Ossiclear Pro Fitting Suite is a fitting software used to program Ossiclear BHD and modify hearing profiles to provide comfortable and usable gain for recipients. Ossiclear SmartApp is a mobile application that provides digital support for patients in controlling and monitoring the Ossiclear BHD.
Ossiclear System is indicated for the following patients aged 12 years and older:
- Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) should be better than or equal to 55 dB HL.
*This summary includes only information that is publicly available and does not contain any confidential information.*
{6}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
3 / 8
- Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally moderate to profound conductive or mixed hearing loss. Symmetrical bone conduction thresholds are defined as less than a 10 dB HL average difference between ears (measured at 0.5 1 2 and 3 kHz), or less than a 15 dB HL difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single sided deafness SSD). Normal hearing is defined as a pure tone average air conduction hearing threshold (measured at 0.5 1 2 and 3 kHz) of better than or equal to 20 dB HL.
- Any patient who is indicated for an air conduction contralateral routing of signals (AC CROS) hearing aid.
# G. Device Comparison Table (Subject vs. Predicate vs. Reference)
This summary includes only information that is publicly available and does not contain any confidential information.
{7}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
4 / 8
| Characteristic | Subject Device (Ossiclear) | Predicate Device (Cochlear BAHA 6 System) | Reference Device (ADHEAR System) | Comments/Equivalence |
| --- | --- | --- | --- | --- |
| Regulatory Name | Ossiclear | Cochlear BAHA 6 System | ADHEAR System | Same intended category |
| 510(k) # | K254296 | K212136 | K172460 | Reference supports adhesive fixation |
| Product Code | LXB | LXB | LXB | Same |
| Class | II | II | II | Same |
| Regulation Number | 874.3302 | 874.3302 | 874.3302 | Same |
| Intended Use | Non-invasive bone conduction hearing aid using adhesive pads; includes Pro Fitting Suite & SmartApp | Bone conduction hearing aid; includes Baha Fitting Software & Smart App | Non-invasive bone conduction hearing aid using adhesive adapter or headband | Same principle; adhesive fixation supported by reference device |
| Indications for Use | Conductive/mixed HL ≤55 dB HL; SSD; AC-CROS | Conductive/mixed HL ≤55 dB HL; SSD; AC-CROS | Conductive HL ≤25 dB HL; SSD | Aligned with predicate; reference covers adhesive use |
| Target Population Age | ≥12 years | All ages | All ages | The subject device's target population is covered by the predicate device, and the same indications for use |
| Surgical Requirement | Non-surgical (adhesive) | Surgical for implant; non-surgical options available | Non-surgical (adhesive) | Subject offers only non-surgical option; supported by reference |
| Prescription Use | Rx Only | Rx Only | Rx Only | Same |
| System Components | External sound processor + adhesives + portable charging station | External sound processor + softband/SoundArc or implant coupling | External sound processor + adhesives | Similar; adhesive solution validated |
This summary includes only information that is publicly available and does not contain any confidential information.
{8}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
5 / 8
| Processor Components | Dual microphones; LED; Battery; Two Axisymmetric electromagnetic transducer; Independent adhesives | Dual microphones; LED; Battery; On device button to change programs; Symmetrical electromagnetic transducer | Single microphone; LED; battery; On device button to change programs; unique vibration system; Independent adhesives | Subject device is equivalent to predicate device except for including 2 axisymmetric electromagnetic transducers & independent adhesives (supported by the reference device). Electroacoustic bench testing per IEC 60118-9:2019 demonstrates comparable output performance. Independent adhesive fixation was validated through peel adhesion strength testing per ASTM D3330. These results confirm that the differences in transducer configuration and fixation method do not raise new questions of safety or effectiveness. |
| --- | --- | --- | --- | --- |
| Battery Type | 1454 Li-ion rechargeable | 312 zinc-air | 13 zinc-air | Different type; safety addressed via IEC 60601-1 |
| Materials | Biocompatible | Biocompatible | Biocompatible | Same |
| Adhesive Lifetime | 1 day | N/A | 3–7 days | Comparable to reference; ISO 10993 compliance |
| Principle of Operation | Bone conduction | Bone conduction | Bone conduction | Same |
| Software | Configurable; Pro Fitting Suite & SmartApp | Configurable; Baha Fitting Software & Smart App | Configurable | V&V testing demonstrated that the Pro Fitting Suite and SmartApp perform their intended functions, including device programming, hearing profile management, device control, and user interaction. Testing confirmed that software differences compared to the predicate device do not affect intended use, performance, safety, or effectiveness and do not raise new questions of substantial equivalence. |
This summary includes only information that is publicly available and does not contain any confidential information.
