The Medline Reusable Blood Pressure Cuffs are intended to be used by medical professionals in conjunction with non-invasive blood pressure measurement systems for the determination of a person's blood pressure. The cuffs are non-sterile, multi-patient use and are tested for up to 10,000 inflation cycles and 100 cleaning/Disinfection cycles. Medline Blood Pressure cuff adapters and connectors are intended to be used by medical professionals in conjunction with non-invasive blood pressure systems and blood pressure cuffs for determination of a person's blood pressure. The Blood Pressure Cuff adapter is for single monitor use only. The Blood Pressure Cuff adapter has been validated for use for up to 10,000 cycles or up to 10,000 connections/disconnections.
Device Story
Medline Reusable Blood Pressure Cuff is a non-invasive accessory for sphygmomanometers; consists of inflatable bladder, air tubing, hook-and-loop closure, and standardized connectors. Used by medical professionals in clinical settings to measure patient blood pressure. Cuff wrapped around patient limb; secured via hook-and-loop; connected to non-invasive blood pressure measurement system via tubing/adapters. Device facilitates pneumatic inflation/deflation cycles controlled by the host monitor. Benefits include multi-patient reusability and compatibility with various monitor types. Validated for 10,000 inflation cycles and 100 cleaning/disinfection cycles.
Clinical Evidence
No clinical testing was performed. Bench testing was conducted per ISO 81060-1 to validate performance. Biocompatibility was evaluated per ISO 10993-1 for surface devices with limited contact duration (≤24 hours) with intact skin.
Technological Characteristics
Non-sterile, reusable cuff with internal bladder, hook-and-loop closure, and air tubing. Materials evaluated per ISO 10993-1. Performance testing per ISO 81060-1. Connectivity via standardized connectors/adapters (ISO 80369-5 compatible). Sizes range from infant to adult (excluding neonatal).
Indications for Use
Indicated for use by medical professionals on infant, pediatric, and adult patients (excluding neonates) for non-invasive blood pressure determination in conjunction with compatible measurement systems.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}
June 22, 2026
Medline Industries, LP
Kelsey Closen
Regulatory Affairs Specialist
Three Lakes Dr.
Northfield, Illinois 60093
Re: K254287
Trade/Device Name: Medline Reusable Blood Pressure Cuff
Regulation Number: 21 CFR 870.1120
Regulation Name: Blood pressure cuff
Regulatory Class: Class II
Product Code: DXQ
Dated: May 29, 2026
Received: June 1, 2026
Dear Kelsey Closen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254287 - Kelsey Closen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254287 - Kelsey Closen
Page 3
Sincerely,
**JENNIFER W. SHIH -S**
for LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K254287
?
Please provide the device trade name(s).
?
Medline Reusable Blood Pressure Cuff
Please provide your Indications for Use below.
?
The Medline Reusable Blood Pressure Cuffs are intended to be used by medical professionals in conjunction with non-invasive blood pressure measurement systems for the determination of a person's blood pressure. The cuffs are non-sterile, multi-patient use and are tested for up to 10,000 inflation cycles and 100 cleaning/Disinfection cycles.
Medline Blood Pressure cuff adapters and connectors are intended to be used by medical professionals in conjunction with non-invasive blood pressure systems and blood pressure cuffs for determination of a person's blood pressure. The Blood Pressure Cuff adapter is for single monitor use only. The Blood Pressure Cuff adapter has been validated for use for up to 10,000 cycles or up to 10,000 connections/disconnections.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
Page 1 of 1
{4}

Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
## 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
### Submitter / 510(k) Sponsor
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
Registration Number: 1417592
### Contact Person
Contact Person: Kelsey Closen, Regulatory Affairs Specialist
Phone: 847-949-2283
Fax: 847-643-4482
Email: KClosen@medline.com
### Summary Preparation Date
December 30, 2025
### Type of 510(k) Submission
Traditional
### Device Name / Classification
Trade Name: Reusable Blood Pressure Cuffs
Common Name: Medline Reusable Blood Pressure Cuffs
Product Code: DXQ
Classification Panel: Cardiovascular
Regulatory Class: II
Regulation Number: 21 CFR 870.1120
### Predicate Device
Minhua Reusable Blood Pressure Cuffs cleared in K233276
### Device Description
Medline Reusable Blood Pressure Cuff is an accessory of non-invasive sphygmomanometers for measurement of blood pressure. The device consists of an inflatable integral cuff bladder (the cuff, including the bladder, is one piece), air hose or tubing (single tube or double tube), hook & loop open-closure and standardized connectors that interface with various types of sphygmomanometers. One or two tubes extend from the bladder and connect to a non-invasive blood pressure measurement system via connectors and adapters. Medline Blood Pressure cuff connectors are attached to the blood pressure cuffs via the tubing. Medline adapters are attached to
K254287
Page 1 of 4
{5}

Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
the monitors. Use Medline blood pressure cuffs only with Medline adapters or ISO 80369-5 adapters. Component materials are listed below in table 1.
Each cuff has rounded corners and finished edges for optimal comfort. The cuffs are color coded and come in a wide range of sizes from Infant, pediatric to adult sizes, but not neonatal. The cuffs are non-sterile, reusable, prescription only devices that are offered in a wide range of sizes.
### **Indications for Use**
The Medline Reusable Blood Pressure Cuffs are intended to be used by medical professionals in conjunction with non-invasive blood pressure measurement systems for the determination of a person's blood pressure. The cuffs are non-sterile, multi-patient use and are tested for up to 10,000 inflation cycles and 100 cleaning/Disinfection cycles.
Medline Blood Pressure cuff adapters and connectors are intended to be used by medical professionals in conjunction with non-invasive blood pressure systems and blood pressure cuffs for determination of a person’s blood pressure. The Blood Pressure Cuff adapter is for single monitor use only. The Blood Pressure Cuff adapter has been validated for use for up to 10,000 cycles or up to 10,000 connections/disconnections.
### **Summary of Technological Characteristics**
**TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES**
K254287
Page 2 of 4
{6}

Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| **Product Name** | Medline Reusable Blood Pressure Cuffs | Reusable Blood Pressure Cuff | N/A |
| **510(k) Reference** | TBD | K233276 | N/A |
| **Product Owner** | Medline Industries, LP | Suzhou Minhua Medical Apparatus Supplies Co., LTD | N/A |
| **Product Code** | DXQ | DXQ | Same |
| **Classification** | II | II | Same |
| **Regulation Number** | 870.1120 | 870.1120 | Same |
| **Intended Use** | The Medline Reusable Blood Pressure Cuffs are intended to be used by medical professionals in conjunction with non-invasive blood pressure measurement systems for the determination of a person's blood pressure. The cuffs are non-sterile, multi-patient use and are tested for up to 10,000 inflation cycles and 100 cleaning/Disinfection cycles. | The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in adult size | Similar **The proposed device is also used on infants and pediatrics. The difference is support with the performance testing supporting these additional patient populations are not a safety concern.* |
| **Patient Population** | Infant, pediatric but not neonatal, and adult sizes. | Adult | Similar **The proposed device is also used on infants and pediatrics. The difference is support with the performance testing supporting these additional patient populations are not a safety concern.* |
| **Tube Configuration** | One Tube | One Tube | Same |
| **Bladder** | Internal Bladder | Internal Bladder | Same |
| **Principle of Operations** | The cuffs are wrapped around a patient's limb and secured with a hook-and-loop closure. The hook-and-loop closure consists of a single layer of material. One tube extends from the bladder and is connected to a non-invasive blood pressure measurement system via connector(s). Reusable multi-patient blood pressure cuffs are intended to be used by medical professionals in conjunction with non-invasive blood pressure measurement systems for determination of a person's blood pressure. There is no defined shelf life. | Bladder is wrapped around the patient's limb and secured by hook and loop closure. Air hose is connected to the noninvasive blood pressure measurement systems | Similar **The proposed device has more verbiage but the principles of operations are identical between the predicate and proposed device. The difference in verbiage does not pose any risk or concerns to the safety or effectiveness of the proposed device.* |
| **Cuff Size** | Infant: 9-14.8cm Infant Long: 9-14.8cm Child: 13.8-21.5cm Child Long: 13.8-21.8cm Small Adult: 20.5-28.5cm Small Adult Long: 20.5-28.5cm | Small adult: 17-22cm Small adult long / Adult: 22-30 cm | Difference **The proposed device cuff sizes differ from the predicate device. Performance testing was conduct which illustrates that there are no safety or performance concerns with the difference in cuff sizes.* |
K254287
Page 3 of 4
{7}

Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
| | Adult: 27.5-36.5cm Adult Long: 27.5-36.5cm Large Adult: 35.5-46cm Large Adult Long 35.5-46cm Thigh: 45-56.5cm | Large Adult: 30-42 cm Large Adult long: 42-48 cm Universal: 22-42 cm | |
| --- | --- | --- | --- |
| Sterile vs. Non-Sterile | Non Sterile | Non Sterile | Same |
| Single Use vs. Reusable | Reusable | Reusable | Same |
## Summary of Non-Clinical Testing
### *Biocompatibility Testing*
The proposed device is categorized according to ISO 10993-1, *Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process* as an surface device with a limited (≤24 hours) contact duration with intact skin.
### *Performance Testing (Bench)*
Performance testing was conducted per ISO 81060-1.
### *Performance Testing (Animal)*
This section does not apply. No animal testing was performed.
### *Performance Testing (Clinical)*
This section does not apply. No clinical testing was performed.
## Summary of Clinical Testing
Not applicable.
## Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Reusable Blood Pressure Cuffs are as safe and as effective for their intended use as the predicate device, Minhua Reusable Blood Pressure Cuffs cleared in K233276
K254287
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.