{9}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
6 / 8
Core similarities in principle, intended use, and essential functionality outweigh implementation differences (fixation via single-use adhesives, rechargeable battery, dual-transducer architecture, software with fewer setting tools). These differences do not affect intended use or raise new questions of safety or effectiveness, as confirmed by verification and validation testing.
### H. Summary of Performance Testing
Testing used production-equivalent devices under approved V&V plans with predefined acceptance criteria.
#### a. Biocompatibility
The Ossiclear System was evaluated for biological safety in accordance with ISO 10993 and FDA guidance. Biocompatibility testing was performed on representative samples for cytotoxicity, sensitization, and irritation, and all results met acceptance criteria, confirming that the device is biocompatible for its intended use.
#### b. Electrical & Thermal Safety / Home-use Usability
The Ossiclear System (BHD and Dock) was evaluated for electrical and thermal safety in accordance with IEC 60601-1, IEC 60601-1-6, and IEC 60601-1-11, and usability was validated per IEC 62366-1. Testing confirmed compliance with applicable safety requirements, and usability validation demonstrated that critical tasks were completed successfully without hazardous use errors. These results confirm that the Ossiclear System is safe for home use and presents no unacceptable residual risks.
#### c. Electromagnetic Compatibility (EMC)
The Ossiclear System was evaluated for electromagnetic compatibility in accordance with IEC 60601-1-2 and applicable FCC requirements. The evaluation included emissions and immunity tests using the methods and acceptance criteria defined by the standards.
All tests were successfully passed, with no degradation of safety or essential performance observed during or after exposure to electromagnetic disturbances. The device meets all relevant international and U.S. requirements for EMC, and Bluetooth radio emissions comply with FCC Part 15 limits, with no unintended emissions outside approved frequency bands.
This summary includes only information that is publicly available and does not contain any confidential information.
{10}
sounduct
510(k) Summary - Ossiclear System
[As Required by 21 CFR 807.92]
7 / 8
d. Bench Testing
The Ossiclear System underwent non-clinical bench testing to verify compliance with design specifications and demonstrate substantial equivalence to the predicate device. Testing included evaluation of key electroacoustic performance characteristics of the Ossiclear BHD, such as output, gain, frequency response, noise, and distortion, using recognized bench test methods for bone conduction hearing devices.
All tests met predefined acceptance criteria. These results confirm that the Ossiclear System performs as intended and supports its substantial equivalence.
e. Software & Cybersecurity (Pro Fitting Suite & SmartApp)
The Ossiclear System includes embedded software within the BHD and DOCK, as well as two external applications: Pro Fitting Suite (for professional fitting) and SmartApp (optional patient mobile app). All software was developed in compliance with IEC 62304, classified as Class A, and assessed as Level of Concern: Minor. Verification confirmed that all software components meet functional and performance requirements and operate as intended.
Cybersecurity was addressed following FDA guidance 'Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions'. The Ossiclear SmartApp was developed under a secure-by-design, monitored-for-life approach, incorporating secure coding, authentication and encrypted communication, controlled update processes, and continuous vulnerability monitoring. All activities confirmed that residual cybersecurity risk is acceptable and that the software supports safe and effective operation of the Ossiclear System.
f. Labeling & UDI / Instruction For Use
The Ossiclear System labeling was developed in compliance with FDA and applicable international requirements. Product and packaging labels include the Unique Device Identifier (UDI) in accordance with regulatory requirements, ensuring traceability and correct identification.
The Instructions for Use (IFU) were validated through usability testing per IEC 62366-1 and include comprehensive guidance for device setup, removal, charging procedures, routine care, and safe disposal. The IFU also addresses troubleshooting, maintenance, and safety information to support proper and safe use in home and clinical environments.
This summary includes only information that is publicly available and does not contain any confidential information.
{11}
| | 510(k) Summary - Ossiclear System [As Required by 21 CFR 807.92] |
| --- | --- |
| | 8 / 8 |
# **g. Clinical Data**
No clinical data required for substantial equivalence; non-clinical performance and technological comparability are sufficient.
# **I. Conclusion**
Based on the indications for use, technological characteristics, comparison to the predicate device, and performance data, the Ossiclear System has been demonstrated to be as safe and effective as the predicate device for its intended use. The differences in design, including the use of consumable adhesive pads for fixation, do not raise new questions of safety or effectiveness and have been addressed through verification and validation testing. Therefore, the Ossiclear System is substantially equivalent to the predicate device, Cochlear™ Baha® 6 System (K212136), with supporting evidence from the reference device ADHEAR System (K172460).
*This summary includes only information that is publicly available and does not contain any confidential information.*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.